Model 2290-10
Output Stored Charge: <20 µC maximum
Keithley Instruments, Inc. 10 kV Voltage Power Supply Characteristics
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
CONDITIONS
This document contains specifications and supplemental information for the Model 2290-10 High Voltage
Power Supply. Specifications are the standards against which the Model 2290-10 is tested. Upon leaving
the factory, the Model 2290-10 meets these specifications. Characteristics, supplemental characteristics,
and typical values are not warranted, apply at 23 °C ± 5 °C, < 70% relative humidity, and are provided
solely as useful information.
CHARACTERISTICS
Voltage range:
Output voltage1
+100 to +10,000 V DC 1.000 mA DC
Voltage set accuracy: ±0.06% of full scale
Voltage display accuracy: voltage set accuracy ±1 V, typical (±2 V, maximum)
Voltage resolution: 1 V (set and display)
Voltage limit range: 0 to 100% full scale
Voltage regulation2:
Line: 0.001% for ±10% line voltage change
Load: 0.04% for 100% load change, typical
Maximum output current
Output ripple (300Hz – 300 kHz):
0.01% of full scale, V RMS
Rise time (from high voltage turn on to final value under full load)3, 4:
<6 seconds to within 1 V of the final value
Discharge time (to 0 V under fulll load)
3, 4
:
<1 second to under 10 V
Discharge time (no load)3:
<6 seconds (to <1% of full scale voltage with no load, typical)
1
The output voltage can be programmed to voltages below 100 V, but accuracy below 100 V is not specified.
2
Regulation specifications apply for greater than 100 V
3
Current limit set to 105% of full scale.
4
Under resistive load.
Specifications are subject to change without not i ce.
SPEC-2290-10 Rev. A / December 2013 Page 1 of 3
Model 2290-10 High Voltage Power Supply Characteristics
CHARACTERISTICS
Voltage range:
Settling time
3, 4
: <4 seconds to within 1 V of the final value
Recovery time
3, 4
: 120 ms for 40% step change in load current (typical)
Current limit range: 0 to 105% of full scale
Current set accuracy: 0.5% of full scale
Current resolution: 1 µA
Current display accuracy: ±1 µA, typical (±2 µA, maximum)
Trip current range: 10 µA to 105% of full scale (excluding stored output charge)
Trip response time: <10 ms
Stability:
Temperature drift: 50 ppm/ºC, 0º to 40º C, typical
Protection: Arc and short circuit protected; programmable voltage and current limits and current trip
Monitor outputs
Output scale: 0 to +10 V for 0 V to full scale
Current rating: 10 mA maximum
Output impedance: <100 Ω
Accuracy: ±0.2% of full scale with a 1 MΩ load, minimum
Update rate: 87.5 Hz
External voltage set
Input scale: 0 to +10 V for 0 V to full scale
Input impedance: 1 MΩ
Accuracy: ±0.2% of full scale
Update rate: 87.5 Hz
Output slew rate (5% to 95% under full load)
Specifications and characteri st ics are sub jec t to change without notice.
Page 2 of 3 SPEC-2290-10 Rev. A / December 2013
: <4 seconds