DATA SHEET > EFT 500N5-series > 20130321
EFT 500N5 SERIES
ELECTRICAL FAST TRANSIENT GENERATOR
FOR TESTS ACCORDING TO ...
> ANSI/IEEE C62.41
> EN 61000-4-4
> EN 61000-6-1
> EN 61000-6-2
> IEC 60601-1-2
> IEC 61000-4-4
> IEC 61326
> IEC 61850-3
EFT 500N5 SERIES - EFT/BURST TESTER
The EFT 500N5 series are the state-of-the-art EFT/Burst testers with a built-in coupling/decoupling network for single phase
or three-phase mains supply lines and a HV terminal to connect a capacitive coupling clamp.
The spike frequency can be varied from as low as 100Hz up to 1,000kHz. The EFT 500N5 is capable to generate 10,000
spikes per second or up to 1,000 spikes per burst.
Its performance fully complies with IEC 61000-4-4 Ed.2:2004. Pre-programmed Standard Test routines and a number of
selected User Test routines make the EFT 500N5 an optimum standalone solution for both testing during development and
final testing of products and sub-assemblies.
HIGHLIGHTS
> 10,000 pulses/sec.
APPLICATION AREAS
INDUSTRY
TELECOM
> 4.8kV Burst voltage
> Spike frequency up to 1MHz
MEDICAL
COMPONENTS
> Built-in 1ph or 3ph coupler up to 50A
> Pulse verification acc. to Ed.2 on each coupling
BROADCAST
path
RESIDENTIAL
www.emtest.com © EM TEST > PAGE 1/3
DATA SHEET > EFT 500N5-series > 20130321
TECHNICAL DETAILS
EFT 500N5 MODELS
EFT 500N5.1 With 3-phase coupler, 3x 440V/32A
EFT 500N5.2 With 3-phase coupler, 3x 440V/50A
EFT 500N5.5 With 1-phase coupler, 250V/32A
EFT 500N5.8 With 3-phase coupler, 3x 690V/32A
Other couplers on request
TEST LEVEL
Voltage range 200V - 4,800V
Into 50ohm 100V - 2,400V
Pulse shape Into 50ohm load
Rise time tr 5ns ± 30%
Pulse width td 50ns ± 30%
Pulse shape Into 1,000ohm load
Rise time tr 5ns ± 30%
Pulse width td 35ns to 150ns
Source impedance 50 Ohm ± 20%
Polarity Positive/negative
TRIGGER
Release of Burst Automatic, manual, external
Burst duration Td = 0.1ms - 999.9ms
Burst repetition
Tr = 10ms - 9,999ms
rate
Spike frequency f = 0.1kHz - 1,000kHz
Test duration T = 0:01 min - 99:59 min
T > 99:59 min - endless
OUTPUT
Direct Via HV coax connector
TEST ROUTINES
Quick Start Immediate start; easy to use and fast
User Test routines Custom made test routine
Change voltage after T
Change frequency after T
Frequency sweep within a burst
Change burst duration after T
Random release of bursts
Synchronous burst mode
Change polarity after T
Standard Test
routines
As per IEC 61000-4-4,
Automatic test Level X - Y
Service Service, set-up, self test
INTERFACES
Serial interface USB
Parallel interface IEEE 488, address 1 - 30
CN interface To control external CNI 503
GENERAL DATA
Dimensions,
weight
EFT 500N5.1: 19"/6HU, 22kg
EFT 500N5.2: 19"/6HU, 40kg
EFT 500N5.5: 19"/3HU, 20kg
EFT 500N5.8: 19"/3HU, 23kg
Supply voltage 115/230V, +10/-15%
Fuses 2 x T 2AT (230V); 2 x T 4AT (115V)
OPTIONS
HFK Coupling Clamp as per IEC
61000-4-4
KW 50 50ohm matching resistor for
verification
KW 1000 1,000ohm matching resistor for
verification
Coupler L, N and PE, all combinations
DUT supply AC single phase: 250V/32A, 50/60Hz
three-phase: 3x440V/32A or 50A,
50/60Hz
DUT supply DC 250V/10A
DUT supply Other DUT supply voltages on
request
CRO Trigger 15V trigger signal for oscilloscope;
burst duration and pulse frequency
can be checked
www.emtest.com © EM TEST > PAGE 2/3
CA EFT kit Kit for burst pulse verification
consisting of KW50, KW1000 and
adapter for DUT port with plastic
case for storage
A6dB 6dB attenuator
CNI 503 Coupling/decoupling networks,
3x440V/32-63-100A
iec.control Software to control the test,
including standard library, test
report facility and data conversion
generator