> EN 301489-1 > EN 301489-17 > EN 301489-24 > EN 301489-7 > EN 61000-3-2 > EN 61000-3-3 > EN 61000-4-15 > EN 61000-4-7 > EN 61000-6-1 > EN 61000-6-2 > IEC 60601-1-2 > IEC 61000-3-11 > IEC 61000-3-12 Ed.2:2011 > IEC 61000-3-2 > IEC 61000-3-3 > IEC 61000-4-15 Ed.2:2010 > IEC 61000-4-7 > IEC 61326 > JIS C 61000-3-2
DPA 500N - FULL-COMPLIANT SINGLE PHASE HARMONICS AND FLICKER ANALYZER
Harmonics and interharmonics are caused by modern electronic power conditioning modules. Such, mostly non-linear, modules to control loads and to reduce power consumption is the source of voltage at unwanted frequencies superposed on the supply voltage. Voltage fluctuations caused by varying load currents may influence the luminance or the spectral distribution of lighting systems. The impression of the unsteadiness of visual sensation induced by this light stimulus is called flicker. The DPA500N harmonics and flicker analyzer is used for single phase applications.
HIGHLIGHTS
> Real-time data acquisition
OBSZARY ZASTOSOWAŃ
INDUSTRY
TELECOM
> Internal hard disk for data storage > 16-Bit A/D converter
MEDICAL
ENERGIA ODNAWIALNA
> Wide-range current input > Wide-range voltage input
BROADCAST
> Built-in lumped flicker impedance
RESIDENTIAL
> High-sophisticated analyzing capability
www.emtest.com © EM TEST > STRONA 1/4
KARTA KATALOGOWA > DPA 500N > 20120125
PARAMETRY TECHNICZNE
BENEFITS
THE FULLY COMPLIANT HARMONICS AND FLICKER ANALYZER
The DPA 500N is a fully compliant analyzer for harmonics and flicker as per the latest IEC/EN 61000-3-2, IEC/EN 61000-3-3 and JIS C 61000-3-2 requirements. It follows the design specifications as per IEC/EN 61000-4-7:2002 and Am.1:2008 (for Class A instruments) and IEC/EN 61000-4-15:2010. Based on a real-time kernel and equipped with its own harddisk it allows to record the measuring data continuously without any gaps or overlapping. The rectangular measurement window is synchronized to each group of 10 or 12 cycles of the mains supply frequency (50Hz or 60Hz) by means of a digital PLL (Phase Lock Loop). A wide-range current input (up to 50A) avoids loss of measured data due to range switching that would occur when using different shunt resistors. The classification of the measurement can be selected at any time, before or after the measurement has been performed. During all measurements the AC supply voltage is measured simultaneously with the current. The built-in flicker impedance, which is automatically selected when doing flicker measurement, makes the DPA 500N a complete single-box unit for both harmonics and flicker analysis. By means of an external current clamp (optional) the current range can be extended up to 140Arms and more.
OTHER MODELS
DPA 500 HARMONICS AND FLICKER ANALYZER SERIES
EM TEST offers two models of analyzers for harmonics and flicker for single phase applications (DPA 500N) and three-phase applications (DPA 503). The three-phase analyzer DPA503 can be used for single phase analysis as well.
AUXILIARY DEVICES
ACS 500N6 - SINGLE PHASE AC VOLTAGE SOURCE 6KVA
The ACS 500N6 single phase AC voltage source is used to provide a pure AC supply voltage for harmonics and flicker analysis as recommended by IEC/EN 61000-3-2, IEC/EN 61000-3-3 and JIS C 61000-3-2. It offers a rated power of 6kVA and the output voltage ranges up to 300V. The ACS 500N6 AC voltage source is controlled by dpa.control.
ACS 500N3 - SINGLE PHASE AC VOLTAGE SOURCE 3KVA
The ACS 500N3 single phase AC voltage source is used to provide a pure AC supply voltage for harmonics and flicker analysis as recommended by IEC/EN 61000-3-2, IEC/EN 61000-3-3 and JIS C 61000-3-2. It offers a rated power of 3kVA and the output voltage ranges up to 300V. This AC voltage source is used for low-power applications. The ACS 500N3 AC voltage source is controlled by dpa.control.
