Ashcroft PT-1 Operating Manual

Page 1
AL
Page 2
2
Thank you for your purchase. Please feel free to contact us if
we may be of assistance in any way.
from the “service” button on the Ashcroft website. Please
Ashcroft Inc.
Tel: (203) 378-8281
For the latest return form and product
Page 3
TABLE OF CONTENTS
Section Description Page
SAFETY PRECAUTIONS
....................................................................................
1.0 General Product Overview
................................................................................
....................................................................
1.2 Optional Product Features
...........................................................................
2.0 Theory of Operation
..........................................................................................
Theory of Operation – Base (Display) Unit
.................................................
2.2
Theory of Operation – Low Pressure Sensor Assemblies
2.3
Theory of Operation – Medium and High Pressure
2.4
Temperature Measurement with Standard RTD Probes
...............................
7
3.0 Unpacking
...........................................................................................................
7
4.0 System Start-Up
5.0 Shift Functions (Shift Key)
................................................................................
6.0 The Enter Function (ENT Key)
7.0 The Port Select Function (PORT Key)
7.1
.....
8.0 Zeroing the Instrument
Zeroing All Installed Pressure Sensor Assemblies On Power UP
...............
8.2
Zeroing All Installed Pressure Sensor Assemblies In Normal Operation
....
8.3
Zeroing Either of Two Installed Pressure Sensor Assemblies
......................
9.0 Selecting Engineering Units
Changing the Engineering Unit On Indicators with One Installed
Pressure Sensor............................................................................................. 12
9.2
Simultaneously Changing the Engineering Unit for Two
Installed Pressure Sensor Assemblies........................................................... 12
9.3
Assemblies.................................................................................................... 13
10.0 The Tare Function
Using the Tare Function On An Indicator with a Single Installed
Pressure Sensor............................................................................................. 14
10.2
Using the Tare Function to Simultaneously Tare Two
Installed Pressure Sensors............................................................................ 14
10.3
Taring Either of Two Installed Pressure Sensor Assemblies
........................
11.0 Damping Function
..............................................................................................
.............................
12.0 Max/Min Tracking Function (Max. & Min. Pressure Measurement Tracking)
Recalling and Maintaining Stored Max/Min Values
....................................
12.2
Recalling Stored Max/Min Values and Erasing After Review
.....................
13.0 RS232 Output Function
RS 232 Connector
........................................................................................
13.1.1
.............................................................
13.1.2
The Remote Protocol
................................................................................
13.1.3
The Print Protocol
.....................................................................................
13.2
.....................................................................
13.3
.........................................................................
Page 4
4
Section Description Page
14.0 The Back Light Function (Optional)
.................................................................
21
14.1
Timed Back Light Function
.........................................................................
14.2
.................................................................
15.0 Battery Check Function
15.1
Using Battery Check Function
.....................................................................
15.2
Low Battery Indicator
..................................................................................
16.0 The Hold Function
..............................................................................................
22
16.1 Using The Hold Function
....................................................................................
17.0 The Print Function
.............................................................................................
23
17.1
Using The Print Key
.....................................................................................
18.0 Temperature Measurement Using RTD probes
23
18.1 Setting Up the System for Temperature Measurement
...............................
18.2 Calibrating/Programming the RTD Interface Assemblies
...........................
19.0 Calibrating the Indicator
25
19.1 Adjusting Calibration Zero and Span Settings
............................................
19.2 Performing A Complete Recalibration
........................................................
19.2.1 Terminal Requirements for Performing A Complete Recalibration
.......
19.2.2 Performing A Recalibration/ Relinearization of the Indicator
................
19.2.3 Remote Zero Adjustment
........................................................................
19.2.4 Remote Full Scale Adjustment
...............................................................
19.2.5 Remote Adjustment of Linearization Values
..........................................
20.0 Installing a Second Pressure or Changing Sensor Assembly
32
20.1 Procedure for Installing or Changing a Pressure Sensor Assembly
................
21.0 Replacement of Optional Cadmium Batteries
.................................................
33
21.1 Procedure for Installing Replacement Nickel Cadmium Batteries
..................
A-1.0 Introduction – Remote Communication Protocol
...............................................
35
A-2.0 Notation Conventions
A-3.0 Numeric Format
.....................................................................................................
A-4.0 Command Reference
A-5.0 Command Library Summary
Appendix A Indicator Serial Port Functions Version 1.0
..............................................
Appendix B End of Message Characters
Appendix C Product Specifi cations
...............................................................................
41
Appendix D Remote Communications Protocol Errata Sheet
Appendix E. RTD Probe Connector
................................................................................
44
Page 5
SAFETY PRECAUTIONS
A few general guidelines for the selection of pressure
measurement instrumentation follow:
– Select a range so that the maximum applied
pressure will never exceed the upper range limit.
Vibration
– Excessive vibration could cause loosening
– Excessive pressure pulsation could result in
fatigue failure of the pressure element.
Temperature
– Operation of the instrument in an envi-
ronment where temperatures are in excess of design
ratings may result in loss of accuracy and failure.
– Some of the sensor types available for the
indicator are designed for use on clean dry air. Use
liquids, wet gases or corrosive gases will reduce system
– Only approved explosion proof
Instruments should
not be used in locations where EMI/RFI conditions may
Electronic pressure instruments must be selected in
ASME B40.100, that apply to elastic pressure elements
ASME B40.100 is available from:
ASME
New York, NY 10017
provides a wide array of pressure measurement and test
nish individuals responsible for pressure measurement
the requirements of both fi eld and lab type applications.
This digital indicator incorporates the latest in pressure
measurement technology. Each instrument is comprised
installed pressure measurement assemblies.
The base display electronics provide:
1) The user interface, via a combination of a liquid
2) Interface connections for two pressure measurement
types and ranges available.
3) Digital interface capability via the full function RS232
interface.
Pressure sensor assemblies are available in two basic confi gu-
rations. These include:
1) Specialized low pressure sensing units based on micro-
machined variable capacitance sensors. These sensor
water through 200 inches of water in gauge, differential
2) Piezo resistive strain gauge based sensor assemblies cov-
The instruments are powered via an AC adapter provided
which each unit. An optional built-in nickel cadmium
rechargeable power pack is available for application in which
AC power is not available.
The instruments are housed in a custom manufactured, injec-
tion molded case. The case is constructed of high impact ABS
plastic. This material is non-conductive and is resistant to
both chipping and cracking. Panel and bench mount confi gu-
rations are available.
Each
Indicator is a fully equipped pressure measurement
instrument. Standard product features include the ability to:
ment display.
Zero:
Zero displayed pressure measurement value(s) at the
push of a button.
programmed engineering units. Dual sensor units can simul-
taneously display both measurement values in independent
Activate any of 3 factory pro-
Tare
Tare the displayed measurement for either one of two
installed sensors or both sensors simultaneously.
Display minimum and maxi-
mum pressure value(s) at the push of a button.
RS232 interface can be confi gured for
transmission of measurement data as well as set up the instru-
ment confi guration over the RS232 interface.
SECTION 1.0 GENERAL PRODUCT OVERVIEW
1.1 Product Capabilities Overview
Page 6
Freeze the measurement values on the dis-
play at the push of a button.
Print the pressure measurement value(s) at
the push of a key over the standard RS232 interface.
Additional capabilities for instruments equipped with the
Display the measured values
Dis-
play measurement data from two sensors in the same or
pressure by automatically subtracting the measurement
value of one sensor from the other.
Two Sensor Addition
two installed sensors. Allows unit with single absolute
pressure sensor assembly to provide measurement data for
both sensors in terms of absolute pressure.
1.2 Optional Product Features
As previously outlined, the pressure indicator is a fully
product features are available to enhance the capabilities of
the unit. These are:
1)
bly can be specifi ed as an option. Any combination of
2)
play can be provided with back-lighting to enhance
readability in low light and high glare areas.
3)
without use of the back lighting option or for up to 5
hours with continuous use of back lighting.
The indicator incorporates the latest in electronic, micro-
processor and sensor technologies. This combination pro-
The instrument includes one or, optionally, two calibrated
ics. All pressure measurement calibration data is stored in
EEPROM (Electrically Erasable Programmable Read Only
Memory) located within the pressure sensor assembly. The
pressure sensor assembly produces an output from 0-1 Vdc
for pressure inputs from the minimum to the maximum
measurement rating of the pressure sensor assembly. The
front end electronics of the main unit “read” the voltage
the EEPROM in the sensor assembly for the correction fac-
tors required to convert the voltage measurement into the
tors in the EEPROM correct for non-linearity and thermal
In addition, the main unit provides the application oriented
user interface to simplify the use of indicator. All higher
level functions, such as: unit select, tare, max/min recall,
push to print and so forth are provided for by the capabili-
ties of the microprocessor in the main unit.
The inches of water pressure measurement assemblies pro-
vide specialized low pressure measurement capabilities.
These pressure measurement assemblies incorporate a
micro-machined silicon variable capacitance sensor. Inside
the sensor there is a micro-machined silicon diaphragm.
This diaphragm is between two non-moving plates on
which metal has been sputtered. The air between the dia-
phragm and the non-moving plates acts as an insulator.
As pressure or vacuum is applied to the sensor, the dia-
phragm moves changing the distance between the dia-
phragm and the fi xed plates. This change in distance
vacuum during the calibration process.
The sensor is connected to an Application Specifi c Inte-
ramp and applies this signal to the top plate of the sensor
while a signal of equal intensity but with opposite polarity
is applied to the bottom plate.
When the measured pressure is balanced, for example
when both ports are opened to atmosphere, the distance
between the diaphragm and both of the fi xed plates is
the same. When this is case, the signal to the top plate
is capacitively coupled to the equal and opposite signal
will fl ow through the center plate (diaphragm). When the
pressure or vacuum, the excess current fl ows through the
into a voltage which is scaled over the full scale range of
the sensor assembly.
Low pressure sensing assemblies incorporating the micro-
machined silicon variable capacitance sensor are available
in ranges from 0.25 inches of water through 200 inches
2.2 Theory of Operation – Low Pressure Sensor Assemblies (Inches of Water Column)
SECTION 2.0 THEORY OF OPERATION
2.1 Theory of Operation – Base (Display) Unit
Page 7
The RT2 assembly is supplied factory programmed to
RT assemblies can be programmed with coeffi cients for
programmed into the RT assembly to provide enhanced
ferent programmed RTD calibration curves. Programming
package ( part number 838X014-01 for the 3 1/2 inch disk
Prior to removing the indicator from the packaging mate-
rial inspect all cartons for shipping damage. Document
must be made against the shipper. After inspection, remove
the instrument, AC adapter, manual and any accessories
purchased from the packaging material. Retain the packag-
ing for use in returning the indicator to the factory for
future recalibration or repair.
Locate, unpack and untie the AC adapter.
Looking at the back of the indicator, locate the AC
back of the instrument) of the rear panel.
Plug the AC adapter jack into the back of the
indicator.
Plug the transformer into an outlet providing the
provided on the label of the AC adapter.
Press the Power key on the far left side of the row
the following start up screens as it performs the initial start
up self tests.
Mfg Date xx/xx/xx
range range
where:
xxxxx = the full scale range of the sensor assembly
installed in the corresponding side of the
indicator
the available low pressure sensor assemblies is provided in
Appendix C- Product Specifi cations.
2.3 Theory of Operation-Medium and High Pressure
The PSI pressure measurement assemblies incorporate a
piezoresistive strain gauge sensor. This technology takes
the fl exing of a diaphragm under pressure or vacuum)
the resistive properties of a piece of silicon will change.
In the manufacturing process resistors are deposited in a
fashion, the output will be near zero when no pressure
with the application of pressure or vacuum. The reverse
When power is applied to the sensor the level of resistance
to the level of pressure or vacuum applied. The output
from the sensor is extremely repeatable and has minimal
hysteresis due to the fact that the sensor substrate is silicon.
