Accessory for YOKOGAWA DL Series Digital Oscilloscopes
100 MHz band differential probe
700924
NEW
Differential Probe Capable of Wide-Band,
High-Voltage Floating Measurements
Allows you to use a single-end input oscilloscope for wide-band differential input measurements
Waveform observed on oscilloscope
Oscilloscope input amplifiers are normally single-ended.
They do not provide accurate measurements when used
in combination with a 10:1 passive probe for measuring
the gate-to-source voltage of a switching power supply,
or potential difference with a common mode voltage in
elements such as Insulated Gate Bipolar Transistors
(IGBTs). Furthermore, there is a risk that current will enter
through the oscilloscope’s ground line and damage the
oscilloscope.
http://www.yokogawa.co.jp/Measurement/English/
Y okogawa Electric WWW server:
Connection to oscilloscope input terminals
700924
Signals floating from ground
With increases in the speed of measured signals in recent
years, it has become necessary to make floating
measurements over a wide frequency band. The 700924
differential probe lets you make wide-band (100 MHz)
differential input measurements. Just connect the probe
to the input of a single-end input digital oscilloscope.
Bulletin 7009-24E
100 MHz band differential probe
700924
●
Overrange (outside-range) indicator
This light turns on when the permitted
differential voltage is exceeded. During
DC measurements, it turns on as an
alert at voltages of 350 V and higher
with a 1/100 range setting, and at
voltages of 1400 V and higher with a
1/1000 range setting.
● Power switch
● AC/DC adapter connector
(on side of probe)
This DC adapter powers the probe. The
probe can also be powered by four dry
cells without using an AC/DC adapter.
● 1/100-1/1000 selector switch
This switch is used to select the input
voltage attenuation ratio.
■ Specifications
Input type: Balanced differential input
Frequency band: DC up to 100 MHz (-3 dB)
Input attenuation ratio: Switched between 1/100 and 1/1000
Input impedance: Approximately 4 MΩ, approximately 10
Gain accuracy: ±2% (common mode voltage % 400 V)
Maximum allowed differential voltage:
Maximum common mode input voltage:
Maximum input voltage: ±1400 V (DC + ACpeak) or 1000 Vrms
Common Mode Rejection Ratio (CMRR):
Output voltage: ±3.5 V
Output impedance: Used with 1 MΩ input impedance
Power supplies: Dry cells, four R6P (SUM-3); when using
Dry cell life: Approximately 2 hours under continuous
External dimensions (approximate):
Total length: Approximately 1.5 meters
Weight: Approximately 800 grams (excluding dry
Approved standards:
EMC: EN55011: 1991 + A1: 1997 + A2: 1996
Safety: EN61010-1: 1993 + A2: 1995
*1: 38 mm is listed as the maximum value.
Note: Dry cells and AC/DC adapter not included.
pF parallel
±3% (common mode voltage % 1000 V)
±350 V (DC + ACpeak) or 250 Vrms (1/
100 range)
±1400 V (DC + ACpeak) or 1000 Vrms
(1/1000 range)
±1400 V (DC + ACpeak) or 1000 Vrms
(both 1/100 and 1/1000 ranges)
(1/1000 range)
-80 dB (60 Hz), -50 dB (1 MHz)
the AC/DC adapter, use an output voltage
of 6 V/200 mA or greater/center plus
use
207 mm ✕ 83 mm ✕ 38 mm*
cells)
EN50082-1: 1992
EN61010-2-031: 1994
1
Example of probe connection with DL1540CL digital oscilloscope
Two or more differential probes can be connected to measure signals with different ground
levels on each channel.
■ Model and Codes
Model name Code Description
Differential probe 700924 DC to 100 MHz band
Each 700924 differential probe comes standard with red and black pincher tips (one each).
■ Spares
Item Part No. Order Q'ty
Red pincher chips B9852MC 3
Black pincher chips B9852MB 3
Use this product in combination with 701856 (high-speed module) when using it with
DL708E or DL716. See the DL Series accessory catalog for information on selecting a
conversion adapter and other components. Bulletin 7009-63E
YOKOGAWA ELECTRIC CORPORATION
Measurement Sales Dept.
Recorders · DAC Sales Dept
YOKOGAWA CORPORATION OF AMERICA Phone: 770-253-7000, Fax: 770-251-2088
YOKOGAWA EUROPE B.V. Phone: 31-33-4-641611, Fax: 31-33-4-631202
YOKOGAWA ENGINEERING ASIA PTE. LTD Phone: 65-783-9537, Fax: 65-786-6650
/Phone: 81-422-52-6614, Fax: 81-422-52-6624
./Phone : 81-422-52-6765, Fax : 81-422-52-6793
Subject to change without notice.
[Ed : 01/b] Copyright ©1999
Printed in Japan, 907(YG)
MS-06E