SOFTWARE
DPA.CONTROL - THE CONTROL AND ANALYSING SOFTWARE FOR HARMONICS AND FLICKER
dpa.control is the software tool needed for the operation of the harmonics and flicker analysing system. It offers all features to control the DPA 500N, to upload the recorded measuring data and for the classification and analysis. It includes analysis as per the latest standards as well as procedures following the former standard requirements. An easy Fail/Pass function allows fast analysis while detailed data is available for extended analysis and EUT evaluation purposes. dpa.control offers a powerful documentation capability with direct export to Word.
www.emtest.com © EM TEST > STRONA 2/4
KARTA KATALOGOWA > DPA 500N > 20120125
PARAMETRY TECHNICZNE
MEASURING SYSTEM
Input channels 2 (1x current & 1x voltage) Frequency range 15Hz - 3,000Hz A/D converter 16 Bit Controller Embedded processor Pentium
200MHz Signal processor Motorola DSP Memory Internal hard disk Category Class I as per IEC/EN 61000-4-7
VOLTAGE INPUT
Input range 10V - 530V rms Overload 4,000V peak Accuracy Better than 0.4% of reading
CURRENT INPUT
Input range
16A continuous, 50A short time internal
Accuracy internal Better than 0.4% of reading
Better than 0.05% rel. to 16A Input range
external
Depending on CT model used.
With optionally delivered CT
max. 140A Accuracy externalCTRelated to 16A
2 turns better than 0.8%
5 turns better than 0.6%
HARMONICS ANALYSIS
As per IEC/EN 61000-3-2
JIS C 61000-3-2 IEC/EN 61000-3-12 with ext. CT
Design as per IEC/EN 61000-4-7 (2002 &
Am.1:2008) Harmonics 1st - 50th order Grouping IEC/EN 61000-4-7 (2002 &
Am.1:2008) for Interharmonics Synchronization PLL; accuracy better than 0.005% Measuring
window
Rectangular window with 8, 10, 12
or 16 periods) Algorithm FFT Smoothing filter 1st order 1,5s digital low pass filter
(on/off), selectable Anti-aliasing filter > 90dB Measurement
duration
More than 30 hours, limited by the
capacity of the hard-disk (approx.
1MB/min of measuring data) Display Vrms, Irms, Ipeak, Vpeak Harmonics V, I, Phase, P, Q, S
(2nd - 50th order) Power information P, Q, S, Power factor, THD(V),
THD(I),
Crest factor(V), Crest factor(I)
FLICKER ANALYSIS
As per IEC/EN 61000-3-3
GENERAL DATA
Temperature 0°C - 40°C Rel. humidity 10% - 90%, non-condensing Power supply 85V - 255V, 47Hz - 63Hz Power Max. 50W Dimension 19" 3HU: 133mm x 449mm x 500mm Weight 13kg Insulation Input to case 3kV rms Interface USB for control and data transfer
Design as per IEC/EN 61000-4-15 (2003 & 2010)
230V , 50/60 Hz and
120V , 50/60 Hz Flicker impedance
(built-in) Accuracy Pst and
Line: 0.24ohm + j0.15ohm
Neutral: 0.16ohm + j0.10ohm
Better than 5% Plt
Accuracy dmax,
0.15%
dc, dt Flicker data Pst and Plt, Vrms, dmax, dc, dt
P50%S, P10%S, P3%S, P1%S,
P0.1% Maximum values Pst, dmax, dc, dt Observation
Min. 1min, selectable period
www.emtest.com © EM TEST > STRONA 3/4
KARTA KATALOGOWA > DPA 500N > 20120125
KOMPETENTNI, GDZIEKOLWIEK BYŚ BYŁ
BEZPOŚREDNI KONTAKT Z EM TEST
Szwajcaria
EM TEST (Switzerland) GmbH > Sternenhofstraße 15 > 4153 Reinach > Switzerland Telefon +41 (0)61/7179191 > Fax +41 (0)61/7179199 Internet: www.emtest.ch > E-mail: sales@emtest.ch
Niemcy
EM TEST GmbH > Lünener Straße 211 > 59174 Kamen > Deutschland Telefon +49 (0)2307/26070-0 > Fax +49 (0)2307/17050 Internet: www.emtest.com > E-mail: info@emtest.de
Francja
EM TEST FRANCE > Le Trident - Parc des Collines > Immeuble B1 - Etage 3 > 36, rue Paul Cézanne > 68200 Mulhouse > France Telefon +33 (0)389 31 23 50 > Fax +33 (0)389 31 23 55 Internet: www.emtest.fr > E-mail: info@emtest.fr
Polska
EM TEST Polska > ul. Ogrodowa 31/35, 00-893 Warszawa > Polska Telefon +48 (0)518 64 35 12 Internet: www.emtest.com/pl > E-mail: info.polska@emtest.de
Wszystkie informacje dotyczące dostaw, wyglądu zewnętrznego i danych technicznych opisujących aktualny stan wraz z datą publikacji mogą ulec zmianie bez wcześniejszych informacji.
USA / Kanada
EM TEST USA Inc. > 9250 Brown Deer Road > San Diego > CA 92121 Telefon +1 (858) 699 1685 > Fax +1 (858) 458 0267 Internet: www.emtest.com > E-mail: tom.revesz@ametek.com