Ashcroft AVC-1000 User manual

Page 1
Installation and Operating Instructions for Ashcroft
®
Pressure Volume Controllers, Models AVC-1000 and AVC-3000
INSTALLATION (see figure) Step 1 –
Step 2 –
Install a valve to the inlet port on the pressure balance valve assembly.
Step 3 –
Install a tee fitting to the rear of the AVC unit.
Step 4 –
Connect a pressure standard, for exam­ple a dial gauge or digital indicator to one side of the tee. To the other side of the tee connect the device to be calibrated or tested.
NOTE (Installation) –
All plumbing connections are made to
1
8 NPTF ports.
OPERATION Step 1 –
Open the pressure balance valve.
Step 2 –
Open the vent valve.
Step 3 –
Tu r
n the control knob to approximately midrange. This will allow for increase and decrease of system pressure. Alternatively, the starting position of the knob may be at the counterclockwise or clockwise stop. This will allow for maximum pressure increase (knob at the counterclockwise stop) or maximum pressure decrease (knob at clockwise stop).
Step 4 –
Close the vent valve.
Step 5 –
Open the inlet valv
e to provide pressure to approximately the desired pressure level.
Step 6 –
When the appropriate pressure has been achieved, close the inlet v
alve.
DEVICE UNDER
TEST
VENT VALVE
ATMOSPHERE
CONTROL KNOB
Step 7 –
Close pressure balance valve.
Step 8 –
Tu r n control knob clockwise to increase system pressure or countercloc decrease system pressure.
NOTE (Operation) –
During operation the pressure balance v
alve will automatically open if the pres­sure across the piston exceeds safe levels. This is factory set at a nominal pressure differential of 150 psi.
PRESSURE
MEASURING
DEVICE
INLET VALVE OR REGULATOR
REGULATOR
PRESSURE
AC
BALANCE VALVE
kwise to
PRESSURE SOURCE
All rights reserved. All sales subject to standard terms and conditions of sale. I&M002-10054-11/00 (A-250-2974B) 1CAMR 2/07
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Model AVC-1000 and 3000 Pressure Volume Controllers
PARTS LIST
ITEM QTY. PART
NO. PER ASSY. NUMBER DESCRIPTION
11 Final Assembly 2 1 603X002-01 Actuator (1000 psi unit) 2a 1 603X002-02 Actuator (3000 psi unit) 3 1 603X003-01 Lead Screw (1000 psi unit) 3a 1 603X003-02 Lead Screw (3000 psi unit) 4 1 603X004-01 Washer, Actuator 5 1 603X005-01 Body 6 1 603X006-01 Valve Body 7 1 603X007-01 End Cap 8 2 603X008-01 Guide Screw
9 3** 854X050-01 Pipe Plug 10 2** 605X003-09 Socket Head Screw (valve) 11 8 605X004-05 Socket Head Screw 12 1 605X005-01 Washer 13 2 607A001-02 O-Ring 14 3 607X001-04 O-Ring 15 3 607X001-05 O-Ring 16 1* 607X001-06 O-Ring 17 1 607X001-08 O-Ring (1000 psi unit) 17a 1 607X001-07 O-Ring (3000 psi unIt) 18 1 607A002-02 Back-up Ring 19 2 607X002-05 Back-up Ring 20 1 607X006-01 Retaining Ring 21 1 609X005-01 Spool Valve 22 1 609X006-01 Retainer Valve 23 1 614X003-01 Piston (1000 psi unit) 23a 1 614X003-02 Piston (3000 psi unit) 24 1 618X004-01 Label, Valve Control 25 1 618X005-03 Label (1000 psi unit) 25a 1 618X005-04 Label (3000 psi unit) 26 1 619X001-02 Knob, Adjustment 27 1 619X002-01 Knob, Balance Valve 28 1 620X001-01 Bearing, Ball Thrust 29 1** 617X006-01 Mounting Kit, Panel 30 1** 621X001-02 Table Mount Bracket(optional) 31 1 610X002-01 Sleeve (3000 psi unit)
**Quantity is 2 in model AVC-3000.
**Not shown.
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE
All rights reserved. All sales subject to standard terms and conditions of sale. I&M002-10054-11/00 (A-250-2974B) 1CAMR 2/07
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...