SL17 Submersible Pressure Transmitter
APPLICATIONS
Because of its small 17mm diameter
and rugged 316L SS construction, the
Ashcroft SL17 submersible pressure
transmitter is ideal for level applications, particularly in situations where
size or chemical compatibilities are an
issue. The SL17 utilizes piezo-resistive
technology (PRT) that is known for its
proven reliability and stability. Notably,
these transmitters can offer all ranges
equivalent from 1.5 psig through 300
psig while all units of measure and
cable lengths are customer-defined for
individualized specifications.
FEATURES
• Complete 316L SS Construction
• IP68 / NEMA 6P Housing
• Excellent Long-Term Stability
• 4-20 mA Output (2-wire)
• Unit of Measure Selection (Psig, Feet
or Meters of H20)
• Optional NIST Traceable Calibration
Chart (9-point)
TYPICAL APPLICATIONS
• Level (Groundwater / Bore)
• Wastewater
• Tank Level Measurement
PERFORMANCE SPECIFICATIONS
Reference Condition: 21°C (70°F)
Humidity: 0-100%RH*
Accuracy: ±0.25% of span, terminal point (>5 psi);
±0.5% of span, terminal point (<5 psi);
Includes effects of linearity, hysteresis
and repeatability
Stability: ±0.25% FS/year
Engineering Units: psi, bar, ftH20, mH20 and customer
defined
Ranges: 1.5-300 psi & equivalent (see table); gauge
pressure only
* Based on cable vented to clean, dry, non-corrosive atmosphere
or proper use of termination desiccant box
ENVIRONMENTAL SPECIFICATIONS
Temperature Limits
Compensated: –10/40°C (14/104°F)
Operating: –10/40°C (14/104°F)
Storage: –20/60°C (14/140°F)
Temperature Effects (–10 to 40°C): ±1% TEB
(>5 psi) ±1.5% span, terminal point (<5 psi)
†
For non-frozen liquid applications only
FUNCTIONALAL SPECIFICATIONS
Overpressure (FS): Proof Burst
1.5-5 psi 300% 500%
10 psi 300% 750%
15-30 psi 200% 500%
75 psi 187% 470%
150 psi 200% 400%
300 psi 200% 300%
Maximum Turndown: 2:1 from original sensor
pressure range
Vibration & Shock: IEC 60068-2-6
‡
Based on sensor range
†
‡
ELECTRICAL SPECIFICATIONS
Output Signal: 4-20mA (2 wire)
Supply Voltage: 10-36Vdc
Insulation Resistance: 50Vdc (100Mohm)
Withstand Voltage: 350Vdc
CE Compliance:
• EMI Class A/EMS Table 2 EN61326-1:2006
• Annex BB (Pressure Transducer)
EN61326-2-3:2007
• ESD EN61000-4-2
• Burst EN61000-4-4
• Surge EN61000-4-5
MECHANICAL SPECIFICATIONS
Pressure Connection: M14 × 1.5
Weight: Sensor (120g)
Nose Cones: standard (13g), weighted (225g)
Enclosure: 316L SS
Labels: Laser-etched with customer part number
Cable External Jacket: Coated (.03˝ ) black polyurethane
Cable Internal Construction: 4-wire 22 AWG conduc-
tors (red, black, green, white)
Vent tube: 0.060˝ ID polyethylene
Semi-rigid PVC with shield and Kevlar strength member
Cable Pull Strength: 180 lb
Cable Voltage Rating: 300V
Load Limitation: 4-20mA Output Only
Wetted Materials: 316L SS and polyurethane
BULLETIN SL17-1
LOOK FOR THIS
AGENCY
All specifications are subject to change without notice.
All sales subject to standard terms and conditions.
© Ashcroft Inc. 2017 07/17
Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
Tel: 203-378-8281 • Fax: 203-385-0408
email: info@ashcroft.com • www.ashcroft.com

Dimensions
Model SL17 Submersible Transducer
VENT TUBE
1.5
TYP.
Ø 0.23
[5.7]
L2
SECTION
A
A
A-A
RED :SUPPLY (+)
BLACK: SUPPLY (-)
L3
L1
D1
[3]
4 x
Ø 0.12
NOTE-1
L4
D1
[6]
0.24
M14 x 1.5
6g
WITHOUT
NOSE FITTING
NOSE FITTING
Without Nose Cone
(SNC) Standard Nose Cone
(P5C) 0.5LBS Nose Cone
L3
L1
DIMENSIONS,
D1
0.67 (17)
0.67 (17)
0.67 (17)
5.90 (150)
6.37 (162)
10.98 (279)
Inches (Millimeter)
L1
L2
N/A
0.35 (9)
4.96 (126)
L2
Per Order
L4
0.47 (12)
0.94 (24)
5.55 (141)
How To Order
Model Type
(SL17)
BULLETIN SL17-1
3
Accuracy Class
(3)
(5)*
*1.5˝ range only
SNCSL17
Nose Fitting
(SNC) Standard NC
(P5C) Weighted NC
T
Joint Type
(T) Threaded
(W) Welded
42
Output Signal
(42)4-20mA
*Available in meters only
†
Available in feet only
‡
Custom cable lengths available,
example for 35 feet, code as (035).
030
Cable Length‡
(001)*
(002)*
(005)
(010)
(015)
(020)
(025)
(030)
(050)
(060)
(075)
(100)
(120)
(150)
(200)
(300)†
(400)†
(500)
(700)†
Measurement
†
All specifications are subject to change without notice.
All sales subject to standard terms and conditions.
© Ashcroft Inc. 2017 07/17
F
of Cable
(F) Feet
(M) Meters
(P) Polyurethane
psi
(1P5#) 1.5
(5#) 5
(10#) 10
(15#) 15
(20#) 20
(30#) 30
(50#) 50
(75#) 75
(100#) 100
(150#) 150
(200#) 200
(300#) 300
P
Cable Type
30#
Pressure Range
Ft H2O
(10FW) 10´H2O
(20FW) 20´H2O
(30FW) 30´H2O
(50FW) 50´H2O
(100FW) 100´H2O
(150FW) 150´H2O
(200FW) 200´H2O
(300FW) 300´H2O
(400FW) 400´H2O
(500FW) 500´H2O
—
Calibration Chart
Mt H2O
(5LW) 5 m H2O
(10LW) 10 m H2O
(15LW) 15 m H2O
(20LW) 20 m H2O
(30LW) 30 m H2O
(50LW) 50 m H2O
(75LW) 75 m H2O
(100LW) 100 m H2O
(125LW) 150 m H2O
(150LW) 200 m H2O
Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
Tel: 203-378-8281 • Fax: 203-385-0408
email: info@ashcroft.com • www.ashcroft.com
XC4
Variations
(XC4)
(0.6BR) 0.6 Bar
(1BR) 1 Bar
(1.6BR) 1.6 Bar
(2.5BR) 2.5 Bar
(4BR) 4 Bar
(6BR) 66 Bar
(10BR) 10 Bar
Bar