Ashcroft AXLdp Specifications

Data Sheet
AXLdp Low Differential Pressure Transmitter
UL94-V0 rated PBT case Calibrated ratiometric output Low voltage power requirement Easy mounting to PCBs Available in pressure ranges below 0.5 IWC
TYPICAL USES
Flow Measurements Room Pressure Control Energy Management Fan Control HVAC and VAV
PERFORMANCE SPECIFICATIONS
Reference Temperature:
Accuracy Class: ±2.0%, ±1.0% of span
Stability: ≤±0.50% span/year
Media Compatibility:
ENVIRONMENTAL SPECIFICATIONS
Temperature Limits:
Thermal Coefficients:
Humidity Effects:
FUNCTIONAL SPECIFICATIONS
Mounting Position: Calibration in vertical position (STD.)
Overpressure Limits:
Max. Static Line Pressure:
ELECTRICAL SPECIFICATIONS
Output Signal (Ratiometric):
Supply Voltage: 4.75 to 5.25 Vdc
70°F ±2°F (21°C ±1°C)
(Terminal Point Method: includes non-linearity, hysteresis, non-repeatability, zero offset and span setting errors)
Clean, dry and non-corrosive gas NOT FOR USE ON LIQUIDS
Storage:
Operating:
Compensated:
Zero & Span: ±0.10% of span/°F (±0.18% Span/°C)
No performance effect at 10-95% R.H. noncondensing
Proof: Burst:
7.25 psid 11.60 psid
11.60 psi
(10% to 90% of the supplied voltage)
0.5 to 4.5 V with a 5 Vdc supply
–40°F to 158°F (–40°C to 70°C) –4°F to 140°F (–20°C to 60°C) 37°F to 127°F (3°C to 53°C)
AXLdp
Pressure Transmitter
KEY BENEFITS
• OEM product
• Broad temperature capability
• High performance ASIC based electronics
• Superior long-term stability and repeatability
• Compact design: 1.18” x 1.65” x 0.37”
PHYSICAL SPECIFICATIONS
Pressure Connection:
Mating Electrical Connection:
Weight: 0.35 oz without leads
Environmental Rating:
³⁄³²˝ I.D. tubing
JST BHR-03VS-1
IP20, NEMA 1(meets UL94-V0)
All speci fications are subject to change without notice.
All sales subject to standard terms and conditions.
©2019 Ashcroft Inc. AXLdp_transmitter_ds1.0, Rev. B, 07/19
of 31
ashcroft.com
info@ashcroft.com
1.800.328.8258
Data Sheet
AXLdp Low Differential Pressure Transmitter
WETTED MATERIAL
Media
Clean, dry air/gases compatible with Silicon, Glass, Gold, Titanium,
Ceramic, Silicone Rubber, Aluminum and PBT
NOT FOR USE ON LIQUIDS
NON-WETTED
Housing
PBT (meets UL94-VO)
Featuring a highly reliable variable capacitance sensor using the patented Ashcroft® Si-Glass™ sensor. This ultra-thin single crystal diaphragm provides inherent sensor repeatability and stability.
Sensor Cross Section
The silicon diaphragm sensor has no glues or
other organics to contribute to drift or
mechanical degradation over time.
ORDERING CODE Example:
Model & Accuracy
AX7 - A XLdp Series, ±1.0% of span AX9 - A XLdp Series, ±2.0% of span AX9
Pressure Connection
MB3 - 3mm tube stube MB3
Output Signal
RM - 0.5-4.5 Vdc ratiometric ou tput RM
Electrical Termination
NC - No lead s ST - Leads with JST Connector ST
Pressure Range
Unidirectional Ranges (differential)
P2IW - 0.20 IWD P25IW - 0.25 IWD P5IW - 0.50 IWD P75IW - 0.75 IWD 1IW - 1.00 IWD 2IW - 2.00 IWD 2IW 2P5IW - 2.50 IWD 4IW - 4.00 IWD 5IW - 5.00 IWD 10IW - 10.00 IWD 15IW - 15.00 IWD 20IW - 20.00 IWD
Bi-directional Ranges
P1IWL - ±0.10 IWD P25IWL - ±0.25 IWD P5IWL - ±0.50 IWD 1IWL - ±1.00 IWD 2IWL - ±2.00 IWD 2P5IWL - ±2.50 IWD 5IWL - ±5.00 IWD 7P5IWL - ±7.50 IWD 10IWL - ±10.00 IWD
AX9 MB3 RM ST 2IW
Consult factory for custom calibration ranges and other temperature compensation ranges
All speci fications are subject to change without notice.
All sales subject to standard terms and conditions.
©2019 Ashcroft Inc. AXLdp_transmitter_ds1.0, Rev. B, 07/19
of 32
ashcroft.com
info@ashcroft.com
1.800.328.8258
Loading...
+ 1 hidden pages