Ashcroft GC52 User Manual

Quick Start Function Summary Instructions for ASHCROFT
®
GC52 Differential Pressure Transmitter Version 6.03 Rev. B
(See Complete I&M Manual for Further Detail)
LOOK FOR THIS
OUR PRODUCTS
© 2010 Ashcroft Inc., 250 East Main St., Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com All sales subject to standard terms and conditions of sale. I&M011-10164-GC52 Ver. 6.03 Rev.B 03/11
2
CH+
CH–
GC52
In H2O
M
b
Differential pressure unit monitor
Test Terminals
d
MODE key
e
DOWN key
f
UP key
c
Scaling - Arbitrary unit monitor
h No function
a
Measured data display
8.8.8.8.8.8.
k Total flow volume
display (X100)
l Total flow volume
display (X1000)
g
Linear scaling mode
j Total flow volume
display (X10)
i Total flow volume
display (X1)
DISPLAY OVERVIEW
DESIGNATION FUNCTION
Measured data display
Differential pressure unit monitor
Scaling; arbitrary unit monitor
MODE key M
DOWN key S
UP key T
Linear scaling mode
No function
to
volume display
Differential pressure, linear scaling value are displayed.
When this unit monitor is ON, the differential pressure (in H
O) is indicated on the measured data display.
2
When this unit monitor is ON, the scaling value of an arbitrary unit (linear scaling), is indicated on the measured data display.
This key is used to switch the setting mode and the measurement mode and to change the setting item.
This key is used to change (decrease) and select the set value.
This key is used to change (increase) and select the set value and to shift from the measurement mode to the zero adjustment mode.
Change differential pressures into quantity
Total flow
None
Accumulated volume calculated based on flow rate
3
11..
Note: Holding the button for more than 3 seconds returns to display mode.
UPON
POWER-UP the unit enters “Measure Mode” – displaying applied pressure.
22..
O FUNCTION is available to the user in “Measure Mode”.
TW
A. Zero Adjustment Mode: In the measurement mode,
the pressure connection is open to the atmosphere and $key is pressed for more than 3 seconds in order to shift to zero adjustment mode for zero point adjustment of the differential pressure sensor
• If the zero point adjustment is performed correctly the message “ADJ” will be displayed for 2 seconds, and the display will return to the meas­urement mode.
B. Key Lock
Function Key Manual Indicator
Setting of key lock MODE+ Release of key lock MODE+$one second UnL (Key invalidity)
Operation during keylock
Function Key Manual Indicator
Zero adjust. mode Hold value reset
33..
FI
VE FUNCTIONS available to the User via “Setting Mode”. To enter the “Set-
$ #
one second LoC (Key invalidity)
$
key greater than 3 sec LoC (Key invalidity) key greater than 3 sec LoC (Key invalidity)
ting Mode” hold (M) key for more than 3 seconds. (See last page for complete Setting Mode menu.)
A. Filter (Damping)
The filter is based on the moving average of the pressure data to de­crease display “bounce” and to smooth the analog output due to system pressure fluctuations at the user’s discretion.
Five selections: (0, 2, 4, 8 and 16 seconds), Use
$#
keys to change value. If “0” is selected the filter is not applied.
B. Re-scaling in “inH
O” units: “Pressure Display Mode” allows for zero
2
(4mA) and span (20ma) adjustment of –10 to +110% Span respectively.
Note: 1. See menu schematic on last page for detail.
2. Must be in “Pressure Display Mode” option within “Setting Mode,”
this is noted on the screen by
-
n non
Use $#keys to move between “Pressure Display Mode” and “Linear Display Mode” which is for re-scaling in “Arbitrary” units.
4
3. To adjust Output Zero Point (4mA) and Output Span Point (20mA) the unit must be in the functional area as noted below while adjustment is via $#keys. The value shown is a percent­age of the pressure range (span) as noted on the product label (ex. If product was supplied as a 0-40 IWC range and the user de­sired the Output Zero Point (4mA) to be “moved” from 0 IWC to 20 IWC then Output Zero Point would be 50.0 which is 50%.
Values shown below are from I&M manual.
Setting LCD Setting Setting
Item Display Description Range
Display Selection of non:pressure
mode pressure display display mode
Output Analog Pressure
zero point output zero point range:–10
.
