Aseptico ARU-HF70 Maintenance manual

P.O. Box 1548 Woodinville, WA 98072-1548
1-800-426-5913 * 425-487-3157 * Fax: 360-668-8722
email: info@aseptico.com * Internet: www.aseptico.com
P/N 420310 Rev. A 04/01
OPERATION AND
MAINTENANCE MANUAL
NSN: Pending
Page 2
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Items Furnished . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Exposure Timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Peak Tube Potential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Tube Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Testing Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Operation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Component Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-16
Technique Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Table of Illustrations
Figure A - Component Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Figure B - Tripod Attachment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Figure C - PC Board M3115 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Figure D - PC Board M9142 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Figure E - Cord Attachments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Figure F - Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Chart A - Testing Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Figure G - Absolute Maximum Rating Charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Figure H - Emission and Filament Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Figure I - Anode Thermal Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Figure J - D -0811 Dimensional Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Figure K - D -0811S Dimensional Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Figure L - Schematic Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Upon Arrival
When the equipment is received, each shipping container should be carefully examined for any evidence of mishandling during shipment. Note its condition, if abnormal, carefully unpack all parts and examine for damage. If any damage is noted, immediately report it to the carrier in the proper manner. All printed matter supplied with the unit should be saved for installation, operation, and future reference.
Page 3
Introduction
Purpose
The ARU-HF70D Command Air X-ray System is a mono-block type high frequency portable dental x-ray unit. This system is compatible with conventional intraoral dental film or with solid­state digital radiographic intraoral sensors and provides a portable and practical way to obtain dental radiographs in the field. To receive the best service and longest life from your Aseptico product, follow the instructions detailed in this manual.
Performance characteristics
Unit Weight: 9.25 lbs. (4.2 kgs.) Unit Dimensions: 8.25”W x 6”H x 10.75”L, plus dental cone
(10.2 cm x 15.2 cm x 20.3 cm) X-ray Output: 10 mA @ 60, 65, 70 kVDC Exposure timer: 0.02 - 1.98 seconds, 99 steps Line Voltage: 110-130V, 50/60 Hz
220-260V, 50/60 Hz Generator Rating: 10 mA at 70 kVDC ±15% Current Draw @ 10 mA, 70 kVDC output: 15 Amps Inverter Frequency: 75 kHz Maximum Deviation from Fixed Factor: ±10% (mA), ±10% (kV),
±20% (Time: 0.02-0.20 sec.),
±10% (Time: 0.22-1.98 sec.) X-ray tube: Toshiba D-0811 or equivalent Focal Spot: 0.8 mm Exposure Switch: Two-stage, deadman with RJ11 connector Exposure Cord: 8 feet (2.8 m) Power cord: 20 feet (6.1 m), 14/3 wire oil resistant Dental Cone (Beam Limiting Device) Yoshida BL -70 or equivalent Focus to Cone Tip Distance: 7.95” (20.2 cm)
Items furnished
X-ray tubehead and control Mounting handle Dental cone Scatter shield Exposure cord Power cord Mil-spec case with shock resistant foam insert ARU-01S tripod Vinyl carrying case for tripod Manual
Page 4
Safety Precautions
WARNING - Careless or improper use of x-ray equipment can be extremely hazardous. WARNING - It is imperative that this equipment be operated and serviced only by trained
personnel familiar with the safety precautions required to prevent excessive exposure to primary x-ray radiation, the dangers of exposure to x-ray radiation, and the proper use of the equipment discussed in this manual.
I. During exposure, the operator must stand as far as possible from the patient being x-
rayed and should wear a lead apron or stand behind a lead shield.
II. The operator must not stand in the primary x-ray beam. III. The operator must wear a monitoring badge while operating this unit. It should be on the
collar, not on an area covered by the lead apron.
IV. X-ray exposure should be as short as possible.
All personnel authorized to operate or service this equipment should be fully acquainted with the established maximum permissible doses, safety recommendations, and procedures derived from the following sources:
A. National Council on Radiation Protection Report No. 33 (Medical X-ray and Gamma Ray
Protection for Energies up to 10 MEV - Equipment Design and Use); from NCRP Publications; P.O. Box 30175, Washington, D.C. 20014.
B. National Bureau of Standards Handbook No. 76 (Medical X-ray Protection up to Three
Million Volts); from the Superintendent of Documents, Government Printing Office, Washington, D.C. 20401.
C. All documents relating to the Performance Standard for Diagnostic X-ray Systems, 21
CFR Subchapter J, Part 1020; obtainable from FDA Center for Devices and Radiological Health, Department of HHS, 2098 Gaither Road, Rockville, MD 20850.
