Atec AR-MT06000 User Manual

Model MT06000
Multistar
Multi-tone RF Radiated Immunity System
80MHz–6GHz
EN/IEC 61000-4-3 EN/IEC 60601-1-2 EN 50130-4 EN 61000-6-1/2 EN 55024
The Model MT06000 (Multistar Multi-tone tester) is a state-of-the-art system
designed to test RF Radiated Immunity faster than ever before possible. By testing multiple frequencies (tones) at once, test times can be reduced by a factor equivalent to the number of tones selected. The number of tones is only limited by the signal generator bandwidth (150MHz) and the size of the amplifier used with the system. The MT06000 contains all the instruments needed to perform radiated immunity testing for IEC 61000-4-3 specification except the required amplifiers, antennas and directional couplers. Amplifiers can be sized and selected based on your required field levels and testing needs. Up to 4 RF amplifiers and directional couplers can be controlled and monitored and power can be delivered to up to 4 antennas to generate the desired fields. The system contains a vector signal generator, a vector signal analyzer, a RF pre-amplifier, a RF field probe and monitor, an RF switch matrix, and automated radiated immunity test software. Everything is contained in a single housing, which eliminates setup issues. The software includes automated routines to calibrate the field and maximize the speed of test (most tones possible) while still meeting the Linearity and Harmonics requirements of the specification. In the event of a EUT failure, margin investigation (thresholding) and traditional single tone testing can be performed causing a slowing of the test only in the areas of concern. This system has the versatility needed for every test laboratory and equipment manufacturer while adding the benefit of reduced test times and greater throughput.
Internal Test Specifications IEC/EN 60601-1-2 IEC/EN 50130-4 IEC/EN 61326 IEC/EN 61000-6-1 IEC/EN 61000-6-2 CISPR 24/EN 55024
Vector Signal Generator Specifications Frequency range Resolution Power Out +10dBm Modulation AM, FM, Pulse, Phase Bandwidth 150MHz Hardware Platform PXIe
IEC 61000-4-3 procedure and levels
80<Hz to 6.6 GHz <4Hz
Vector Signal Analyzer Specifications Frequency Range Resolution Dynamic Range 80dB Bandwidth 50Mhz Hardware Platform PXIe
RF Preamplifier Specifications Frequency Range 50MHz-8GHz Gain 26dB Hardware Platform PXIe
Field Monitor/Probe Spec Channels 4 Probe 1 Type Isotropic, Laser powered Frequency 100kHz-6GHz Range 0.5–800 V/m
10MHz-6.6GHz 533nHz
081011
160 School House Road Souderton, PA 18964-9990 • 215-723-8181 • www.ar-worldwide.com Page 1 of 2
Connections
RF signal input 2 SMA Male (rear)
For optional signal generators
RF Signal Out 4 SMA Male (rear) to RF
amplifiers
High power RF in 4- Type N Male (rear) from RF
amplifiers
High power RF out 4 Type N male (rear) to
antennas/loads
Monitor port In 4- SMA Male (rear)
4 for forward power
Control PC Computer Intel Core 2 Duo, 2.53GHz Operating system Windows 7 RAM 4GB DDR2 Hardware Platform PXIe
General
Power 115/230 VAC
50/60 Hz, single phase 16A Breaker 2 pole, 20A Cooling active cooling, air ventilation Environmental conditions 10°C - 40°C Dimensions, 50.3 x 47.2 x 61 cm
19.8 x 18.6 x 24 in
Weight 22.7 kg (50.0 lb)
Accessories DAQ Module (AR P/N 10025318) (NI USB-6212)
Low Pass Filter, 80MHz­1GHz Low Pass Filter, 80MHz-
4.2GHz Low Pass Filter, 80MHz­6GHz
Data acquisition module is used to collect data, identify user defined failures, report user defined error messages, and reset EUT after failure Absorptive filter used to remove harmonics Absorptive filter used to remove harmonics Absorptive filter used to remove harmonics
Page 2 of 2