Associated Research 3670 Data Sheet

Page 1
Dielectric Withstand and Ground Bond Testers
For Safety Agency Production Line Compliance Testing
Safety agency listed.
U.S. patent#: 5,548,501, 6,054,865. Other patents pending.
Page 2
Clear test parameters can be displayed
on a single screen
Configurable test result screen can be
set up to show detailed test results
A COMPLETE MID-RANGE SYSTEM TO MEET THE MOST COMMON
ELECTRICAL SAFETY COMPLIANCE TESTS
With its 2U 1/2 rack mount cabinet design the Hypot®III family of products can be interconnected to the HYAMP®III, model 3130 Ground Bond tester that is also designed in a 2U 1/2 rack mount style cabinet. These two instruments can then be rack mounted side by side in a single 19” 2U rack space. The interconnect capability of these two instruments
also provides a single DUT connection and automates the test sequence or sequences chosen by the operator such as
AC Hipot, DC Hipot, IR test and Ground Bond test.
INTERCONNECT CAPABILITIES
ENHANCED GRAPHIC LCD
The enhanced graphic liquid crystal display on the Hypot®III and the HYAMP®III provides the operator with complete test
setup and results in an easy-to-use interface. The graphic LCD
makes testing safer, easier and more reliable than ever before.
(Above display is shown at actual size.)
Page 3
USER-FRIENDLY FEATURES
NEW TO HYPOT®III
FEATURE
Exclusive SmartGFI™function
BENEFIT
The patent pending SmartGFI provides maximum operator protection to the user. If the circuit detects excessive leakage to the ground it shuts down the high voltage in less than 1 millisecond. SmartGFI is automatically activated if the DUT is not grounded so that the operator does not need to make the decision whether to activate the SmartGFI.
FEATURE
Max output current 20 mA AC and 7.5 mA DC
BENEFIT
Hypot®III is a true hipot tester with enough output current to test capacitive loads in AC mode and allows the instrument to comply with the UL “120 K ohm” requirements.
FEATURE
Switchable Floating and Grounded Return
BENEFIT
This feature is included within the software of the instrument. The instrument automatically detects if the DUT is floating or connected to Earth Ground. If the DUT is grounded the smart GFI circuit turns off the GFI detector. If the DUT is floating the GFI circuit is enabled.
FEATURE
Digitally controlled arc detection system
BENEFIT
The arc detection system allows the operator to select whether low-level arcs should be detected and provides the operator with the ability to digitally select and program multiple sensitivity levels.
FEATURE
Built-in adjustable Continuity test mode
BENEFIT
Hypot®III meets ground continuity test requirements called out by UL and other safety agencies.
FEATURE
Electronic ramping (up & down)
BENEFIT
Electronic ramping provides a gradual and timed method to increase or decrease output voltage to the DUT effectively minimizing any damage from quick high voltage changes to sensitive DUTs.
FEATURE
Resistance Range of 1-9999 M
BENEFIT
Model 3670 AC/DC/IR includes an increased maximum resistance limit, which will allow for increased resistance measurement of the insulation of a DUT.
• Tamper Proof Front Panel Controls To Limit User Access
• User Selectable Output Voltage & Frequency
• Complete User Configurable System Capability
• Built-In Security Settings
• Multiple Test Memories & Steps
• Electronic Dwell Setting
• No Load Setup Of Voltage Current
• Minimum & Maximum Trip Settings
NEW TO HYPOT®III & HYAMP®III
FEATURE
Enhanced graphic LCD
BENEFIT
The enhanced graphic LCD provides the operator with an advanced user-interface. This backlit graphic display makes viewing and interpreting test results and data easier than ever before. The LCD can display text and symbols eliminating the need to decipher abbreviations.
FEATURE
Built-in VERI-CHEK™self verification system
BENEFIT
Regular verification is required by some agencies to validate that the instrument is functioning correctly. The VERI-CHEK feature, by prompting the user through the correct steps, allows for quick and easy validation. This built-in feature eliminates the additional cost of having to purchase an external box for verification.
FEATURE
Remote Safety Interlock
BENEFIT
The Remote Safety Interlock prevents the high voltage from being activated without the interlock enabled. The interlock works as a safety feature by using a set of contacts to enable or disable the instrument’s high voltage output. If the interlock contacts open during a test the high voltage output is shut down.
FEATURE
Built-in CAL-ALERT™function
BENEFIT
This feature automatically alerts the user when the instrument is due for re-calibration. This eliminates the need for manual tracking of calibration dates.
