A
Specifications
Specifications
Table A-1 lists the specifications of the dc source. Unless otherwise noted, specifications are warranted
over the ambient temperature range of 0 to 55 °C. Specifications apply with typical cellular phone
capacitive loads from 0µF to 12,000µF. Sensing is at the rear terminals of the power supply after a 30minute warm-up period. Sense terminals are externally jumpered to their respective output terminals.
Table A-1. Performance Specifications
Parameter Agilent 66111A Agilent 66311B/D;
Agilent 66309B/D
output 1 only
Output Ratings
Programming Accuracy
(@ 25°C ±5°C)
DC Measurement Accuracy
(via GPIB or front panel meters with
respect to actual output @ 25°C
±5°C)
Ripple and Noise
(in the range of 20 Hz to 20 MHz
High Current range
+20 mA to + rated I:
−20 mA to − rated I:
Low Current range
−20 mA to +20 mA:
Voltage (rms/p-p):
Voltage:
Current:
Peak Current:
Voltage:
+Current:
Voltage:
Current:
Current (rms):
0.03% + 5 mV
see below
4
0.2% + 9 mA
0.2% + 9 mA
0 – 15 V
0 – 3 A
5 A
0.05% + 10 mV
0.05% + 1.33 mA
NA
1 mV/6 mV
2 mA
1
3
0.03% + 5 mV
see below
0.2% + 0.5 mA
0.2% + 1.1 mA
0.1% + 2.5 µA
7
5
6
with outputs ungrounded or with
either terminal grounded)
Load Regulation
(change in outp ut voltage or current
Voltage:
Current:
2 mV
0.75 mA
for any load change within ratings)
Line Regulation
(change in outp ut voltage or current
Voltage:
Current:
0.5 mV
0.75 mA
for any line change within ratings)
Transient Response Time
< 35 µs
8
(for the output voltage to recover to
20 mV below its final setting)
1
Peak current for up to a 7 millisecond time period. Average current cannot exceed 3 A.
2
Peak current for up to a 1 millisecond time period. Average current cannot exceed 1.5 A.
3
0.05% +2.3mA when programming between zero and 0.03% of full scale current.
4
Agilent 66111A units do not have a low current range, the dc measurement accuracy applies from 0 amperes to the rated current.
5
Applies with current detecto r set to DC. ACDC mode accuracy is 0.2% + 3mA for sourcing and 0. 2% + 3.6 mA for sinking.
6
This specification may degrade slightly when th e unit is subjected t o an RF field ≥3 V/meter.
7
Specification is for phone capacitance greater than 6µF.
8
Applies to the negative portion of the output voltage transient waveform, following a 0.1 A to 1.5 A load current change.
(Applies with unit set to High capacitance compensation range with 5
9
Following a 0.75 A to 1.5 A load change.
µF minimum output load capacitance.)
Agilent
66309B/D
output 2 only
0 – 12 V
0 – 1.5 A
2.5 A
0.2% + 40 mV
0.2% + 4.5
mA
0.2% + 15 mV
0.2% + 3 mA
NA
NA
NA
1 mV/6 mV
2 mA
1.6 mV
0.375 mA
0.4 mV
0.25 mA
< 400 µs
2
7
9
143
A - Specifications
Supplemental Characteristics
Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical
performance determined either by design or type testing.
Table A-2. Supplemental Characteristics
Parameter Agilent 66111A
Output Programming
Range
Voltage:
Current:
0 – 15.535 V
0 – 3.0712 A
OVP:
Average Programming
Resolution
Voltage:
Current:
OVP:
OVP Accuracy
2.4 % + 240 mV NA
Agilent
66311B/D,
Agilent 66309B/D
output 1 only
22 V
3.75 mV
0.75 mA
100 mV
Agilent
66309B/D
output 2 only
0 – 12.25 V
0 – 1.52 A
NA
14 mV
1.75 mA
NA
Average Current
Measurement Resolution
Maximum Current
High Range:
Low Range:
3.5 mA
NA
213 µA
0.6 µA
55 µA
NA
7 A 1.8A
Measurement
Sink Current - 2 A @ 7.5 V
Programming Accuracy
Temperature Coefficient
(change/C°)
Readback Accuracy
Temperature Coefficient
(change/C°)
2
Drift
Remote Sense Capability
Current (low range):
Voltage:
Current:
OVP:
Voltage:
Current (dc):
Current (acdc):
Voltage:
Current: +
0.01% + 0.5 mV
0.01% + 15 µA
0.015% + 4 mV
0.01% + 300 µV
0.02% + 15 µA
NA
NA
0.01% + 1 mV
0.01% + 30 µA
up to 4 V
1
0.01% + 300 µV
0.02% + 15 µA
0.05% + 120 µA
0.01% + 0.3 µA
3
- 0.03 A @ 7.5 V
0.03% + 1.5 mV
0.03% + 0.4 mA
NA
0.02% + 1.5 mV
0.02% + 0.4 mA
NA
NA
0.03% + 3 mV
0.03% + 0.8 mA
up to 4 V
(voltage per load lead)
Output Voltage Rise Time
Output Voltage Fall Time
Output Voltage Settling
5
Time
Dynamic Measurement
System
1
2
3
4
5
6
current pulse of 1.4A with a rise time constant of 50µs, the error in measurement of a single data point during the pulse rise time
is typically 10mA.
6
The sink current decreases linearly from 2.8A @ 0 V to 1.2 A @ 15 V. Sink current does not track the programmed current.
Following a 30 minute warm-up, the change in output over 8 hours, under constant ambient, load and line conditions.
Add 2 mV to the load regulation for each 1 V chan ge in th e positive output lead.
Add 3 mV to the load regulation for each 1 V change in the negati ve output lead.
To settle within 1 LSB or 3.75mV of the final value for Output 1.
For a pulse waveform, the accuracy of any individual data point in the b uffer is dependent on the rise t ime of the pulse. For a
10% to 90%:
90% to 10%:
0 to rated voltage:
rated voltage to 0:
Buffer Length:
Sample Rate Range:
NA
< 200 µs
< 200 µs
0.5 ms
1 ms
0 − 4096 points
15.6µs − 31200s
< 200 µs
< 35 ms
NA
NA
NA
4
144