Ashcroft A2X User Manual

A2X EXPLOSION/FLAME PROOF
.887˝
4.78˝
3.065˝
.675˝
1.06˝
DIA.
(22.5)
(121)
(78)
(17)
(27)
.887˝
4.78˝
3.065˝
.885˝
1.06˝ (27) Hex. Common For All Pressure Fittings
(22)
(78)
(121)
(22.5)
1.06˝
DIA.
(27)
PRESSURE TRANSMITTER INSTRUCTION SHEET
WARNING! READ

BEFORE INSTALLATION
denly stopped, as with quick closing solenoid valves. Surges occur when
low is suddenly begun, as when a
f pump is turned on at full power or a valve is quickly opened.
1. GENERAL:
A failure resulting in injury or dam-
ge
ay be caused by excessive
a
m overpressure, excessive vibration or pressure pulsation, excessive instru­ment temperature, corrosion of the pressure containing parts, or other misuse. Consult Ashcroft Inc., Stratford, Connecticut, USA before installing if there are any questions or concerns.
2. OVERPRESSURE:
Pressure spikes in excess of the rated overpressure capability of the trans­ducer may cause irreversible electri- cal and/or mechanical damage to the pressure measuring and containing elements.
Fluid hammer and surges can destroy any pressure transducer and must always be avoided. A pressure snub­ber should be installed to eliminate the damaging hammer effects. Fluid ham-
Liquid surges are particularly damag­ing to pressure transducers if the pipe is originally empty. To avoid damaging surges, fluid lines should remain full (if possible), pumps should be brought up to power slowly, and valves opened slowly. To avoid damage from both fluid hammer and surges, a surge chamber should be installed.
Symptoms of fluid hammer and surge's damaging effects:
• Pressure transducer exhibits an out­put at zero pressure (large zero offset).
• Pressure transducer output remains constant regardless of pressure
• In severe cases, there will be no output.
FREEZING:
Prohibit freezing of media in pressure port. Unit should be drained (mount in vertical position with electrical termina­tion upward) to prevent possible over­pressure damage from frozen media.
mer occurs when a liquid flow is sud-
3. STATIC ELECTRICAL CHARGES:
Any electrical device may be suscepti-
le to damage when exposed to static
b electrical charges. To avoid damage to the transducer observe the following:
• Operator/installer should follow the proper ESD (electrostatic discharge) protection procedures before handling the pressure transducer.
• Ground the body of the transducer BEFORE making any electrical connections
• When disconnecting, remove the ground LAST!
Note: The shield and drain wire in the cable (if supplied) is not connected to the transducer body, and is not a suit­able ground.
Ashcroft®A2X Pressure Transmitter, Typical Dimensions and Construction*
© 2011 Ashcroft Inc. All rights reserved. 250 East Main Street, Stratford, CT 06614 USA Tel: 203-378-8281, Fax: 203-385-0402 www.ashcroft.com All sales subject to standard terms and conditions of sale.
Dimensions and construction details may vary based on product specified.
*
I&M011-10167-A2X. Rev. 02/14
A2X EXPLOSION/FLAME PROOF
+
+
POWER SUPPLY
(+)
(–)
T
RANSDUCER
V+
V–
P
IN 1
P
IN 2
4-20mA
M
ETER
O
R OTHER DEVICE
+
+
POWER SUPPLY
METER
OR OTHER DEVICE
TRANSDUCER
(Common)
(+ Power)
(+ Output)
PIN 1
PIN 2
PIN 3
3-Wire Voltage
.830
FLYING LEADS
E LE CT RI CA L TE RM IN AT IO N
CONDUIT - 1/2 NPT MALE
(C2), (C5)
1.062
.830
PIGTAIL
E LE CT RI CA L TE RM IN AT IO N
CONDUIT - 1/2 NPT MALE
(C1), (P7)
1.062
PRESSURE TRANSMITTER INSTRUCTION SHEET
Mounting
The A2X transmitter requires no special mounting hardware, and can be mounted in any plane with negligible position error.
Although the unit can withstand normal vibration without damage or significant output effects, it is always good practice to mount the transducer where there is minimum vibration.
For units with NPT type pressure fittings apply Teflon
®
tape or an equivalent
sealant to the threads before installing.
When tightening, apply a wrench to the hex wrench flats located just above the pressure fitting. DO NOT tighten by using a pipe wrench on the housing.
Power Supply
Power Supply Voltage
Output Signal Min Max
0-5Vdc 10V 30V 1-5Vdc 10V 30V 1-6Vdc 10V 30V 0-10V 14V 30V 4-20mA* 12V 30V**
* For transmitters with 4-20mA output
signal, the minimum voltage at the temi­nals is 12Vdc. However, the minimum supply voltage should be calculated using the following graph and formula.
** For Intrinsically Safe Installations max,
supply voltage is 30Vdc. Refer to Ashcroft drawing #825A022 for wiring and installation requirements.
Noise
or minimum noise susceptibility, avoid
F running the transducer’s leads in a con­duit that contains high current AC power cables. Where possible avoid running the cable near inductive equipment.
Sintered Metal Filter
All units less than 500 psi include a small metal sintered filter at the top of the unit.
his is necessary to equalize the internal
T pressure with atmospheric pressure but can be a point of moisture ingress.
A2X Wiring Diagrams
A2X transducer has internal transient protection: for safety, limit line-to­ground voltage to 36 Vdc max.
HAZARDOUS AREA CERTIFICATIONS
Wire Voltage Current
olor Output Output
C
Red (+) Power (+) Power White (+) Output None Black (-) Power (-) Power
* Installation of conduit leads should be limit-
ed to 3m to maintain CE compliance
Wire Voltage Current
Color Output Output
Red (+) Power (+) Power White (+) Output None Black (-) Power (-) Power
Loop Supply Voltage vs. Loop Resistance
1000
30 Vdc max for
800
Intrinsically Safe Installations
600
400
Operating
200
Loop Resistance (Ohms)
Region
0
10
20 30
36
Loop Supply Voltage (Vdc)
Vmin = 12V+ (.022A x RL) (includes a 10% safety factor) RL = RS + RW RL = Loop Resistance (ohms) RS = Sense Resistance (ohms) RW = Wire Resistance (ohms)
© 2011 Ashcroft Inc. All rights reserved. 250 East Main Street, Stratford, CT 06614 USA Tel: 203-378-8281, Fax: 203-385-0402 www.ashcroft.com All sales subject to standard terms and conditions of sale.
Explosion Proof* – cUL: Specify A2X Class I, Div. 1 & 2, Groups A, B, C and D Class II, Div. 1 & 2, Groups E, F and G
Flame Proof* – ATEX: Specify A2X
Compliance with standards
EN 60079-0: 2012
and
EN 60079-1: 2007
II 2 G
Ex d IIC T4 Gb
-40°C Tamb. 125°C
DEMKO 04 ATEX 0237942
Intrinsically Safe (applies to 4-20mA) FM/CSA:
Intrinsic Safety: Class I, II and III Div.1 and 2,
Non-Incendive: Class I, II and III Div.1 and 2,
Groups A, B, C, D, F and G per entity requirements see Ashcroft drawing # 825A022
Groups A, B, C, D, F and G, no barriers needed
* Model A2X enclosure is intended for
installation using metallic conduit and requires installer to comply with appropri­ate codes to complete proper installation to meet the assigned hazardous area designation.
I&M011-10167-A2X. Rev. 02/14
SHIELDED CABLE (PIG TAIL)
ELECTRICAL TERMINATION
CONDUIT - 1/2NPT MALE
(C1), (C6), (C7), (P7)