Page 4
HYPOT®III SPECIFICATIONS
INPUT
VOLTAGE 115/230 V selectable
± 10% variation
FREQUENCY 50 - 60 Hz ± 5%
FUSE
115 VAC, 230 VAC - 3.15 A fast acting 250 VAC
DIELECTRIC WITHSTAND TEST MODE
OUTPUT Rating: AC 0 - 5000 V, 20 mA
DC 0 - 6000 V, 7.5 mA
Regulation: ± (1% of output + 5 V)
VOLTAGE SETTING Rating: 0 V - Max output rating, 10 V/step
Accuracy: ± (2% of setting + 5 V)
(relative to displayed output) Can be adjusted during operation via UP and DOWN arrow keys.
OUTPUT AC models: 50/60 Hz selectable FREQUENCY AC & DC models: DC and 50/60 Hz selectable
WAVE FORM Sine wave Distortion: < 2% THD
RIPPLE 5% at 6 KV DC/7.5 mA, Resistive Load
DWELL TIME AC: 0 and 0.3 - 999.9 secs, 0.1 sec/step SETTING DC: 0 and 0.4 - 999.9 secs
(0 = Constant)
RAMP TIMER Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec
DC 1.0 - 999.9 sec 0=OFF
Resolution: 0.1 sec
FAILURE AC Max. Limit: 0.00 - 20.00 mA, 0.01 mA/step SETTINGS Min. Limit: 0.0 - 9.999 mA,
0.001 mA/step (0 = OFF)
Accuracy: ± (2% of setting + 2.0 mA)
DC Max. Limit: 0.00 - 7.50 mA, 0.001 mA/step
Min. Limit: 0.0 - 999.9 µA,
0.0001 mA/step (0 = OFF)
Accuracy: ± (2% of setting + 2 mA)
DISCHARGE TIME 200 ms The maximum capacitive load vs.
output voltage:
0.20 µF < 1 KV 0.050 µF < 4 KV
0.10 µF < 2 KV 0.040 µF < 5 KV
0.06 µF < 3 KV 0.015 µF < 6 KV
VOLTAGE DISPLAY Range: 0.00 - 5.00 KV AC
0.00 - 6.00 KV DC Resolution: 0.01 KV Accuracy: ± (2% of reading + 0.01 KV)
CURRENT DISPLAY Auto Range
AC Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 20.00 Resolution: 0.01 mA
DC Range 1: 0.0 µA - 350.0 µA
Resolution: 0.1 µA
Range 2: 0.300 mA - 3.500 mA Resolution: 0.001 mA
Range 3: 3.00 mA - 7.50 mA Resolution: 0.01 mA
ACCURACY All Ranges: ± (2% of reading + 2 counts)
TIMER DISPLAY Range: 0.0 - 999.9 secs
Resolution: 0.1 sec Accuracy: (0.1% of reading + 0.05 secs)
GROUND Current: DC 0.1 A ± 0.01 A, fixed CONTINUITY CHECK Resistance: 0.01 - to 1.50
GROUND GFI Trip Current: 450 µA max (AC or DC) FAULT INTERRUPT HV Shut Down Speed: < 1 ms
INSULATION RESISTANCE TEST MODE
(Model 3670)
OUTPUT VOLTAGE Range: 30 - 1000 V DC
Resolution: 1 V Accuracy: ± (2% of reading + 5 V)
RESISTANCE Range: 1 - 9999 M(4 Digit, Auto Ranging) DISPLAY Resolution: 500 V DC - 100 V DC
M M
0.001 1.000 - 9.999
0.01 10.00 - 99.99
0.1 100.0 - 999.9 1 1000 - 9999 Accuracy: ± (2% of reading + 2 counts)
at test voltage 500 - 1000 V and 1 - 999.9 M
± (5% of reading + 2 counts) at test voltage 500 - 1000 V and 1000 - 9999 M
± (8% of reading + 2 counts) at test voltage 30 - 500 V and 1 - 1000 M
IR VOLTAGE Range: 0 - 1000 V DISPLAY Resolution: 1 V/step
Accuracy: ± (2% of reading + 2 V)
MAXIMUM
Range: 0,1 - 9999 M(0 = OFF)
RESISTANCE LIMIT
MINIMUM
Range: 1 - 9999 M
RESISTANCE LIMIT
IR DELAY Range: 0, 0.5 - 999.9 sec (0 = Constant) SETTING Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
GROUND GFI Trip Current: 450 µA max FAULT INTERRUPT HV Shut Down Speed: < 1 ms
GENERAL SPECIFICATIONS
REMOTE CONTROL The following input and output signals are & SIGNAL OUTPUT provided through two 9 pin D type connectors:
1. Test, Reset, and Remote Interlock
2. Remote recall of memory program #1, #2 & #3
3. Outputs: Pass, Fail, Test-in-Process, & Reset.
PROGRAM MEMORY 10 Sets 3 steps per set-up with ability to link
tests in any order.
