Model CI00400A,
M1 through M7
RF Conducted Immunity System
100 Watts nominal
Complete testing solutions to the following standards:
MIL-STD-461D & E CS114
DO160D & E BCI Testing
EN/IEC 61000-4-6
IEC 60601-1-2
EN 50130-4
EN 61000-6-1/2
EN 55024
The Model CI00400A is a fully self-contained state of the art system
designed to test RF Conducted Immunity. The system contains a signal
generator, 3 channel power meter, 100W nominal AR amplifier 10kHz to
400MHz, directional coupler, and emcware® control software. Everything is contained in a single housing, which
eliminates setup issues. This system provides the versatility needed for every test laboratory and equipment manufacturer.
The RF amplifier and the signal generator can be used independently of the system. If special needs arise or standards
were to change a larger amplifier can be connected to the system. The use of spectrum analyzers and monitoring
equipment may also be controlled by the software.
Option 1 adds an additional power head for testing automotive standards such as Ford, GM, Chrysler, BMW, Renault,
and ISO11452-4. Option 2 adds data acquisition capability. Option 3 provides a laptop PC preloaded with the test
software. Option 4 is a lower cost solution providing all the functions and features of the full CI00400A base model
without the cost of the internal amplifier.
The export classification for this equipment is EAR99. These commodities, technology or software are controlled for export
in accordance with the U.S. Export Administration Regulations. Diversion contrary to U.S. law is prohibited.
Internal Test Specifications (See NOTE 1)
CI00400A CI00400A w/Option 1
MIL-STD-461D (CS114)
MIL-STD-461E (CS114) ISO 11452-4
DO160D (Sec 20 BCI Test) GMW 3097
DO160E (Sec 20 BCI Test) Ford ES-XW7T-1A278-AC
IEC/EN 60601-1-2 DaimlerChrysler DC-11224
IEC/EN 50130-4 BMW GS95002
IEC/EN 61326 Renault 36-00-8081-G
IEC/EN 61000-6-1
IEC/EN 61000-6-2
IEC/EN 61000-4-6
CISPR 24/EN 55024
NOTE 1: Specifications can be met using AR-specified external accessories (injection probes, monitor probes, cal fixtures, CDN’s,
attenuators, etc.) Contact AR for further information.
NOTE 2: The use of a spectrum analyzer may be necessary on some of the low level bulk current injection tests. This is especially true on
power and I/O lines with a great amount of ambient noise.
All standards of CI00400A plus:
Signal Generator Specifications
Frequency range
resolution
Power range
resolution
Modulation AM, FSK. FM, Phase, External
Power Meter Specifications (See NOTE 2)
Channels 3
Power heads 2
Type diode
Frequency 10kHz to 8GHz
Range -60 to +20 dBm
9 kHz to 1.2 GHz
1Hz
-140 to +13 dBm
0.1dB
Pulse
10 kHz–400 MHz
Approved for public release by AR RF/Microwave Instrumentation 092412
160 School House Road Souderton, PA 18964-9990 • 215-723-8181 • www.arworld.us Page 1 of 2
RF Amplifier Specifications
Frequency range 10 kHz to 400 MHz
Power rating 100 Watts Nominal
1dB compression 75 Watts Nominal
Harmonic Distortion -20dBc at 50 Watts
Mismatch tolerance 100% of rated power without
fold back. Will operate without
damage or oscillation with any
magnitude of source and load
impedance.
Gain 51dB minimum
Connections
RF Out Type N (front)
Monitor Port In Type N (front)
Signal Generator Out Type N (rear)
Directional Coupler In Type N (rear)
Amplifier Out/In Type N (rear)
Pulse In BNC (rear)
Communication GPIB (IEEE 488) (rear)
Directional Coupler Fwd Out Type SMA (rear)
Directional Coupler Fwd In Type SMA (rear)
Directional Coupler Rev Out
(with Option 1)
Directional Coupler Rev In
(with Option 1)
Monitor Port Out Type SMA (rear)
Monitor Port In Type SMA (rear)
Power Meter Calibration Port Out Type SMA (rear)
Application Model Description
IEC 61000-4-6
BCI Clamp
50 System
IEC 61000-4-6
BCI Clamp
150 System
IEC 61000-4-6
EM Clamp
50 System
MIL-STD-461
DO160
Automotive TK3000 Conducted immunity test kit containing all the attenuators, injection
Type SMA (rear)
Type SMA (rear)
TK1000 Conducted immunity test kit containing all the attenuators, injection
probes, calibration fixtures, calibration resistors, and termination
resistors necessary for IEC testing.
TK1001 Conducted immunity test kit containing all the attenuators, injection
probes, calibration fixtures, calibration resistors, and termination
resistors necessary for IEC testing.
TK1002 Conducted immunity test kit containing all the attenuators, injection
probes, calibration fixtures, calibration resistors, and termination
resistors necessary for IEC testing.
TK2000 Conducted immunity test kit containing all the attenuators, injection
probes, calibration fixtures, calibration resistors, and termination
resistors necessary for MIL-STD-461 and DO160 testing.
probes, calibration fixtures, calibration resistors, and termination
resistors necessary for automotive testing.
General
Power 115/230 VAC
50/60 Hz, single phase 16A
Breaker 2 pole, 20A
Cooling active cooling, air ventilation
Environmental conditions 10°C - 40°C
Dimensions, 50.3 x 42.2 x 52.1 cm
19.8 x 16.6 x 21.7 in
Weight 22.7 kg (50.0 lb)
Weight CI00400AM5 17.3 kg (38.2 lb)
Control Software
emcware® Software Suite
PC Requirements
Computer 2GHz Single Core Processor
Operating system Windows XP, Windows Vista,
Windows 7
RAM 1 GB Minimum
Screen Resolution 1024 x 768
Ports 2 available USB 2.0 ports
Options
1 Additional power head to add the ability to monitor
reverse power
2 Data acquisition card.
3 Laptop PC with software preinstalled
4 Amplifier removed; requires use of external amplifier
MODEL CONFIGURATIONS
MODEL DESCRIPTION
CI00400AM1 Includes Option 3
CI00400AM2 Includes Options 2 and 3
CI00400AM3 Includes Option 2
CI00400AM4 Includes Option 1
CI00400AM5 Includes Option 4
CI00400AM6 Includes Options 1 and 4
CI00400AM7 Includes Options 1 and 3
ACCCESSORY KITS