HYPOT®III SPECIFICATIONS
INPUT
VOLTAGE 115/230 V selectable
± 10% variation
FREQUENCY 50 - 60 Hz ± 5%
FUSE
115 VAC, 230 VAC - 3.15 A fast acting 250 VAC
DIELECTRIC WITHSTAND TEST MODE
OUTPUT Rating: AC 0 - 5000 V, 20 mA
DC 0 - 6000 V, 7.5 mA
Regulation: ± (1% of output + 5 V)
VOLTAGE SETTING Rating: 0 V - Max output rating, 10 V/step
Accuracy: ± (2% of setting + 5 V)
(relative to displayed output)
Can be adjusted during operation via
UP and DOWN arrow keys.
OUTPUT AC models: 50/60 Hz selectable
FREQUENCY AC & DC models: DC and 50/60 Hz selectable
WAVE FORM Sine wave Distortion: < 2% THD
RIPPLE ≤ 5% at 6 KV DC/7.5 mA, Resistive Load
DWELL TIME AC: 0 and 0.3 - 999.9 secs, 0.1 sec/step
SETTING DC: 0 and 0.4 - 999.9 secs
(0 = Constant)
RAMP TIMER Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec
DC 1.0 - 999.9 sec 0=OFF
Resolution: 0.1 sec
FAILURE AC Max. Limit: 0.00 - 20.00 mA, 0.01 mA/step
SETTINGS Min. Limit: 0.0 - 9.999 mA,
0.001 mA/step (0 = OFF)
Accuracy: ± (2% of setting + 2.0 mA)
DC Max. Limit: 0.00 - 7.50 mA, 0.001 mA/step
Min. Limit: 0.0 - 999.9 µA,
0.0001 mA/step (0 = OFF)
Accuracy: ± (2% of setting + 2 mA)
DISCHARGE TIME ≤ 200 ms The maximum capacitive load vs.
output voltage:
0.20 µF < 1 KV 0.050 µF < 4 KV
0.10 µF < 2 KV 0.040 µF < 5 KV
0.06 µF < 3 KV 0.015 µF < 6 KV
VOLTAGE DISPLAY Range: 0.00 - 5.00 KV AC
0.00 - 6.00 KV DC
Resolution: 0.01 KV
Accuracy: ± (2% of reading + 0.01 KV)
CURRENT DISPLAY Auto Range
AC Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 20.00
Resolution: 0.01 mA
DC Range 1: 0.0 µA - 350.0 µA
Resolution: 0.1 µA
Range 2: 0.300 mA - 3.500 mA
Resolution: 0.001 mA
Range 3: 3.00 mA - 7.50 mA
Resolution: 0.01 mA
ACCURACY All Ranges: ± (2% of reading + 2 counts)
TIMER DISPLAY Range: 0.0 - 999.9 secs
Resolution: 0.1 sec
Accuracy: (0.1% of reading + 0.05 secs)
GROUND Current: DC 0.1 A ± 0.01 A, fixed
CONTINUITY CHECK Resistance: 0.01 Ω - to 1.50 Ω
GROUND GFI Trip Current: 450 µA max (AC or DC)
FAULT INTERRUPT HV Shut Down Speed: < 1 ms
INSULATION RESISTANCE TEST MODE
(Model 3670)
OUTPUT VOLTAGE Range: 30 - 1000 V DC
Resolution: 1 V
Accuracy: ± (2% of reading + 5 V)
RESISTANCE Range: 1 - 9999 MΩ (4 Digit, Auto Ranging)
DISPLAY Resolution: 500 V DC - 100 V DC
MΩ MΩ
0.001 1.000 - 9.999
0.01 10.00 - 99.99
0.1 100.0 - 999.9
1 1000 - 9999
Accuracy: ± (2% of reading + 2 counts)
at test voltage 500 - 1000 V
and 1 - 999.9 MΩ
± (5% of reading + 2 counts)
at test voltage 500 - 1000 V
and 1000 Ω - 9999 MΩ
± (8% of reading + 2 counts)
at test voltage 30 - 500 V
and 1 - 1000 MΩ
IR VOLTAGE Range: 0 - 1000 V
DISPLAY Resolution: 1 V/step
Accuracy: ± (2% of reading + 2 V)
MAXIMUM
Range: 0,1 - 9999 MΩ (0 = OFF)
RESISTANCE LIMIT
MINIMUM
Range: 1 - 9999 MΩ
RESISTANCE LIMIT
IR DELAY Range: 0, 0.5 - 999.9 sec (0 = Constant)
SETTING Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
GROUND GFI Trip Current: 450 µA max
FAULT INTERRUPT HV Shut Down Speed: < 1 ms
GENERAL SPECIFICATIONS
REMOTE CONTROL The following input and output signals are
& SIGNAL OUTPUT provided through two 9 pin D type connectors:
1. Test, Reset, and Remote Interlock
2. Remote recall of memory program
#1, #2 & #3
3. Outputs: Pass, Fail, Test-in-Process, & Reset.
PROGRAM MEMORY 10 Sets 3 steps per set-up with ability to link
tests in any order.
SECURITY Lockout capability to avoid unauthorized access
to test set-up program.
DISPLAY 128 x 64 dot resolution with front panel
contrast setting.
BUZZER Alarm with volume control.
LINE CORD Detachable 7 ft. (2.13 m) power cable
terminated in a three prong grounding plug.
TERMINATIONS Detachable 5ft. (1.52 m) high voltage and return
leads (2) with clips and a standard U.S. style
(NEMA 5 - 15) remote receptacle box for
testing items terminated with a line cord.
International receptacles also available. Front
and Rear outputs standard.
MECHANICAL Tilt up front feet. 2U 1/2 rack.
Dimensions: 8.5 x 3.5 x 14.5 inches
(215 x 89 x 370 mm)
Weight: approx. 22 lbs. (10 kgs)
ENVIRONMENTAL Operating Temperature: 32° - 104° F (0 - 40° C)
Relative Humidity: 0 to 80%
CALIBRATION Traceable to National Institute of Standards and
Technology (NIST). Calibration controlled by
software. Adjustments are made through front
panel keypad in a restricted access calibration
mode. Calibration information stored in nonvolatile memory.
Models 3605 AC, 3665 AC/DC and 3670 AC/DC/IR