The sensor output is then amplifi ed by circuitry within the
pressure sensor assembly to a level of 0-1 Vdc for the full
mable Read Only Memory) within the pressure sensor
Pressure sensor assemblies for PSI measurement ranges
These pressure sensor assemblies can be confi gured to
provide the capability to measure gauge, absolute and com-
pound pressures as well as vacuum. PSI sensor assemblies
in ranges up to 300 psi are designed for use on clean,
0/10 through 0/300 psi. Ranges from 500 psi through
The RTD interface assembly allows the base unit, when
used with RTD temperature probes, to provide precision
temperature measurement data. The interface assembly is
into the Switchcraft TA4F connector on the module. The
RT1 assembly comes factory programmed with the curves
for Pt 100 (385 & 392), Cu 10 and Ni 120 RTD probes. It
(Single Port Sensor Assemblies)
SECTION 3.0 UNPACKING
SECTION 4.0 SYSTEM START-UP
2.4 Temperature Measurement with Standard RTD Probes
Page 8
EU = the primary or default engineering unit for
the sensor assembly installed in the cor-
responding side of the indicator.
ing unit data will appear centered on the display during
normal instrument pressure measurement operations.
2-
If your instrument is equipped with the optional
rechargeable nickel cadmium batteries the battery pack
was charged prior to shipment. While in storage, nickel
mended that the nickel cadmium batteries be recharged
prior to use of the indicator. When the indicator is
ies will automatically recharge. While connected to line
power, the batteries will be recharged whether the instru-
ment is in use or not. Battery recharge time is 12-14
hours from a completely discharged state to a fully charged
indefi nitely without damaging the batteries.
SECTION 5.0 SHIFT FUNCTIONS (Shift Key)
There are two classes of functions accessible through the
front panel of the indicator. These are primary functions
providing access to two capabilities. Secondary functions
the key that will activate the desired secondary function.
In all cases, the primary function of a key is the function
performed when the key is pressed without fi rst pressing
the shift key. Single function keys and the primary position
required digital indicator functions. Functions are accessed
by pressing the key or combination of keys indicated below.
The shift key need only be pressed once during a given
required, to highlight the desired set up confi guration. The
indicator will stay in the shift mode until the desired selec-
tion is made through the use of the enter key.
Power Turn unit on POWER
Zero Zero sensor(s) ZERO
functions
Unit Select engineering UNIT/PORT
unit(s)
Port Select Select port SHIFT then
Tare Tare pressure TARE/DAMP
preload
Damp Set Level of SHIFT then
Damping TARE/DAMP
Max/Min Recall Max/Min MAX/MIN
values OUTPUT
Backlight Lighting LITE
Display display
Battery Checking battery SHIFT then
Enter accepting set ENT
up option high-
lighted on display
Freezing Locking measured HOLD
Display values on display
Push To Print Printing measured PRINT
values on demand
protocol and the indicator must be confi gured in the
measurement data for the push to print function to oper-
included in Section 13.3 Set Up for Print Protocol.
There are various set up functions which require a selec-
tion from a menu of available performance parameters.
These set up functions include selection of the desired:
Engineering Unit(s)
Port or Display Confi guration
Level of Damping
The ENT key is used to select the highlighted confi gura-
tion or operating parameter. Proper use of the ENT key is
included, as required, in each of the following sections.
As an optional feature the indicator can be equipped with
two pressure sensor assemblies.
SECTION 6.0 THE ENTER FUNCTION (ENT Key)
SECTION 7.0 THE PORT SELECT FUNCTION (PORT Key)
Page 9
In units with the dual installed sensor option the UNIT/
PORT key allows the operator to confi gure the instrument
both sensors. For use of the PORT function in indicators
with two installed sensors proceed to Section 7.1 Setting
Instrument Confi guration with the Port key.
The standard indicator is supplied with a single pressure
pressed followed by pressing the PORT key the indicator
will display:
module required
After displaying the above message the indicator will auto-
matically resume pressure measurement.
7.1 Setting Instrument Confi guration with the
This section only applies to indicators equipped with the
Press the SHIFT key one time. After pressing
the shift key an upward pointing arrow will appear on the
^
Press the PORT key. If the indicator has the
the PORT key will result in the following message display:
L-R R-L L+R
The current confi guration will be the default con-
Press the PORT key the number of times required
to activate the desired measurement display confi guration.
The SHIFT key should not be pressed after the port selec-
tion menu has been activated. Each time the PORT key is
pressed the next port confi guration will be highlighted on
the display and will fl ash.
Left: activates the left sensor and side of the indicator
Right: activates the right sensor and side of the indicator
Both: activates both sensors and both sides of the dis-
play.
L-R:
produces display of the differential of the left
minus the right measured values.
R-L: produces display of the differential of the right
minus the left measured values.
R+L produces the display of the sum of the measure-
ments of the two installed sensors.
With the desired display confi guration fl ashing
press the enter key (ENT). After pressing the enter key the
indicator will begin displaying the measurement values in
the desired confi guration.
The left side of the display presents the pressure
measurement data generated from the pressure sensor
from the sensor assembly installed in the right side of the
indicator.
Upon instrument start up it is possible that gauge, compound,
may display a slight zero offset. In most cases this offset
will diminish or even disappear as the instrument warms
up.
zeroed prior to use or calibration.
Upon power up, abso-
lute pressure sensors will display measured values repre-
pressure sensor assemblies can be accomplished by adher-
ing to the methods detailed in the subsequent sections.
must be vented to atmosphere prior to initiating the zero
process. Be certain any pressure connections to the sensor
to be zeroed are vented to atmosphere prior to zeroing the
pressure sensor assembly.
Absolute sensors must be connected to an evacuated refer-
blies. A vacuum pump and a vacuum gauge are required to
4%
the following error message will appear.
too much
This protects from accidental zeroing of absolute pressure
is appropriately vented to atmosphere (for gauge, vacuum,
have been damaged due to exposure to improper media
brated, if possible, or returned to the factory for evaluation
The zero procedures for the different display confi gura-
Port Select Key
(PORT Key)
SECTION 8.0 ZEROING THE INSTRUMENT
Page 10
tions of the indicator appear in the sections as shown
below. In order for these procedures to provide the desired
results the operator must have already set up the instru-
ment sensor assemblies to be zeroed in accordance with
the instructions outlined in the appropriate subsection of
8.1 For zeroing any installed sensors on power
up.
8.2 Zeroing any installed sensor assemblies in
normal operation.
8.3 Zeroing either one of two installed sensors
in indicator.
zero the indicator.
values from both of the installed assemblies are dis-
8.1 Zeroing All Installed Pressure Sensor
When initially turned on the indicator will commence mea-
blies. The sensor assembly is installed in the left side of
the instrument if one sensor assembly is ordered. As such,
pressure measurement data, the display will look as follows.
PSI
If the indicator was purchased with the optional second
pressure sensor assembly the display, on power up, will
look as follows.
+12345 +12345
The indicator will always commence pressure mea-
pressure measurement assembly. Default engineering units
pressure port) and inches of water column for low pressure
To zero all installed pressure sensor assemblies
press the ZERO key and hold for 2 seconds after the
initial start-up sequence is completed. As shown below,
the indicator will respond by zeroing out any displayed
pressure measurement.
00000 00000
This section only applies to indicators equipped with the
As outlined in Section 8.0, the measured values of a
two installed sensors has been de-activated the deactivated
for the zero procedure to work for that sensor.
The following procedure details the method used to restore
both installed sensor assemblies to the display for zeroing.
Press the SHIFT key one time. After pressing the
^
Press the PORT key. If the indica-tor has the
the PORT key will result in the following message display.
Left Right Both
The current confi guration will be the default con-
the indicator is only displaying the measurement data from
the pressure sensor assembly installed in the left side of
the indicator.
Press the PORT key the number of times required
to highlight the word ‘‘both’’ in the confi guration menu. In
this example the PORT key would be pressed two times.
The word both will fl ash when highlighted. Each time the
PORT key is pressed the next display confi guration will be
highlighted on the display and will fl ash.
L-R R-L L+R
With the word both fl ashing on the display press
the enter key (ENT).After pressing the enter key the indicator
will begin displaying the measurement for both installed
PSI PSI
To zero both installed pressure sensor assemblies,
press the ZERO key and hold for 2 seconds while the
measured outputs from both sensors are shown on the
installed sensors shown on the display will simultaneously
Assemblies On Power UP
8.2 Zeroing All Installed Pressure Sensor Assemblies In Normal Operation
Page 11
0.0000 0.0000
After the zero procedure has been completed the indicator
will automatically resume pressure measurement activities.
8.3 Zeroing Either of Two Installed Pressure
The following section applies only to indicators equipped
with the optional second pressure sensor assembly.
The zero function will only zero the measurement values
that are shown on the display. If the objective is to zero one
measured values for the sensor assembly not being zeroed.
The following steps detail the methodology for zeroing one
Determine which sensor assembly is to be zeroed.
The measured value from a sensor assembly always
installed.
Press the Shift key to allow access to the port
indicator will respond by adding an upward pointing arrow
to the display as follows.
^ PSI PSI
Press the PORT key to activate the port select
function. Pressing the PORT key will result in the follow-
ing confi guration options being displayed on the indicator.
Left Right Both
values from both installed pressure sensor assem-blies the
word “both” will fl ash on the display.
With the port select confi guration options dis-
played, press the port key the number of times required to
leave only the measured output for the sensor assembly to
be zeroed visible on the display. For example, if the left
not to be zeroed press the PORT key four (4) times, the
number of times necessary to highlight the word left from a
L-R R-L L+R
With the appropriate selection high-lighted and
The indicator will respond by changing the display so that
the indicator is displayed.
+12345 --------
With the measurement data for the pressure
for 2 seconds.
The indicator will respond by zeroing the displayed pres-
PSI--------
To return the measurement values for the sensor
installed in the right side of the instrument case to the
by adding the upward pointing arrow to the display as
follows.
^ PSI ---------
With the shift function active, as shown by the
L-R R-L L+R
Press the PORT key the number of times required
to highlight the word “both”. When highlighted the word
Left Right Both
With the word “both” fl ashing press the enter
key (ENT). The indicator will respond by displaying the
measured values from both installed sensor assemblies.
The measured value from the left sensor will be zeroed
the previous zero offset.
PSI PSI
The digital pressure indicator provides the ability to chose
with the zero function, the engineering unit selection func-
tion will only effect the measured values shown on the
blies as well as for the selection of different engineering
units for each installed pressure sensor assembly.
Sensor Assemblies
SECTION 9.0 SELECTING ENGINEERING UNITS
Page 12
9.1 Changing the Engineering Unit On Indicators
The following procedure should be followed for changing
the engineering unit of measure in indicators that have one
installed pressure sensor assembly.
With the indicator on and displaying pressure
measurement data press the key labeled PORT/UNIT.
The indicator will respond by displaying the factory pro-
Bar mBar kPa mPa
the second screen accessed after scrolling through the fi rst
mmHg cmWC mmWC kgcm
where:
Unit Description
psi pounds per square inch
˝ Hg inches of mercury (@ 0
˝ W Cinches of water (@ 20
ftSW feet of seawater Bar Bar
mBar millibar
kPa kilopascals
mPa megapascal
mmHg millimeters of mercury (@ 0
mmWC centimeters of Water (@ 20
The engineering unit library is automatically presented
with the current engineering unit highlighted (fl ashing) on
the display of the indicator. In this example the engineering
unit currently in use is psi.
To select the desired engineering unit, repeatedly
press the UNIT key to scroll through the on-board engi-
neering unit library. In this example, we will change from
psi to inches of water by pressing the unit key three times
to highlight the text “˝ WC”.
Bar mBar kPa mPa
2nd screen
mmHg cmWC mmWC kgcm
With the text ˝ WC fl ashing press the enter key.
The indicator will respond by resuming the function of
pressure measurement in the selected engineering unit.