Output Analog Pressure
span point output span point range:–10
Note: For setting of zero point and span point in the analog output, input the percent value over the pressure range.
C. Re-scaling in “Arbitrary” units, “Linear Display Mode.” This function
allows the user to establish a linear relationship from the standard “inH unit to any user defined unit.
Note: See menu schematic on last page. Unit must be in “Linear Display Mode” option within “Setting Mode”, this is noted on the screen by
-
n non
a 10.0
a 90.0
mode : non Lin: linear
(4mA) : 10.0 (%Span) to 110% Span
(20mA) : 90.0 (%Span) to 110% Span
display mode
O”
2
-
n L;n
Use $#keys to move between “Linear Display Mode” and “Pressure Display Mode.”
5
Setting LCD Setting Setting
Item Display Description Range
Display mode
Min. differen­tial pressure
Max. differen­tial pressure
Decimal point position
OFFSET
FULL SCALE
Output zero point
-
n00l0n
-
p0 -20
p0-120
d0-001
d0-00.0
d0-50.0
a0-50.0
Selection of linear display mode: Lin
Min. differential pressure corresponding to OFFSET
:20.0(inH
Max. differential pressure corresponding to FULL SCALE
Display after decimal point Number of digits: 1(digit)
OFFSET corresponding min. differential pressure
: 0.0 (m)
FULL SCALE correspon­ding to max. differential pressure
Analog output zero point : (4mA): 0.0 (%Span)
O)
2
:120(inH
:50.0 (m)
O)
2
non: Differential pressure display mode; Lin: Linear display mode
Differential pres­sure range: 0 to 75% Span
Differential pressure range: 25 to 100%Span
0,1,2,3 digit
–1999 to 1999
–1999 to 1999
Max. display span: –10 to 110% Span
Output span point
a0100.0
Analog output span point : (20mA): 100.0 (%Span)
Max. display span: –10 to 110%Span
Note: Values shown are from example in I&M manual.
D. Flow Measurement/Square Root Extraction Mode
Mode Square Square Root Display Mode
Combining the sensing elements (orifice and pitot tube) with this product, the mode is used for the display of the momentary flow rate, integrated volume and for analog output corresponding to the momentary flow rate.
(1) Momentary flow rate
The momentary flow rate display has the maximum display span from zero to the maximum momentary flow rate. The Momentary flow rate display span depends on the setting of the zero point and span point of the analog output as shown in (2) analog output. It can display 0 to 105%F.S. display span. The scaling method of momentary flow rate can be performed only by setting the maximum momentary flow rate and then generate differential pressure using the following square root formula.
Momentary flow rate Dx is expressed by the square root formula (a), and can be calculated only by measuring the generated differential pressure Px (percent value over the differential pressure range).
x(%)
(a)
Dx = k x
P
100(%)
6
In addition, the coefficient k is determined by substituting the maximum
999999
0
0
overflows
clr
momentary flow rate Dm, which is measured from the formula (a), and the then generated differential pressure Pm into the formula (b).
m00
(b)
• The differential pressure generated during the maximum momentary flow rate can be set in the range of 25 to 100%F.S. of the differential pressure range.
• The setting range for values of the maximum momentary flow rate is 0 to 1999. Note: The decimal point can be set arbitrarily.
• When the display resolution lowers and the wobbling of momentary flow rate increases in the low flow domain of the differential pressure flow meter, the domain (below the set value) will be forcedly indicated as zero by means of the low-cut for momentary flow rate.
Moreover, the analog output has a fixed value of 4mA at the zero point. For setting of low-cut, input the percent value over the maximum display span. Its range is 0 to 30%F.S. and the decimal point position can be set up to one digit after decimal point as fixed point.
(2) Analog output
The zero point (4mA) and span point (20mA) of analog output can be set in the range of 0 to 110%F.S. of the maximum display span (0 to maximum momentary flow rate). The span between the zero point and the span point in this analog output is the momentary flow rate display span.
(3) Integrated volume
• The units of integrated volume include two standards: Time factor and flow rate volume factor.
k =
D Pm
100
• The number of digits of integrated volume display is a maximum of 6 figures (999999); the display will return to 0 once the maximum reading has been met.
• The zero reset of an integrated volume is executed by pressing S key for
more than 3 seconds and displaying "cLr" (clear) for 2 seconds.
