Ashcroft 2089, 2086, 2084 Specifications

Data Sheet
FEATURES
 Full Scale Total Error Band (TEB) includes the
effects of linearity, hysteresis, repeatability and
temperature from 0°F to 150°F (-18°C to 63°C)
 Weatherproof IP65  Large LCD display  Seven languages  Field calibration capability
TYPICAL USES
 Test bench / gauge repair facilities  Calibration laboratories  Mobile calibration labs  Test measurement and validation
SPECIFICATIONS
Accuracy: 2089: ±0.05% of span accuracy
2086: ±0.10% of span accuracy 2084: ±0.25% of span accuracy Terminal Point, Total Error Band (TEB) Including Hysteresis, Linearity, Repeatability and Temperature (0°F to 150°F)
Size:
Case Material: 300 Stainless Steel
Case Finish: Tumble polished
Case: Weatherproof ABS gauge carrying case
Process Connection:1⁄4 NPT Male
Process Connection Location:
Mounting: Stem
Ranges: Gauge, compound, absolute and vacuum thru
Temperature Corrected: Ye s
Calibration Chart: 10 NIST Traceable chart included with 2089,
Enclosure Rating: Weatherproof, IP65
WETTED COMPONENTS
Model Process Connection and Sensor
2084, 2086, 2089 316 SS
NON-WETTED COMPONENTS
Model Case
2084, 2086, 2089 304 SS
MIN/MAX TEMPERATURE LIMITS
Version Process Storage
2084, 2086, 2089
Lower, left or right
7,000 psi (see engineering units on page 3 for other units of measurement)
optional for 2084 and 2086
0°F to 150°F
(–18°C to 65°C)
–40°F to 180°F
(–40°C to 82°C)
2084, 2086 & 2089
3˝ dial size
KEY BENEFITS
• Highly accurate test instruments
• Silicone MEMS sensor technology provides accuracy and long term stability
• Rugged and portable design
• Twelve engineering units
DISPLAY
Type: LCD
Display Digits: 5
Character Height: 0.66˝
Backlight: Off by default, programmable
Manually turns backlight on and off
Bar Graph: 10 segment
Battery Life:
1,000 hrs. (3 AAA alkaline batteries)
All specifications are subject to change without notice.
All sales subject to standard terms and conditions.
©2018 Ashcroft Inc. 2084-2086-2089_ds_RevA_Ltr_05-19
ashcroft.com
info@ashcroft.com
1.800.328.8258
of 41
Data Sheet
PRESSURE RATINGS
Overpressure: Proof: Burst:
Vac to ≤300 psi
≥300 to ≤3,000 psi
≥5,000 to ≤7,000 psi
AGENCY APPROVALS
FM and CSA Intrinsically Safe. FM Class I to III, Div. 1 Groups A-G CSA Class I, Div. 1, Groups A-D, Class II Div. 1, Groups E-G, Class 2 Div. 2 Groups F and G, Class III
Note: FM/CSA approval not valid on vacuum and 15# & vacuum ranges
2 X Range 2 X Range 2 X Range
3 X Range 5 X Range 2 X Range
ORDERING CODE Example:
Dial Size
30 - 3.0˝ 30
Model (Accuracy)
2089 - ±0.05% of span (includes NIST traceable calibration certificate) 2089
2086 - ±0.10% of span
2084 - ±0.25% of span
System
S - 316 Stainless Steel S
Case
D - Dry D
Process Connection Size
02 - 1⁄4 NPT Male 02
Process Connection Location
L - Lower L
E - Left (9 o’clock)
D - Right (3 o’clock)
T - Top
Range (coding example, see range table below for all standard ranges)
1,000# - 0-1,000 psi 1000#
Options (if choosing an option(s) must include a “X”) -X__
6B - Cleaned for oxygen service
6D - Cleaned for liquid oxygen service
C4 - Individual certified calibration chart (included on 2089 no code required)
S7 - Weatherproof ABS gauge carrying case S7
FF - Flange for panel mounting
B1 - Protective rubber boot (black)
B2 - Protective rubber boot (orange)
NH - Metal tag wired to case
NN - Paper tag bonded to case
MF - Free from mercury certificate
TU - Throttle plug
MN - Mylar tag bonded to case
30 2089 S D 02 L 1000# -XS7
Consult factory for calibration charts, throttle plugs, free from mercury certificate, and more range options
All specifications are subject to change without notice.
All sales subject to standard terms and conditions.
©2018 Ashcroft Inc. 2084-2086-2089_ds_RevA_Ltr_05-19
ashcroft.com
info@ashcroft.com
1.800.328.8258
of 42
Loading...
+ 2 hidden pages