+12345
This section applies only to indicators equipped with the
If the measured values from both installed pressure sensor
If either of the installed sensor assemblies has been deacti-
vated (removed from the display via the port select func-
tion) it must be restored prior to initiating the change
restore both measured values to the display and simultane-
With the indicator on and displaying pressure
measurement data, press the SHIFT key to activate the
indicator will respond by adding an upward pointing arrow
to the data currently seen on the display.
+xxxxxx
With the upward pointing arrow visible on the
the port select function. Pressing the Port key will result
in the following confi guration options being displayed on
the indicator.
L-R R-L L+R
The current confi guration in use will automatically fl ash
when the port select menu is activated. In this example, the
measurement values from the left pressure sensor are being
PORT key the required number of times to highlight the
word both from the list of port select options.
For this example the PORT key would be pressed two
times to highlight the word “both”.
Left Right Both
With the word "both" highlighted and fl ashing
press the enter key (ENT). The indicator will respond by
rently in use. In this case, the measurement values from
both of the installed pressure sensor assemblies are being
+12345 +12345
with One Installed Pressure Sensor
9.2 Simultaneously Changing the Engineering Unit for Two Installed Pressure Sensor Assemblies.
Page 13
To change the engineering unit for both the dis-
played values press the key designated UNIT with both
measured outputs shown on the display. The indicator will
respond by displaying the factory programmed engineering
unit library as follows.
Bar mBar kPa mPa
2nd screen
mmHg cmWC mmWC kgcm
where:
Unit Description
psi pounds per square inch
˝ Hg inches of mercury (@ 0
O
˝ WC inches of water (@ 20
O
C)
ftSW feet of seawaterBarBar
mBar millibar
kPa kilopascals
mPa megapascal
mmHg millimeters of mercury (@ 0
O
C)
O
C)
mmWC centimeters of Water (@ 20
O
C)
The unit currently in use will fl ash on the display. In this
With the engineering unit library activated select
the desired unit of measure by pressing the UNIT key the
number of times necessary to highlight the engineering
unit of choice. When highlighted the unit of measure will
the list will be highlighted.
For the example here we will select ˝ Hg as the engineering
unit of choice. The UNIT key would be pressed one time to
Bar mBar kPa mPa
2nd screen
mmHg cmWC mmWC kgcm
With the desired engineering unit highlighted, ˝
Hg in this example, and fl ashing, press the enter (ENT)
key. As shown, the indicator will respond by altering the
+12345 +12345
9.3 Changing the Engineering Unit for Either of
This section only applies to indicators equipped with the
With the indicator on and displaying pressure
measurement data, determine which sensor assembly is to
be changed to the new unit of measure.
The measured value from a sensor always appears on
the side of the display that corresponds to the side of the
instrument in which the sensor assembly is installed.
Press the SHIFT key to activate the shift function
respond by adding an upward pointing arrow to the data
+xxxxxx +xxxxxx
Press the PORT key to activate the port select
function. Pressing the PORT key will result in the follow-
ing confi guration options being displayed on the indicator.
L-R R-L L+R
In this example, since the measurement values from both
pressure sensor assemblies are actively being shown on the
port select menu is activated.
With the port select function activated, press the
PORT key the required number of times to deactivate (re-
move from the display) the measured output from the sensor
For example, if the left side of the display is to be changed
from psi to kPa and the right side is to remain in psi the
PORT key would be pressed three times (the number of
times necessary to change the highlighted word from the
word “both” to the word left from the list of port select
Left Right Both
L-R R-L L+R
With the appropriate selection highlighted and
The indicator will respond by displaying following.
psi -----
With the side of the display to be changed to the
new engineering unit active (shown on the display) and the
UNIT key. The indicator will respond by displaying
Two Installed Sensor Assemblies
Page 14
Bar mBar kPa mPa
2nd screen
mmHg cmWC mmWC kgcm
With the engineering unit library on the display
to be selected is highlighted.In this case pressing the UNIT
key 4 times will change the highlighted engineering unit
from psi to Bar.
Bar mBar kPa mPa
2nd screen
mmHg cmWC mmWC kgcm
With the desired engineering unit high-lighted
indicator will respond by converting the pressure displayed
+12345 -----
To restore the right side of the display to the
The indicator will respond by adding an upward pointing
^ Bar-----
Press the PORT key. The indicator will respond
by displaying.
Left Right Both
lighted and fl ashing when the port confi guration menu is
Press the PORT key the number of times required
to activate the desired display confi guration (2 times in this
Left Right Both
With the desired confi guration displayed press the
the following.
Bar psi
Using the engineering unit and the port select functions
in tandem any combination of available units and display
location can be arranged.
SECTION 10.0 THE TARE FUNCTION
Use of the tare function allows the indicator to be set up to
is often used for applications such as:
platform or container is to be subtracted from the measure-
ment
used for short term gauge pressure measurement activities.
As with the zero function, the tare function will only effect
the measured values shown on the display. To tare out a
pressure preload the pressure sensor measuring the preload
must be active (shown on the display). This allows for
well as for the taring of either one of two installed pressure
The full scale range of a sensor is not
For example, if a 100 psi sensor is tared with a pressure
preload of 15 psi the remaining usable measurement span
is 85 psi. With 15 psi tared out on a 100 psi sensor the
measured values will exceed the rated range of the sensor
when the measured pressure exceeds 110% of the sensor
measured value exceeds 95 psi (85 psi + 10% of the 100
psi range of the sensor which is 10 psi).
With the indicator on and displaying pressure
measurement data press the key labeled DAMP/TARE. The
indicator will respond by adding a small T to the display
to indicate that the instrument is in the tare mode. When
initially placed in the tare mode, the display of the indica-
tor will show a "zero" pressure measurement until the
measured value changes from the measurement value at the
time the tare function was activated.
+000000
To exit the tare mode press the TARE key again.
The indicator will resume the function of basic pressure
measurement (untared).
This section only applies to instruments with the optional
If the measured values from both installed pressure sensor
10.1 Using the Tare Function On An Indicator with a Single Installed Pressure Sensor
10.2 Using the Tare Function to Simultaneously Tare Two Installed Pressure Sensors
Page 15
TARE key one time will simultaneously tare both sides of
the pressure display.
If either of the installed sensor assemblies have been deac-
tivated (removed from the display via the port select func-
tion) it must be restored prior to initiating the tare function.
Follow the steps below to restore both measured values to
the display and simultaneously tare both sensor assemblies.
If both pressure sensor assemblies are active (measurement
values are visible on the display) skip to step 5 of this
procedure.
With the indicator on and displaying pressure
measurement data press the SHIFT key to activate the
indicator will respond by adding an upward pointing arrow
to the data currently seen on the display.
Press the PORT key to activate the port select
function. Pressing the PORT key will result in the follow-
ing confi guration options being displayed on the indicator.
Left Right Both
With the port select function activated press the
PORT key the required number of times to highlight the
word both from the list of port select options. Each time
the PORT key is pressed the next port select confi guration
will be highlighted and be available for selection.
Left Right Both
With the word "both" highlighted and fl ashing
press the enter key (ENT). The indicator will respond by
12345 12345
To tare both the displayed values press the TARE
key with both measured outputs shown on the display. The
will respond by displaying.
PSI
PSI
The T will appear to the upper left of the pressure
measurement being conducted in the tare mode for as long
To discontinue the tare mode of opera-tion press
the TARE key a second time. The tare function will be
the value of the pressure preload from the displayed pres-
+12345 +12345
10.3 Taring Either of Two Installed Pressure Sensor
This section only applies to indicators equipped with the
Determine which sensor assembly is to be tared.
The measured value from a sensor always appears
the instrument in which the sensor assembly is installed.
With the indicator on and displaying pressure
measurement data press the SHIFT key to activate the
functions. The indicator will respond by adding an
upward pointing arrow to the data currently seen on the
With the upward pointing arrow visible on the
function. Pressing the Port key will result in the following
L-R R-L L+R
With the port select function activated press the
PORT key the required number of times to leave activated
be tared press the PORT key the number of times neces-
Left Right Both
With the appropriate selection highlighted and
The indicator will respond by displaying following.
+12345 -----
With the measurement value to be tared on the
respond by taring out the displayed value. The tared value
will fl ash on the display.
+00000 -----
The T will appear to the left of the pressure mea-
the pressure measurement value from the right sensor can
be returned to the display untared through the following
process.
Press the SHIFT key to activate the shift function
Assemblies
Page 16
indicator will respond by adding an upward pointing arrow
to the data currently seen on the display.
+00000 -----
With the upward pointing arrow visible on the
function. Pressing the Port key will result in the following
this example, the word "left" will be highlighted and fl ash
the left pressure sensor assembly was active (measurement
values visible on the display)
L-R R-L L+R
With the port select function activated press the
PORT key the required number of times, 2 for this exam-
ple, to highlight the word “both”.
L-R R-L L+R
To activate (bring to the display) the measure-
ment values from both installed pressure sensor assem-
blies with the word “both” fl ashing, press the enter key
following information.
+00000 +12345
When exiting the tare mode, only the measured
values currently being tared can be on the display. There-
fore, to exit the tare mode, when one of two installed
that is not being tared must be disabled (removed from dis-
play) using the PORT function. When, only the measure-
ment value that is in the tare mode is active, press the
TARE key. As shown below, the indicator will disable the
tare function, including removing the T from upper left
hand corner of the display.
psi -----
After the tare function has been disabled the second sensor
SECTION 11.0 DAMPING FUNCTION
The damping function provides the ability to stabilize the
bility. Damping is a secondary function and, therefore,
requires the use of the shift key to set up and implement.
The level of damping is set up through a damping menu
TARE/DAMP key.
Damping accomplishes the stabilization of the pressure
measurement value by producing an averaged pressure
value which is shown on the display and available for
transmission over the RS232 interface. When damping is
not activated the indicator takes a new pressure value over
required, this same measurement data is also available for
transmission over the RS232 interface.
When damping is activated the measurement value dis-
played and/or transmitted is an average value. With damp-
ing invoked the displayed value will be the average mea-
readings.
The following are the available levels of damping.
Level Effect
New every 130 mS, not averaged
LOW Averages 4 consecutive readings
Represents readings over 0.4 seconds
MEDIUM Averages 8 consecutive readings
Represents readings over 0.8 seconds
HIGH Averages 16 consecutive readings
Represents readings over 1.6 seconds
Programming and initiating the damping function is
This note only applies to instruments equipped with
the optional 2nd pressure sensor assembly.
Damping is a universal function. This means that the level
Damping cannot be enabled for one of two installed pres-
Press the SHIFT key one time. The indicator will
respond by adding an upward pointing arrow to whatever
information is currently presented on the display of the
indicator. For example, if the indicator is a dual sensor
instrument with the measurement values of both sensors
when the shift function is activated.
+12345 +12345
11.1 Setting, Activating & Disabling The Damping Function
Page 17
Press the TARE/DAMP key to activate the damp-
ing menu. The indicator will respond by displaying the
programmable levels of damping allowed.
OFF LOW
where:
Low = averaging of 4 consecutive readings
Medium = averaging of 8 consecutive readings
High = averaging of 16 consecutive readings
When the damping menu is activated the current level of
Press the DAMP key, as required, to highlight the
was pressed 2 times to highlight the medium damping level.
OFF LOW
MEDIUM HIGH
With the desired level of damping highlighted
resume the function of pressure measurement with the
ment value. Damping is also applied to any pressure mea-
effect until damping is turned off as detailed in Step 5
To disable damping press the SHIFT key one
time. The indicator will respond by adding an upward
pointing arrow to the display as shown.
+12345 +12345
Press the DAMP key to activate the damping
menu. The indicator will respond by displaying the pro-
in effect, if any, will fl ash. In this case, medium damping
is in effect.
MEDIUM HIGH
Press the DAMP key, as required, to highlight the
was pressed 2 times to highlight off damping.