• As backup in case of POWER OFF, the integrated volume value is stored in the nonvolatile memory for every hour. After power returns, integration starts from the integrated volume value stored in the memory.
• Integration is halted during the "FFF" display at the time of differential pressure range OVER .
7
• The indicated value which is blinking is integrated during the "blink" display at the time of momentary flow rate display span OVER.
Flow Measurement/Square Root Extraction Mode
No.
Setting Item LCD Display Setting Description Setting Range
Display mode
Maximum differential
1
selection Flow rate deci-
mal pt. position
n~ roT
p. 400
D 0
Selection for flow measure­ment/ square root extraction mode
Maximum differential pres­sure relating to the flow rate
Displays of value after deci­mal point, # of digits
non: Differential pressure display mode Lin: Linear display mode Rot: Square root display mode
25 to 100% F.S. of sensor range
0,1,2,3 digit
Default
non
100.0%
of sensor
range
0
Max. momen-
tary flow Low cut
Output zero point
Output span point
Time factor
Flow rate vol-
21
ume factor
Display switch set-
22
ting Switch time
interval
23
Loop check
24
(1) In the setting of a differential pressure the decimal point position is fixed for each differential pressure range.
The max. differential pressure can be set from the value which is 25%F.S of the differential pressure range above the minimum differential pressure.The values under 25%F.S. cannot be increased or decreased by T, S key.
(2) For setting zero point and span point of the analog output, input the percent value over the maximum display span (be-
tween OFFSET and FULL SCALE). Its decimal point position can be set up to one digit after the decimal point (xx.x).
(3) Regardless of generated differential pressure or low-cut, the loop check can be changed arbitrarily linking the
momentary flow rate display with the analog output using the T, S keys. This example of LCD display shows the display set to span point.
2
2
D 1000
l 0.0
A 0.0
A100.0
U SE[
U 1
s bT
T 5
3
c 0.0
Max. momentary of flow using arbitrary units
Forces display and output to zero
Momentary flow rate of ana­log output zero point (4mA):
100.0% F.S. Momentary flow rate of ana-
log output zero point (20mA): 100.0% F.S.
Measurement of max. mo­mentary flow rate over time selected
Flow rate x time selected 1,10,100,1000
Selection of display switching method of momentary flow rate and integrated volume
Selection for ti: automatic Displays switching time in­terval in seconds
Output check using arbitrary value – displays pressure correlating to the 4 to 20mA signal
0 to 1999
0.0 to 30.0% F.S. of max. display span
–10 to 100% F.S. of max. display span
–10 to 100% F.S. of max. display span
Seconds, minutes or hours
ti = automatic bt = manual
1 to 10 seconds (10 stage)
Display: momentary flow rate display span Analog output: 4 to 20mA
0.0 to 100.0%
1000
0.0
0.0
100.0
Sec
1
bt
5
0
(4.0mA)
8
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
a
b
e
c
d
f
g
h
i
j
k
l
u 6.00
f 2
n non
n l n
a 10.0
a 90.0
[ 1.00[ 1.00
p 20.0
p 120.0
d 1
d 0.0
d 50.0
a 0.0
a 100.0
[ 50.0
-
­`
-
-
Linear display
mode setting
Differential pressure
display mode setting
Differential pressure display mode
Output span point pressure
Output zero point pressure
Loop check (zero point)
Version display
Filter
Linear display mode
Min. pressure
(Re-scaling in
“inH
2
O” units)
(Re-scaling in arbitrary
user defined units)
Max. pressure
Decimal point position
OFFSET
Full scale
Output zero point
Output span point
Loop check (span point)
Basic key operation
The setting item is changed by
u
M key.
The set value is changed or selected by key or key. When changing the value, it is increased or decreased by pressing key or key, respectively. (Refer to next page)
key for more than 3 seconds
Setting Mode
Measurement Mode
m
Flow
Measurement
Mode
See
Page 6
(4) Display switching method of momentary flow rate and integrated
volume
Display switching methods of momentary flow rate and integrated vol­ume include the automatic switching display method to display them by turns at intervals of fixed time (1 to 10 seconds) and the manual switching display method to change the display by pressing (M) key.
