Ashcroft 2279 User Manual

Installation and Maintenance Instructions for
A
PS
4-20mA
RED
R (250)
BLACK
+
+
+
X
V
PS
1-5V
RED
BLACK
+
X
SUPPLY VOLTAGE
LOAD RESISTANCE OHMS
1400
1
000
5
00
0
0 12 24 36 40
ASHCROFT®Type 2279 Duratran®Pressure Transmitter
igure 1 –
F
X = TRANSMITTER PS = POWER SUPPLY (16-40V max) R=250 RESISTOR A=AMMETER (to 20mA) V=VOLTMETER (to 5V)
Figure 2 –
OPERATING PRINCIPLE
This instrument is based on pressure measuring systems that have made Ashcroft Duragauge
®
gauges the standard of perfor-
mance throughout the world.
The Bourdon tube in this instrument drives the geared mecha­nism and pointer in traditional pressure gauge fashion. The indi­cating portion is therefore calibrated and maintained in accor­dance with good pressure measurement practice.
The same Bourdon tube also positions a unique self-regulating photo-optic position sensing system to produce a dc electronic signal proportional to the pressure value. This signal is condi­tioned to a standard 4-20mA dc output over two wires, powered by a dc supply between 12 and 40 volts. Since there is no mechanical sliding or other frictional interaction within the non­contacting photo-optic system, the transmitting function is highly reliable and not subject to wear.
INSTALLATION PRECAUTIONS
To get the maximum value and performance there are a few good installation practices to follow:
1) Pulsation, Vibration and Shock – The excellent sensitivity
of the photo-optic system will reliably sense and transmit any motions of the Bourdon tube, whether caused by pressure changes or installation seismic effects. Therefore, please mount the instrument on a stable surface or piping system free from vibrations and mechanical shock and – if pressure pulsations are present that you don’t want to transmit – use a suitable snubber or throttling valve to exclude these effects. Metal tubing is recommended to pipe pressure to the instrument. The Ashcroft 1115A armoured capillary line with fittings is available to remote mount the Duratran.
© 2007 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com All sales subject to standard terms and conditions of sale.
I&M00810043-9/01 (Form No. 250-2598) Rev. 10/12
2) Process Fluid Effects – Make sure the process fluid will not
corrode or overheat the wetted materials of the pressure measuring systems. These materials are clearly marked on the dial face. Ashcroft chemical seals are readily available for corrosive fluids.
3) Ambient Temperature – Most accurate results will be
obtained if the instrument is mounted indoors, although it is well protected with hermetic sealing to operate outdoors over a range of –40° to 160°F with some accompanying error (see specifications).
4) Electrical fitting – Avoid disturbing the hermetic seal on the
electrical connection by holding the wrench flats against rota­tion when installing electrical fittings.
5) Pressure Fitting – Always use the wrench flats provided when connecting piping to the transmitter – DO NOT USE
THE HOUSING AS A HANDLE.
OUTPUT CONNECTIONS – See Figure 1.
Voltage – Internal regulation permits accurate transmission
within widely varying applied dc charge – consult the load resis­tance vs voltage curve in Figure 1.
Load Resistance – Maximum allowable load resistance for a given supply voltage can also be determined from Figure 2.
Polarity – Although the transmitter is protected against reverse polarity, there will be no usable signal unless proper dc polarity is observed. NOTE THAT THE RED LEAD IS POSITIVE (+).
Installation and Maintenance Instructions for ASHCROFT®Type 2279 Duratran®Pressure Transmitter
CALIBRATION
If it is necessary to rezero the instrument, (i.e. to correct for head effect in the piping), this may be done by removing the front ring and window. The indicating pointer is set by the micrometer adjustable hub. Zero and span adjustments are accessible from the front through holes in the dial; use a small screwdriver for this. The case should only be opened in a dry and clean environment, preferably in an instrument shop.
Complete recalibration may be required after extended service
or if the instrument has been overpressured.
Calibration should be checked against a test gauge and pres­sure source, or a deadweight tester, appropriate to the range of the transmitter. ANSI B40.1, “Gauges – Pressure Indicating Dial Type – Elastic Element,” recommends that the pressure stan­dard be at least four times as accurate as the instrument being tested. A power supply and milliameter, or voltmeter and resis­tor, of suitable accuracy are also required to test the transmitter output. These are to be connected in series with the output leads, observing correct polarity, as in Figure 1.
With the input pressure at zero, the micrometer pointer adjust­ment and the “zero” potentiometer should be adjusted for zero indication and 4mA output respectively. The pressure should then be raised to full scale and, if the output is not 20mA, the “span” potentiometer used to correct the span error. The zero should then be rechecked. If an indication span error is found, it is necessary to open the back of the case and adjust the slide. More information on indicator calibration is available in Ashcroft Maintenance and Repair Manual. visit www.ashcroft.com for the most up to date version. When the indication span and zero are correct, replace the back of the case prior to final calibration of the transmitter. Caution: do not adjust the “zero” and “span”
potentiometers while the back of the case is open; ambient light may cause erratic results. Mechanical zero/span adjust-
ment should normally not require any electrical adjustment.
The front window and rear cover must be firmly and evenly tightened to prevent moisture or corrosives from damaging the circuit board.
© 2007 Ashcroft Inc., 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com All sales subject to standard terms and conditions of sale.
I&M00810043-9/01 (Form No. 250-2598) Rev. 10/12
Loading...