OFF LOW
MEDIUM HIGH
With the word "off" highlighted and fl ashing
press the enter (ENT) key. Damping has now been dis-
to any displayed pressure measurement value. In indicators
with two sensor assemblies the programmed level of damp-
ing will apply to each of the displayed installed sensor
SECTION 12.0 MAX/MIN TRACKING FUNCTION
When measuring pressure the indicator constantly tracks
the minimum and maximum pressure measurement values
for each active pressure port. At the end of each measure-
ment cycle, after the electrical measurement from the pres-
mal errors, the indicator compares the measured value
with those stored in both minimum and maximum memory
locations maintained for each active pressure port.
If the measured value is found to exceed the previous
maximum value the reading is stored in memory, over-
writing the last stored value. Likewise, if the value, when
it will be stored in the memory over-writing the previous
minimum value for the given pressure port.
The values stored in the maximum and minimum memory
locations for each installed sensor can be reviewed at any
time by pressing the MIN/MAX/OUTPUT key. The fol-
lowing sections detail the method for reviewing and dis-
the current maximum and minimum pressure values for
below assume the indicator has two installed pressure
pressure sensor assembly the measurement values for that
Press the MAX/MIN/OUTPUT key. The indicator
will respond by displaying the maximum pressure and
minimum pressures measured since the memory locations
were cleared for each active sensor. The display will look
+12345 +12345
If the indicator is equipped with only one pressure
is active (measured outputs appearing on the display) the
indicator will display the maximum and minimum values
for a single sensor assembly. If a pressure sensor assembly
is not active minimum and maximum pressure measure-
ment data are not tracked and will not be displayed when
the MAX/MIN key is pressed.
(Maximum &Minimum Pressure Measurement Tracking)
12.1 Recalling and Maintaining Stored Max/Min Values
Page 18
To return to pressure measurement
without
eras-
ing the values stored in the maximum/minimum memory
locations press any key other than the enter (ENT) key.
The indicator will respond by returning to the basic pres-
+12345 +12345
12.2 Recalling Stored Max/Min Values and
below assume the indicator has two installed pressure
pressure sensor assembly the measurement values for that
Press the MAX/MIN/OUTPUT key. The indicator
will respond by displaying the maximum pressure and
minimum pressures measured since the memory locations
were cleared for each active sensor. The display will look
+12345 +12345
To erase the stored maximum and minimum
values and return to basic pressure measurement press the
measurement
+12345 +12345
If the indicator is equipped with only one sensor or only
mum values for a single sensor assembly. If a pressure
pressure measurement data will not tracked or displayed
when the MAX/MIN key is pressed.
values stored in the max/min memory locations will be
lost when the instrument is turned off.
2
will not appear even if checked immediately after clear-
ing the max/min memory locations. These values are
reset to +999999 for the minimum value and -999999
for the maximum value when the memory location is
mS after the memory has been cleared will produce a
new value in these memory locations. Even with no
pressure applied to a gauge, compound, vacuum or dif-
ferential pressure sensor assembly the high resolution of
the indicator will display slight zero offsets as measur-
tor to a true pressure change, the averaging software
used for damping is bypassed for the capture of maxi-
mum and minimum pressure values.
The output key provides access to the RS232 set up menus.
fore, requires the use of the shift key to initiate. The cali-
brator supports 3 communications protocols (print, remote
4800 & 9600).
testing and calibration and is not documented for customer
use. The following sections detail the set up of the remote
When the RS232 interface is not in use it should be
particularly important for instruments being powered by
the optional built-in nickel cadmium batteries. Activating
the RS232 interface results in an approximately 20-25%
increase in power consumption. This increase in power
Following, is detailed information on the RS232 interface.
The interface as defi ned here is standard on all indicators.
Pin Function
2 Transmit data
3 Receive data
5 Ground
The RS232 requires a straight through cable type. This
will provide a 9 pin connector to the serial port on the
Pins
5 4 3 2
Pins
Erasing After Review
SECTION 13.0 RS232 OUTPUT FUNCTION
13.1 RS232 Connector
13.1.1 Connecting the RS232 Interface
Page 19
to select the same baud rate on the device as that to be
for the COM port on the computer, recorder or acquisition
For parity select none
For fl ow control select none
To summarize:
none
none
through
13.1.2 The Remote Protocol
The remote protocol allows the operator to query instru-
ment status and confi gure the indicator remotely over the
RS232 interface. When set up in the remote protocol mode
the operator can query and/or change the following instru-
ment set up and operating parameters over the RS232
interface.
protocol.
Query Set Up
Measurement value(s) yes n/a
Max/Min yes n/a
Last Error yes n/a
Battery Power Level yes n/a
Level Damp yes yes
Engineering Unit yes yes
Hold yes yes
Port Selection yes yes
Tare yes yes
The print protocol simplifi es the process of confi guring the
indicator for communication with an RS232 compatible
printer, dumb terminal or data acquisition system. Through
the print protocol the indicator can be set up to print
instrument front panel is pressed, at one of the factory
preprogrammed print intervals or when a CR (carriage
return) signal is received.
The set up function for the print protocol provides the abil-
ity to select the desired baud rate as well as the content of
The following procedure details how to set the indicator up
for use in the remote mode.
With the indicator on and displaying pressure
measurement value(s) press the shift key. In response to
pressing the shift key the indicator will display an upward
pointing arrow.
^
With the upward pointing arrow visible on the
respond by displaying the following.
ISO1745 PRINT
The RS232 confi guration currently in use will fl ash on the
With the selection of available interface protocols
interface option will be highlighted (fl ash). With the word
respond be displaying.
300 600 2400
These numbers represent the baud rates (communication
with which the indicator is to communicate. As with the
RS232 protocol menu, the current baud rate selection will
Press the Output key the required number of times
to highlight the desired baud rate. Each time the Output
key is pressed the next baud rate option will be highlighted
two times to change the baud rate from 2400 to 9600. With
the desired baud rate fl ashing press the enter key.
300 600 2400
The indicator will display the eight available end of mes-
is set for the device. The display of the available end of
message characters looks as follows.
CRLF CR EOT ,
ETX HTAB ; NULL
The decimal and hexadecimal representations of each
Appendix B.
13.2 Set Up of the Remote Protocol
13.1.3 The Print Protocol
Page 20
20
The indicator will automatically display the avail-
rently active fl ashing.
Press the Output key, as required, to highlight the
text character is being changed from CRLF (carriage return
with a line feed) to a CR (carriage return only). With the
key. Although it is diffi cult to see, the indicator will briefl y
Storing
This appears as the indicator writes the selected confi gura-
tion data to the Electrically Erasable Programmable Read
the indicator will resume pressure measurement activities
with the display confi gured as it was prior to entering the
+12345 +12345
The setup confi guration, since it is stored in
EEPROM, will be retained in memory until a new confi gu-
ration is programmed into the indicator.
The indicator is now ready for use in the remote communi-
notation conventions and numeric formatting, is provided
in Appendix A of this manual.
13.3 Set Up of the Print Protocol
The following steps should be followed to use the indicator
for use with an RS232 compatible printer, dumb terminal
With the indicator on and displaying pressure
measurement value(s) press the shift key. In response to
pressing the shift key the indicator will display an upward
pointing arrow.
With the upward pointing arrow visible on the
will respond by displaying the following.
OFF REMOTE
The indicator will automatically display the protocol
3
With the selection of available interface protocols
protocol options. Each time the OUTPUT key is pressed
the next interface option will be highlighted (fl ash). In this
OFF REMOTE
With the word “print” fl ashing press the enter key.
The indicator will respond by displaying.
300 600 2400
These numbers represent the baud rates (communication
with which the indicator is to communicate. As with the
RS232 protocol menu, the current baud rate selection will
Press the OUTPUT key the required number of
times to highlight the desired baud rate. Each time the OUT-
PUT key is pressed the next baud rate option will be high-
lighted (fl ash). In this example the OUTPUT key will be
pressed four times to scroll from 300 baud to 9600 baud.
With the desired baud rate fl ashing press the enter key.
300 600 2400
After the desired baud rate has be entered, the indicator
will display the eight available end of message characters.
As with the baud rate the end of message character must
be selected to be consistent with the requirements of the
CRLF CR EOT ,
ETX HTAB ; NULL
The decimal and hexadecimal representations of each
Appendix B.
The indicator will automatically display the avail-
Press the OUTPUT key as required, to highlight
the desired end of text selection. In this example the OUT-
PUT key will be pressed seven times to scroll from CR (car-
riage return) to CRLF (carriage return with a line feed).
CRLF CR EOT ,
ETX HTAB ; NULL
With the end of text character to be used fl ashing
press the enter key (ENT). The indicator will respond by
Pressure
This allows for the selection of the content to be included
Page 21
1m 15m 1h Cont
As with the RS232 protocol menu, baud rate, end of text
frequency currently in use will automatically fl ash on the
where:
Man the indicator will transmit the programmed data
Pace the indicator will transmit the programmed data
received.
Press the OUTPUT key to scroll through the
Man Pace 1s 15s
With the desired output frequency highlighted and
indicator will briefl y display:
Storing
This appears as the indicator writes the selected confi gura-
tion data to the Electrically Erasable Programmable Read
the indicator will resume pressure measurement activities
with the display confi gured as it was prior to entering the
+12345 +12345
The indicator is now ready for use in the print mode.
The indicator can, optionally, be equipped with a back-
lit liquid crystal display. The back lit display provides
in the data string transmitted by the indicator. As with the
RS232 protocol menu, baud rate and end of text charac-
ter the data string content command currently in use will
following format: hold, damp, left tare, left data, left units,
right tare, right data, right units, end of message
Each fi eld can have the following data:
with Function
Function Disabled Invoked
hold blank space “H”
2 for medium
3 for high
left tare blank space “T”
left data no data with displayed value
sensor inactive for left port
left units no data with engineering unit
sensor inactive of left port
right tare blank space “T”
right data no data with displayed value
sensor inactive for left port
right units no data with engineering unit
sensor inactive of left port
End of n/a end of text
message character
selected
No data will be transmitted for an inac-
tive port. If the indicator is confi gured for differential pres-
replaced with the differential measurement indication for
that port.
All: The indicator will include all the above param-
Pres+EU: The indicator will transmit only the numeric
pressure value and the current engineering unit
in use.
Pressure: The indicator will only transmit the numeric
value of the pressure measurement values.
In this example, the data string will be changed
from All to pressure plus engineering unit. Therefore, the
All Pres+EU
With the desired data string output highlighted
respond by displaying the menu of available output fre-
quency options. The output frequency menu will appear as
follows.
Section 14.0 The Back Light Function (Optional)
Page 22
22
high glare conditions.
Use of the backlighting option greatly
increase the power consumption of the indicator. The
useful life of the optional built-in nickel cadmium power
pack will be greatly reduced by continuous use of the back
light function. It is estimated that the power pack will pro-
vide 5 hours of operation with continuous use of the back
lighting function.
The operation of the back lighting option is included in the
following two sections.
14.1 Timed Back Light Function
The back light function can be initiated so that it will
To use the timed back light function, with the indicator
BATT/LITE key once. The back light will be turned on and
will stay on for 1 minute. After the 1 minute interval is
light function and return to displaying data and measure-
ment information in the standard non-backlit mode.
The back light function can be initiated so that it will stay
function in the continuous mode the following procedure
With the indicator on and displaying pressure
measurement data press the BATT/LITE key once. The
indica-tor will respond by lighting the liquid crystal dis-
play.
To implement continuous back lighting, press the
LITE key while the back light is on. The back light func-
tion will now stay on until turned off by the operator.
In order to disable continuous back lighting
mode.
The battery check function provides a means to determine
the remaining power level of the nickel cadmium power
pack. In addition, the battery check function provides a
means at looking at the internal voltage regulation of the
indicator when the instrument is being powered off the
To use the battery check function follow the procedure
below.
With the indicator on and displaying pressure
measurement data press the SHIFT key. The indicator will
respond by adding an upward pointing arrow to the display
to indicate the instrument is in the shift mode.
