E-FIELD PROBE
EF0391
Measuring electric fields
from 100 kHz to 3 GHz
using instruments in the NBM-500 family
General public and occupational field
exposure from broadcasting, telecoms and
industrial equipment
Isotropic (non-directional) measurement
64 dB dynamic range without changing
measurement range
Maximum sensitivity starts at 0.2 V/m
The probe contains three orthogonally arranged dipoles with detector
diodes. The three voltages, corresponding to the spatial components,
are available individually at the probe output. The NBM basic unit
calculates the resulting isotropic field strength.
APPLICATIONS
The probe detects electric fields from 100 kHz to 3 GHz, covering the
fields that occur in broadcasting, telecoms, and industry. The high
sensitivity of 0.2 V/m and excellent linearity make it ideal for measuring
human safety limit values in the general public domain.
PROPERTIES
The probe is designed with mechanical and electrical properties ideal for
field use. The probe head is made of foam material to provide effective
protection for the sensors, while having excellent RF characteristics.
The electric destruction limits for pulsed and continuous wave signals
are several times higher than any of the human safety limit values.
CALIBRATION
The probe is calibrated at several frequencies. The correction values are
stored in an EPROM in the probe and are automatically taken into
account by the NBM instrument. Calibrated accuracy is thus obtained
regardless of the combination of probe and instrument.
NSTS 0209-E0233B-0.0 1 / 2 Subject to change
SPECIFICATIONS a
Probe EF0391 Electric (E-)Field
Frequency range
Type of frequency response Flat
Measurement range
Dynamic range 64 dB
CW damage level 800 V/m 170 mW/cm²
Peak damage level
Sensor type Diode based system
Directivity Isotropic (Tri-axial)
Readout mode / spatial assessment 3 separate axes
UNCERTAINTY
Flatness of frequency response
Calibration uncertainty not included
Calibration uncertainty
@ 0.01 mW/cm² (6.14 V/m)
Linearity
Referred to 0.01 mW/cm² (6.14 V/m)
Isotropic response
Temperature response +0.2/ -1 dB (±0.025 dB/K)
GENERAL SPECIFICATIONS
Calibration frequencies
Recommended calibration interval 24 months
Temperature range
Operating
Non-operating (transport)
Humidity 5 to 95 % RH @ ≤25 °C ≤23 g/m³ absolute humidity
Size 318 mm x 66 mm Ø
Weight 90 g
Compatibility NBM-500 series meters
Country of origin Germany
Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 25% to 75%, sinusoidal signal
(a)
(b)
Cutoff frequency at approx. -3 dB
(c) Pulse length 1µsec, duty cycle 1:100
Frequency response can be compensated for by the use of correction factors stored in the probe memory
(d)
(e)
Accuracy of the fields generated to calibrate the probes
Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe
(f)
(b)
(c)
(d)
(e)
(f)
±1 dB
100 kHz to 3 GHz
0.2 to 320 V/m (CW)
0.2 to 10 V/m (True RMS)
10 nW/cm² to 27 mW/cm² (CW)
10 nW/cm² to 0.027 mW/cm² (True RMS)
8 kV/m 17 W/cm²
±1 dB (1 MHz to 1 GHz)
±1.25 dB (1 GHz to 2.45 GHz)
±1 dB (<400 MHz)
±1.5 dB (400 MHz to 1.8 GHz)
±1 dB (≥ 1.8 GHz)
±0.5 dB (1.2 to 200 V/m)
±0.7 dB (200 to 320 V/m)
±0.5 dB (0.00038 to 10.6 mW/cm²)
±0.7 dB (10.6 to 27 mW/cm²)
0.1/ 0.2/ 0.3/ 1/ 3/ 10/ 27.12 MHz
0.1/ 0.2/ 0.3/ 0.5/ 0.75/ 1/ 1.8/ 2.45/ 2.7/ 3 GHz
0 °C to +50 °C
-40 °C to +70 °C
ORDERING INFORMATION
Probe EF0391, E-field for NBM, 100 kHz – 3 GHz, isotropic 2402/01
Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7
72793 Pfullingen, Germany
Phone: +49 (0) 7121-97 32-777
Fax: +49 (0) 7121-97 32-790
E-Mail: support@narda-sts.de
www.narda-sts.de
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. - Trade names are trademarks of the owners.
NSTS 0209-E0233B-0.0 2 / 2 Subject to change
Narda Safety Test Solutions
435 Moreland Road
Hauppauge, NY 11788, USA
Phone: +1 631 231-1700
Fax: +1 631 231-1711
E-Mail: NardaSTS@L-3COM.com
www.narda-sts.com
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano), Italy
Phone: +39 02 2699871
Fax: +39 02 26998700
E-mail: support@narda-sts.it
www.narda-sts.it
Part number