Atec 7000 User Manual

Series
7000 Precision LCR Meters
USES:
• Production Testing of LCR Components
• Frequency Response Characterization of Components
• Component Screening
• Quality Assurance Testing
• Measuring Dielectric Constant Using Standard Test Cell
FEATURES:
• Frequency Range 10 Hz to 2 MHz (7600), 500 kHz (7400)
• 0.05% Basic Measurement Accuracy
• 7 Digit Measurement Resolution
• Programmable Test Voltage and Current
• Auto Ranging
• Test Setup and Measurement Data Storage
• Four-Terminal Kelvin Connections
• IEEE-488, RS-232, Handler, and Parallel Printer Interfaces
• Graphical and Tabular Display of Swept Frequency, Voltage and Current Measurements
• Sequence Testing of Up To 6 Individual Tests
• Load Correction
• Binning (15)
• Built-in Automatic Calibration Routine
Introduction
The 7000 LCR Meter is designed to perform precision impedance measurements over a wide frequency range, 10 Hz to 500 kHz for the model 7400 and 10 Hz to 2 MHz for the
7600. The instrument is capable of measuring 14 different parameters with 0.05% accu­racy to meet today’s requirements for component and material testing. The ease of use and user friendly menu programming makes the 7000 Series ideal for applications in product development, incoming inspections, or production line testing.
Description
14 Different Impedance Parameters Measure and display any two parame-
ters simultaneously to achieve coverage and flexibility not previously available.
Automatic Test SequencingRun up to six different tests in sequence with a sin-
gle push of the start button. Each test can have different conditions and limits.
Swept Measurements Fast and accurate swept parameter measurements, graphi-
cal or tabular, for verification of component and material response to changes in AC test frequency, AC test voltage or AC test current, without the need for complex programming or an external controller.
Program and Data Storage Test setups can be stored and recalled from either
internal memory or from standard DOS formatted 3 1/2” floppy disks. The front panel can be locked out, with password protection, to ensure procedures are run the same way every time. Measured data can be stored on a floppy disk and then transferred to PC for data reduction and analysis.
Load Correction Substantially improves instrument accuracy by performing meas-
urements on a known standard and applying correction to subsequent measurements. Ideal for repetitive testing of identical devices at like test conditions.
Automatic Calibration ProcedureThe 7000 can be calibrated without return-
ing the unit to the factory using the NIST traceable QuadTech Calibration Kit, reducing downtime and calibration costs.
Easy to UseLarge LCD graphics display and user friendly menu driven interface allows
the 7000 to be put on line fast, providing useful measurements by operators with little or no training.
1-800-253-1230
Fax 1-508-485-0295
Intl. 1-508-229-0806
For more information about special purchase,
rent & lease options, call
For more detailed specifications, visit
www.quadtech.com
CE Marked Impedance Analyzer
HIPOT • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS • DECADES • STANDARDS
7400/7600
Measured Parameters: Any two of 14 parameters measured and
displayed simultaneously, user selectable
Test Frequency: 7400 Range: 10 Hz to 500 kHz, continuous
Resolution: 0.1 Hz from 10 Hz to 10 kHz, 5 digits>10kHz Accuracy: +/-(0.002% + 0.02Hz)
7600 Range: 10 Hz to 2 MHz, continuous
Resolution: 0.1 Hz from 10 Hz to 10 kHz, 5 digits>10kHz Accuracy: +/-(0.25% + 0.02Hz)
Measurement Speed: 7400 Basic: 40 meas/sec
Enhanced: 8 meas/sec Extended: 1 meas/sec
7600 Fast: 25 meas/sec
Medium: 8 meas/sec Slow: 1 meas/sec
Ranging: Automatic, Range Hold or user selectable Trigger: Internal (automatic), External (RS-232,
IEEE-488.2 or Handler interfaces) and Manual
AC Test Signal: Voltage: 20 mV to 5.0 V (open circuit)
up to 500kHz in 5 mV steps
7600 20 mV to 1.0 V (open circuit)
500kHz-1MHz in 5 mV steps 20 mV to 0.5 V (open circuit) >1MHz in 5 mV steps
Current: 250 µA to 100 mA (short circuit) in 50 µA steps (Max. Compliance 3V < 500kHz.