^ psi psi
With the upward pointing arrow visible on the
respond by displaying the current battery voltage. The bat-
tery voltage measurement will look as follows.
Battery Voltage
+x.xx
The voltage measurement will remain on the display for
the display confi guration that was in use prior to perform-
ing the battery voltage check.
A low battery icon is provided on indicators equipped
with the battery option. The unit will display a fl ashing
battery icon indicating that the batteries have discharged
to the point of having insuffi cient voltage (5.6 Vdc) to
The hold function allows the operator to freeze the dis-
played measurement data at the push of a key. The hold
function is considered an “immediate need” function. As
The hold function is a universal function. Use of the
hold function freezes the entire display of the indicator.
The indicator cannot freeze the data for one of two dis-
played pressure measurement assemblies.
The following procedure details the use of the hold function.
With the indicator on and displaying pressure
measurement data, to freeze the displayed data simply
press the HOLD key one time. The indicator will respond
by adding a small “H” to the upper left side of the display.
When in the hold mode the displayed data will look as
follows.
+12345 +12345
To exit the hold mode press any key.
The indicator will remain in the hold mode until a
key is pressed.
14.2 Continuous Back Light Function
15.2 Low Battery Indicator
Section 16.0 The Hold Function
SECTION 15.0 BATTERY CHECK FUNCTION (for use in indicators with optional built-in batteries)
15.1 Using Battery Check Function
16.1 Using The Hold Function
Page 23
23
SECTION 17.0 THE PRINT FUNCTION
The print key, when used with the RS232 interface allows
the operator to print, on command, pressure measurement
function. If the indicator is not set up in the print mode the
instrument will display the following error message when
the PRINT key is pressed.
PRINT protocol
After displaying the above error message the indicator
will resume basic pressure measurement activities. If it is
this operation.
To print pressure measurement data follow the steps below.
Be certain the confi guration settings are
tion system with which the indicator is to communicate.
With the measurement data to be “printed” press
the PRINT key. The measurement data as confi gured in
ing device.
As long as the indicator is set up with the RS232 interface
in the print protocol it will remain ready to transmit mea-
receiving device.
When the RS232 interface is not in use
it should be disabled by selecting the word “off” from
the RS232 protocol menu. When the off option is selected
the circuitry that supports the interface is powered down.
This conserves on power consumption. Reducing power
being powered by the optional built-in nickel cadmium
batteries. Activating the RS232 interface results in an
This increase in power consumption will produce a cor-
responding 20-25% decrease in battery life. See section
The system can be used with most standard RTDs (Resis-
tance Temperature Detectors) to provide precision temper-
installed in one of the sensor assembly positions in order
for the system to perform temperature measurement with
the system base unit and the RTD probe.
The capabilities of the interface assemblies are summa-
rized below:
-200 to 550° C:
0.15° C 550 to 850° C:
0.2° C
80 to 260° C:
0.1° C
70 to 150° C:
0.6° C
+0.01% reading +0.02 ohms
184 to 275° C:
0.15° C
Fahrenheit, Celsius, Kelvin, Rankine or ohms
RT1 module:0/400 ohms
RT2 module: 0/4000 ohms
The interface assembly incorporates a Switchcraft TA4M
type connector and requires an RTD with the appropriate
the appropriate mating connector is part number 828X136-
01. The pinout assignments are as follows:
Turn the system on by pressing the On/Off key.
To access the temperature setup menu press the
The will respond by displaying the following:
^
With the upward pointing arrow displayed press
the PORT/UNIT key one time to activate the port select
function. The indicator will respond by displaying:
Left Right Both
The Left minus right (L-R), right Minus left (R-L)
Pin Function
1 2 Wire +
2 4 Wire +
3 4 Wire –
4 2 Wire –
Pin Function
1 2 Wire +
2 4 Wire +
3 4 Wire
4 2 Wire
17.1 Using The Print Key
SECTION 18.0 TEMPERATURE MEASUREMENT USING RTD PROBES
18.1 Setting Up the System for Temperature Measurement
Page 24
24
with two installed pressure sensors are disabled when a
temperature measurement assembly is installed in one of
the two sensor locations.
As required, press the PORT/UNIT key until the
port that the RT interface assembly is installed in fl ashes
in the left sensor location.
Left Right Both
With the text “Left” fl ashing press the enter key
The system will respond by displaying:
2W 3W 4W
This represents the confi guration of the RTD to be used,
where the number equals the number of wires the RTD
uses.
The use of a 4 wire RTD is recommended for opti-
mum accuracy.
Using the PORT/UNIT key select the appropriate
below the 4 wire confi guration was selected.
2W 3W 4W
With the desired wiring confi guration selected press
the enter key (ENT).
The system will respond by displaying the available units
Engnring units:
Using the PORT/UNIT key select the desired
With the desired engineering unit selected
The system will respond by displaying the following:
Probe: 12345678
probe will appear selected. Use the PORT/UNIT key to
programmed for the following RTDs.
Position Probe
1 �385 -200/850 degrees C
Ni Minco
4 Cu 10 ohm
5-8 available for programming using the optional
RT1 module, part number 838X014-01 for 3 ½
inch disks.
with output resistance values of 0 to 400 ohms. Use of the
module with higher resistance RTD probes will result in
The RT2 is pre-preprogrammed for the following RTDs
Position Probe
1 Pt
2-8 available for programming using the optional
RT2 module, part number 838X014-01 for 3 ½
inch disks.
with output resistance values of between 1000 and 4000
tance RTDs accuracy will be greatly diminished. In addi-
tion, use of the module with higher resistance RTD probes
will result in diminished accuracy and possible damage to
the module electronics.
Using the left/right arrow keys select the RTD
probe type for the measurements to be made. In the
PT 100 type 385
a385 -200/850
With the desired probe type selected and
fl ashing press the enter key (ENT).
The system will respond by displaying the following:
1 .1 .01 0.001
The maximum resolution that the combination of the indi-
the RTD module is 0.001% of the full scale resistance of
the reference resistor used in the module. The RT1 incor-
porates a 400 ohm reference resistor and the RT2 incorpo-
rates a 4000 ohm reference resistor.
Page 25
25
SECTION 19.0 CALIBRATING THE INDICATOR
The indicator provides the ability to perform calibration
terminal is required to perform a complete recalibration
values. In most cases zero and span adjustment will be all
that is required to recalibrate the instrument.
The following sections detail the methodologies for zero/
This procedure details the method for performing a calibra-
tion level zero and span of any installed pressure sensor
Be certain that the indicator is set up in accordance with
the requirements outlined in Section 8 Zeroing Indicator
for the pressure sensor assembly to be zeroed and spanned.
With the indicator off press and hold the PRINT
key. While holding the PRINT key press the POWER key.
After the unit begins the start up process the PRINT key
pressure sensor assemblies, after the start up screens have
following.
Left Right
The left port will be the default calibration port and will
To select:
The Left Port (sensor assembly in left side of indicator):
Press the enter key (ENT)
The Right Port (sensor assembly in left side of indicator):
Press the PRINT key one time. The indicator will respond
by changing the highlighted port selection to the right port.
Pressing the PRINT key a second time will return the left
With the port corresponding to the pressure sensor to be
The indicator will respond by displaying the following.
psi “Hg “WC ftSW
Using the RT1 as the example the maximum resolution is:
0.001% of 400 ohms
= 0.00001 x 400 = 0.004 ohms
For a Pt 100 the temperature resolution can be determined
0.004 ohms/(100 ohms x 0.00385 ohms/degree C) = 0.008
Therefore, rounding to the next highest digit results in a
resolution of 0.01 degrees C.
The resolution supported for the factory programmed
RTDs can be found below:
Probe type Resolution
Pt 100 0.01 degrees C or F
(385 & 392)
Ni 120 0.1 degrees C or F
Cu 10 0.1 degrees C or F
Probe type Resolution
Pt 100 01.0 degrees C or F
Resolution must be selected in conjunction with the full
Select the desired resolution using the PORT/
UNIT key. In this case, since we are using the Pt100 RTD
with a maximum resolution of 0.01 degrees C the 0.01
Resolution:
With the desired resolution selected and fl ashing
press the enter key (ENT).
The system will briefl y show internal limit settings and
will then commence temperature measurement.
The measurement parameters can be changed at any time
through the activation of the RT set up menu via selection
The RT can be programmed in the fi eld. Programming,
plished with the Calibration Programming Utility disk part
number 838X014-01 (3 ½ inch disk). Please note, in order
to improve the accuracy of the measurement system it
will be necessary to generate a signifi cant number of very
temperature baths. In general, the simplest way to enhance
the accuracy of the temperature measurement is to simply
re-zero the probe while immersed in an ice bath.
19.1 Adjusting Calibration Zero and Span Settings
18.2 Calibrating/Programming the RT RTD Assemblies
Page 26
26
2nd screen
mmHg cmWC mmWC kgcm
The indicator will automatically default to the
primary engineering unitfor the pressure sensor to be
key until the desired unit isfl ashing on the display. With
the desired unit fl ashing on the display press the enter key
The indicator will respond by displaying the following:
Zero/Full Scale
The zero/full scale option allows for re-zeroing and re-
REMOTE option provides the ability to re-zero, re-span
The Calibration Remote option is detailed in the next sec-
tion of this manual.
To perform a re-zero and re-spanning of the pressure
press the enter key (ENT). The indicator will respond by
X.XX MODE
where:
X.XX: represents the zero offset.
Unit: represents the engineering unit select
To re-zero the pressure sensor assembly apply
vacuum sensor to atmosphere or pull a hard vacuum for an
With the proper zero applied press the ZERO key. The
indicator will respond by displaying the following
000 MODE
Indicating that any zero offset has been eliminated
Apply full scale pressure and allow the measured
value shown below the engineering unit designation to
Adjust the displayed pressure measure-ment value
to equal that of the applied pressure.
Use the:
TARE key: To decrease the displayed pressure value.
MAX/MIN Key: To increase thedisplayed pressure value.
The required key should be pressed the number of times
required to make the displayed pressure value equal that of
the applied pressure value.
When the displayed pressure value andthe applied
pressure input are the samepress the enter key (ENT).
The indicator will respond by displaying:
Storing Data
This will appear on the instrument display for approxi-
mately 5 seconds
Next, the indicator will display:
Yes No
Use the PRINT key to make the desired selection. With the
19.2 Performing A Complete Recalibration
This section details the process for performing a complete
recalibration of the indicator.
If the pressure sensor assembly has not been zeroed and
point should be the zero pressure input and the last pres-
just been re-zeroed and re-spanned, per the instructions
in section 18.1, then these points can be omitted from
the linearization process. In this case the fi rst linearization
point will be the fi rst non-zero pressure input and the last
linearization point will be the pressure input prior to the
full scale input.
The following are the minimum recommended number of
linearization points:
F
enhanced temperature performance
Page 27
The indicator can accommodate a maximum of 20 calibra-
tion points, including zero and span. Enhanced tempera-
ture performance sensors meet accuracy specifi cation
though temperature range of 20-120 degrees F. Tem-
perature error will not result in instrument performance
19.2.1 Terminal Requirements for Performing A
In order to calibrate pressure sensors in the indicator the
indicator must be connected to a terminal for data entry
Baud Rate: 2400
Data Bits: 8
Parity: None
Flow Control: Xon/Xoff
In addition the indicator RS232 Remote interface should
be set up as follows:
Baud Rate: 2400
End of
Text Message: CRLF
The RS232 interface does not have to be active in the
remote mode when initiating the calibration function as the
indicator will activate the Remote mode operation when
the Calibration-REMOTE selection is made from the main
This procedure details the method for performing a calibra-
tion zero and span of any installed pressure sensor assembly.
With the indicator off, prior to initiating the calibration
process, connect the indicator to the desired serial port of the
number 838X011-01) will connect the indicator to a stan-
port connector on the terminal or computer.
