DATA SHEET > CWS 500N4 > 20121113
CWS 500N4
SIMULATOR FOR CONDUCTED, COMMON-MODE
DISTURBANCES, AS PER IEC 61000-4-16, IEC 61000-4-19
ANNEX C, 0HZ (DC) TO 165KHZ
FOR TESTS ACCORDING TO ...
> EN 61000-4-16
> IEC 60255-22-7
> IEC 60533
> IEC 61000-4-16
> IEC 61000-4-19, Annex C
> IEC 61326
> IEC 61543
> IEC 61850-3
> prTR 50579
CWS 500N4 - TESTING FOR IMMUNITY TO CONDUCTED, COMMON-MODE DISTURBANCES
The CWS 500N4 is the state-of-the-art solution in a compact single-box design to test for immunity to conducted, common
mode disturbances in the frequency range 0Hz (DC) to 165kHz. Such test requirements are specified in IEC 61000-4-16 and
cover continuous mode testing as well as short term testing with DC, 16 2/3Hz, 50Hz and 60Hz with 4 test levels each plus a
sweep mode from 10Hz to 165kHz.
Additionally, the CWS 500N4 can be used for testing electricity metering equipment as per prTR 50579 and Draft standard
IEC 61000-4-19, Annex C.
HIGHLIGHTS
> Most compact equipment for common-mode
disturbances as per IEC 61000-4-16
> Built-in LF signal generator and LF amplifier
> Sweep mode capability 10Hz to 165kHz
> Built-in rectifier module for DC testing
> Option for Electricity Meter testing acc IEC
61000-4-19 Annex C
APPLICATION AREAS
INDUSTRY
RESIDENTIAL
www.emtest.com © EM TEST > PAGE 1/5
DATA SHEET > CWS 500N4 > 20121113
TECHNICAL DETAILS
TESTING AS PER IEC/EN 61000-4-16
OUTPUT CHARACTERISTICS CONTINUOUS MODE
Test frequencies DC, 16 2/3Hz, 50Hz and 60Hz
Signal level 0.1 - 35Vrms or DC
Test level 1 1V continuous
Test level 2 3V continuous
Test level 3 10V continuous
Test level 4 30V continuous
Output impedance 50ohm ± 10%
Total harmonics
distortion
Ripple on DC < 5%
OUTPUT CHARACTERISTICS SHORT-TERM MODE
Test frequencies DC, 16 2/3Hz, 50Hz and 60Hz
Signal level 1 - 330Vrms or DC
Test level 1 10V for 1s duration
Test level 2 30V for 1s duration
Test level 3 100V for 1s duration
< 10% (sinusoidal waveform)
TESTING AS PER IEC/EN 61000-4-16
OUTPUT CHARACTERISTICS 15HZ - 165KHZ
Frequency range 10Hz to 165kHz
Signal level 0.1 to 35Vrms
Test level 1 1V - 0.1V - 1V
Test level 2 3V - 0.3V - 3V
Test level 3 10V - 1V - 10V
Test level 4 30V - 3V - 30V
Output impedance 50ohm ± 10%
Total harmonics
distortion
MEASUREMENT
Voltage internal rms measurement
TEST ROUTINES
Quick Start Immediate start; easy to use and fast
Service Service, Set-up
< 1% (sinusoidal waveform)
discrete frequency testing
Test level 4 300V for 1s duration
Output impedance 50ohm ± 10%
Total harmonics
distortion
Phase
synchronisation
Ripple on DC < 5%
Fall/rise time between 1us and 5us
Note: These tests require either the
< 10% (sinusoidal waveform)
0° ± 5%
optional MV 2606N2.2 motor variac
or the optional programmable AC
voltage source ACS 500N2.3.
TESTING OF ELECTRICITY METERING
EQUIPMENT AS PER TR-EN 50579
INTERFERENCES DISTURBANCES TESTING AS PER TR-EN
50579 AND IEC 61000-4-19 ANNEX C (OPTION)
Test frequencies 2kHz to 150kHz
Signal level Max. 3A up to 150kHz
Total harmonic
distortion (THD)
Accuracy better than ± 5%
Frequency step Max. 1%
Output impedance 1.0ohm (external decoupling
<5% at maximum level
impedance SH 1R)
www.emtest.com © EM TEST > PAGE 2/5