MODEL KX THIN FILM PRESSURE TRANSDUCER INSTRUCTION SHEET
WARNING! READ
m m
BEFORE INSTALLATION
1. GENERAL:
A failure resulting in injury or damage may be caused by excessive overpres­sure, excessive vibration or pressure pulsation, excessive instrument tempera­ture, corrosion of the pressure containing parts, or other misuse. Consult Ashcroft Inc., Stratford, Connecticut, USA before installing if there are any questions or concerns.
2. OVERPRESSURE:
Pressure spikes in excess of the rated overpressure capability of the transducer may cause irreversible electrical and/ or mechanical damage to the pressure measuring and containing elements.
Fluid hammer and surges can destroy any pressure transducer and must always be avoided. A pressure snubber should be installed to eliminate the damaging hammer effects. Fluid hammer occurs when a liquid flow is suddenly stopped, as with quick closing solenoid valves. Surges occur when flow is suddenly be­gun, as when a pump is turned on at full power or a valve is quickly opened.
Liquid surges are particularly damag­ing to pressure transducers if the pipe is originally empty. To avoid damaging surges, fluid lines should remain full (if possible), pumps should be brought up to power slowly, and valves opened slowly. To avoid damage from both fluid hammer and surges, a surge chamber should be installed.
Symptoms of fluid hammer and surge's
damaging effects:
• Pressure transducer exhibits an output
at zero pressure (large zero offset).
• Pressure transducer output remains
constant regardless of pressure
• In severe cases, there will be no output.
FREEZING:
Prohibit freezing of media in pressure port. Unit should be drained (mount in vertical position with electrical termination upward) to prevent possible overpressure damage from frozen media.
3. STATIC ELECTRICAL CHARGES:
Any electrical device may be susceptible to damage when exposed to static elec­trical charges. To avoid damage to the
transducer observe the following:
• Ground the body of the transducer
BEFORE making any electrical connections.
• When disconnecting, remove the
ground LAST!
Note: The shield and drain wire in the cable (if supplied) is not connected to the transducer body, and is not a suitable ground.
Mounting Although the unit can withstand normal vi­bration without damage or significant output effects, it is always good practice to mount the transducer where there is minimum vibration.
Power Supply The supply voltage for the 1-5 and 1-6 Vdc output transducers must be within the range of 10 to 36 Vdc. The maximum supply voltage for a 4-20mA current output transducer is 36 Vdc while the minimum supply voltage is dependent upon the loop resistance of the circuit. The figure below shows the minimum supply voltage (V required for a given loop resistance (R
min
LOOP
)
).
OPERATING REGION
V
= 10V+ (.022A x RL)
min
R
= RS + R
L
R
L
R
S
R
W
W
= Loop Resistance (ohms)
= Sense Resistance (ohms)
= Wire Resistance (ohms)
Noise
For minimum noise susceptibility, avoid running the transducer's cable in a conduit that contains high current ac power cables. Where possible avoid running the cable near inductive equipment.
Shield Wiring Connect the braided shield to the guard terminal on the reading instrument (meter, etc.) if available or to ground or to the power supply negative terminal.
Adjustment Potentiometers The zero and span pots are accessible through the top of the case. Loosen the col­lar and separate the top carefully. The zero pot is marked with a white dot.
Vent Tube The cable will have a clear Teflon vent tube that's required at pressure below 500 psi to provide atmospheric reference. The open end should be placed in a dry area.
© 2011 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614, USA • Tel: 203-378-8281, Fax: 203-385-0402 • email: info@ashcrfoft.com, www.ashcroft.com. All specifications subject to change without notice. All sales subject to standard terms and conditions of sale. I&M011-10156 (250-A152) Rev. A 5/11
Page 2
MODEL KX THIN FILM
6.1
5.2
1/2 NPT CONDUIT FITTING
O-RING SEAL
5/16 18 UNC THREAD
1.032
4.9
4.2
1/2 NPT (C1)
1.06
.8
4.9
4.2
1/2 NPT (C1)
1.06
.8
+
+
POWER SUPPLY
(+)
(–)
METER
TRANSDUCER
V+
V–
PIN 1
PIN 2
4-20 mA
+
+
POWER SUPPLY
METER
TRANSDUCER
(Common)
(+ Power)
(+ Output)
PIN 1
PIN 2
PIN 3
3-Wire Voltage
PRESSURE TRANSDUCER INSTRUCTION SHEET
KX Series – Electrical Connections
Voltage Output Units 1-5, 1-6 Vdc
Cable Type C1
Red = + Power Black = Common White = Output
3
2
DIN Type
PIN-1 = + Power
1
PIN-2 = Common PIN-3 = Output
Current Output Units 4-20mA
Cable Type C1
Red = + Power Black = – Power
3
2
DIN Type
1
PIN-1 = + Power PIN-2 = – Power
KX Series – Wiring Diagrams
KX Series – Recalibration Instructions
1. Apply 0% Full Scale Pressure.
2. Set the output using the Zero adjustment potentiometer.
3. Apply 100% Full Scale Pressure.
4. Set the output using the Span adjustment potentiometer.
5. Repeat steps 1 thru 4 as necessary.
KX Series Dimensions
© 2011 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614, USA • Tel: 203-378-8281, Fax: 203-385-0402 • email: info@ashcrfoft.com, www.ashcroft.com. All specifications subject to change without notice. All sales subject to standard terms and conditions of sale. I&M011-10156 (250-A152) Rev. A 5/11