Atec 8104, 8106 User Manual

Patented SmartGFI®safety circuit protects the operator from shock hazards
Patented VERI-CHEK®feature prompts users through steps to validate the instrument’s operation
Patented prompt and hold function provides a safe and easy method for performing steps during a test cycle
Patented CAL-ALERT®alerts the operator that the OMNIA is due for re-calibration.
True Line Leakage tester with 7 different measuring devices and RMS or PEAK leakage measurements
4 wire measurement and milliohm offset for accurate Ground Bond test results
Features and Benefits
AC and DC Hipot, Insulation Resistance, Ground Bond/ Continuity With Optional HV and HC Scanner
AC and DC Hipot, Insulation Resistance, Ground Bond/ Continuity and Functional Run Test
AC and DC Hipot, Insulation Resistance, Ground Bond/ Continuity, Functional Run Test and Line Leakage Test
Model 8105
Model 8104
Model 8106
13860 W. Laurel Dr., Lake Forest, IL USA 60045-4546 P: +1.847.367.4077 F: +1.847.367.4080 info@asresearch.com
U.S. Patents: 6,744,259, 6,549,385, 6,515,484, 6,054,865, 6,011,398, 5,936,419, 5,828,222, 5,548,501. Other patents pending.
Safety agency listed.
www.asresearch.com
1.800.858.TEST [8378]
Real Current measurement allows operators to monitor reactive and real current on a single screen
50 memories with 30 steps per memory that can be stored and recalled in any alphanumeric combination
Patented RAMP HI and CHARGE LO testing for more effective DC Hipot testing
Perform Hipot/Line Leakage without changing test leads
Patented Graphic LCD and intuitive menu system to simplify the entire testing process from set-up to results
RS-232 or GPIB automation interfaces available. Use your own software or our stand alone test software
Input Specifications
Voltage 115 / 230 V selectable, ± 10% variation
Frequency 50/60 Hz ± 5%
Fuse 115 VAC, 230 VAC - 6.3 A Slow-Blo 250 VAC
Dielectric Withstand Test Mode
Output Rating 5 KV @ 40 mA AC
5 KV @ 20 mA DC 50/60 Hz user selectable
HI and LO-Limit/ AC Total Range: 0.000 mA - 40.00 mA Current Display AC Real Range: 0.000 mA - 40.00 mA
DC Range: 0.0 - 20,000 µA
Ramp HI >20 mA peak maximum, ON/OFF selectable
Charge LO Range: 0.000 - 350 µA or Auto Set
DC Output Ripple 4% Ripple RMS at 5 KV DC @ 20 mA, Resistive Load
Discharge Time 200 ms
Maximum 1 µF < 1 KV 0.08 µF < 4 KV Capacitive Load 0.75 µF < 2 KV 0.04 µF < 5 KV DC Mode 0.5 µF < 3 KV
AC Output Waveform Sine Wave, Crest Factor = 1.3 - 1.5
Output Regulation ± (1% of output + 5 V) from no load to full load and over
input voltage range.
