Atec DM70 User Manual

DM70 Hand-held Dewpoint Meter
for Spot Checking and Field Verification
FEATURES/BENEFITS
 New DRYCAP® polymer-based sensor provides
accuracy, reliability, long-term stability and
fast response  Unique automatic self-calibration  ± 2° C accuracy down to -58 °F (-50 °C )  Data logging and transferring capabilities  Multilingual, menu-based interface  Numeric and graphical display  Sample cell adapter for pressurized air lines
The DM70 provides a new level of fuctionality in hand­held dewpoint meters. It serves as an independent data logger, and an integrated calibration reference for the DMT242 and DMP246 transmitters.
The DM70 measures dewpoint temperature accurately in a wide measurement range: from -58 °F to +140 °F (-50 °C to +60 °C). It features a fast response time between start-up and low or high process dewpoints. The DM70 incorporates Vaisalas advanced DRYCAP polymer-based technology, which enables reliable and high performance dewpoint measurement. The probe withstands condensation and may be placed directly in pressurized spaces.
®
DM70 Hand-held Dewpoint Meter with
optional Sample Cell Adapter
Numerical and graphical measurement data
Measurement data can be displayed numerically or graphically in the LCD display. One key press changes the readout to a real time graphics display, so the user will know when the dewpoint plateau is reached. The graphic history shows the data curve from the time the device was turned on. It is easily scalable in the display.
Data-logging & data-transferring capability
The DM70 can be used as an on-site data-logger. To simplify operation, the DM70 has a sophisticated MI70 Link Windows® software program. It is a practical tool for transferring logged data and real time measurement data from the DM70 to a PC.
New Dewpoint Technology
The DRYCAP® measurement technology, with the patented auto-calibration feature, provides revolutionary long-term stability that results in a long calibration interval. Another advantage is the sensors durability against condensation.
The DM70 is calibrated at the factory against internation­ally traceable standards, and delivered with a calibration certificate. It can also be sent to a Vaisala Service Center for a traceable re-calibration.
Various Applications
Spot-checking instrument
The DM70 has a fast response time for temporary dewpoint measurements. It can be used in industrial dewpoint spot-checking applications, such as com­pressed air lines and plastic molding processes.
Field-calibration instrument
The DM70 can perform the one-point calibration check in the process for Vaisalas DMT242 and DMP248 fixed dewpoint transmitters. Transmitters can be connected to DM70 with optional connection cables.
DSC74 Sample Cell Adapter
The DM70 can easily be connected to pressurized processes by using the optional Sample Cell Adapter/ Quick Connector DSC74.
DM70KIT
The DM70 Kit includes:
 DM70 indicator + probe with rechargeable battery
and 115 VAC adapter  MIK70 Link and analog output signal cable  DMP74 Probe  DSC74 Sample cell and quick connector  Carrying case  Certificate of calibration, traceable to NIST
100 Commerce Way, Woburn, MA 01801 Tel: 1-888-VAISALA (824-7252) Fax: (781) 933-8029 Email: incsales@vaisala.com Online: www.vaisala-usa.com
TECHNICAL DATA - DM70
DMP74 Dewpoint Measurement Probe Measured variables
Dewpoint
Measurement range (typical) -76+140 °F (-60...+60 °C) Accuracy -50...+60 °C ±3.6 °F (±2 °C)
see the graph below
70 60 50 40 30 20 10
0
-10
-20
-30
-40
-50
-60
-70
-80
Dewpoint temperature ( C)
-10
c
a
r
u
c
c
A
n
i
o
p
w
e
D
0
20
Process temperature ( C)
d
T
C
2
±
<
y
t
i
m
i
l
r
e
w
o
l
e
g
n
a
r
t
n
e
m
e
r
u
s
a
e
m
t
40
3010
60
50
Response time 63% (90%) at flow rate 0.2 m/s, 1 bar pressure, +68 °F (+20 °C)
-58 to 32 °F T 32 to -4 °F T 32 to -40 °F T 32 to -58 °F T
(-50 to 0 °C Td) 10 s [35 s]
d
(0 to -20 °C Td) 10 s [35 s]
d
(0 to -40 °C Td) 15 s [120 s]
d
(0 to -50 °C Td) 35 s [300 s]
d
Temperature
Measurement range 14 °F 140 °F (-10....+60 °C)
Accuracy at +68 °F (+20°C) ±0.36 °F (±0.2 °C) Typical temperature dependence of electronics ±0.005 °C/°C Temperature sensor Pt100 IEC751 1/3 class B
MI70 Measurement Indicator Indicator general
Menu languages English, German, French,
Finnish and Spanish
Display LCD with backlight
Graphic trend display of any parameter
Character height up to 16 mm Probe inputs 1 or 2 Power Supply Rechargeable NIMH battery pack
with AC-POWER or
4xA4 size alkalines, type IEC LR6 Analog output 0...1 VDC Output resolution 0.6 mV Data interface RS232C Data-logging capacity 900...2700 points with real time Alarm Audible alarm function
Operating temperature range 14 °F 104 °F (-10....+40 °C)
Operating humidity range non-condensing Housing classification IP54 Weight 400 g
Battery operation time
Continuous use 48 h typical at +68 °F (+20°C) Data logging use up to 30 days
DMP74 Probe + MI70 Indicator = DM70 Hand-held Dewpoint Meter General
Housing materials ABS/PC blend Storage temperature -40 158 °F (-40...+70 °C) Storage humidity 0..100 %RH non condensing Total weight 750 g
ppm volume concentration
Measurement range (typical) 40...200 000 ppm Accuracy at +68 °F (+20°C) 7.3 ppm + 8.3% of reading
Relative humidity
Measurement range 0...100 %RH Accuracy at +68 °F (+20°C)
RH < 10% ±0.025 %RH + 8.75% of reading RH > 10% ±0.7 %RH + 2% of reading
Measurement environment
Temperature 14 °F 140 °F (-10....+60 °C)
Pressure 0...20 bar Sample flow rate no effect
Probe general
Sensor DRYCAP180S Probe material (wetted parts) Stainless steel (AISI 316L) Sensor protection Sintered filter (AISI 316L)
part no: HM47280
Mechanical connection G1/2" ISO228-1 thread
with bonded seal ring (U-seal) Housing classification IP65 (NEMA4) Weight 350 g
Electromagnetic compatibility
Complies with the following standard: EN 61326-1:1997 + Am1:1998, Electrical equipment for measurement, control and laboratory use - EMC requirements
- Generic Environment.
Accessories and DM70KIT
DSC74 sample cell adapter
sampling system for pressurized processes Pressure range 0...10 bar Weight 300 g
Connection cables
for DMT242 Transmitter order no: 27160ZZ
for DMP248 Transmitter order no: 27159ZZ Analog output cable order no: 27168ZZ Windows
Carrying Case MI70Case DM70 Calibration Kit order no: DM70KIT
Specifications subject to change without prior notice. DRYCAP® is a registered trademark of Vaisala. ©Vaisala Oyj
®
software including PC connection cable
order no: MI70LINK
includes DM70 indicator with rechargeable battery and 115 VAC
adapter; MIK70 link & analog output signal cable; DMP74 probe;
DSC74 Sample cell and quick adapter; carrying case; certificate of
calibration, traceable to NIST.