SECURITY Lockout capability to avoid unauthorized access
to test set-up program.
DISPLAY 128 x 64 dot resolution with front panel
contrast setting.
BUZZER Alarm with volume control.
LINE CORD Detachable 7 ft. (2.13 m) power cable
terminated in a three prong grounding plug.
TERMINATIONS Detachable 5ft. (1.52 m) high voltage and return
leads (2) with clips and a standard U.S. style (NEMA 5 - 15) remote receptacle box for testing items terminated with a line cord. International receptacles also available. Front and Rear outputs standard.
MECHANICAL Tilt up front feet. 2U 1/2 rack.
Dimensions: 8.5 x 3.5 x 14.5 inches
(215 x 89 x 370 mm)
Weight: approx. 22 lbs. (10 kgs)
ENVIRONMENTAL Operating Temperature: 32° - 104° F (0 - 40° C)
Relative Humidity: 0 to 80%
CALIBRATION Traceable to National Institute of Standards and
Technology (NIST). Calibration controlled by software. Adjustments are made through front panel keypad in a restricted access calibration mode. Calibration information stored in non­volatile memory.
Models 3605 AC, 3665 AC/DC and 3670 AC/DC/IR
Page 5
HYAMP®III SPECIFICATIONS
GROUND BOND TEST MODE
INPUT VOLTAGE 115/230V selectable, ± 10% variation
50/60 Hz, ± 5%
FUSE 6.3 A fast blow 250 VAC
OUTPUT Current: AC 1 - 30 A, 0.01 A/step
Regulation: ± (2% of setting + 0.02 A) Voltage: AC 6 V fixed
OUTPUT FREQUENCY
50/60 Hz selectable
DWELL TIME 0 and 0.5 - 999.9 secs, 0.1 sec/step SETTING 0 for continuous running
Accuracy: ± (0.1% of setting + 0.05 secs)
FAILURE Max. limit:
0 - 120 mfor 1 - 30 A, 1 mΩ/step
SETTINGS
0 - 510 mfor 1 - 10 A, 1 mΩ/step
Accuracy: ± (2% of setting + 2 mΩ)
Min. limit:
0 - 120 mfor 1 - 30 A, 1 m/step 0 - 510 mfor 1 - 10 A, 1 mΩ/step
Accuracy: ± (2% of setting + 2 mΩ)
MILLIOHM Offset Capability: 0 - 100 m,1 m/step OFFSET Accuracy: ± (2% of setting + 2 mΩ)
METERING Ammeter (3 digits)
Range: 0 - 30 A Resolution: 0.01 A/step Accuracy: ± (3% of reading + 0.03 A)
Ohmmeter (3 digits)
Range: 0 - 510 m Resolution: 1 mΩ/step Accuracy: ± (2% of reading + 2 mΩ)
Timer (4 digits)
Range: 0 - 999.9 secs Resolution: 0.1 secs/step Accuracy: ± (0.1% of reading + 0.05 secs)
GENERAL SPECIFICATIONS
VERIFICATION Built-in software driven verification SYSTEM menu to test fault detection circuits.
REMOTE CONTROL The following input and output signals & SIGNAL OUTPUT are provide through two 9 pin D type
connectors;
1. Remote control: Test, Reset, Interlock and Withstand Processing
2. Outputs: Pass, Fail, Test in Process, Start Out, and Reset Out
PROGRAM MEMORY 10 Memories, 3 Steps per memory.
SECURITY Key Lock capability to avoid unauthorized
access to all test parameters. Memory Lock capability to avoid unauthorized access to Memory locations.
LINE CORD Detachable 7 ft. (2.13m) power cable
terminated in a three prong grounding plug.
DISPLAY 128 x 64 dot resolution with front panel
contrast setting.
ALARM VOLUME Front panel adjustable volume setting with SETTING 10 set points.
