Differential Probe Capable of Wide-Band,
High-Voltage Floating Measurements
Allows you to use a single-end input oscilloscope for wide-band differential input measurements
Waveform observed on oscilloscope
Oscilloscope input amplifiers are normally single-ended. They do not provide accurate measurements when used in combination with a 10:1 passive probe for measuring the gate-to-source voltage of a switching power supply, or potential difference with a common mode voltage in elements such as Insulated Gate Bipolar Transistors (IGBTs). Furthermore, there is a risk that current will enter through the oscilloscope’s ground line and damage the oscilloscope.
http://www.yokogawa.co.jp/Measurement/English/
Y okogawa Electric WWW server:
Connection to oscilloscope input terminals
700924
Signals floating from ground
With increases in the speed of measured signals in recent years, it has become necessary to make floating measurements over a wide frequency band. The 700924 differential probe lets you make wide-band (100 MHz) differential input measurements. Just connect the probe to the input of a single-end input digital oscilloscope.
Bulletin 7009-24E
100 MHz band differential probe
700924
Overrange (outside-range) indicator
This light turns on when the permitted differential voltage is exceeded. During DC measurements, it turns on as an alert at voltages of 350 V and higher with a 1/100 range setting, and at voltages of 1400 V and higher with a 1/1000 range setting.
Power switch
AC/DC adapter connector
(on side of probe)
This DC adapter powers the probe. The probe can also be powered by four dry cells without using an AC/DC adapter.
1/100-1/1000 selector switch
This switch is used to select the input voltage attenuation ratio.
Specifications
Input type: Balanced differential input Frequency band: DC up to 100 MHz (-3 dB) Input attenuation ratio: Switched between 1/100 and 1/1000 Input impedance: Approximately 4 M, approximately 10
Gain accuracy: ±2% (common mode voltage % 400 V) Maximum allowed differential voltage:
Maximum common mode input voltage:
Maximum input voltage: ±1400 V (DC + ACpeak) or 1000 Vrms Common Mode Rejection Ratio (CMRR): Output voltage: ±3.5 V
Output impedance: Used with 1 M input impedance Power supplies: Dry cells, four R6P (SUM-3); when using
Dry cell life: Approximately 2 hours under continuous External dimensions (approximate): Total length: Approximately 1.5 meters
Weight: Approximately 800 grams (excluding dry Approved standards:
EMC: EN55011: 1991 + A1: 1997 + A2: 1996
Safety: EN61010-1: 1993 + A2: 1995
*1: 38 mm is listed as the maximum value.
Note: Dry cells and AC/DC adapter not included.
pF parallel
±3% (common mode voltage % 1000 V) ±350 V (DC + ACpeak) or 250 Vrms (1/
100 range) ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range)
±1400 V (DC + ACpeak) or 1000 Vrms (both 1/100 and 1/1000 ranges)
(1/1000 range)
-80 dB (60 Hz), -50 dB (1 MHz)
the AC/DC adapter, use an output voltage of 6 V/200 mA or greater/center plus
use 207 mm 83 mm 38 mm*
cells)
EN50082-1: 1992
EN61010-2-031: 1994
1
Example of probe connection with DL1540CL digital oscilloscope
Two or more differential probes can be connected to measure signals with different ground levels on each channel.
Model and Codes
Model name Code Description
Differential probe 700924 DC to 100 MHz band
Each 700924 differential probe comes standard with red and black pincher tips (one each).
Spares
Item Part No. Order Q'ty
Red pincher chips B9852MC 3
Black pincher chips B9852MB 3
Use this product in combination with 701856 (high-speed module) when using it with DL708E or DL716. See the DL Series accessory catalog for information on selecting a conversion adapter and other components. Bulletin 7009-63E
YOKOGAWA ELECTRIC CORPORATION
Measurement Sales Dept. Recorders · DAC Sales Dept
YOKOGAWA CORPORATION OF AMERICA Phone: 770-253-7000, Fax: 770-251-2088 YOKOGAWA EUROPE B.V. Phone: 31-33-4-641611, Fax: 31-33-4-631202 YOKOGAWA ENGINEERING ASIA PTE. LTD Phone: 65-783-9537, Fax: 65-786-6650
/Phone: 81-422-52-6614, Fax: 81-422-52-6624
./Phone : 81-422-52-6765, Fax : 81-422-52-6793
Subject to change without notice.
[Ed : 01/b] Copyright ©1999
Printed in Japan, 907(YG)
MS-06E
Loading...