Model CI00250A,
Internal Test Specifications*
IEC 61000-4-6 procedure and
Signal Generator Specifications
Power Meter Specifications
-60 to +20 dBm RF Amplifier Specifications
100% of rated power without
M1 through M12
RF Conducted Immunity System
75 Watts
10 kHz–250 MHz
Complete Testing Solutions to the following standards:
• EN/IEC 61000-4-6
• IEC 60601-1-2
• EN 50130-4
• EN 61000-6-1/2
• EN 55024
• CISPR 14-2
The Model CI00250A is a fully self-contained state of the art system
designed to test RF Conducted Immunity. The CI00250A contains all the
instruments needed to perform conducted immunity testing to the IEC
61000-4-6 specification. The system contains a signal generator, 3 channel power meter, 75W minimum AR amplifier
10kHz to 250MHz, and emcware® control software. Everything is contained in a single housing, which eliminates setup
issues. This system will have the versatility needed for every test laboratory and equipment manufacturer. The RF amplifier
and the signal generator can be used independently of the system. If special needs arise or standards were to change a
larger amplifier can be connected to the system. The use of spectrum analyzers and monitoring equipment may also be
controlled by the software.
The export classification for this equipment is EAR99. These commodities, technology or software are controlled for export
in accordance with the U.S. Export Administration Regulations. Diversion contrary to U.S. law is prohibited.
IEC/EN 50130-4
IEC/EN 61326
IEC/EN 61000-6-1
IEC/EN 61000-6-2
*Specifications can be met using AR-specified external
accessories (injection probes, monitor probes, cal fixtures,
CDN’s, attenuators, etc.) Contact AR for further information.
resolution
1Hz
External Pulse
levels
Approved for public release by AR RF/Microwave Instrumentation 111412
160 School House Road Souderton, PA 18964-9990 • 215-723-8181 • www.arworld.us Page 1 of 2
0.01 Hz
Internal/External
fold back. Will operate without
damage or oscillation with any
magnitude of source and load
115/230 VAC
50/60 Hz, single phase 16A
active cooling, air ventilation
USB to GPIB adaptor included
(NI-GPIB-USB-HS)
Directional Coupler Fwd Out
(with Option 1)
Directional Coupler Fwd In
(with Option 1)
Directional Coupler Rev Out
(with Option 2)
Directional Coupler Rev In
(with Option 2)
Type SMA (rear)
Monitor Port Out
Power Meter Calibration Port Out
2 GHz Single Core Processor
Windows, XP, Windows Vista,
Windows 7
2 available USB 2.0 ports Options
Directional coupler and additional power head to level
Additional power head to add the ability to monitor
reverse power
Laptop PC with software preinstalled
100 kHz to 6 GHz Signal Generator SG6000
(Replaces standard Signal Generator SG1200)
ISO 17025-accredited calibration for system
Amplifier removed; requires use of external amplifier.
Includes Options 1, 3 and 4
Conducted immunity test kit containing all the attenuators, injection
resistors necessary for IEC testing.
Conducted immunity test kit containing all the attenuators, injection
resistors necessary for IEC testing.
Conducted immunity test kit containing all the attenuators, injection
resistors necessary for IEC testing.
19.8 x 16.6 x 21.7 in
on and monitor forward power
BCI Clamp
BCI Clamp
EM Clamp
probes, calibration fixtures, calibration resistors, and termination
probes, calibration fixtures, calibration resistors, and termination
probes, calibration fixtures, calibration resistors, and termination
Page 2 of 2