ISOPROBE
Electrostatic Voltmeter
Features:
• ±3kV Range
• Measurement Without Physical
Contact
• Accuracy Better Than 0.1% at Almost
any Probe-to-Surface Spacing
• Full Complement of State-of-the-art
High Ferquency Probes
• Provision for Master/Slave Operation
High performance with a proven
track record
model 244A
Manufactured with the same care and attention
that have made Monroe ISOPROBE® Electrostatic
Voltmeters the industry st andard, the Model 244A
measures surface potential of insulating,
semiconductive or conductive materials on a small
0.04” (1mm) diameter spot or over a large area using
patented techniques which assure high-accuracy
drift-free measurements almost independent of probe/
surface separation.
Uses 1017A Series Probes:
Monroe Electronics T ype 1017AE (end-viewing) or
1017AS (side-viewing) (probes are 0.35” [9mm] x 0.35”
[9mm] x 2.85” [72.5mm] in length. Add 0.8” [20mm]
length for cable at minimum bend radius.) Contains
1kHz tuning fork chopping driver and onboard hybrid
microcircuit preamp. Useable from -50°C to +80°C.
Optional probe configurations are available for high or
low resolution and transparent probes for light decay
measurements. Length of probe cable is 10 ft. (3.0
meters). Provision has been made for air or inert gas
purging of probe. Unit is calibrated independent of
probe and includes certificate of NIST traceability .
Interchangeable Model 1017A probe (type specified by
customer) is sold separately . See Model 1017data
sheet for more details and options.
The fifth generation from the first
name in non-contact
measurement
Model 244A takes advantage of Monroe’s years of
experience in design of reliable instruments for NONCONTACTING measurement of electrost atic potential
combined with modern semiconductor technology. A
full spectrum of proven-design interchangeable probes
exposes broad new areas for exploratory research as
well as providing a precision instrument for routine
applications in electrostatic measurements. Some
typical and potential applications include:
• Electrophotographic and Xerographic
Measurements
• IC Manufacruring and Handling
• Radiation Effects on Insulators and
Semiconductors
• Electret Research
• Static Electrification, Electric Field
Studies
• Process Monitoring and Control
ISOPROBE® Electrostatic Voltmeter
Specifications:
Range: ±3000 volts, auto polarity .
Accuracy:* 0.1%, ±0.003%/°C over +20 to +40°C
range (at recorder output). Useable to
+50°C.
Speed of
response:*
For typical input step: <3ms (10% to 90%)
for 1kV step: <2½ms (10% to 90%)
for 2kV step: <3½ms (10% to 90%)
for 3kV step: <4ms (10% to 90%)
Settling time: <3½ms to 1% of 1kV step input.
Frequency Small signal frequency response
response: typically ±3db to >300Hz.
Drift:* <0.01V/hr after 1 hr warm-up (0.003V/
hr typical). Not measurably affected by
10°C temperature variation or changes
between 10% and 90% relative
humidity .
Noise:* <0.3V rms or 2 volts peak-to-peak wide
band (bandwidth restricted to<1kHz)
referred to input.
Surface Determined by probe aperture size and
resolution: surface-probe separation. Standard
type 1017AE & AS probes with 0.07”
(1.75mm) aperture will resolve a 0.10”
(2.5mm) spot at 0.02” (0.5mm)
separation.
Recorder Compressed analog output is input
output: divided by 1000 (optional Model 244AK
has divide by 200) for external loads
greater than 1KΩ. Output connector is
BNC.
Output filter: Bessel low pass filter with 0.7ms
constant delay .
Size: 4 x 8½ x 14 inches (10 x 21.8 x 35.6
cm), 1.75” rack mounting available (1 or
2 per rack).
Calibration:
Monroe Electronics instruments are factorycalibrated prior to shipment. Recalibration should be
performed annually , or more frequently if specified by
contract or company policy . Your instrument should
also be recalibrated any time it has been repaired or
tampered with. We are happy to recalibrate your
instrument for you at a reasonable cost, or provide
information and procedures on calibration upon request.
Warranty:
Monroe Electronics, Inc., warrants that each
instrument and sub-assembly manufactured by them
shall be free from defects in material and workmanship
for a period of one year after shipment from the factory .
This warranty is applicable to the original purchaser
only.
The finest Electrostatic instrumentation and
support:
For more than 40 years - ever since we invented the
feedback--nulled electrostatic voltmeter , Monroe has
been the technology and quality leader in electrostatic
detection and measurement instrumentation. Today we
offer the world’s most complete array of fieldmeters,
voltmeters, and resistivity meters. Our customers
include the leading makers of photocopiers and laser
printers, converters and microelectronics worldwide.
We know you need quality support as well as
quality products. We pride ourselves on providing our
customers with the most knowledgeable applications
and installation support — as well as superior customer
service.
How can we help?
Contact your Monroe Electronics representative for
price and delivery information on this and other ME
products, to schedule a no-obligation demonstration at
your convenience. For the name of your nearest dealer,
or for technical or applications assistance, contact
Monroe Electronics directly at the address and numbers
below.
Weight: 7 lb. (3.2 kg)
Power 100, 1 15, 230 V AC, ±10%, 50/60Hz, 15
requirement: watts.
Accessories Manual on CD, line cord, phono plug
Included
Certifications: CE Mark compliant
www.monroe-electronics.com
100 Housel Ave. | Lyndonville | NY | 14098
800-821-6001 | 585-765-2254 | fax 585-765-9330
Specifications subject to change without notice. | Printed in USA | Copyright© 2002 | Monroe Electronics Inc. | 244A ds 083106.p65
* Dependent on specific probe model, probe-surface
separation and environment. Specifications shown are
for standard T ype 1017AE or AS probes in a normal
laboratory atmosphere. Separation for accuracy and
response speed tests is 1/8” (3mm) and for noise and
drift tests, 0.13mm (0.005”). Performance generally
improves in controlled environments and may be
degraded under exceptional dirty or dusty conditions or
in ambiance of unstable gaseous constituents.