Model 2290-5
Keithley Instruments, Inc. 5 kV Voltage Power Supply Characteristics
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
CONDITIONS
This document contains specifications and supplemental information for the Model 2290-5 High Voltage
Power Supply. Specifications are the standards against which the Model 2290-5 is tested. Upon leaving the
factory, the Model 2290-5 meets these specifications. Characteristics, supplemental characteristics, and
typical values are not warranted, apply at 23 °C ± 5 °C, < 70% relative humidity, and are provided solely as
useful information.
CHARACTERISTICS
Voltage range:
Output voltage1 Maximum output current Conditions
50 to +5000 V dc 5.000 mA DC No filter
50 to +3000 V DC 5.000 mA DC Filter 1
50 to +5000 V dc 3.000 mA DC Filter 2
Voltage set accuracy2: ±(0.01% of setting + 2.5 V)
Voltage display accuracy: voltage set accuracy ±1 V, typical (±2 V, maximum)
Voltage resolution: 1 V (set and display)
Voltage limit range: 0 to 100% full scale
Voltage regulation3:
Line: 0.001% for ±10% line voltage change
Load: 0.005% for 100% load change, typical
Output ripple (10 kHz – 100 kHz)4:
0.002% of full scale, V RMS, maximum - No filter
1.0 mV RMS @ 1 kV – Filter 1 or Filter 2
2.0 mV RMS @ 3 kV – Filter 1 or Filter 2
3.0 mV RMS @ 5 kV – Filter 2
Rise time (full load)
5, 6
:
1.5 seconds for 0 V setting to within 1 V of 5000 V DC – No filter
3 seconds for 0 V setting to within 1 V of 3000 V DC – Filter 1
4 seconds for 0 V setting to within 1 V of 5000 V DC – Filter 2
1
The output voltage can be programmed to a voltage lower than 50 V, however, performance below 50 V is not specified.
2
Add ±2.5 V DC when Filter 1 or Filter 2 is enabled.
3
Regulation specifications apply for greater than 25 V DC (with full load), or 50 V DC (with no load). Below these values, the unit may not
regulate correctly.
4
Peak-to-peak values are within five times the RMS value.
5
Current limit set to 105% of full scale.
6
Under resistive load.
Specifications are subject to change without not i ce.
SPEC-2290-5 Rev. A / December 2013 Page 1 of 3
Model 2290-5 High Voltage Power Supply Characteristics
CHARACTERISTICS
Voltage range:
Discharge time (full load)
1 second for 5000 V DC to 1 V DC – No filter
3 seconds for 3000 V DC to 1 V DC – Filter 1
4 seconds for 5000 V DC to 1 V DC – Filter 2
Discharge time (no load)5:
<12 seconds (to <50 V DC) – No filter
<30 seconds (to <50 V DC) – Filter 1 or Filter 2
Output stored charge: <0.9 mC maximum
5, 6
:
Settling time
Recovery time
: From 0 to programmed voltage; to within 99.9% of final value within 3 seconds
5, 6
: 120 ms for 40% step change in load current (typical)
Current limit and trip range Filter
0 mA to 5.25 mA No filter or Filter 1
0 mA to 3.25 mA Filter 2
Current set accuracy7: ±(0.01% of setting + 2.5 µA)
Current resolution: 1 µA
Current display accuracy: ±1 µA, typical (±2 µA, maximum)
Stability: ±0.01% per hour; <0.03% per eight hours
Temperature drift: 50 ppm/ºC, 0º to 40º C, typical
Protection: Arc and short circuit protected; programmable voltage and current limits and current trip
Monitor outputs
Output scale: 0 to +10 V to full scale
Current rating: 10 mA maximum
Output impedance: <1 Ω
Accuracy: ±0.2% of full scale with a 100 kΩ load, minimum
Update rate: 8 Hz
7
Add 2.5 µA offset when Filter 1 or Filter 2 is enabled.
Specifications and characteri st ics are sub jec t to change without notice.
Page 2 of 3 SPEC-2290-5 Rev. A / December 2013