In addition to the 2 GHz RF generator, 
the IFR-1900 CSA also provides full 
audio/data generator capabilities, full level 
control and measurement facilities and 
precision power control features for 
enhanced sensitivity and high accuracy 
testing needs. 
Analog paging encoding/decoding, 
DTMF, tone coded squelch, digital squelch, 
AM modulation/demodulation along with 
two separate AF generators and cross band 
duplex gives added test versatility in a 
variety of wireless systems.
Software Options Make Complex 
Testing Simple
As with every IFR test set, you get the 
advantage of IFR applications engineering 
support.
Our comprehensive portfolio of 
application software options are designed 
to automate and expand the functionality 
of your instrument.
Plus, on-going software support means 
that you can easily upgrade your 
IFR-1900 CSA when test and service 
requirements change.
AC1009W - EasySpanTMis a Windowsbased software utility which extracts 
spectrum analyzer and tracking generator 
traces from the IFR-1900 CSA.
AC1017 - AutoCell-Series II is a 
comprehensive program for FCC 
compliance testing of Lucent Series II cell 
sites.
AC1019 - EasySweepTMis a swept 
measurement utility designed to test 
antennas and transmission lines.
AC1020D - AutoCell NTD provides 
automated testing of Northern Telecom cell 
sites.
AC1021 - CellScanTMcellular utility 
software simplifies combiner alignment, 
monitoring RF levels and base stations on 
DAMPS, NT400 or PCS cellular channel 
sets. 
AC1027 - AutoCell-882/884 is an 
autotest program for performing 
acceptance tests on Ericsson 
884/882/882D/882M/882DM base 
stations.
EasySpan Software
Dynamic IQ Constellation Display 
Simplifies Analysis
The IFR-1900 CSA provides you with a 
dynamic constellation display for precise 
RF modulation analysis of DQPSK digitally 
modulated waveforms from 10 MHz to 
2010 MHz. This unique IFR-1900 CSA 
feature gives a near real time display for 
testing and troubleshooting, an ability that 
points out the cause of the trouble in digital 
radios.
IQ Constellation display allows for comprehensive
digital modulation testing 
Complex Functionality Made Easy
Even with its elaborate capabilities, the 
IFR-1900 CSA was developed to execute 
complex tests simply and with minimal 
operator training.
Using field-proven front panel and user 
man-machine interfaces, the IFR-1900 CSA 
gives you the performance and ease-of-use 
features that reduces your testing and 
training time.
Its test macro command programming 
language (TMAC) can be easily configured 
to perform automatic base station testing 
and remote terminal, singl
e and multimode, single and multiband equipment. 
This powerful capability allows you to 
create and save simple one button test 
routines for future use. This flexibility 
means you can create and execute 
complex and repeatable routines no matter 
what your level of expertise.
A new color VGA display gives 
IFR-1900 CSA users vivid screen clarity. 
Extensive use of softkeys reduces your 
complex cellular / PCS parametric and 
protocol tests to fast, simple and 
manageable routines.
Specification
(T/R) AND DUPLEX CONNECTOR 
Frequency Range
10 MHz to 2010 MHz
Resolution
100 Hz
Accuracy
Same as Master Oscillator
Range
-127 dBm to + 10 dBm into 50 Ω 
(T/R Connector: -30 dBm maximum with reverse 
power present)
Resolution
0.1 dB 
Accuracy
± 1.5 dB (≥-110 dBm)
Duplex Connector Input Protection
Alarm sounds when level exceeds +20 dBm.
EXTERNAL 
Generator IF Output Frequency Range
88 MHz to 90 MHz
Level Range
-30 dBm to +25 dBm into 50 Ω
Residual FM
<10 Hz RMS
SSB Phase Noise
<-94 dBc/Hz (20 kHz Offset)
Residual AM
<0.3% RMS (50 Hz to 15 kHz BW)
Non-Harmonics
<-50 dBc 
Nominal Input Frequency for Generator IF In
90 MHz
Input Level
-26 dBm to +28 dBm into 50 Ω
Signal Bandwidth
8.5 MHz
System Gain
28 dB ± 7dB, from GEN IF Input to T/R Ports
INTERNAL FM 
Range
Off and ±100 Hz to ±100 kHz Dev.
Accuracy
± 5% (1 kHz to 20 kHz Dev, 1 kHz rate) 
± 10% (1 kHz Dev and >20 kHz Dev, 1 kHz rate)
Resolution
100 Hz
Modulation Rate
1 kHz to 10 kHz - 5% accuracy
Waveforms
Sine, Square, Triangle
INTERNAL PHASE/QUADRATURE (IQ) 
RF Ranges
10 MHz to 2010 MHz
IQ Error Vector Magnitude
5% from ideal DQPSK waveform 
(TIA/EIA-136)
IQ Origin Offset
<-30 dBc
AF Generators #1 and #2 
Range
10 Hz to 40 kHz
Resolution
0.1 Hz ≤2 kHz 
1 Hz >2 kHz
Accuracy
±0.1%
Waveshapes
Sinewave, Square, Triangle, Ramp, Pulse
Frequency Range
10 Hz to 200 kHz (in 4 decade ranges)
Accuracy
Same as Master Oscillator
Resolution
0.1 Hz (10 Hz to 2 kHz) 
1 Hz (>2 kHz to 20 kHz) 
10 Hz (>20 kHz to 40 kHz)
Input Waveform
Sine or Square
External Level
0.5 VRMS to 30 VRMS (SINAD/BER input)
0.1 VRMS to 3.5 VRMS (EXT MOD input)
Audio Frequency Counter
AF Signal Generators
Modulation
RF Signal Generator
IFR-1900 CSA
IFR-1900 CSA
2
http://www.ifrinternational.com