AC Output Voltage: Range: 50V to 5000V AC, 1V resolution
Regulation: <(1% of setting +5V) at Rated Load
Freq: 50-600 Hz, programmable in 1Hz steps
50~60Hz : 1% + 5 cnts, > 60Hz: 3%+10 cnts
Waveform: Sinusoidal
Voltage Display: Accuracy: ±(1% of reading + 5V)
Resolution: 1Volt
AC Current Display: Total current
Range: 0.001 to 40mA AC
Resolution: 1 or 10µA steps
Accuracy: ±(1% of reading + 5cnt)
High/Low Limit Test: 1µA to 40mA AC
Low limit can be turned OFF
Arc Detection: Programmable 1-20mA,
Pulse Width 40µ, 20µ, 10µ or 4µsec
DC Output Voltage: Range: 50V to 6000V DC, 1V resolution
Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: ±(1% of reading + 5V)
Resolution: 1Volt
DC Current Display: Range: 0.1µ to 12mA DC
Resolution: 0.1, 1 or 10µA steps
Accuracy: ±(1% or reading + 5cnt)
High/Low Limit Test: 0.0001mA to 12mA DC
Low limit can be turned OFF
Arc Detection: Programmable 1-10mA
Pulse Width 40µ, 20µ, 10µ or 4µsec
Insulation Resistance: Range: 50V - 1000V DC
Accuracy: ±5% to ±15% depending upon
voltage and resistance
Voltage Range: 100kΩ - 50GΩ
Voltage Display: ±(1% of reading + 5V)
High/Low Limit Test: 100kΩ - 50GΩ
Low limit can be turned OFF
Ground Bond:
Output Current: Range: 1.0 to 30.0A AC, setting .01A/step; *40A
Regulation: ±(1% of setting + 0.3A)
Accuracy: ±(1% of reading + 3 counts)
Frequency: 50 or 60Hz Selectable
No Load Voltage: 6 to 15 V Programmable
Resistance: Range: 0.1mΩ - 510mΩ , 4 digits
Accuracy: ±(1% of reading + 3 counts)
Resolution: 0.1mV
High Limit: 0.1mΩ to 510mΩ
Start Wait: 0.1 - 99.9sec
Common Features:
Open/Short Circuit Mode: Check for Open & Short against a standard
capacitance value (Cs); <100V, 600Hz
Offset Function: 0 to 100mΩ offset, user selectable
AC Programmable Offset, 0-25mA
Test Time: Ramp: 0.3 to 999s (±20ms), (AC/DC/IR)
Dwell: 0 to 999s (±20ms), (DC only)
Start Wait: 0.1 to 99.9s (±20ms), (GB only)
Test: 0.3 to 999s (±20ms) or Continuous (ALL)
Ramp, Dwell & Start Wait can be set to OFF
Remote Control: Inputs: Start,Stop
Characteristics: Optically Isolated, Low,
Pulse Width >1ms.
Outputs: Pass/Fail/Under Test
Characteristics: Dry Contact relay
Electrical Characteristics: 120V 100mA max.
Logic: Closed if True
Connector: Terminal Strip and 9 pin D Series
Test Setups: 100 Test Setups with 50 Steps each,
13 character Alpha-Numeric Label
Connectors: Front and Rear 4- terminal connection
Display: 320 x 240 enhanced LCD with status indicators
Front Panel Lockout: Key Lock with Password
Fail Lock with Password
Safety Features: Fast Cutoff (<0.4ms) & Fast Discharge (0.2s)
Adjustable Discharge: .05-5.1kV DC
Miscellaneous: Fail Retest; Continue on Fail
PAUSE Mode: Program pause between steps
Indication: Pass/Fail lights, audible sound; Remote, Lock,
Offset & Error status indicators
Buzzer Level: 1,2,3 and Off
Standard Interfaces: RS-232, Remote I/O
Data bits: 8, Stop bits: 1; Parity: None
Selectable Baud Rate: 300 - 19200
Optional Interfaces: IEEE-488, Printer
Dimensions: (w x h x d):17x5.25x18.5in (430x133x470mm)
Weight: 53 lbs (24kg) - Net, 60 lbs (27kg) Shipping
Environmental: Operating: 0 to + 40° C,
Humidity: <75%
Storage: - 20 to + 70° C
Warm-up Time: 1minute
Power: • 90 - 130V AC • 50 or 60Hz
• 200 - 250V AC • 800W max
Guardian 6000 Plus Electrical Safety Analyzer
Includes:
150798 Instruction Manual
S02 HV Lead Set, 1m
G15 Ground Continuity Lead Set
G14 Power Entry Adapter
4200-0300 AC Power Cable
520157 8A 250V Power Line Fuse
520053 4A 250V Power Line Fuse
N/A Calibration Cert. Traceable to NIST
Optional Accessories
N/A Calibration Data
6000-01 Scanner, 8 Channel 5HV,3GC
6000-02 Scanner, 8 Channel 3HV,5GC
6000-03 Scanner, 8 Channel HV
S04 HV Lead Set 2m
S05 Foot Switch
S08 Gun Probe
S09 HV Lead, 1M, Unterminated
S10 HV Lead, 1M, Unterminated
G13 Corded Product Adapter (115V)
G14 Power Entry Adapter
G16 International Power Strip
G24 Scanner Cable (5000 scanners)
G25 Corded Product Adapter (240V)
G38 Printer Interface (replaces IEEE)
G39 IEEE-488 Interface
G43 Rack Mount Kit
G44 Barcode Scanner
G45 40A Ground Bond Transformer
Ordering Information
Guardian 6000 Plus
P r o d u c t S p e c i f i c a t i o n s
*GB to 40A with optional transformer
For more detailed information on specifications, pricing and
special purchase, rent and lease options, contact us at:
www.quadtech.com or 800-253-1230
P/N: 030145/A6