Atec Dito User Manual

DATA SHEET > dito > 20100706
DITO
ESD SIMULATOR
FOR TESTS ACCORDING TO ...
> Bellcore GR-1089-Core > Chrysler PF-9326 > DaimlerChrysler PF-10540 > EN 300329 > EN 300340 > EN 300342-1 > EN 300386 V1.3.2 > EN 301489-1 > EN 301489-17 > EN 301489-24 > EN 301489-7 > EN 55024 > EN 61000-4-2 > Fiat 9.90110 > Ford EMC-CS-2009.1 > Ford ES-XW7T-1A278-AB > Ford ES-XW7T-1A278-AC > Ford WDR 00.00EA > IEC 61000-4-2 > IEC 61326 > ISO 10605 > ...
DITO - THE ULTIMATE ESD TESTER
Electrostatic discharges either from a human body to any other part or between two different objects can cause persistent disturbances or even destruction to sensitive electronics or controls. Voltages of several 1,000 volts are generated. dito is the most advanced ESD tester to simulate ESD pulses as per EN/IEC 61000-4-2 and other standards requirements.
HIGHLIGHTS
> Most ergonomic design
APPLICATION AREAS
AUTOMOTIVE
RESIDENTIAL
> Light weight (940g) > Interchangeable discharge tips
INDUSTRY
MILITARY
> Interchangeable R/C networks
MEDICAL
AVIONICS
> Power booster to perform long duration tests > Library of standard test routines
BROADCAST
COMPONENTS
> Test voltage measurement and control > Various test routines for testing and evaluation
TELECOM
RENEWABLE ENERGY
www.emtest.com © EM TEST > PAGE 1/4
DATA SHEET > dito > 20100706
TECHNICAL DETAILS
TEST LEVELS
CD mode 500V - 10,000V +/-5% AD mode 500V - 16,500V +/-5% Resolution 100V steps Holding time More than 5 seconds Polarity Positive, negative Test voltage measured, indicated in the display
TRIGGER
Single mode Release of single pulses Continuous mode Continuous discharge up to
20Hz (AD) and 10Hz (CD)
Pulse release via software
Remote control from computer via fiber optic link (USB)
OUTPUT
CD mode EN/IEC 61000-4-2; 150pF/330ohm Rise time 0.8ns +/- 25% Current 1st peak 7.5A +/- 15% @2kV
15.0A +/- 15% @4kV
22.5A +/- 15% @6kV
30.0A +/- 15% @8kV
Current at 30ns 4.0A +/- 30% @2kV
8.0A +/- 30% @4kV
12.0A +/- 30% @6kV
16.0A +/- 30% @8kV
Current at 60ns 2.0A +/- 30% @2kV
4.0A +/- 30% @4kV
6.0A +/- 30% @6kV
8.0A +/- 30% @8kV
OPERATION
Manual Via function keys Remote control Via built-in optical interface
VOLTAGE SUPPLY
Battery (ACCU) More than 50,000 pulses @ 16.5kV;
automatic power-off after 10 minutes if not operated
Mains supply Via mains supply adapter (MAD)
GENERAL DATA
Weight 940 gramms incl. battery Case material ABS; non-inflammable as
per UL 94VO Temperature 5°C - 40°C (operation conditions) Rel. humiditiy 20% - 80% (operation conditions) Battery charger 100V - 240V +/-10%; 50/60Hz
Charging time less than 1.5 hours
ACCESSORIES
DM1 150pF/330ohm discharge module
with built-in contact discharge relay;
Discharge tips for AD and CD ACCU Rechargeable battery ACH Battery charger
Transportation case, grounding lead
2m, test certificate and operation
manual
TEST ROUTINES
Easy Zap Test based on standard levels Quick Zap All test parameters are adjustable
during test; fast and easy-to-use
Evaluation Air discharge with 20Hz; for test
point evaluation on the DUT
Standard Test routines
As per EN/IEC 61000-4-2, EN 61000-6-1/-6-2
Set-up Parameters, standard levels, etc.
DISPLAY AND CONTROLS
Display LCD graphic display; 128 x 64 pixels Operation 5 function keys, 2 cursor keys, pulse
trigger
www.emtest.com © EM TEST > PAGE 2/4
Loading...
+ 2 hidden pages