Ashcroft ST-2A Specifications

Type ST-2A Digital Indicator
FEATURES
• Measurement of pressure, tempera­ture, voltage and current
• Accuracy ratings of ±0.1, 0.05% and
0.025% F.S. (pressure)
• Pressure ranges from 0.25 inches of water to 10,000 psi
• Interchangeable pressure and temper-
ature modules
• Multiple engineering units
• Temperature measurement with most common RTDs and thermocouples
• Programmable damping
• Tare capability
• Display hold
• RS232 two-way communications
• Standard N.I.S.T.-traceable certificate of calibration
The ST-2A is the perfect bench compan-
ion product to the Ashcroft
®
ATE-100 field handheld calibrator. This bench top (or panel mounting) package shares the same pressure and temperature modules and interfaces with the same software packages as the Ashcroft ATE-100. An intuitive menu- driven user interface puts all of the ST-2A’s power at the simple press of a key. It uses the AQS (Ashcroft Quick-Select™) modular sensor system to provide the ultimate in measurement flexibility.
Now you can purchase a calibration system to serve your field and bench mounting requirements while enjoying the significant cost savings gained through the availability of common pressure mod­ules and software.
SPECIFICATIONS
Standard Display: Alphanumeric LCD, 0.37-inch
height, 2 lines, 16 characters/line
Optional Display: Backlit LCD alphanumeric with 2
lines and 16 characters/line
Display Resolution: ±0.002% of span, 60,000
Display Update Rate: 130ms (nominal) with one
sensor installed
Engineering Units: psi, in.H
mbar, kPa, MPa, mmHg, cmH2O, mmH
2
kgcm
and any single user-programmable
engineering unit
Damping: Programmable averaging from zero
through 16 consecutive readings
Standard Operating Range: 32° to 120°F (0 to
49°C)
Compensated Range: 20° to 120°F (–7° to +49°C) Reference Temperature: 70 ±3°F Standard Temperature Effect: ±0.004% of span per
degree Fahrenheit over the compensated range
Optional: AQS-2 Quick-Select modules are available
calibrated to maintain rated accuracy over the 20° to 120°F (–7° to +49°C) compensated temperature range.
Storage Limits: –4°F to +158°F (–20° to +70°C) Electrical Sourcing Specification: 24 Vdc at 25mA Electrical Measurement Specification:
Input (volts) 0/10 Vdc ±0.025% F.S. 0/30 Vdc ±0.10% F.S. Input (mA) 0/20mA ±0.03% F.S
Temperature Effects Electrical Measurement:
±0.001% of span per °F over the compensated range
Weight: ST-2A Base unit: 3.0 lb (1.4 kg) Housing: Molded, high-impact ABS case Electrical Connections: Standard banana jacks Optional Battery Power Supply: 5 “AA” nicads with
built-in charger or 5 “AA” alkaline (non­rechargeable)
External Power Supply: AC adaptor 9 Vdc, 500mA Portable Operation: 20 hours with optional back-
light off, 2 hours with backlight on
Warm-up: 5 minutes for rated accuracy (max-
imum), 30 minutes for complete stability
RS232 Serial Interface: with 9-pin D type at 300,
1200, 2400, 9600 baud.
Field Calibration: Calibration module and proper
pressure and electrical standards are required.
Optional Data Logging Capacity:
Standard measurements: Date/time stamped measurements: Provides prompts for storage of calibration history information, including instrument tags, As Found/As Left data and corresponding ambient temperature
O, in.Hg, ftSW, bar,
2
Accuracy
Accuracy
714 records
O and
2
384 records
BULLETIN ST-2A
All specifications are subje ct to change without notice. All sales subject to standard terms and conditions. © Ashcro ft Inc. 2008 11/08
Ashcroft Inc., 250 Eas t Main Street, Stratfo rd, CT 06614 USA Tel: 203-378-8281 • Fax: 203-385-0408 email: info@ ashc roft .com • www.ashcroft.com
0.93
0.97
1.44
3.47
4.00
9.24
0.56
6.74
6.33
3.31
1.511.25
1.10
Type ST-2A Digital Indicator
STANDARD RANGES
AQS-2 AQS-1
psi ˝H2O
(gauge & (gauge/ Other Engineering Units**
absolute differential
pressure) pressure)
100 150 200 250 300 500
600 1000 1500 2000 2500 3000 5000 6000 7500
10,000
vacuum
compound
±10
±15
–15/+30 –15/+60
* Non-isolated, for clean dry gas only
5
0.25*
10
0.5*
15
1.0*
20
2.0*
25
3.0*
30
5.0*
50
10*
60
15* 25*
50* 100* 150* 200*
**Note: Engineering units
psi
in. H2O
in. Hg
ftSW
Bar
mBar
kPa
mPa mmHg cmH2O
mmH2O
2
kg/cm
User Selectable
identified above are accessible through the unit select feature. How­ever, readout will default to the primary unit of measure on start-up. Sensor modules scaled in primary units other than
O (AQS-1) or psi
in. H
2
(AQS-2) are also available. Consult factory.
5 10 15
±0.125*
±5
±0.25* ±0.5* ±1.0* ±1.5* ±2.5* ±5.0*
±7.5* ±12.5* ±25* ±50* ±75*
±100*
BULLETIN ST-2A