Submersible Liquid Level Sensors
AST4500 AST4510
The AST4500 and AST4510 submersible liquid level sensors are approved to UL/cUL913 (CSA 157)
Class I Div 1, Groups C and D for use in intrinsically safe areas with an approved barrier. It
is also certied for ATEX Class I Zone 0 Exia IIB T4 Ga (Ta = -40°C to +80°C). For pressure ranges
from 0-1 to 0-100 PSI that require a wide range of media compatibility, the submersible series is an
excellent solution to level monitoring for indoor and outdoor applications.
The AST4500 and AST4510 level sensors are completely sealed for submersion, yet vented through
the cable to correct for barometric pressure changes. The welded housing is tested in-house via a
helium leak tester to ensure proper protection. The conductors of the cable are also isolated from the
outside environment to keep the sensor operational for long-term use.
With a removable nose cone, the AST4500 and AST4510 series can be also be installed outside of the
tank through a 1/4” NPT pipe connection. In this conguration, the sensor continuously monitors the
tank level through a threaded connection outside the tank, yet remains fully submersible for applications
with ood prone environments or severe wash-down conditions. Available with voltage or 4-20mA output
signals, AST can provide a cost effective solution for level monitoring for a variety of applications.
For UL certied barrier
drawing, see A01657.
For CSA certied barrier
drawing, see A08949.
Environmental Data
Temperature
Operating -40 to 80°C (-40 to 176°F)
Storage -40 to 100°C (-40 to 212°F)
Thermal Limits
Compensated Range 0 to 55°C (32 to 131°F)
TC Zero <±1.5% of FS (<±2.5%, typ. for 1PSI)
TC Span <±1.5% of FS (<±2.5%, typ. for 1PSI)
Other
Shock 100G, 11 msec, 1/2 sine
Vibration 10G peak, 20 to 2000 Hz.
EMI/RFI Protection: Yes
Rating: IP-68
Electrical Data
Output 4-20mA 1-5VDC
Excitation 10-28VDC 10-28VDC
Output Impedance >10k Ohms <100 Ohms, Nominal
Current Consumption: 20mA, typical <10mA
Bandwidth (-3dB): DC to 250 Hz (-3dB): DC to 1kHz
Output Noise: - <2mV RMS
Zero Offset: <±1% of FS (<±4% 1PSI) <±1% of FS (<±4% 1PSI)
Span Tolerance: <±2% of FS (<±4% 1PSI) <±1.5% of FS (<±4% 1PSI)
Output Load: 0-800 Ohms@10-28VDC 10k Ohms, min
Reverse Polarity Protection Yes Yes
Benefits
High Strength Stainless Steel Construction
No Internal O-rings
Wide Operating Temperature
Pressures up to 100 PSI
Low Static and Thermal Errors
Unparalleled Price and Performance
New Conduit Fitting at Electrical Connection
Survives Harsh Environments
Compatible with Wide Variety of Liquids
EMI/RFI Protection
ABS (American Bureau of Shipping) Approved
Applications
Ground Water Level
Bio-Fuels
Salt Water Holding Tanks
Gasoline & Diesel Fuel Tanks
Fertilizer Tanks
Earthen & Concrete Dams
Irrigation Equipment
Ballast Tanks
Oil Tanks
Waste Water Canals
Performance @ 25°C (77°F)
Accuracy* < ±0.25% BFSL (<±0.5% BFSL for 0-1 PSI)
Stability (1 year) ±0.25% FS, typical
Over Range Protection 2X Rated Pressure
Burst Pressure 5X or 1,250 PSI (whichever is less)
Pressure Cycles > 50 Million
American Sensor Technologies • 450 Clark Dr., Mt. Olive, NJ 07828 • phone (973) 448-1901 • fax (973) 448-1905 • email: info@astsensors.com

Ordering Information
AST4510 L 00005 P 4 1
Series Type
Process Connection
L= Cone
Pressure Measurement
Insert 5-digit pressure code
Pressure Unit
B= Bar
H= Inches H
O
2
K= kg/cm
P= PSI
Outputs
3= 1-5V
4= 4-20mA (2 wire loop powered)
Electrical
N= Conduit tting, Cable 6 ft.
P= Conduit tting, Cable 10 ft.
X= Optional Length (see options)
Wetted Material
1= 316L / 304 / Hytrel Cable / Kynar Cord Grip
2
N
000
-SS
Options Cable Lengths:
140= 15 ft. (4.6 m)
075= 20 ft. (6.1 m)
074= 25 ft. (7.6 m)
004= 35 ft. (10.7 m)
130= 40 ft. (12.2 m)
065= 50 ft. (15.2 m)
003= 100 ft. (30.5 m)
050= 150 ft. (45.7 m)
Approval
(Left Blank)= UL ANSI/ISA 12.12.01 Class I Div 1 Intrinsically Safe Groups C, D (formerly UL913)
-SS= Add “-SS” for CSA157 Class I Div 1 Groups C, D Intrinsically Safe and ANSI/ISA 12.27.01 Single Seal Approval and SIRA ATEX Exia IIB
Class I, Zone 0, T4
Note: CSA approved products require case/earth ground electrical connection.
See wiring installation sheet for further details
Pressure Ranges
PSIG Pressure Code Feet of Water Column @ 4ºC (approx.)
0-100 00100 230.67
0-50 00050 115.33
AST4500
AST4510
+
All pressures between 0-1 PSI and 0-100 PSI are available. Please consult factory. *2.5 and 7.5 PSI Sensor must be ordered in inches of H2O.
0-30 00030 69.20
0-20 00020 46.13
0-15 00015 34.60
0-10 00010 23.07
0-7.5* 00208* 17.30
0-5 00005 11.53
0-2.5* 00069* 5.77
0-1 00001 2.31
+
www.astsensors.com
© 2014 American Sensor Technologies Inc. AST4500/4510 06/18/2014.
All specications subject to change without notice.