CALIBRATION Traceable to National Institute of Standards
and Technology (NIST). Calibration controlled by software. Adjustments are made through front panel keypad in a restricted access calibration mode. Calibration information stored in non­volatile memory.
MECHANICAL Tilt up front feet. 2U 1/2 rack.
Dimensions: 8.5 x 3.5 x 14.5 inches
(215 x 89 x 370 mm)
Weight: 19.65 lbs. (8.9 Kgs)
ENVIRONMENTAL Operating Temperature: (0˚ - 40˚ C)
Relative Humidity - 0 to 80%
Unless otherwise stated, accuracies are relative to a laboratory standard measurement.
ACCESSORIES
ACCESSORIES:
Rack Mount Kit Connects the instruments & provides the necessary flanges for mounting the system into a 19” rack. 38321
Resistor Kit Precision 1% resistor is ideal as a load to set current trip points. Meets 120 k ohm requirements. 35534
Black Stationary Probe Provides an insulated fixed metallic contact point to a DUT for a return path. 35539
Foot Switch Ideal for applications where an operator needs to start a test “hands off” the instrument. 35822
Retractable Probe High voltage retractable 5 KV AC probe for safe testing & application of high voltage. 38081
Retractable Return Probe This retractable probe is used on the return side of the Hypot
®
III. 38082
Safe-T-Probe
®
Test gun with trigger that controls retractable probe & activates instruments high voltage circuit. 38083
Adjustable Resistor Bank Provides test loads from 120 K to 2.148 k ohms. 6 terminals with 12 settings. Helps verify regulation. 36956
Interface Kit Includes cables & adapter box so multiple tests can be performed through a single DUT. 260-01
Model 3130 Ground Bond Tester
Page 6
13860 West Laurel Dr.
Lake Forest, IL U.S.A. 60045-4546
Tel: 847-367-4077 • Fax: 847-367-4080
E-mail: info@asresearch.com
For more information visit our web site at:
www.asresearch.com
and ORDER ON-LINE!
Or contact us toll-free at:
1-800-858-TEST(8378
)
30 Amp Ground Bond tester that works as a stand-alone instrument or can be interconnected to the HypotPLUS®II to form a semi-automated test system with a single DUT connection.
The first semi-automated microprocessor controlled Dielectric Withstand testers available in AC or AC/DC versions. All models include enhanced PLC control, remote memory recall, advanced failure detection systems. Available optional 10 V analog signal and real current.
The 905D functional run test system powers and runs a product while measuring current, voltage, power factor and watts. It can be interconnected to our safety testers so all tests can be performed through a single DUT connection. Available with standard GPIB or RS-232 interface.
4-in-1 Electrical Safety Compliance Analyzers include AC/DC Hipot, Insulation Resistance, Ground Bond tester and optional built-in switching matrix in a single instrument. Complete with IEEE-488 (GPIB), RS-232 or printer interface. Available in 500 VA output versions.
Designed to automate line leakage testing in production line or lab environments. The 510L is a stand-alone system and the 520L can interconnect with other AR safety testers to form a complete automated testing system.
The HS-8A is an 8 channel and the HS-16 is a 16 channel matrix. Both models are high voltage and high current matrices for multi-point or multi­product testing. (For use with QUADCHEK
®
II,
HypotULTRA®III or OMNIA™)
SWITCHING MATRIX
Software designed to store, analyze and retrieve data on automated Associated Research instruments, while performing Line Leakage, Insulation Resistance, Dielectric Withstand, Ground Bond and Functional Run tests. Autoware also allows for bar coding inputs and provides basic statistical analysis graphs.
The first fully-automated, multi-function Electrical Safety Compliance Analyzer with an enhanced graphic LCD. OMNIA provides 4-in-1, 5-in-1 and 6-in-1 models that include AC Hipot, DC Hipot, Insulation Resistance, Ground Bond, Functional Run and Line Leakage testing. All models available with IEEE-488 (GPIB), RS-232 or printer interface.
The Industry’s Most Complete Line of Instruments
for Electrical Safety Compliance Testing
Catalog # HYPOT III 1/03
State-Of-The-Art Electrical Safety Compliance Testing Instruments
24-Hour Turn-Around On All Instrument Calibrations
Informative Industry Seminars, Professional Training And Educational Programs
Knowledgeable Customer Support And Expert Technical Services At 1-800-858-TEST
Guaranteed Product Satisfaction And A 5-Year Extended Product Warranty*
*With Annual Calibration