Be certain that the indicator is set up in accordance with
the requirements outlined in Section 8, Zeroing Indicator,
for the pressure sensor assembly to be zeroed and spanned.
Be certain the indicator has been connected to the
terminal or computer prior to powering up the indicator
With the indicator off press and hold the PRINT
key. While holding the PRINT key press the POWER
key. After the unit begins the start up process the PRINT
key can be released. If the indicator is equipped with two
pressure sensor assemblies, after the start up screens have
following.
Calibrate Port?
If the indicator is equipped with a single pressure
To select:
The Left Port (sensor assembly in left side of indicator):
Press the enter key (ENT)
The Right Port (sensor assembly in left side of indicator):
Press the PRINT key one time. The indicator will respond
by changing the highlighted port selection to the right port.
Pressing the PRINT key a second time will return the left
With the port corresponding to the pressure sensor to be
The indicator will respond by displaying the following.
psi “Hg “WC ftSW
Bar mBar kPa mPa
2nd screen
The indicator will automatically default to the
primary engineering unitfor the pressure sensor to be
key until the desired unit isfl ashing on the display. With
the desired unit fl ashing on the display press the enter key
The indicator will respond by displaying the following:
Zero/Full Scale
Calibrate REMOTE
The zero/full scale option allows for re-zeroing and re-
REMOTE option , as detailed in this section provides the
using a terminal or computer. The Zero/Span option is
To perform a complete recalibration/relinearization of the
highlight the text "Calibration- REMOTE". With the text
After pressing the enter key the following text
will appear on the screen of the terminal.
Complete Recalibration
19.2.2 Performing A Recalibration/ Relinearization of the Indicator
Page 28
28
Z zero
F full scale
L linearity
X exit
In some cases the indicator may miss getting the
typing the letter for the desired selection and pressing the
ENTER key on the computer/terminal keyboard.
The procedures for performing each of these functions
Remote Zero: 19.2.3
Remote Span: 19.2.4
Remote Linearity: 19.2.5
19.2.3 Remote Zero Adjustment
The indicator must be connected to the computer or termi-
nal being used for data input and review as outlined in
If the indicator is already in the Remote Calibration
mode skip directly to step 6
With indicator off connect unit to a computer or
terminal as outlined is section 18.2.1
Press the PRINT key on the indicator and while
holding the PRINT key press the POWER key. After the
indicator display will show the following query.
Left Right
Using the PRINT key highlight the sensor to be
The indicator will respond by displaying the following.
psi “Hg “WC ftSW
2nd screen
The indicator will automatically default to the
primary engineering unitfor the pressure sensor to be
key until the desired unit isfl ashing on the display. With
the desired unit fl ashing on the display press the enter key
The indicator will respond by displaying.
Calibrate REMOTE
To adjust the calibration zero via the remote mode
press the PRINT key one time to highlight the text "Cali-
brate REMOTE".
With the text "Calibrate REMOTE" fl ashing press
the ENTER key. The indicator will respond by transmitting
the main calibration menu to the computer/terminal display
Z zero
F full scale
L linearity
X exit
the letter Z. With the letterZ typed, pressed the Enter key
lowing information for display on the screen.
Apply ZERO pressure and press ENTER
Apply the appropriate zero pressure. Be certain a good
isand absolute unit.
With the correct zero pressure applied press the Enter key
The will respond by transmitting the following information
for display on the screen of the computer/terminal.
The indicator will respond by displaying 8 readings. These
readings represent the amplifi ed voltage output of the
nominal amplifi ed output range for all sensor assemblies is
0/1 Vdc for zero to full scale pressure. After the completion
value of the reading to establish the new correction value
for the zero linearization point.
The sequence of events will appear as follows.
+0.000001
-0.000000
+0.000001
+0.000000
Sensor ZEROED
After completion of the zeroing procedure the indicator
Page 29
29
will once again display the main calibration menu as follows.
Z zero
F full scale
L linearity
X exit
To complete the zero adjustment process and exit
the calibration menu Type X on the keypad and press the
Enter key. For information on adjusting the full scale value
proceed to Section 19.2.4
The computer/terminal will respond by displaying:
y- yes n-no
To Save calibration data type "y" and press the
respond by displaying
Storing data
19.2.4 Remote Full Scale Adjustment
The indicator must be connected to the computer or termi-
nal being used for data input and review as outlined in
If the indicator is already in the Remote Calibration
mode skip directly to step 6
With indicator off, connect the unit to a computer
Press the PRINT key on the indicator and while
holding the PRINT key press the POWER key. After the
released. At the end of the start up sequence the indicator
Left Right
Using the PRINT key highlight the sensor to be calibrated.
With the sensor to be calibrated fl ashing on the
by displaying.
Zero/Full Scale
Calibrate REMOTE
To adjust the calibration zero via the remote mode
press the PRINT key one time to highlight the text "Cali-
brate REMOTE".
With the text "Calibrate REMOTE" fl ashing press
the ENTER key. The indicator will respond by transmitting
the main calibration menu to the com- puter/terminal dis-
play as follows.
Z zero
F full scale
L linearity
X exit
the letter F. With the letterF typed, pressed the Enter key on
thekeyboard. The will respond by transmitting the follow-
ing information for display on the screen.
Apply FULL-SCALE pressure and then enter applied pres-
Apply the appropriate full scale pres-sure to the sensor
being adjusted.
With the correct full scale pressure applied to
the sensor and the corresponding value typed into the
the computer orterminal device being used. The indicator
will respond by transmitting for display measurement 8
measurement readings for the applied full scale pressure.
These readings represent the amplifi ed voltage output of
the sensor assembly at the applied pressure or vacuum. The
nominal amplifi ed output range for all sensor assemblies is
0/1 Vdc for zero to full scale pressure. After the completion
full scale linearization point. The sequence of events will
+1.000001
+1.000000
+1.000001
+1.000000
Sensor FULL-SCALE
After the above message has been displayed the indicator
will then return to the main calibration menu as follows.
Z zero
F full scale
L linearity
X exit
To complete the full scale adjustment process and
the Enter key. For information on adjusting the lineariza-
tion values proceed to section 18.2.6. Resetting of calibra-
tion zero was covered in Section 19.2.4.
The computer/terminal will respond by displaying:
Page 30
y- yes n-no
To Save calibration data type "y" and press the
respond by displaying
19.2.5 Remote Adjustment of Linearization Values
The indicator must be connected to the computer or termi-
nal being used for data input and review as outlined in
If the pressure sensor assembly has not been zeroed and
point should be the zero pressure input and the last pres-
just been re-zeroed and re-spanned, per the instructions
in Section 19.1, than these points can be omitted from
the linearization process. In this case the fi rst linearization
point will be the fi rst non-zero pressure input and the last
linearization point will the pressure input prior to the full
The Following are the minimum recommended number of
linearization points:
enhanced temperature performance
The indicator can accommodate a maximum of 20 calibra-
tion points, including zero and span. Enhanced temperature
performance sensors meet accuracy specifi cation though
temperature range of 20-120 degrees F. Temperature error
will not result in instrument performance exceeding rated
The indicator must be connected to the computer or termi-
nal being used for data input and review as outlined in
If the indicator is already in the Remote Calibration
mode skip directly to step 6
With indicator off connect unit to a computer or
terminal as outlined is section 19.2.1
Press the PRINT key on the indicator and while
holding the PRINT key press the POWER key.
After the start up sequence has commenced the PRINT
keycan be released. At the end of the start up sequence the
indicator display will show the following query.
Calibrate
Using the PRINT key highlight the sensor to be calibrated.
With the sensor to be calibrated fl ashing on the
Zero/Full Scale
Calibrate REMOTE
To adjust the calibration zero via the remote mode
press the PRINT key one time to highlight the text "Cali-
brate REMOTE".
With the text "Calibrate REMOTE" fl ashing press
the ENTER key.
the letter L. With the letter L typed, press the Enter key
information for display on the screen.
C- cancel
P- Proceed
Typing "C" on the keyboard of the computer/terminal and
pressing will result in escaping from the linearization pro-
Typing "P" on the keyboard of the computer/terminal will
Enter key on the keyboard of the computeror terminal. In
response to the "P" entry the indicator will transmit the
following information for display on the screen of the
Linearity point 1:
Apply pressure then enter applied pressure and
press Enter.
The fi rst applied pressure can be either zero or, if the
indicator has just been rezeroed and respanned, can be
the fi rst non-zero calibration point. In general, it is recom-
mended that both zero and full scale calibration points be
incorporated in the linearization process.
In response to the request for the fi rst linearity point the
brated.
into the terminal/computer press the Enter key on the
Page 31
keyboard.If zero is used the indicator will respond by
These readings represent the amplifi ed voltage output of
the sensor assembly at the applied pressure or vacuum.
The nominal amplifi ed output range for all sensor assem-
blies is 0/1 Vdc for zero to full scale pressure. After the
rection. The sequence of events will appear as follows.
+0.000001
-0.000001
+0.000001
+0.000002
If the fi rst linearization point is zero the displayed voltage
values will be close to zero. The voltage output displayed
will increase in direct proportion to the level of applied
pressure or vacuum as a percent of the full scale range of
the sensor assembly. After presenting the 8 readings the
indicator will calculate the average value and use it to
the indicator will display the following requesting the next
linearization point..
Linearity point 2: Apply pressure then enter applied pres-
In response to the request for the second linearity
point the operator should:
brated.
the terminal/computer press the Enter key on the Key-
board. At a minimum the following linearization points
enhanced temperature performance
In the case where the fi rst non-zero linearization point
is taken at 10% of the sensors range the indicator will
respond by displaying 8 readings of approximately 10% of
the 1 Vdc full scale output or 0.10000. After the comple-
tion of the eight readings the indicator will use the average
value of the reading to establish the new correction value
for the second linearization. The sequence of events will
The difference between the displayed measured
voltage and the ideal voltage of 0.100000 represents the
non-linearity of the sensor. This linearization process will
ity.
+0.100020
+0.100023
+0.100022
+0.100021
Linearity point 3: Apply pressure then enter
Repeat the process, as outlined in step 7, for each
linearization point to be done. To optimize measurement
very close the full scale range of the sensor being cali-
brated.
If the last linearization point is done at the full scale value
0.1% of the full scale value) will use the this point as
0.1% of
full scale the indicator will retain the last span setting
point be within +0.1% of the full scale range of the sensor
To complete the calibration process, without entering a
numeric value for a new linearization point, press the enter
key on the keyboard of the terminal/computer being used
transmitting the following information to the screen of the
terminal/computer.
Linearization Done
S- SAVE
C- Cancelled
transmitted to the screen of the computer/terminal.
wait
Storing Data
transmitted to the screen of the computer/terminal
Linearization cancelled
old lin data retained
Page 32
Typing and entering "C" on the key-board of the terminal/
ization process and retain old linearization data in the
EEPROM memory
It is not recommended that a calibration point by
placed within 1% of either the zero or span values for the
within the
pressure measurements beyond the range limits of the
instrument could subject to very large inaccuracies.
Regardless of the placement of calibration points, it is not
recommended that indicator sensors be used outside the
**********IMPORTANT NOTE**********
After saving the new linearization data the calibration
from the main calibration menu. Failure to do this will
After entering your selection to step 8
the indicator will respond by transmitting the main calibra-
tion menu to the screen of the terminal/computer being
used for the calibration interface.
The main calibration menu will appear as follows.
Z zero
F full scale
L linearity
X exit
To exit the calibration mode type and enter "X" from the
keyboard of computer/terminal being used as the calibra-
tion interface.
