PROFLINE 2100 - OVERVIEW
HARMONICS AND FLICKER MEASURING SYSTEM
System
ProfLine 2100 is an accurate and fl exible system designed to measure harmonics and fl icker in accordance with IEC 61000-3-2 and IEC 61000-3-3. This is a measuring system only and no AC source is supplied. It can therefore be used with any suitable, clean AC source to measure and record any harmonics 
and fl icker created by the EUT. A wide range of AC sources are available from Teseq, please see ProfLine 
2103, 2105, 2115, 2130, 2145 datasheets for suitable systems including power sources.
1- or 3-phase
ProfLine 2100 can be supplied as either a 1-phase or 3-phase measuring system and can measure 
harmonics and fl icker up to 16 Amps per phase. Measurements are made using precision, no burden, 
active hall-effect current transformers connected via a dedicated cable to a multichannel fast Data Acquisition Card (DAQ) fi tted inside a PC. Three current and one voltage measuring channels are dedicated 
to each supply phase allowing simultaneous measurement of both current and voltage on all phases. 
Calculations are made using dedicated Teseq software (WIN 2100) to determine harmonics (classes A-D), 
 Single phase version 
 Three phase version 
 Complete measurement system 
 Use existing power sources or 
mains
inter-harmonics, fl icker, dc, dt, dmax, Pst, Plt, inrush current and 24 x dmax.
Impedance
Measurement of fl icker requires a fi xed, stable source impedance as specifi ed in IEC 61000-3-3 (0.24 Ω 
+ j0.15 Ω in the line and 0.16 Ω + j0.1 Ω in the neutral). When supplied as a complete system (see ProfLine 
2103, 2105, 2115, 2130 or 2145) which include a precision AC source, the impedance is provided by a 
combination of a controlled source impedance and physical impedances supplied as part of the system. 
Users of the ProfLine 2100 measurement system must take care to provide the necessary source impedance. Hardware lumped impedances are available from Teseq, see options list.

PROFLINE 2100 - OVERVIEW
HARMONICS AND FLICKER MEASURING SYSTEM
Technical information
  ProfLine 2100-1  ProfLine 2100-3
System contents  CCN 1000-1 1-phase  CCN 1000-3 1-phase 
  coupling unit  coupling unit 
  DAQ card  DAQ card 
  Interface cable  Interface cable 
  WIN 2100 test software  WIN 2100 test software
Coupling unit 
Number of phases  1  3 
Measurement channels  4  12 
EUT connector: Front panel  CEE 77  None 
EUT connector: Rear panel  Terminal block  Terminal block 
Max. voltage: Front panel  240 Vac  N/A 
Max. current: Front panel  16 Arms  N/A 
Max. voltage: Rear panel  300 Vac  480 Vac 
Max. current: Rear panel  40 Arms/Ph.  40 Arms/Ph.
(200 A Pk for 10 ms)  (200 A Pk for 10 ms) 
Supply power: Voltage  115/230 Vac +/- 10%  115/230 Vac +/- 10% 
Supply power: Current  <0.5 A  <0.7 A 
Supply power: Frequency  50/60 Hz  50/60 Hz 
Dimensions (H x W x D)  89 x 427 x 560 mm  89 x 427 x 560 mm 
Weight  5 kg  5 kg
DAQ card and cable
Interface  PCI  PCI 
Size  Standard height  Standard height 
Resolution  16 bit  16 bit 
Speed  250 kSamples/s  1.25 MSamples/s 
Cable length PC to CCN  2 m  2 m
Teseq AG
Nordstrasse 11F 4542 Luterbach Switzerland 
T + 41 32 681 40 4 0 F + 41 32 681 40 4 8 
sales@teseq.com www.teseq.com
© Februa ry 2011 Teseq® 
Specifications subject to change without notice. 
®
Tes eq
 is an ISO-registered company. Its products 
are designed and manufactured under the strict 
quality and environmental requirements of the ISO 
9001. This document has be en c arefully checked. 
However, Teseq 
errors or inaccuracies.
691-262A Fe bruary 2011
®
 does not assume any liability for 
Options
 INA 2151, 1-phase IEC 61000-3-3 16 Amp reference impedance 
 INA 2154, 3-phase IEC 61000-3-3 16 Amp per phase reference impedance 
 CCN 1000-1L, 1-phase coupling unit with integrated IEC 61000-3-3 16 Amp impedance 
 INA 2188, Rack mounting kit for CCN 1000-1, CCN 1000-1L or CCN 1000-3 
 WIN 2106, Test software for IEC 61000-3-11 and IEC 61000-3-12