Atec 2100 User Manual

PROFLINE 2100 - OVERVIEW
HARMONICS AND FLICKER MEASURING SYSTEM
System
ProfLine 2100 is an accurate and fl exible system designed to measure harmonics and fl icker in accor­dance with IEC 61000-3-2 and IEC 61000-3-3. This is a measuring system only and no AC source is sup­plied. It can therefore be used with any suitable, clean AC source to measure and record any harmonics and fl icker created by the EUT. A wide range of AC sources are available from Teseq, please see ProfLine 2103, 2105, 2115, 2130, 2145 datasheets for suitable systems including power sources.
ProfLine 2100 can be supplied as either a 1-phase or 3-phase measuring system and can measure harmonics and fl icker up to 16 Amps per phase. Measurements are made using precision, no burden, active hall-effect current transformers connected via a dedicated cable to a multichannel fast Data Acqui­sition Card (DAQ) fi tted inside a PC. Three current and one voltage measuring channels are dedicated to each supply phase allowing simultaneous measurement of both current and voltage on all phases. Calculations are made using dedicated Teseq software (WIN 2100) to determine harmonics (classes A-D),
Single phase version Three phase version Complete measurement system Use existing power sources or
mains
inter-harmonics, fl icker, dc, dt, dmax, Pst, Plt, inrush current and 24 x dmax.
Impedance
Measurement of fl icker requires a fi xed, stable source impedance as specifi ed in IEC 61000-3-3 (0.24 Ω + j0.15 Ω in the line and 0.16 Ω + j0.1 Ω in the neutral). When supplied as a complete system (see ProfLine 2103, 2105, 2115, 2130 or 2145) which include a precision AC source, the impedance is provided by a combination of a controlled source impedance and physical impedances supplied as part of the system. Users of the ProfLine 2100 measurement system must take care to provide the necessary source imped­ance. Hardware lumped impedances are available from Teseq, see options list.
PROFLINE 2100 - OVERVIEW
HARMONICS AND FLICKER MEASURING SYSTEM
Technical information
ProfLine 2100-1 ProfLine 2100-3
System contents CCN 1000-1 1-phase CCN 1000-3 1-phase coupling unit coupling unit DAQ card DAQ card Interface cable Interface cable WIN 2100 test software WIN 2100 test software
Coupling unit
Number of phases 1 3 Measurement channels 4 12 EUT connector: Front panel CEE 77 None EUT connector: Rear panel Terminal block Terminal block Max. voltage: Front panel 240 Vac N/A Max. current: Front panel 16 Arms N/A Max. voltage: Rear panel 300 Vac 480 Vac Max. current: Rear panel 40 Arms/Ph. 40 Arms/Ph.
(200 A Pk for 10 ms) (200 A Pk for 10 ms) Supply power: Voltage 115/230 Vac +/- 10% 115/230 Vac +/- 10% Supply power: Current <0.5 A <0.7 A Supply power: Frequency 50/60 Hz 50/60 Hz Dimensions (H x W x D) 89 x 427 x 560 mm 89 x 427 x 560 mm Weight 5 kg 5 kg
DAQ card and cable
Interface PCI PCI Size Standard height Standard height Resolution 16 bit 16 bit Speed 250 kSamples/s 1.25 MSamples/s Cable length PC to CCN 2 m 2 m
Teseq AG
Nordstrasse 11F 4542 Luterbach Switzerland T + 41 32 681 40 4 0 F + 41 32 681 40 4 8 sales@teseq.com www.teseq.com
© Februa ry 2011 Teseq® Specifications subject to change without notice.
®
Tes eq
is an ISO-registered company. Its products are designed and manufactured under the strict quality and environmental requirements of the ISO
9001. This document has be en c arefully checked. However, Teseq errors or inaccuracies.
691-262A Fe bruary 2011
®
does not assume any liability for
Options
INA 2151, 1-phase IEC 61000-3-3 16 Amp reference impedance INA 2154, 3-phase IEC 61000-3-3 16 Amp per phase reference impedance CCN 1000-1L, 1-phase coupling unit with integrated IEC 61000-3-3 16 Amp impedance INA 2188, Rack mounting kit for CCN 1000-1, CCN 1000-1L or CCN 1000-3 WIN 2106, Test software for IEC 61000-3-11 and IEC 61000-3-12
Loading...