AST AST5100 Datasheet

TM
Low Pressure Differential Transmitter
AST5100
The AST5100 Wet - Wet Differential Pressure transmitter is your accurate pressure sensing device for low differential pressure. With a differential pressure range as
low as 0 to 5” water column (12.5mbar), this product can be used to measure ow across an orice, differential across a lter, tank level, or gauge pressure. Using
LVDT technology and AST’s advanced electronics, the AST5100 delivers accurate, repeatable measurements.
Applications
Liquid Level Control including Bubbler
systems
Climate Control
Energy Management Air-fuel Ratio including Measurement for
Benefits
Accurate Low Pressure Measurement
Excellent Repeatability
Wide Range of Liquids and Gases including: Water, Natural Gas,
Hydrocarbon Fuels, Air and Non-Corrosive Gases
Vapor Recovery
Leak Detection Air or liquid Filtration Flow Measurement
Wetted Materials
Nickel Alloy 52, Ni-Span C, Viton, 304 Stainless Steel,
Aluminum 6061, RoHS Solder, Loctite 680 (meets NSF61)
Environmental Data
Temperature
Operating -40 to 80°C (-40 to 176°F )
Storage -40 to 100°C (-40 to 212°F)
Thermal Limits
Compensated Range 0 to 55°C (32 to 131°F)
TC Zero <±1.5% of FS
TC Span <±1.5% of FS
Electrical Data
Output
Excitation 10-28VDC 10-28VDC
Output Change with Input Voltage Change
Current Consumption: - < 10mA
Bandwidth 5Hz 5Hz
Output Noise: < 0.0035mA, RMS < 1mV, RMS
Zero Offset: < ± 1% FS < ± 1% FS
Span Tolerance: < ± 1.5% FS < ± 1.5% FS
Output Load: 0-800 Ohms@10-28 VDC 5k Ohms, min.
Reverse Polarity Yes Yes
Installation Guidelines
The AST5100 must be mounted on a at surface within ± 15° to the ideal 0° plane to maintain specications. Do not Overtighten the pressure connections or insert any ob-
jects in P1 or P2 to avoid damaging the sensing element. When using isolation valves, both should be mounted close to the sensor. For liquid level and wet applications, install bleed screw adapters close to P1 and P2 so that trapped air can be purged if needed. For optimum performance, always make sure pressure is equalized within the pressure range chart ranges. The AST5100 has asymmetric protection on P1 and P2.
Performance @ 25°C (77°F)
Accuracy <±1.0% of FS
Stability (1 year) ±0.5%FS, typ
Burst Pressure 2000 PSI
Pressure Cycles >100,000 Cycles
* Accuracy includes non-linearity, hysteresis & non-repeatability
4-20mA 0-5V Three Wire
- <0.1% from 10 to 32 VDC
American Sensor Technologies 450 Clark Dr., Mt. Olive, NJ 07828 phone (973) 448-1901 fax (973) 448-1905 email: info@astsensors.com
TM
Ordering Information
4
E
Series Type
Process Connection
J= 1/8” NPT Female
Pressure Range
See Chart
Pressure Unit
H= Inches H2O P= PSI
Outputs
2= 0-5V 3-wire 4 = 4 - 2 0 m A ( 2 w i r e l o o p p o w e r e d )
Electrical (for wiring information visit: http://www.astsensors.com/wiring.php)
Y= M12x1
Wetted Material
5= Nickel Alloy 52, Ni-Span C, Viton, 304 Stainless Steel, Aluminum 6061, RoHSSolder, Loctite 680 (meets NSF61)
Options
000= No Options
Mating PUR 22 AWG Cable Assembly
Part Number Cable Length
A10089 4 feet (1 m)
A10090 10 feet (3 m)
Pins
Pin 1 Brown +V +V
Pin 2 White N/C N/C
Pin 3 Blue -V -V
Pin 4 Black V Out N/C
Conductor
Colors
0-5V
3-wire
4-20mA
The over-pre ssure specication is the maximum pressure the AST5100 can
see without damage. Any pressure appli ed over the listed numb ers will likely
damage the s ensor and will, at minimum, cause a permanent zero shift.
Over-pressur e between 2X span and the numbers listed applied to port P1
will likely cause no permanent harm. Over-pressure of between 2X span and
the numbers listed applied to port P2 may cause a temporary zero shift. To
recover from a zero shift caused by negative over-pressure to P2 within the
listed limits, apply a positive over-pressure P1 to just under the listed limit
for a durati on of 5 minutes. Remove the over-pressure and check the zero
with no pres sure applied. If the zero has not recovered, repeat the positive
over-pressur e and recheck zero. If it has not recovered after the second try,
the zero has been permanently shifted. Contact the factory.
5YH00050JAST5100
000
Pressure Ranges
Differential
Pressure
0-5 inch H
0-10 inch H2O (25 mbar) 00010H 5 PSI 3 PSI
0-20 inch H2O (50 mbar) 00020H 8 PSI 5 PSI
0-50 inch H2O (125.5 mbar) 00050H 15 PSI 10 PSI
0-100 inch H2O (249 mbar) 00100H 35 PSI 25 PSI
0-200 inch H2O (498 mbar) 00200H 35 PSI 25 PSI
0-15 PSID (1034 mbar) 00015P 75 PSI 50 PSI
O (12.5 mbar) 00005H 5 PSI 3 PSI
2
Pressure
Code
Proof
Pressure
(P1>P2)
Pressure
(P2>P1)
www.astsensors.com
Proof
Dimensional Data
© 2014 American Sensor Technologies Inc. AST5100 03/13/2014.
All specications subject to change without notice.
Loading...