Source Impedance: 25, 400, 6.4k, or 100k, range dependent DC Bias Voltage: Internal: 2.0 V
External: 0 to +/-200V
0 to +/-500V on 7400A/7600A
Display: LCD Graphics with back light and adjustable
contrast
Result Formats: Engineering or scientific notation
% Deviation from nominal of primary parameter Deviation from nominal of primary parameter Pass/Fail Binning summary No Display for maximum throughput
Sweep Result: Primary parameter vs. frequency, voltage or
current Graphical or Tabular Format Up to 200 measurement points per sweep
Sequencing Result: Displays up to 6 sequential test results, primary
and/or secondary
AutoAcc: Automatic calculation and display of overall
instrument accuracy for selected settings, test conditions and device under test
Standard Interfaces: IEEE-488.2, RS-232, Handler, Printer Port,
3.5" Disk Drive
Charged Capacitor Protection:
8/C for Vmax 250 V;
2/C for Vmax ≤ 1000V
C = Capacitance in farads of the device under test
Additional Fuse Protection on 7400C
Measurement Delay: Programmable from 0 - 1000 ms in 1 ms steps Averaging: Programmable from 1 - 1000
Median value mode
Data Storage: 40,000 measurements/disk
ASCII format DOS compatible
Program Storage: 125 per disk 25 setups internal
Password protected ASCII format
Calibration: Recommended Calibration Interval 1 year
Complete NIST Traceable Calibration using QuadTech 7000-09 Cal Kit Built-in automatic calibration procedure
Usage & Cal Data: Displays last calibration date, standard values
used in calibration and # of hours operation
Self-Test Routine: Verifies critical instrument operation at power-
up or when selected from menu
Contact Check: Time to detect, 2ms Test Terminals: Front panel, four terminal (BNC) guarded Mechanical: Bench mount with tilt bail
Dimensions: (w x h x d): 16 x 6 x 14in
(410 x 150 x 360mm) Weight: 17 lbs (8kg) net, 23 lbs (10.5kg) shipping
Environmental: Meets MIL-T-28800E, Type 3, Class 5,
Style E & F Operating: 0 to + 50˚ C Humidity: <75% for 11 to 30˚ operating Storage: - 40 to + 71˚ C
Power: • 90 - 250Vac • 47/63Hz
• 100W max
For more detailed specifications, visit www.quadtech.com • For more information about special purchase, rent & lease options, call
1-800-253-1230 Fax 1-508-485-0295 Intl. 1-508-229-0806
5 Clock Tower Place, 210 East, Maynard, MA 01754
Product Specifications
PN030087/A1
7400/7600 Precision LCR Meters
7600 LCR Meter 7600-CE LCR Meter, CE Marked 7600A LCR Meter, +/-500V External Bias 7600A-CE LCR Meter, +/-500V External Bias,
CE Marked 7400 LCR Meter 7400-CE LCR Meter, CE Marked 7400A LCR Meter, +/-500V External Bias 7400A-CE LCR Meter, +/-500V External Bias,
CE Marked
7400C LCR Meter, Charged Capacitor
Protection
7400C-CE LCR Meter, Charged Capacitor
Protection, CE Marked
Includes:
Instruction Manual Power Cord Calibration Certificate Traceable to NIST
Optional Accessories:
7000-00 Rack Mount Kit
7000-01 BNC Cable Set, 1 meter 7000-02 BNC Cable Set, 2 meters 7000-03 Kelvin Clip Leads 7000-04 Alligator Clip Leads 7000-05 Chip Component Tweezers 7000-06 Axial/Radial Lead Component
Test Fixture 7000-07 Chip Component Test Fixture 7000-08 High Voltage Test Fixture 7000-09 Calibration Kit
Ordering Information
Cs, Cp 00000.01 fF to 9.999999 F ±0.5% ±0.25% ±0.05% Ls, Lp 0000.001 nH to 99.99999 H ±0.5% ±0.25% ±0.05% D .0000001 to 99.99999 ±0.005 ±0.0025 ±0.0005 Q .0000001 to 999999.9 ±0.005 ±0.0025 ±0.0005 Z , Rs, Rp, 000.0001 mto 99.99999 M ±0.5% ±0.25% ±0.05% ESR, Xs Y , Gp, Bp 00000.01 µS to 9.999999 MS ±0.5% ±0.25% ±0.05% Phase Angle -180.0000 to +179.9999 degrees ±1.8˚ ±0.9˚ ±0.18˚
Basic Measurement Accuracy*
Parameter Measurement Range Speed
Fast Medium Slow
*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty.
Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q) Factors, Impedance Z, Admittance Y , Phase Angle (0-), Equivalent Series Resistance (ESR), Conductance (Gp), Reactance (Xs), Susceptance (Bp)
Note: s = series, p = parallel, ESR equivalent to Rs