Complete Setting Mode Menu
9
u
M
u
M
u
M
u
M
u
M
N ROT
d 0 d i d Z d 3
a 100.0
L 10.0
P. 4.00
a 0.0
U SE1
U SE( U houU n~In
U 10 U 100 U 1000
Flow Measurement Mode
(Square Root Extraction Settings)
Refer to Section 14.3
b Flow Measurement / Square Root Extraction Mode
n Maximum Differential Selection
o Flow Rate Decimal Point Position
5 Bt 5 tI
u22 Display Switch
Setting
u
23 Switch Time
Interval
u
24 Loop Check
Returns to Setting Mode
q Low Cut
s Output Span Point
t Time Factor
u
M
u
M
u
M
u
M
u
M
u
M
u
M
u
M
D1000.0
p Maximum Momentry Flow
D 100.0
C D
C D
C D
C D
C D
t 5
c 0.0
u21 Flow Rate Volume Factor
C D
C D
C D
C D
u
M
r Output Zero Point
u22
E. Loop Check: Use to send a 4-20mA signal meant to simulate applied
pressure, can be accessed either through Pressure Display Mode or Linear Display Mode. See “Complete Setting Mode Menu”; Loop Check is noted
on the screen with a prefix starts at the zero (4mA) point.
. The display is indicating in actual units and
(
If $button is pressed, the linear display will auto increment by linkage be­tween the linear display and the analog output. By continuing to press #but­ton, auto decrement will occur. Release the button at the desired indication.
Flow Measurement Mode (Square Root Extraction) Setting
10
4. WIRING
Terminal box
Display (board)
Shield
Power source
Receiver
Transmission cable
CASE
DISPLAY
CAP
Inside sensor Line
LCD holder
Notch
Transmission cable (Twist)
Sheath
Power supply requirements, 12-36Vdc, note installation recommendations as follows:
Terminal Strip: SMKDSP1.5/2-5.08 Phoenix contact
A. Cable Requirements
• Two core shielded cable
• Cable outer diameter: 0.35˝ to 0.47˝ (9-12mm) Required for proper installation with cable gland option
• Wire Gauge: 14-22 AWG (multi-strand or solid)
B. Wiring Instructions
• Do not run pressure transmitter cable / wires within the same conduit as high
voltage (line power) line to reduce the potential for noise (interference). Use dedicated conduit on GC52 cables / wires for optimum results.
• Cable diameter, specified above, must be maintained when using the Cable
Gland termination to retain environmental ratings.
• When connecting shield / drain wire, only connect one end which should be at
the received ground.
• Wire stripping instructions; remove cable jacket 2-3˝ and strip wires 0.25˝.
Shield / drain wire should not be exposed at the pressure transmitter termination.
11
Display
Wire terminals
Wire
Turn with a screwdriver
LCD clips (4)
Display board
Power supply terminal block
• Remove cover and carefully remove the display to access the terminal strip, take care not to mishandle the display and associated electronics.
• Turn display over to expose terminal strip, make positive and negative connec­tions; insert wire equal to the recommended strip length (0.25˝).
• After completing connections, align the retaining clips of the display with the housing’s notches and carefully place into the housing. Be sure that the inter­nal sensor ribbon cable does not cross the power supply lines just installed.
• Be sure to properly tighten the sealing grommet when using the Cable Gland before applying tension to the cable; the cable gland provides strain relief and environmental sealing.
• Tighten GC52 cover to maintain environmental rating.
• Connect to power source and receiver, than apply power to confirm correct wiring.
• Power Supply Requirements: Although the 4-20mA signal can travel over long distances, a very common issue to arise involves inadequate power at the pressure transmitter – this results in voltage drop across the loop. Be sure to review the accompanying table to determine whether the 12-36Vdc has been received at the pressure transmitter.
12
1000
750
500
250
0
0 322010
1020
30
Load Limitations 4-20mA Output Only
12 24
545
OPERATING
REGION
Loop Resistance (V)
LOOP SUPPLY VOLTAGE
min
V
= 12V+[.022A*RL)]
*Includes a 10% safety factor
RL= RS+ R RL= Loop Resistance (ohms) R R
W
S
= Sense Resistance (ohms)
W
= Wire Resistance (ohms)
13
14
15
© 2010 Ashcroft Inc., 250 East Main St., Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com All sales subject to standard terms and conditions of sale. I&M011-10164-GC52 Ver. 6.03 Rev. B 03/11
16
Loading...