Atec AR-CI00250A User Manual

Model CI00250A,
Internal Test Specifications*
IEC/EN 60601-1-2
CISPR 24/EN 55024
IEC 61000-4-6 procedure and
Signal Generator Specifications
SG1200
SG6000 (optional)
Frequency range
9 kHz to 1.2 GHz
100kHz to 6 GHz
Power range resolution
-140 to +13 dBm
0.1dB
-110 to +7 dBm
0.02 dB
Modulation
AM, FSK. FM, Phase,
AM, FM, Phase,
Pulse
Power Meter Specifications
Channels
3
Power heads
1
Type
diode
Frequency
10kHz to 8GHz
Range
-60 to +20 dBm RF Amplifier Specifications
Frequency range
10 kHz to 250 MHz
Power rating
75Watts Minimum
1dB compression
50 Watts Minimum
Harmonic Distortion
-20dBc at 50 Watts
Mismatch tolerance
100% of rated power without
impedance.
Gain
49dB minimum
M1 through M12
RF Conducted Immunity System
75 Watts
Complete Testing Solutions to the following standards:
EN/IEC 61000-4-6
IEC 60601-1-2
EN 50130-4
EN 61000-6-1/2
EN 55024
CISPR 14-2
The Model CI00250A is a fully self-contained state of the art system designed to test RF Conducted Immunity. The CI00250A contains all the
instruments needed to perform conducted immunity testing to the IEC 61000-4-6 specification. The system contains a signal generator, 3 channel power meter, 75W minimum AR amplifier 10kHz to 250MHz, and emcwarcontrol software. Everything is contained in a single housing, which eliminates setup issues. This system will have the versatility needed for every test laboratory and equipment manufacturer. The RF amplifier and the signal generator can be used independently of the system. If special needs arise or standards were to change a larger amplifier can be connected to the system. The use of spectrum analyzers and monitoring equipment may also be controlled by the software.
The export classification for this equipment is EAR99. These commodities, technology or software are controlled for export in accordance with the U.S. Export Administration Regulations. Diversion contrary to U.S. law is prohibited.
IEC/EN 50130-4
IEC/EN 61326 IEC/EN 61000-6-1 IEC/EN 61000-6-2
*Specifications can be met using AR-specified external accessories (injection probes, monitor probes, cal fixtures, CDN’s, attenuators, etc.) Contact AR for further information.
resolution
1Hz
External Pulse
levels
Approved for public release by AR RF/Microwave Instrumentation 111412
160 School House Road Souderton, PA 18964-9990 • 215-723-8181 • www.arworld.us Page 1 of 2
0.01 Hz
Internal/External
fold back. Will operate without
damage or oscillation with any magnitude of source and load
General
Power
115/230 VAC 50/60 Hz, single phase 16A
Breaker
2 pole, 20A
Cooling
active cooling, air ventilation
Environmental conditions
10°C - 40°C
Dimensions,
50.3 x 42.2 x 52.1 cm
Weight
20.5 kg (45.0 lb)
Control Interface
GPIB Adaptor
USB to GPIB adaptor included (NI-GPIB-USB-HS)
Connections
RF Out
Type N Male (front)
Monitor Port In
Type N Male (front)
Signal Generator Out
Type N Male (rear)
Amplifier In
Type N Male (rear)
Pulse In
BNC Male (rear)
Communication
GPIB (IEEE 488) (rear)
Directional Coupler Fwd Out (with Option 1)
Type SMA (rear)
Directional Coupler Fwd In (with Option 1)
Type SMA (rear)
Directional Coupler Rev Out (with Option 2)
Type SMA (rear)
Directional Coupler Rev In (with Option 2)
Type SMA (rear) Monitor Port Out
Type SMA (rear)
Monitor Port In
Type SMA (rear)
Power Meter Calibration Port Out
Type SMA (rear)
Control Software
emcware® Software Suite
PC Requirements
Computer
2 GHz Single Core Processor
Operating system
Windows, XP, Windows Vista,
Windows 7
RAM
1 GB Minimum
Screen Resolution
1024 x 768
Ports
2 available USB 2.0 ports Options
1
Directional coupler and additional power head to level
2
Additional power head to add the ability to monitor reverse power
3
Data acquisition card
4
Laptop PC with software preinstalled
5
100 kHz to 6 GHz Signal Generator SG6000 (Replaces standard Signal Generator SG1200)
6
ISO 17025-accredited calibration for system
7
Amplifier removed; requires use of external amplifier.
MODEL CONFIGURATIONS
MODEL
DESCRIPTION
CI00250AM1
Includes Option 4
CI00250AM2
Includes Option 1
CI00250AM3
Includes Options 1 and 4
CI00250AM4
Includes Options 1 and 3
CI00250AM5
Includes Option 5
CI00250AM6
Includes Options 3 and 5
CI00250AM7
Includes Option 6
CI00250AM8
Includes Options 3 and 6
CI00250AM9
Includes Options 3 and 4
CI00250AM10
Includes Options 1, 3 and 4
CI00250AM11
Includes Option 7
CI00250AM12
Includes Options 4 and 6
Accessory Kits
Application
Model
Description
IEC 61000-4-6
50Ω System
TK1000
Conducted immunity test kit containing all the attenuators, injection
resistors necessary for IEC testing.
IEC 61000-4-6
150Ω System
TK1001
Conducted immunity test kit containing all the attenuators, injection
resistors necessary for IEC testing.
IEC 61000-4-6
50Ω System
TK1002
Conducted immunity test kit containing all the attenuators, injection
resistors necessary for IEC testing.
19.8 x 16.6 x 21.7 in
on and monitor forward power
BCI Clamp
BCI Clamp
EM Clamp
probes, calibration fixtures, calibration resistors, and termination
probes, calibration fixtures, calibration resistors, and termination
probes, calibration fixtures, calibration resistors, and termination
Page 2 of 2