DATA SHEET > CWS 500N2 > 20120604
CWS 500N2
CONTINUOUS WAVE SIMULATOR
FOR TESTS ACCORDING TO ...
> BMW GS 95002 (2001) > DaimlerChrysler DC-10614 > DaimlerChrysler DC-11224 > DO 160 Section 20 > EN 61000-4-6 > EN 61000-6-1 > EN 61000-6-2 > Fiat 9.90110 > Ford ES-XW7T-1A278-AB > Ford ES-XW7T-1A278-AC > GMW 3097 (2001) > GMW 3097 (2004) > IEC 60601-1-2 > IEC 61000-4-6 > IEC 61326 > IEC 61850-3 > ISO 11452-4 > ISO 11452-5 > MBN 10284-2 > MIL STD 461 D CS 114 > MIL STD 461 E CS 114 > ...
BULK CURRENT INJECTION (BCI) TESTING
Bulk Current Injection (BCI) is a test procedure to proof the immunity to electrical disturbances by narrowband electromagnetic energy. The test signal is injected by means of a current injection probe physically being a current transformer laid around the wiring harness. Immunity tests are performed varying the level and the frequency of the injected test signal. The BCI test methode is widely known in the automotive industry as well as in the military/aircraft industry to test single components of a complex system. The CWS 500N2 is designed to be used also for tests with CDNs and EM clamps for tests according to EN/IEC 61000-4-6 and related standards.
HIGHLIGHTS
> Most compact equipment
APPLICATION AREAS
AUTOMOTIVE
INDUSTRY
> Supports BCI testing as per various standard
requirements
TELECOM
MEDICAL
> Basic frequency range 9kHz up to 400MHz > Extendable frequency range up to 1GHz > Built-in 100W class A amplifier up to 400MHz
AVIONICS
MILITARY
BROADCAST
RESIDENTIAL
www.emtest.com © EM TEST > PAGE 1/4
DATA SHEET > CWS 500N2 > 20120604
TECHNICAL DETAILS
BULK CURRENT INJECTION AS PER ISO 11452-4
Output level As required in ISO 11452-4, using
closed loop or substitution method
BULK CURRENT INJECTION AS PER MIL 461 CS 114
Output level As required in MIL 461 CS 114, using
the closed loop method
MEASUREMENTS FOR BULK CURRENT INJECTION
Directional coupler
Included to measure forward power
and reverse power Forward power Internal power meter #1 Reverse power Internal power meter #2 Injected current Internal power meter #3
TEST ROUTINES FOR BULK CURRENT INJECTION
ISO 11452-4 Operation via icd.control MIL 461 CS114 Operation via icd.control
IEC 61000-4-6
Output level 1V - 30Vrms (emf)
all standard test levels are
guaranteed with all coupling
methods
MEASUREMENTS, IEC 61000-4-6
Cal in (BNC) Integrated power meter to record the
calibration data of a coupling device
OUTPUT
RF output N connector at the front panel Ouptut power Built-in amplifier
50dBm (9kHz - 400MHz) at 1dB
compression. Gain amplifier > 50dB Output impedance 50ohm Harmonic
< -20dBc at max. power distortion
Insertion loss Approx. 1dB (directional coupler +
RF relay)
TIME PARAMETERS
Dwell time td = 0.3s - 9,999s Pause time tr = 0/0.3s - 9,999s
TEST FREQUENCIES
Frequency range 9kHz - 400MHz (built-in amplifier)
9kHz - 1,000MHz (ext. amplifier) Unmodulated
CW (continuous wave) signal
Amplitude modulation
1kHz, 80% AM (EN/IEC 61000-4-6)
1kHz, 95% AM (automotive)
400Hz, 80% AM
50Hz, 80% AM (automotive)
2Hz, 80% AM (IEC 60601-1-2) Pulse modulation 1Hz, 50% duty cycle (EN 50130-4)
1kHz, 50% duty cycle (MIL 461)
Injected current Measured by internal power meter RF indicator LED indicating the RF output status LCD Display of the test level and the
preselected frequency value
INTERFACE
Serial interface USB Parallel interface IEEE 488, addresses 1 - 30 Fail 1 BNC input; test will be stopped
(active low)
TEST ROUTINES FOR IEC 61000-4-6
Quick Start Immediate start; easy-to-use and
fast test routine
Fail 2 BNC input; test status will be saved
(max. 10 events) when active low.
Test continues
User Test routines Frequency sweep Service Service, Set-up
www.emtest.com © EM TEST > PAGE 2/4