Specifications
Specifications
SpecificationsSpecifications
NOMINAL CHARACTERISTICS
Specifications
Specifications
SpecificationsSpecifications
Nominal characteristics describe parameters and attributes that
have are guaranteed by design, but do not have associated
tolerances.
Input Configuration: True Differential (+ and –
Inputs); with shield Ground
connector.
1
Effective Gain
Input coupling: DC. AC Coupling obtained by
Differential Mode Range
(with 10X Gain): ±40 mV (÷1 Attenuation)
: X10, X1, ÷10, ÷100
installing AC Coupling Adapter.
±400 mV (÷10 Attenuation)
±4V (÷100 Attenuation
2
2
)
(with 1X Gain): ±400mV (÷1 Attenuation)
±4V (÷10 Attenuation)
±40 V (÷100 Attenuation
Common Mode Range: ±4.2 V (÷1 Attenuation)
±42 V (÷10 Attenuation)
±42 V (÷100 Attenuation
Maximum Input Voltage: ±42 V either input from ground.
1
Notes:
AP033-OM-E Rev F ISSUED: October 2001 9
From combinations of gain, internal and external
attenuation.
2
Use external plug-on ÷10 attenuator for ÷100.
2
2
9–1111
99
)
)
WARRANTED ELECTRICAL CHARACTERISTICS
Warranted characteristics are parameters with guaranteed
performance. Unless otherwise noted, tests are provided in the
Performance Verification Procedure for all warranted
specifications.
LF Gain Accuracy: 2% into 50.0 Ω load
Common Mode Rejection Ratio
3
Notes:
Output impedance is 50 Ω, intended to drive 50 Ω.Add
uncertainty of termination impedance to accuracy.
4
LeCroy measures CMRR with a fixture that connects
the probe tip ground to the signal source ground. This
method is necessary to obtain a reproducible CMRR
measurement.
Often, users leave the probe tip ungrounded when
measuring high frequency signals. Not grounding the
probe tip can actually improve CMRR by allowing
some of the common mode signal to be impressed
across the entire length of the probe cable instead of
from probe tip to probe ground. The CMRR
improvement obtained without grounding the probe tip
depends on proximity to probe cable ground, and is
therefore nonreproducible.
LeCroy has chosen to use a reproducible method of
measurement, rather than obtain a more optimistic
measurement.
AP033 Active Probe
AP033 Active Probe
AP033 Active ProbeAP033 Active Probe
3
measured at 1 kHz with 0 volt
offset
4
: (Probe head grounded, DC
Coupled, ÷1 Attenuation,
without external attenuator)
70 Hz ≥ 3160:1 (70 dB)
1MHz ≥ 1000:1 (60 dB)
250 MHz ≥ 5:1 (14 dB)
,
9999–2222 ISSUED: October 2001 AP033-OM-E Rev F