D. State and local regulations governing radiation protection and the use of diagnostic x-ray
equipment.
E. Requirements of the user's in-house radiation protection program. F. Instructions and precautionary notices of this manual.
Although this equipment incorporates protective design features for limiting both the direct (primary) x-ray beam and the secondary radiation produced by this beam, design factors alone cannot prevent human carelessness, negligence, or lack of knowledge. This apparatus is sold with the understanding that the user assumes sole responsibility for radiation safety and that MinXray, Inc. (the manufacturer) and Aseptico, Inc. (the distributor), its agent and representatives, do not accept any responsibility for:
1. Injury or danger to patient or other personnel from x-ray exposure.
2. Overexposure due to poor operating techniques or procedures.
3. Equipment not properly serviced or maintained in accordance with this manual.
4. Equipment which has been modified or tampered with in any way.
5. Equipment that has been connected to input voltage other than the voltage specified on the rating label affixed to the x-ray unit.
Page 5
X-ray tubehead and control Mounting handle Dental cone Scatter shield Exposure cord Power cord ARU-01S tripod
Includes: #510311 5/16 - 18 wing-nut #510322 5/16 fender washer
Vinyl carrying case Mil-spec case with shock
resistant foam insert
1
2
3
4
1
2
3
4
Figure A - Component Identification
Assembly
1. Remove the ARU-HF70D x-ray tubehead and control, dental cone, scatter shield, exposure cord, and power cord from the carrying case. (See Fig. A above)
2. If the ARU-HF70D will be attached to the ARU-01S tripod: A. Remove the tripod from its vinyl carrying case. B. Rotate the carrying handle of the ARU-HF70D around the
rear of the unit so it is positioned under the unit.
C. Insert the threaded stud on the tripod center post through
the hole in the center of the ARU-HF70D carrying handle. D. Place the washer over the carrying handle on the threaded stud. E. Affix the ARU-HF70D to the tripod by screwing the wing-nut
onto the threaded stud. (See Fig. B at right) F. Adjust the length of each telescoping tripod leg and lock it. G. Adjust the height of the center post and lock it.
3. Carefully screw the dental cone into the coneholder on the front of the ARU-HF70D. Avoid crossthreading the cone when attaching it to the unit.to prevent leakage radiation and distortion of the x-ray field. NEVER OPERATE THE ARU-HF70D UNLESS THE DENTAL CONE IS
ATTACHED.
4. Connect the leaded scatter shield to the end of the dental cone. The scatter shield should be used if the ARU-HF70D is to be configured for handheld operation.
5. Connect the AC power cord to the socket on the back of the ARU-HF70D. Note the rated voltage of the unit before connecting it to a power source
6. Connect the exposure cord to the RJ11 telephone-type socket on the back of the ARU-HF70D.
7. Plug the AC power cord into a grounded wall outlet. The unit is now assembled and ready for testing.
5
6
7
8
5
6
7
8
9
9
Figure B - Tripod Attachment
Page 6
Performance Verification
The following tests must be conducted without fail before the ARU-HF70D Command Air X-Ray System can be used for radiology. Fill out the checklist for these tests when installation of x-ray unit is complete.
1. Exposure Timer
1-1 Test method:
1. Set the exposure timer to 0.02 seconds.
2. Set output to 70 kV.
3. Set NERO to the following settings: SID: 18 inches; Wheel range: 50 ~ 85 kV; Sensitivity: L ow; Phase select: 1Ø;
Measurement mode: STAT; X-ray values: 70 kV, 0.02 seconds.
4. Make exposure and read the measured value of exposure time on the NERO.
5. Measure value of exposure time at the following settings with the same procedure:
0.02, 0.04, 0.10, 0.20, 0.50, and 1.00 seconds.
1-2 Instruments: NERO Model 6000M X-ray Beam Analyzer
Manufactured by Victoreen, Inc. or equivalent
1-3 Rejection limit: ±20% (0.02 sec. ~ 0.2 sec.), ±10% (0.22 sec. ~ 1.98 sec.)
If the timer needs adjustment:
1. Remove the 4 screws on the side of the top cover of the ARU-HF70D and lift the top off of the body of the unit.
2. Adjust the potentiometer “VR1” on the printed circuit board M3115 on the inside of the top cover (see Fig. C below) so that the radiation exposure time equals the indicated time value, taking into consideration the accuracy tolerances.
3. Replace the top cover and the 4 screws.
Figure C - PC Board M3115. Exposure Timer Adjustment (VR 1)
VR 1
Loading...
+ 11 hidden pages