In addition to the 2 GHz RF generator,
the IFR-1900 CSA also provides full
audio/data generator capabilities, full level
control and measurement facilities and
precision power control features for
enhanced sensitivity and high accuracy
testing needs.
Analog paging encoding/decoding,
DTMF, tone coded squelch, digital squelch,
AM modulation/demodulation along with
two separate AF generators and cross band
duplex gives added test versatility in a
variety of wireless systems.
Software Options Make Complex
Testing Simple
As with every IFR test set, you get the
advantage of IFR applications engineering
support.
Our comprehensive portfolio of
application software options are designed
to automate and expand the functionality
of your instrument.
Plus, on-going software support means
that you can easily upgrade your
IFR-1900 CSA when test and service
requirements change.
AC1009W - EasySpanTMis a Windowsbased software utility which extracts
spectrum analyzer and tracking generator
traces from the IFR-1900 CSA.
AC1017 - AutoCell-Series II is a
comprehensive program for FCC
compliance testing of Lucent Series II cell
sites.
AC1019 - EasySweepTMis a swept
measurement utility designed to test
antennas and transmission lines.
AC1020D - AutoCell NTD provides
automated testing of Northern Telecom cell
sites.
AC1021 - CellScanTMcellular utility
software simplifies combiner alignment,
monitoring RF levels and base stations on
DAMPS, NT400 or PCS cellular channel
sets.
AC1027 - AutoCell-882/884 is an
autotest program for performing
acceptance tests on Ericsson
884/882/882D/882M/882DM base
stations.
EasySpan Software
Dynamic IQ Constellation Display
Simplifies Analysis
The IFR-1900 CSA provides you with a
dynamic constellation display for precise
RF modulation analysis of DQPSK digitally
modulated waveforms from 10 MHz to
2010 MHz. This unique IFR-1900 CSA
feature gives a near real time display for
testing and troubleshooting, an ability that
points out the cause of the trouble in digital
radios.
IQ Constellation display allows for comprehensive
digital modulation testing
Complex Functionality Made Easy
Even with its elaborate capabilities, the
IFR-1900 CSA was developed to execute
complex tests simply and with minimal
operator training.
Using field-proven front panel and user
man-machine interfaces, the IFR-1900 CSA
gives you the performance and ease-of-use
features that reduces your testing and
training time.
Its test macro command programming
language (TMAC) can be easily configured
to perform automatic base station testing
and remote terminal, singl
e and multimode, single and multiband equipment.
This powerful capability allows you to
create and save simple one button test
routines for future use. This flexibility
means you can create and execute
complex and repeatable routines no matter
what your level of expertise.
A new color VGA display gives
IFR-1900 CSA users vivid screen clarity.
Extensive use of softkeys reduces your
complex cellular / PCS parametric and
protocol tests to fast, simple and
manageable routines.
Specification
(T/R) AND DUPLEX CONNECTOR
Frequency Range
10 MHz to 2010 MHz
Resolution
100 Hz
Accuracy
Same as Master Oscillator
Range
-127 dBm to + 10 dBm into 50 Ω
(T/R Connector: -30 dBm maximum with reverse
power present)
Resolution
0.1 dB
Accuracy
± 1.5 dB (≥-110 dBm)
Duplex Connector Input Protection
Alarm sounds when level exceeds +20 dBm.
EXTERNAL
Generator IF Output Frequency Range
88 MHz to 90 MHz
Level Range
-30 dBm to +25 dBm into 50 Ω
Residual FM
<10 Hz RMS
SSB Phase Noise
<-94 dBc/Hz (20 kHz Offset)
Residual AM
<0.3% RMS (50 Hz to 15 kHz BW)
Non-Harmonics
<-50 dBc
Nominal Input Frequency for Generator IF In
90 MHz
Input Level
-26 dBm to +28 dBm into 50 Ω
Signal Bandwidth
8.5 MHz
System Gain
28 dB ± 7dB, from GEN IF Input to T/R Ports
INTERNAL FM
Range
Off and ±100 Hz to ±100 kHz Dev.
Accuracy
± 5% (1 kHz to 20 kHz Dev, 1 kHz rate)
± 10% (1 kHz Dev and >20 kHz Dev, 1 kHz rate)
Resolution
100 Hz
Modulation Rate
1 kHz to 10 kHz - 5% accuracy
Waveforms
Sine, Square, Triangle
INTERNAL PHASE/QUADRATURE (IQ)
RF Ranges
10 MHz to 2010 MHz
IQ Error Vector Magnitude
5% from ideal DQPSK waveform
(TIA/EIA-136)
IQ Origin Offset
<-30 dBc
AF Generators #1 and #2
Range
10 Hz to 40 kHz
Resolution
0.1 Hz ≤2 kHz
1 Hz >2 kHz
Accuracy
±0.1%
Waveshapes
Sinewave, Square, Triangle, Ramp, Pulse
Frequency Range
10 Hz to 200 kHz (in 4 decade ranges)
Accuracy
Same as Master Oscillator
Resolution
0.1 Hz (10 Hz to 2 kHz)
1 Hz (>2 kHz to 20 kHz)
10 Hz (>20 kHz to 40 kHz)
Input Waveform
Sine or Square
External Level
0.5 VRMS to 30 VRMS (SINAD/BER input)
0.1 VRMS to 3.5 VRMS (EXT MOD input)
Audio Frequency Counter
AF Signal Generators
Modulation
RF Signal Generator
IFR-1900 CSA
IFR-1900 CSA
2
http://www.ifrinternational.com