Dwell Timer Range: AC 0.4 - 999.9 sec (0 = Constant)
DC 0.3 - 999.9 sec (0 = Constant)
Ramp Timer Range: Ramp-Up: AC 0.1 - 999.9 sec, DC 0.4 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec, DC 0.0,1.0 - 999.9 sec
Ground Continuity Current: DC 0.1 A ± 0.01 A, fixed
Max. Ground Resistance: 1 ± 0.1 , fixed
Ground Fault Interrupt GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1ms
Continuity Test Mode
Output Current DC 0.1A ± 0.01A, fixed
Resistance Display/ Range: 0.00 - 10.00 HI and LO-Limit
Dwell Timer Range: 0.0, 0.3 - 999.9 sec (0 = Constant)
Milliohm Offset Range: 0.00 - 2.00
Ground Bond Test Mode
Output Voltage Range: 3.00 - 8.00 VAC (Open Circuit Limit) 50/60 Hz, user selectable
Output Current Range: 1.00 - 40.00 A, Resolution: 0.01 A
Output Regulation Accuracy: ± (1% of output + 0.02 A)
Within maximum load limits, and over input voltage range
Maximum Loading 1.00 - 10.00 A, 0 - 600 m
10.01 - 30.00A, 0 - 200 m
30.01 - 40.00A, 0 - 150 m
Ohmmeter Display/ 0 - 150 mfor 30.01 - 40.00 A HI and LO Limit 0 - 200 mfor 10.01 - 30.00 A
0 - 600 mfor 6.00 - 10.00 A Accuracy: ± (2% of reading + 2 mΩ) 0 - 600 mfor 1.00 - 5.99 A Accuracy: ± (3% of reading + 3 mΩ)
Milliohm Offset Range: 0 - 200 m
Insulation Resistance Test Mode
Voltage Setting Range: 50 - 1000 VDC
Charging Current Maximum >20 mA peak
Resistance Display/ Range: 0.05 M- 50000 M(4 Digit, Auto Ranging) HI and LO Limit Accuracy:
50 - 499 V
0.05 M- 999.9 M, ± (7% of reading +2 counts)
500 - 1000 V
0.05 M- 999.9 M, ± (2% of reading +2 counts) 1000 M- 9999 M, ± (5% of reading +2 counts) 10000 M- 50000 M, ± (15% of reading +2 counts)
Insulation Resistance Test Mode (Continued)
Charge LO Range: 0.000 - 3.500 µA or Auto Set
Ramp Timer Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: 0.0, 1.0 - 999.9 sec
Delay Timer Range: 1.0 - 999.9 sec (0 = Constant)
Ground Fault Interrupt GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1 ms
General Specifications
Mechanical Bench or rack mount with tilt up front feet
Dimensions 3U (WxHxD) 17 x 5.8 x 20.3 in.
(432 x 147 x 515 mm)
Weight 51.68 lbs (23.44 kgs) varies with option
Interface RS-232 standard or select GPIB or Printer
Port with time and date stamp.
Safety Built-in Smar tGFI
®
circuit
Memory 50 memories, 30 step/memory
OMNIA 8105 and 8106 Functional Run Test Mode
Delay Time Setting Range: 0.2 - 999.9 seconds
Dwell Time Setting Range: 0.1 - 999.9 seconds (0 = constant)
Trip Point Settings and Metering
Voltage Volt-Hi/Volt-LO Range: 0.0 - 277.0 VAC (one hot and one neutral)
Current Amp-HI/Amp-LO Range: 0.1 - 15.00 AAC
Watts Watt-HI/Watt-LO Range: 0 - 4200 W
Power Factor
PF-HI/PF-LO Range: 0.000 - 1.000
Leakage Current
Leak-HI/Leak-LO Range: 0.00 - 10.00 mA (0 = OFF)
Timer Display Range: 0.0 - 999.9 seconds
OMNIA 8106 Line Leakage Test Mode (DUT Power)
Voltage 0 - 277 VAC
Current 15 AAC max. continuous
Short Circuit Protection 23 AAC, Response Time < 3 sec
Leakage Current
Current Display Range 1: 0.0 µA - 999.9 µA RMS or PEAK Range 2: 1000 µA - 6000 µA Accuracy RMS DC to 100 kHz, ± (1.5% of reading+ 3 counts)
>100 k to 1 MHz ± 5% of reading, (10.0 µA - 6000 µA)
Accuracy PEAK DC to 1 MHz, ± (10% of reading + 2 µA)
Measuring Device A UL544 Non Patient
B UL544 Patient C IEC601-1, UL2601, EN60601-1 D UL1563 E IEC1010, UL3101, IEC950, UL1950,
IEC60990 FIG. 4 - U2 H IEC60990 FIG. 5 - U3 I IEC60990 FIG. 3 - U1
MD A-D and H-I components Accuracy: Resistance ± 1%, Capacitance ± 5% MD E components Accuracy: Resistance ± 0.1%, Capacitance ± 1% MD Voltage Limit Maximum 30 V peak or 30 VDC
HI and LO-Limit Range: 0 - 6000 µA (0 = OFF)
Delay Timer Range: 0, 1.0 - 999.9 sec (0 = Constant)
At Associated Research, Safety Compliance Testing Is Our Only Focus.
13860 W. Laurel Dr., Lake Forest, IL USA 60045-4546 P: +1.847.367.4077 F: +1.847.367.4080 info@asresearch.com
www.asresearch.com
1.800.858.TEST [8378]