Ashcroft AUF User Instructions

Page 1
Ashcroft® Model AUF Comp act Local Electronic Display
User Instructions
Page 2
1.0 Note
Rev. Jan. 2002
Prior to unpacking and installation, please read the operating instructions and follow them carefully. These units are to be used, serviced, and repaired only by individuals who are familiar with the operating instructions and the applicable regulati ons for operational safety and accident prevention.
2.0 Intended Use
The plug-in display unit model AUF is a universally applicable, on-site display unit for a variety of measuring transducers. The measuring transducer is merely required to have an analog output and a plug-type connection in accordance with DIN 43 650. The plugin display unit is simply attached between plug and socket and is immediately ready for operation. Scale, the position of the decimal point, attenuation, and the operating position (option) are adjusted with two keys (upon removal of the cover). The individual menu steps ar e displayed on the LED display . The di splay rotates by 90°.
3.0 Control of Units
The units are checked before shipment and shipped in good condition. If you detect a visible defect on the unit, we recommend that you carefully check the packing material. In the event of a defect, please notify the mail service/fre ight forwarder, as the shipper is responsible for sh ipping damage.
4.0 Mechanical Connection
The plug-in display unit is simply connected between plug and socket and is immediately ready for operation. Please be sure to nstall the provided gasket between the sensing instrument connector and the AUF. The attachment scr ew must be replaced by t he longer s cr ew included in the shipment.
5.0 Electrical Connection
Attach the plug-in display unit as shown
+V Supply............PIN 1
-V Supply.............PIN 2
Switch outlet.........PIN 3
For units with Open Colle ctor output, connect PIN 2 and PIN 3.
Page 3
6.0 Setting
Rev. Jan. 2002
A
A=>downward and select i on of menu items
B=>upward and selection of menu items
A+B=>select menu item to set or to confirm setting
6.1 Decimal Point
Press B key until - d P - is displayed =>A+B menu item to set:- -.- ­=>B or A for up and down:- - -.­=>A+B to confirm setting and r et ur n t o menu item - d P –
6.2 Zero Point (value to be displayed at 4 mA)
Press B until - Z P 4 is displayed =>A+B menu item to set:- 0.0 0 for example (0 psig) =>B or A for up and down =>A+B to confirm setting and r et ur n t o menu item -Z P -
Page 4
6.3 Range (value to be displ ayed at 20 mA)
Rev. Jan. 2002
Press B key until E P 2 0 is displayed =>A+B menu item to set - 0 0 0 =>B or A for up and down 6 0.0 0 (60 psig) =>A+B to confirm setting and r et ur n t o menu item - E P –
6.4 Attenuation
Press B key until F I L t is displayed =>A+B menu item to set 0 0 0. 3 ( M in.= 0.3 sec.;Max.=20.0 sec.) =>B or A for up or down - - 1.5 (1.5 sec.) =>A+B to confirm setting and r et ur n t o menu item F I L t
6.5 Range Exceeded (dis played if below 4 mA or abov e 20 m A)
Display: “HI”=>range exceeded or “L0”=>below range Press B key until H I L 0 is displayed =>A+B menu item for setting: - 0 F F display inactive =>B or A for up or down: - 0 n - displ ay active =>A+B to confirm setting and r et ur n t o menu item H I L 0 Report - h I - = measuring r ange exceeded, - L 0 - = below measuring range Please note: If H I L 0 display is inactive, measuring range ex ceeded (-1999 to 9999) is indicated by “E r 0 6”
6.6 Operating Position (O pt ion)
Press B until - S P t is displayed =>A+B menu item for setting - 0.0 0 =>B or A for up or down: 2 0.0 0 (20 psig) =>A+B to confirm setting and return to menu item - S P t Please note: Standard hys t er esi s is the set point minus 3 digits Upon customer request, t he hysteresis is adjustabl e at t he plant as per request
Page 5
7.0 Operating Direction (Option)
Rev. Jan. 2002
Willy Instrumentos de Medição e Controle Ltda.
Press B key until d I r - is displayed =>A+B menu items for setting: - - d n (hysteresis below operat in g position) =>B or A for up or down: - - u P (hysteresis is above oper ating position) =>A+B to confirm setting and r et ur n t o menu item d I r -
8.0 Return to Measuring Mode
By pressing A or B key 1 to 6 times de pending on the selected me nu item.
(Uma Empresa Ashcroft Inc.) Rua João Pessoa, 620 – São Caetano do Sul – SP – Brasil – CEP: 09520-000 Phone: (55 11) 4224-7400 ● Fax: (55 11) 4224-7477 E-mail: vendas@ashcroft.com Web Site: www.ashcroft.com.br