2801 and 2802 Advanced Power Analyzers
The 2801 and 2802 Advanced Single- and Two-Channel Power Analyzers combine an unprecedented feature set with an ideal combination of precision, speed and ease-of-use in an instrument so economical it can be on every bench.
With an extended measurement range from micro-amps to hundreds
of amps and millivolts to kilovolts, the 2801 and 2802 are the ideal
analyzers for standby power or Energy Star testing. In addition to
numerical results, the 2801 and 2802 captures waveforms with true
512-point precision. Results and waveforms can be displayed, read
via the communication ports, or sent directly at full resolution to a USB
printer. Power and amplitude measurements with an accuracy of <0.1%
are automatically synchronized to the fundamental frequency. Peak,
rms, rectifi ed, and DC measurements of voltage, current and power are
provided including continuous, inrush, startup and history modes, plus
an integration mode for W-Hr, A-Hr, VA-Hr, as well as integrated average
power. (Channels can be operated synchronously or independently on
the 2802) The 2801 and 2802 also provide waveform peak and glitch
capture modes.
Quality and Reliability
XiTRON Technologies, founded in 1990, is the premier source of
precision power testing and measuring equipment for industrial and
consumer product development and manufacturing. XiTRON’s
sophisticated technology provides companies the edge in design
verifi cation and product manufacturability. XiTRON is ISO9001-2000
certifi ed, EN46001 registered and FDA (GMP820) compliant.
Highest Performance-to-Cost ratio in the industry
»
Up to 2000V peak and 150A peak measurable internally
»
(external CT capable)
Two channels for independent or synchronous measurements,
»
including input/output effi ciency and loss (2802 only)
Base accuracy <0.08%. Current and voltage accuracies specifi ed
»
to less than 1mArms and 1Vrms respectively (<0.2%)
Measures and displays volts, current, power, frequency, harmonics
»
Provides PASS/FAIL tests to user limits*
»
Integrated line switch and inrush waveforms
»
Source or load measurements with wiring loss and voltage
»
burden compensation
DC charge and discharge
»
measurements
Frequency Range: DC and
»
20mHz – 200kHz
Graphics display shows
»
numerical results, wave­forms, bar graphs, startup & history charts with zoom & scroll features
16-bit A/D converters at 235ksps
»
12 User-confi gurable digital I/O
»
USB Flash drive support for data logging*
»
Communications interfaces include GPIB (IEEE488), RS-232, USB
»
(host and device), and Ethernet*
Suitable for AC, DC, 1-phase 3-wire, 2-phase 3-wire†, 3-phase 3-wire†,
»
in/out synchronous†, or independent measurements (†2802 only).
(*planned)
XiTRON XVIEW Software
While all XiTRON precision test equipment is designed to
be used in a completely stand-alone manner, there are
times when external tools can aid or enhance the operation
of an instrument. XView software tools and drivers are designed to help
easily confi gure an instrument from a single screen, or are used to view a
complete set of measurements in a single screen. Other XView tools are
designed for data collection where results can be recorded in an Excel-
compatible fi le for post-processing, insertion into reports, or simply for
archival purposes.
These precompiled applications can be downloaded from the XiTRON
web site at www.xitrontech.com/support.html. Often, XiTRON can provide
the source code for a particular application, and it can be used as a
convenient ‘starting point’ for a custom software application.
9770-A Carrol Centre Road | San Diego, CA 92126 | 858.530.8099 phone | 858.530.8077 fax
www.xitrontech.com | sales@xitrontech.com | support@xitrontech.com
2801 2802
1 year Specifi cation, AC Voltage, Autorange, CF <1.5, Line or LF Measurements
Line Measurements: Add 0.05% per KHz LF Measurements: Add 0.3% per KHz
1 year Specifi cation, AC Current, Autorange, CF <2.5, Line or LF Measurements
Line Measurements: Add 0.05% per KHz LF Measurements: Add 0.3% per KHz
Current, Voltage and Power Harmonics Barcharts (to 100th)
Current and Voltage Inrush Charts
Numerics Results Screens
CONDENSED SPECIFICATIONS
Accuracy specifi cations valid for 1 year for ambient temperatures within 5° C of calibration temperature.
(Contact XiTRON for complete specifi cations.)
VOLTAGE RANGE (MEASURABLE)
2000Vpk (850Vrms) max. continuous
INTERNAL CURRENT RANGE (MEASURABLE)
150Apk (30Arms) max. continuous
EXTERNAL CURRENT TRANSDUCERS (MEASURABLE)
35Vpk (15Vrms) max. continuous
VOLTAGE AND CURRENT ACCURACY
See charts to the left.
FREQUENCY MEASUREMENT ACCURACY
0.005%
CREST FACTOR ACCURACY
V: (50mV + 0.01% of pk rng)/RMS value A: (50uA + 0.01% of pk rng)/RMS value
HARMONIC ACCURACY (VOLTAGE & CURRENT)
0.02%
WAVEFORMS
Actual, Peak Capture, Distortion, and Glitch capture
CHARTS, WITH ZOOM AND SCROLL
History, Startup and Inrush
LINE SWITCH
Max Open Voltage: 720Vpk (480Vrms) Max Surge Current: TBD Max On Current: 10Arms Turn On Phase: 1° resolution
DIGITAL INTERFACES
GPIB (IEEE-488), USB (host and device), RS-232, Digital IO, Ethernet* (*planned)
PHYSICAL
Power Input: 12VDC @ 1.5A minimum output Size: (HxWxD) 4.7”x 13.8”x 9.5”
(11.94cm x 35.05cm x 24.13cm)
Weight: 7.5 lbs (3.4 kg) Operating Range: 0ºC to 55ºC, <95% RH non-condensing Storage Range: -20ºC to 70ºC <95% RH non-condensing
Unit is supplied with one XiTRON Technologies T5 universal external power supply, 100-240Vrms, 50-60Hz, with a 2.5mm 12VDC output plug and a three prong IEC320 AC inlet receptacle, plus a three-prong AC power cord.
WARRANTY
Two years
9770-A Carrol Centre Road | San Diego, CA 92126 | 858.530.8099 phone | 858.530.8077 fax