After typing and entering the "X" the indicator will
transmit the following information to the screen of the
yes
no
Type and enter "Y" on the keyboard of the computer/
terminal to save data. When "Y" is entered the text
"Storing data"
REMOTE Calibration SAVED
will appear on both the display of the indicator as well
result in the text
being displayed on the screen of the computer/terminal
After entry of either "Y" or "N" the will resume normal
pressure measurement activities
SECTION 20.0 INSTALLING A SECOND PRESSURE OR
This section details the procedure for adding a second
pressure sensor assembly in an instrument that previously
had a single sensor or changing an installed sensor assem-
bly. indicator pressure sensor assemblies can be added or
requirements of any specifi c measurement exercise to be
performed. The calibration data for each pressure sensor
Programmable Read Only Memory) located within the
pressure sensor assembly. As such, pressure sensor assem-
blies can be installed in any instrument, originally sold for
the required accuracy, and provide the rated accuracy of
the pressure sensor assembly.
To install a second pressure sensor assembly or change an
installed sensor assembly follow the procedures detailed in
prior to initiating any of the procedures detailed in this
performed by individuals experienced with working on
precision electrical test equipment.
initiating this process. Failure to do so could result in
instrument damage due to electrostatic discharge.
Power down the indicator and unplug from the ac
Place the indicator top down on a workbench and
remove the four (4) feet by removing the phillips head
halves together, pick up the indicator. With the instrument
right side up separate the two case halves.
If a sensor is to be installed in a sensor location
that currently is empty proceed to step 7.
To remove a pressure sensor assembly fi rst disconnect the
ribbon cable between the main board and pressure sensor
Looking at the rear panel, remove and retain
the three (3) phillips head screws that hold the sensor in
position. When removing the last screw, hold the pressure
instrument.
CHANGING SENSOR ASSEMBLY
20.1 Procedure for Installing or Changing a Pressure Sensor Assembly
Page 33
After the last screw has been removed, carefully
lift the pressure sensor assem-bly from the instrument
housing by lifting out and up the end with the ribbon cable
If a pressure sensor assembly is being installed
in the second sensor location for the fi rst time remove
the pressure connection cover by removing the three (3)
phillips head screws that hold it in place. After the screws
have been removed, set aside the cover plate and retain for
future use. Install sensor through the rear panel with the
label side up.
Align the sensor so that the three (3) phillips head
For fi rst time installations, connect the small ten
the larger 10 pin connector to the connector on the printed
If the sensor has been installed for the fi rst time
front board set (switch and display board set) up and out
housing. The ribbon cable between this board set and the
main board can remain connected.
For installation of replacement sensors only: If
the sensor installed is replacing a different range sensor
remove the range indication label on the front panel by
front panel. Retain range label with removed sensor for
future use.
For installation of all other sensors: Locate the range indi-
which the pressure sensor assembly has been installed.
Lift the front board set up and slide back into the
retaining groove in the bezel in the front of the case of
the indicator. Be certain board set is fi rmly seated in the
Replace the top cover of the instrument, carefully
front board set in the front of the indicator.
Carefully, holding the two case halves together,
place the indicator top side down on a workbench.
Re-install the four (4) rubber feet. Do not over-
tighten the phillips head screwsthat hold feet in position.
Installation of the pressure sensor assembly is now com-
plete. General use of the indicator can now be resumed.
Section 21.0 Replacement of Optional
the ability to recharge suffi ciently to provide any long
term use capability. When this occurs the nickel cadmium
power pack must be replaced. The following section
in rechargeable power pack.
prior to initiating any of the procedures detailed in this
performed by individuals experienced with working on
precision electrical test equipment.
initiating this process. Failure to do so could result in
instrument damage due to electrostatic discharge.
The optional built-in power pack uses 5 standard AA
nickel cadmium rechargeable batteries. Use of any other
battery type could cause damage to the indicator as well
AA RECHARGEABLE NICKEL CADMIUM BATTERIES
Power down the indicator and unplug from the ac
Place the indicator top down on a workbench and
remove the four (4) feet by removing the phillips head
halves together, pick up the indicator. With the instrument
right side up separate the two case halves.
Disconnect, at the main board, all ribbon cable
These include;
main board
Cadmium Batteries
21.1 Procedure for Installing Replacement Nickel Cadmium Batteries
Page 34
4- cable between the switch board and the main board (left
front)
Remove the front board set by lifting upward out
Remove four phillips head screws that retain main
board. One screw in each of the rear corners of the main
board. One on the left side of the main, just in from the cut
board just behind the cut out in the board.
to bend the pin connections between the main board and the
power supply board. This will expose the battery holders.
To replace the batteries, lift off each battery
retaining clip by placing a small screwdriver in the gap
between the battery contact and the end clip. Gently angle
the screwdriver handle toward the center of the individual
battery holder to disengage clip. Repeat procedure on clip
With all the battery retaining clips removed, care-
fully remove each of the installed batteries.
Install 5 new AA rechargeable batteries. Be cer-
tain to follow polarity instructions
provided in each of the battery receptacles.
After all 5 batteries have been replaced, re-install
the 5 battery retaining clips by aligning the clip over the
battery recep-tacle and pressing down evenly on each end
Replace main board, carefully aligning the pin
Hints to help when changing batteries:
pins is positioned to the front of the indicator during
main board installation2)Aligning the screw holes with
the main board mounting posts will facilitate the align-
ment of the pin connection between the main board and
the power supply board.
the four (4) retaining screws used to hold the main board
in position.
Re-connect the three ten pin connectors in the rear
panel to receptacle on the left rear of the main board.
tacle directly in front of it on the main board.
boards) into the groove in the bezel on the front of the
instrument case. Make sure the board set is seated all the
way in the groove.
Reconnect the cables between the main board
right side of the front board set to the eight (8) pin (single
row) connection on the front right side of the main board.
pin connector on the front left side of the main board.
Replace the top cover of the instrument, carefully
front board set in the front of the indicator.
place the indicator top side down on a workbench.
Re-install the four (4) rubber feet. Do not over-
tighten the phillips head screws that hold feet in position
Installation of the replacement batteries is now complete.
SECTION A-1.0 INTRODUCTION
The remote RS232 interface is designed to provide the user
with:
tion regarding the various user programmable set up
terminal or computer.
tor receives an unrecognized command it will respond with
Important Note:
When the indicator is in the remote
mode local (keypad activity) is locked out. The indica-
tor must be removed from the remote operating mode to
restore access to front panel functionality.
Page 35
SECTION A-2.0 NOTATION CONVENTIONS
This appendix uses the following conventions:
Bold, uppercase letters indicate a com-
mand name
italicized letters indicate command
parameters
parameter
USER I/O Gothic letters indicate literal user I/O
Numeric arguments may contain the following characters;
will be truncated to an integer value. For example, a
numeric input of 2.7 will be truncated to 2. There is no
requirement in the indicator communication protocol for
the use of numeric values that are not whole numbers.
Each allowable entry is listed alphabetically. The informa-
tion contained herein will provide details on the action
performed as the result of each allowable command as
well the proper syntax to use. A remarks section, providing
included, as required, for each command. The expected
return values (response of the indicator to the specifi c
SECTION A-3.0 NUMERIC FORMAT
SECTION A-4.0 COMMAND REFERENCE
Page 36
SECTION A-5.0 COMMAND LIBRARY SUMMARY
Where input fi elds for two numeric values are provided for independent setting can be accommodated for both installed
pressure sensor assemblies in indicators with optional second sensor.
Command Expected Result
? Indicator transmits numeric pressure value for active installed
pressure sensor assemblies
BATCK? Indicator transmits battery level in volts (numeric only)
DAMP? Indicator transmits numeric value from 0-3 representing level of
damping invoked
DAMP_0 Sets level of damping off
DAMP_1 Sets level of damping to low
DAMP_2 Sets level of damping to medium
DAMP_3 Sets level of damping to high
EUNIT? Indicator transmits numeric value for engineering unit in use for
active sensor(s) where:
Response Represents
-1 leave as is
1 psi
2 “Hg
3 “WC
4 ftSW
5 bar
6 mbar
7 kPa
8 MPa
10 cmWC
11 mmWC
12 kgcm
EUNIT X [,X] Command to change engineering unit(s) numeral in [ ] would be for
optional second (right) sensor.
HOLD? returns 0 or 1 where:
0 = hold off
1 = hold active
HOLD_0 Disable hold function
HOLD_1 Activates hold function
KEYLOCK? Requests status of keypad lockout function returns 0 or 1 where:
Response Represents
0 Keypad is disabled
1 Keybpad is enabled
KEYLOCK_0 Enables keypad
KEYLOCK_1 Disables keypad.
LASTERR? Provides information on last error.
MIN/MAX? Indicator transmits minimum & maximum value of any
MIN/MAX [ ] [, ] Indicator transmits minimum & maximum value of any active
sensor(s)
use: 0 or -1 to review and not reset values
use: 1 to review and reset values
PORT? Indicator transmits numeric value of
0-4 representing current confi guration.
POR_ 0 Set indicator for left port active only
PORT_1 Set indicator for right port active only
PORT_2 Set indicator for left & right ports active
PORT_3 Set indicator for left - right (differential)
PORT_4 Set indicator for right - left (differential)
TARE? Indicator transmits numeric value representative
of current tare status.
TARE X [,X ] where: numeric value -1 = leave as is
numeric value 0 = deactivate tare
numeric value 1 = activated tare
ZERO X [,X] where: numeric value -1 = leave as is
numeric value 1 = establish new zero value
Page 37
2. If there is no response, check cable connec-
tions, cable type (modem vs.thru), Com Port,
3. If communication is established proceed to
III Verifi cation Procedure - Baud Rate
A. Verify communication at all selectable baud
rates.
B. Confi gure the serial port on the indicator
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “4800” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF (car-
riage return & line) is fl ashing
7. Press “Enter”. The indicator’s serial port is
now confi gured for 4800 baud communication
with CRLF appended to data packet.
C. Confi gure the com port in the PC Terminal
application.
1. Click on the
2. Click on 4800 Baud, 8 Data Bits, 1 Stop Bit,
Parity None, Flow Control None ,select the
Terminal application is now confi gured to
Remote Mode.
D. Verify Communication at confi gured Baud Rate
1. Type a question mark character and press
0.004469,-0.000227
2. If there is no response, check cable connec-
tions, cable type (modem vs.
thru), Com Port, and repeat steps B and C
3. If communication is established repeat steps
B through D, changing the baud rates to 2400,
is encountered note it in the notes column and
APPENDIX A – INDICATOR SERIAL PORT FUNCTIONS
The following procedure can be employed for verifi cation
understanding of the command library for integration into
I. Equipment Requirements.
A. Indicator with one or two sensors
B. DB 9 Pin serial connector cable wired for straight
through operation
C. HVC, fi ttings and plumbing to generate pressure
D. PC with Windows, Terminal application and
open com port
II. Test setup
A. Connect cable to Indicator serial port and
PC com port.
B. Power up Indicator by pressing and holding
play the Firmware Revision currently installed in
the unit. Record the number in the block labeled
C. Confi gure the serial port on the indicator
as follows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF (car-
riage return & line) is fl ashing
7. Press “Enter”. The ndicator’s serial port is
now confi gured for 9600 baud communication
with CRLF appended to data packet.
E. Confi gure the com port in the PC Terminal
1. Click on the
2. Click on 9600 Baud, 8 Data Bits, 1 Stop Bit,
Parity None, Flow Control None, select the
minal application is now confi gured to com-
municate with the Indicator in the Remote
Mode.
