Atec AT2000R User Manual

AT-2000R/RQ Specifications
Note: All specifications apply over the standard 0oC to + 50oC operating temperature range, after a minimum of 2 hours of storage within the operating temperature range, if the unit is not fully at ambient temperature. The AT-2000R/RQ meets all it’s specifications within 1 minute after it is turned on, providing that the AT-2000R/RQ is within the one year calibration cycle. Avantron's unique AutoCal feature assures accuracy by periodically self-testing itself and triggering a non­obstructive calibration, as required.
FREQUENCY
Frequency range----------------------------------------------------------- 1 MHz - 1 GHz Frequency Reference:
Aging-------------------------------------------------------------- ± 1 PPM/Yr Temperature Stability------------------------------------------- 1 PPM (0ºC to 50ºC)
Frequency counter:
Accuracy---------------------------------------------------------- ±1 PPM ± 1count Resolution -------------------------------------------------------- 10 Hz
Stability (Noise sidebands offset from CW signal)----------------- 85 dBc @ ± 10kHz
SPAN
Frequency Span:
Range-------------------------------------------------------------- Variable from Max Span 1000 MHz to
100 kHz & Zero Span.
Accuracy---------------------------------------------------------- 2 PPM
Sweep Time:
Range-------------------------------------------------------------- 3mS, 6mS, 15ms, 30mS, 100mS,
300mS, 1000mS, 3000mS
Stability ----------------------------------------------------------- 2 PPM
Sweep Trigger ------------------------------------------------------------- Automatic (Sweep Mode only), Free
Run, Linelock and TV Frame.
Resolution Bandwidth:
Range-------------------------------------------------------------- 1 MHz, 300 kHz, 30 kHz & 10 kHz Accuracy---------------------------------------------------------- ± 5% Selectivity (60 dB/3dB Ratio)--------------------------------- 5.3:1, 3:1, 2:1, 2:1
Video Bandwidth:
Range-------------------------------------------------------------- 1 MHz, 100 kHz & 10 kHz
Avantron AT-2000R/RQ Spectrum Analyzer - User Manual 4.0 (preliminary) 8
AMPLITUDE
Response Flatness--------------------------------------------------------- ± 0.75 dB (5 - 1000 MHz) Sensitivity ------------------------------------------------------------------ -65 dBmV to +65 dBmV
Level accuracy------------------------------------------------------------- ± 0.75 dB @ 25oC Level resolution ----------------------------------------------------------- 0.1 dB Impedance at RF input --------------------------------------------------- 75 ohm Input Return Loss --------------------------------------------------------- > 16 dB (>10 dB attenuation) Maximum safe input------------------------------------------------------ + 68 dBmV
Noise figure ---------------------------------------------------------------- 10 dB max. 0 dB attenuation Spurious free dynamic range-------------------------------------------- > 70 dB
Vertical scale--------------------------------------------------------------- 10, 5, 2 dB/Division
Input attenuator------------------------------------------------------------ 0 - 65 dB in 5 dB steps Internal calibrator --------------------------------------------------------- 38.5 MHz @ -5 dBmV Temperature Readout Range-------------------------------------------- 40ºC to 100ºC ±1.5ºC
Display type---------------------------------------------------------------- TFT Active Matrix Color LCD Display size ---------------------------------------------------------------- 162.5 mm (6.4 inches)
POWER
Battery type ---------------------------------------------------------------- Rechargeable lead acid, 12 Volt 5 Ah. Charger Type -------------------------------------------------------------- Battery Charger 12 Volt 2 A Charge Time --------------------------------------------------------------- Approx. 3 hours Operating Time------------------------------------------------------------ Approx. 2.5 hours
Temperature range:
Operating --------------------------------------------------------- 0°C to +50°C Non-operating---------------------------------------------------- -20°C to +55°C
Size -------------------------------------------------------------------------- 304.8mm x 177.8mm x 355.6mm
(12"W x 7"H x 14"D)
Weight (w\battery)-------------------------------------------------------- 8.6 Kg (19.6 pounds)
Avantron AT-2000R/RQ Spectrum Analyzer - User Manual 4.0 (preliminary) 9
CATV Measurement Specifications
Channel Selection --------------------------------------------------------- Frequency, Channel Video or Channel Audio Channel Plans-------------------------------------------------------------- 4 custom plans, NTSC, PAL or other.
Maximum of 250 positions.
Tuning Range-------------------------------------------------------------- 1 MHz to 1 GHz TV Channel amplitude range-------------------------------------------- -40 dBmV to +65 dBmV ± 0.75dB for
S/N > 30dB
TV Visual Frequency:
Accuracy-------------------------------------------------------------------- Carrier Frequency, ± 1 PPM Resolution ------------------------------------------------------------------ 10 Hz
Visual/Aural Delta Frequency:
Range------------------------------------------------------------------------ 1 - 10 MHz Accuracy-------------------------------------------------------------------- ± 200 Hz Resolution ------------------------------------------------------------------ 10 Hz Visual/Aural Delta Amplitude ------------------------------------------ ± 0.75 dB for S/N > 30 dB
Digital Carrier Measurement
Amplitude Range --------------------------------------------------------- -30 to +60 dBmV Resolution ----------------------------------------------------------------- 0.1 dB Absolute Accuracy ------------------------------------------------------- ± 1.5 dB Bandwidth Range -------------------------------------------------------- 200 kHz to 200 MHz
FM Deviation:
Range------------------------------------------------------------------------ ± 150 kHz Accuracy-------------------------------------------------------------------- ± 2 kHz, 1 - 75 kHz, ± 5 kHz to 150
kHz
HUM/Low Freq. Disturbances:
Modes ----------------------------------------------------------------------- CW or Video (In-Service) Range------------------------------------------------------------------------ 1 - 10 % Accuracy---------------------------------------------------------------------± 0.5 % from 1 to 5 %, ± 1 % from 5 to
10 %
Modulation Depth
AM Range ------------------------------------------------------------------ 40 to 95% Resolution ------------------------------------------------------------------ 0.1% Accuracy -------------------------------------------------------------------- ± 1.5% (C/N > 40 dB) Signal type ------------------------------------------------------------------ Use VITS line with white reference
In-Channel Response
Range ------------------------------------------------------------------------ ± 10 dB Resolution ------------------------------------------------------------------0.1 dB Accuracy --------------------------------------------------------------------± 0.25 dB Signal type ------------------------------------------------------------------Use VITS line with full amplitude
multiburst signal
Avantron AT-2000R/RQ Spectrum Analyzer - User Manual 4.0 (preliminary) 10
Carrier-to-Noise Ratio
Optimum Range -----------------------------------------------------------0 dBmV to +10 dBmV with 0 dB
Attenuation
Maximum C/N -------------------------------------------------------------60 dB with ± 1 dB accuracy, 65 dB
with ± 3 dB accuracy
Resolution ------------------------------------------------------------------0.1 dB
CSO/CTB
Optimum Range ---------------------------------------------------------- -4 dBmV to +4 dBmV with 0 dB
Attenuation
Maximum ------------------------------------------------------------------CSO/CTB 64 dB with ± 1.5 dB
accuracy 71 dB with ± 4 dB accuracy Resolution 0.1 dB
Note: C/N, CSO and CTB specifications Without Preselector & 77 Channel Loading
Avantron AT-2000R/RQ Spectrum Analyzer - User Manual 4.0 (preliminary) 11
Loading...