DATA SHEET > EFT 500N5-series > 20130321
EFT 500N5 SERIES
ELECTRICAL FAST TRANSIENT GENERATOR
FOR TESTS ACCORDING TO ...
> ANSI/IEEE C62.41 > EN 61000-4-4 > EN 61000-6-1 > EN 61000-6-2 > IEC 60601-1-2 > IEC 61000-4-4 > IEC 61326 > IEC 61850-3
EFT 500N5 SERIES - EFT/BURST TESTER
The EFT 500N5 series are the state-of-the-art EFT/Burst testers with a built-in coupling/decoupling network for single phase or three-phase mains supply lines and a HV terminal to connect a capacitive coupling clamp. The spike frequency can be varied from as low as 100Hz up to 1,000kHz. The EFT 500N5 is capable to generate 10,000 spikes per second or up to 1,000 spikes per burst. Its performance fully complies with IEC 61000-4-4 Ed.2:2004. Pre-programmed Standard Test routines and a number of selected User Test routines make the EFT 500N5 an optimum standalone solution for both testing during development and final testing of products and sub-assemblies.
HIGHLIGHTS
> 10,000 pulses/sec.
APPLICATION AREAS
INDUSTRY
TELECOM
> 4.8kV Burst voltage > Spike frequency up to 1MHz
MEDICAL
COMPONENTS
> Built-in 1ph or 3ph coupler up to 50A > Pulse verification acc. to Ed.2 on each coupling
BROADCAST
path
RESIDENTIAL
www.emtest.com © EM TEST > PAGE 1/3
DATA SHEET > EFT 500N5-series > 20130321
TECHNICAL DETAILS
EFT 500N5 MODELS
EFT 500N5.1 With 3-phase coupler, 3x 440V/32A EFT 500N5.2 With 3-phase coupler, 3x 440V/50A EFT 500N5.5 With 1-phase coupler, 250V/32A EFT 500N5.8 With 3-phase coupler, 3x 690V/32A
Other couplers on request
TEST LEVEL
Voltage range 200V - 4,800V Into 50ohm 100V - 2,400V Pulse shape Into 50ohm load Rise time tr 5ns ± 30% Pulse width td 50ns ± 30% Pulse shape Into 1,000ohm load Rise time tr 5ns ± 30% Pulse width td 35ns to 150ns Source impedance 50 Ohm ± 20% Polarity Positive/negative
TRIGGER
Release of Burst Automatic, manual, external Burst duration Td = 0.1ms - 999.9ms Burst repetition
Tr = 10ms - 9,999ms
rate Spike frequency f = 0.1kHz - 1,000kHz Test duration T = 0:01 min - 99:59 min
T > 99:59 min - endless
OUTPUT
Direct Via HV coax connector
TEST ROUTINES
Quick Start Immediate start; easy to use and fast User Test routines Custom made test routine
Change voltage after T Change frequency after T Frequency sweep within a burst Change burst duration after T Random release of bursts Synchronous burst mode Change polarity after T
Standard Test routines
As per IEC 61000-4-4, Automatic test Level X - Y
Service Service, set-up, self test
INTERFACES
Serial interface USB Parallel interface IEEE 488, address 1 - 30 CN interface To control external CNI 503
GENERAL DATA
Dimensions, weight
EFT 500N5.1: 19"/6HU, 22kg EFT 500N5.2: 19"/6HU, 40kg EFT 500N5.5: 19"/3HU, 20kg
EFT 500N5.8: 19"/3HU, 23kg Supply voltage 115/230V, +10/-15% Fuses 2 x T 2AT (230V); 2 x T 4AT (115V)
OPTIONS
HFK Coupling Clamp as per IEC
61000-4-4 KW 50 50ohm matching resistor for
verification KW 1000 1,000ohm matching resistor for
verification
Coupler L, N and PE, all combinations DUT supply AC single phase: 250V/32A, 50/60Hz
three-phase: 3x440V/32A or 50A,
50/60Hz DUT supply DC 250V/10A DUT supply Other DUT supply voltages on
request CRO Trigger 15V trigger signal for oscilloscope;
burst duration and pulse frequency
can be checked
www.emtest.com © EM TEST > PAGE 2/3
CA EFT kit Kit for burst pulse verification
consisting of KW50, KW1000 and adapter for DUT port with plastic
case for storage A6dB 6dB attenuator CNI 503 Coupling/decoupling networks,
3x440V/32-63-100A iec.control Software to control the test,
including standard library, test
report facility and data conversion
generator