F. Verify Communication between and PC
1. Type a question mark character and press
0.004469,-0.000227
VERSION 1.0
Page 38
IV. Test Procedure - Verify Remote Command Library
The following library commands must be typed in
terminal and the results verifi ed. This is done by simply
typing the commands exactly as shown, followed by press-
ing the enter key, and comparing the display results to
those listed under each command. Where spaces
required in the command string, they are indicated by
underscore marks (don’t type underscore hit space bar)
A. Battery Check Function
1. BATCK?
2. Response: 6.71 (or similar value)
B. Set/Get Dampening Value
1. DAMP_2
a. Response: OK
2. DAMP?
a. Response: 2
C. Set/Get Engineering Unit Values
1. Type: EUNIT_1,_2
a. Response: OK
2. Type: EUNIT?
a. Response: 1,2
3. Type : EUNIT_0,_0
a. Response: OK
D. Verify engineering unit conversion factors
1. Apply slight pressure to the right sensor and
lock HVC system
2. Type: EUNIT_0
a. Response: OK
3. Observe and record the value displayed for
the right sensor on the
4. Type: EUNIT_1
a. Response: OK
b. The right sensor data is now displayed in
Inches of Water
c. Refer to drawing 822B122 “Conversion
Factors for Units of Pressure”
d. Using the PSI value recorded in step E-3,
ing calculate the inches of water value
e. Assuming they are equivalent, proceed, if
not consult engineering
5. Repeat steps E-3 through E-4-e, substituting
the following values in the “type EUNIT_1
2 (inches HG - psi x 2.03602)
3 (inches water @ 20C: psi x 27.703)
4 (ftSW - psi x 2.2457)
5 (Bar - psi x .0689476)
6 (mBar - psi x 68.9476)
7 (Kpa - psi x 6.89476)
8 (MPa - psi x 68.9476x10-4)
9 (mmHg - psi x 51.7149)
10 (cmWC@ 20C - psi x 70.433)
11 (mmWC - psi x 684.82)
12 (Kg/cm2 - psi x .07030696)
E. Set/Check Hold Value
1. Type: HOLD_1
a. Response: OK (display indicates H)
2. Type: HOLD?
a. Response: 1
3. Type: HOLD_0
a. Response: OK (display H disappears)
F. Set/Check Keylock Value
1. Type: KEYLOCK_1
a. Response: OK
2. Type: KEYLOCK?
a. Response: 1 (keyboard is locked)
3. Type: KEYLOCK0
a. Response: OK (keyboard is unlocked)
G. Check Last error code
1. Type: ABC
a. Response: Err01
2. Type: LASTERR?
a. Response: Err01
3. Type: LASTERR?
a. Response: None
I. Reset/Get MinMax Values
1. Type: MINMAX?
a. Response:
2. Type MINMAX?_0,_0
a. Response:
from step 1)
J. Get/Reset Port Confi guration
1. Type: PORT?
a. Response: 2
2. Type: PORT_0
a. Response: OK (display left port only)
3. Type: PORT?
a. Response: 0
4. Type: PORT_1
a. Response: OK (display right port only)
Page 39
11. Press “Enter”. The serial port is now con-
key.
B. Press “Print” Key
1. Response: ,2 -0.0001 kPa +0.000 PSI
VI. Check Data Packet Contents
A. Confi gure the serial port on the indicator as fol-
lows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF (car-
riage return & line) is fl ashing
7. Press “Enter”. The indicator’s serial port is
now confi gured for 9600 baud communication
with CRLF appended to data packet.
8. Press “Output” repeatedly until “Press+EU”
is fl ashing
9. Press “Enter”
10. Press “Output” repeatedly untill “Man” is
11. Press “Enter”. The serial port is now con-
pressure and engineering unit data
B. Press “Print” Key
1. Response: -0.0001 kPa +0.000 PSI (or
C. Confi gure the serial port on the indicator as fol-
lows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF (car-
riage return & line) is fl ashing
7. Press “Enter”. The indicator’s serial port is
now confi gured for 9600 baud communication
with CRLF appended to data packet.
8. Press “Output” repeatedly until “Pressure” is
9. Press “Enter”
10. Press “Output” repeatedly untill “Man” is
5. Type: PORT?
a. Response: 1
6. Type: PORT_2
a. Response: OK (display both ports)
7. Type: PORT?
a. Response: 2
8. Type: PORT_3
a. Response: OK (display left-right port)
9. Type: PORT?
a. Response: 3
10. Type: PORT_4
a. Response: OK (display right-left port
11. Type: PORT?
a. Response: 4
12. Type: PORT_5
a. Response: OK (display left+right port)
13. Type: PORT_2
K. Get/Set Tare Value
1. Type: TARE?
a. Response: 0,0
2. Type: TARE_1,_1
a. Response: OK (display shows two T’s)
3. Type: TARE?
a. Response: 1,1
4. Type: TARE_0,_0
a. Response: OK (display T’s gone)
5. Type: TARE?
a. Response: 0,0
L. ZERO Ports
1. Type ZERO_1,_1
a. Response: OK
1. Type ZERO_0,_1
a. Response: OK
1. Type ZERO_1,_0
a. Response: OK
V. Verify Print Function
A. Confi gure the serial port on the indicator as follows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF (car-
riage return & line) is fl ashing
7. Press “Enter”. The indicator’s serial port is
now confi gured for 9600 baud communication
with CRLF appended to data packet.
8. Press “Output” repeatedly until “All” is fl ash-
ing
9. Press “Enter”
10. Press “Output” repeatedly untill “Man” is
Page 40
40
11. Press “Enter”. The serial port is now confi g-
ured to respond to activity on the “Print” key
with only pressure data.
D. Press “Print” Key
1. Response: -0.0002 -0.000 (or similar data)
VII. Verify Output Options
A. Pace Mode
1. Confi gure the serial port on the indicator as
follows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF (car-
riage return & line) is fl ashing
7. Press “Enter”. The indicator’s serial port is
now confi gured for 9600 baud communication
with CRLF appended to data packet.
8. Press “Output” repeatedly until “All” is fl ash-
ing
9. Press “Enter”
10. Press “Output” repeatedly untill “Pace” is
11. Press “Enter”. The serial port is now con-
key.
2. Press the “Enter” key on the computer keyboard
1. Response: ,2 -0.0002 kPa -0.001 PSI (or
A. 1s Mode
1. Confi gure the serial port on the indicator as
follows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF
7. Press “Enter”. The indicator’s serial
port is now confi gured
for 9600 baud communication with
8. Press “Output” repeatedly until “All” is
9. Press “Enter”
10. Press “Output” repeatedly untill “1s” is
11. Press “Enter”. The serial port is now
2. Press the “Enter” key on the computer keyboard
1. Response: ,2 -0.0002 kPa -0.001 PSI
B. 15s Mode
1. Confi gure the serial port on the indicator as
follows:
1. Press “Shift”
2. Press “Output” repeatedly until the menu
3. Press “Enter”.
4. Press “Output” repeatedly until “9600” is
5. Press “Enter”
6. Press “Output” repeatedly until CRLF
7. Press “Enter”. The indicator’s serial port
is now confi gured for 9600 baud com-
munication with CRLF appended to data
packet.
8. Press “Output” repeatedly until “All” is
9. Press “Enter”
10. Press “Output” repeatedly untill “15s”
is fl ashing
11. Press “Enter”. The serial port is now
2. Press the “Enter” key on the computer key-
board
1. Response: ,2 -0.0002 kPa -0.001 PS (or
C. Repeat steps B1 through B2.1 substituting “1m”,
Page 41
CRLF 13 10 0D 0A
CR 13 0D
EOT 4 04
, 44 2C
ETX 3 03
HTAB 9 09
; 59 3B
NULL 0 00
APPENDIX B - END OF MESSAGE CHARACTERS
Weight
from both bays.
–4° to 158°F
Available: ac adapter provided for 220Vac/50 Hz
ac adapter provided for 100Vac/60 Hz
Pack*
vating RS232 results in approximately 30% reduction in battery life.)
Available Ranges
Available Accuracies
2
2
Temperature Effect
ture range of 70° ±3°)
2
2
–15 to 50 psi
/
8
NPT female
/
8
British standard
Available Ranges
Available Accuracies
Temperature Effect
APPENDIX C – PRODUCT SPECIFICATIONS
PHYSICAL SPECIFICATIONS
OPERATING SPECIFICATIONS
PRESSURE SENSOR SPECIFICATIONS
Page 42
42
RTD INTERFACE ASSEMBLY
Accepts TA4F type RTD connector
with 2 pressure sensors installed
(gauge and (gauge/
absolute differential
pressure) pressure)
5
10
15
30
50
60
100
150
200
250
300
500
600
1000
1500
2000
2500
3000
5000
6000
7500
10,000
vacuum
5
10
15
compound
±5
±10
±15
–15/+30
–15/+60
0.25*
0.5*
1.0*
2.0*
3.0*
5.0*
10*
15*
25*
50*
2
ftSW
2
2
accessible through
the unit select feature.
able. Consult factory.
±0.125*
±0.25*
±0.5*
±1.0*
±1.5*
±2.5*
±5.0*
±7.5*
±12.5*
±25
±50
±75
±100
/
8
NPT female
/
8
female with fl ush port
(ranges 10 psi and over)
G
/
8
British standard
G
/
8
British standard with fl ush port
(ranges 10 psi and over)
Welded VCR fi tting with standard
fi nish
This errata sheet describes changes were made to the
These changes were made at Firmware Version 2.04 and
Version 2.03.
All characters received by the unit, when
in Remote protocol, will be echoed back to the host.
After the End Of Message (EOM) sequence is received
by the fi nal EOM sequence.
Version 2.04.
An additional menu is added if Remote
Protocol is selected. The user can select between ‘Terminal
Mode’ or ‘Computer Mode’. In Terminal Mode, the char-
query command will not be echoed and an additional EOM
End of Message Character(s).
The End Of Message (EOM) character sequence is confi g-
ured from the menus when Remote protocol is selected.
Version 2.03.
If the unit is confi gured for the CRLF EOM character
mits a two character sequence CRLF (0Dh 0Ah) at the
STANDARD RANGES
. . .
.
APPENDIX D – REMOTE COMMUNICATIONS PROTOCOL ERRATA SHEET
Page 43
43
unit will (correctly) transmit the two character CRLF(0Dh
0Ah) EOM sequence on transmissions to the host.
If the unit is confi gured for any single character EOM
transmit a two character EOM sequence consisting of the
EOM character followed by a NULL, on transmissions to
the host.
If the comma is selected as the EOM character, the unit
will not recognize it as an EOM character since the comma
is recognized (fi rst) as a delimiter character.
The EOM sequences and the resultant receive and transmit
Menu Receive Transmit
EOT 04 04 00
ETX 03 03 00
NULL 00 00 00
The following examples illustrate these characteristics.
host running HyperTerminal. Note that the characters are
The host will transmit: EUNIT 1,2[CR]
The unit will reply:
EUNIT 1,2[CR][CR][LF]OK[CR][LF]
EOM=CRLF, MSG=EUNIT 1,2 Data is shown in Hex.
Host 45 55 4E 49 54 20 31 2C 32 0D
Unit 45 55 4E 49 54 20 31 2C 32 0D 0D 0A
Unit 4F 4B 0D 0A
EOM sequences (CR).
The host will transmit: HOLD?[CR]
The unit will reply:
HOLD?[CR][CR][NULL]0[CR][NULL]
EOM=CR, MSG=HOLD? Data is shown in Hex.
Host 48 4F 4C 44 3F 0D
Unit 48 4F 4C 44 3F 0D 0D 00 30 0D 00
Version 2.04.
The unit will correctly recognize and process, one and two
The EOM sequences and the resultant receive and transmit
Host 48 4F 4C 44 3F 0D
Unit 30 0D
Remote calibration is initiated by holding the PRINT key
is activated, the communications protocol is forced to
Remote-Terminal Mode, regardless of any mode that was
previously selected from the front panel menus.
The terminal (or computer) that is attached to the unit
Baud Rate: 300/600/2400/4800/9600
Data Bits: 8
Stop Bits: 1
Parity: None
Flow control: None
Comm Port: per port being used.
The indicator should be set up as follows:
Baud Rate: 300/600/2400/4800/9600 (must
match terminal rate)
End of Msg: CRLF
Note: Any EOM character
References in the manual to Left Port and Right Port
Page 44
44
2
PIN FUNCTION
2 4W+
3 4W–
4 2W–
PIN FUNCTION
2 4W+
3 4W–
4 2W–
Page 45
45
Page 46
Ashcroft Inc.
250 E. Main Street
Tel: (203) 378-8281
© Ashcroft Inc. 11/06
Loading...