Atec AP033 User Manual

!
Specifications
Specifications
SpecificationsSpecifications
Specifications
Specifications
SpecificationsSpecifications
Nominal characteristics describe parameters and attributes that have are guaranteed by design, but do not have associated tolerances.
Input Configuration: True Differential (+ and –
Inputs); with shield Ground connector.
1
Effective Gain Input coupling: DC. AC Coupling obtained by
Differential Mode Range
(with 10X Gain): ±40 mV (÷1 Attenuation)
: X10, X1, ÷10, ÷100
installing AC Coupling Adapter.
±400 mV (÷10 Attenuation) ±4V 100 Attenuation
2
2
)
(with 1X Gain): ±400mV (÷1 Attenuation)
±4V 10 Attenuation) ±40 V (÷100 Attenuation
Common Mode Range: ±4.2 V 1 Attenuation)
±42 V (÷10 Attenuation) ±42 V (÷100 Attenuation
Maximum Input Voltage: ±42 V either input from ground.
1
Notes:
AP033-OM-E Rev F ISSUED: October 2001 9
From combinations of gain, internal and external
attenuation.
2
Use external plug-on ÷10 attenuator for ÷100.
2
2
9–1111
99
)
)
WARRANTED ELECTRICAL CHARACTERISTICS
Warranted characteristics are parameters with guaranteed performance. Unless otherwise noted, tests are provided in the Performance Verification Procedure for all warranted specifications.
LF Gain Accuracy: 2% into 50.0 load
Common Mode Rejection Ratio
3
Notes:
Output impedance is 50 Ω, intended to drive 50 .Add uncertainty of termination impedance to accuracy.
4
LeCroy measures CMRR with a fixture that connects the probe tip ground to the signal source ground. This method is necessary to obtain a reproducible CMRR measurement.
Often, users leave the probe tip ungrounded when measuring high frequency signals. Not grounding the probe tip can actually improve CMRR by allowing some of the common mode signal to be impressed across the entire length of the probe cable instead of from probe tip to probe ground. The CMRR improvement obtained without grounding the probe tip depends on proximity to probe cable ground, and is therefore nonreproducible.
LeCroy has chosen to use a reproducible method of measurement, rather than obtain a more optimistic measurement.
AP033 Active Probe
AP033 Active Probe
AP033 Active ProbeAP033 Active Probe
3
measured at 1 kHz with 0 volt offset
4
: (Probe head grounded, DC
Coupled, ÷1 Attenuation,
without external attenuator)
70 Hz 3160:1 (70 dB)
1MHz 1000:1 (60 dB)
250 MHz 5:1 (14 dB)
,
9999–2222 ISSUED: October 2001 AP033-OM-E Rev F