This Quick Reference Guide is a brief
summary of the operation and routine
maintenance of the AUTION ELEVEN
Semi-Automated Urine Analyzer.
Read the Operating Manual before
using the instrument.
Wear protective gloves to prevent
exposure to pathogenic microbes.
Page 3
System Overview
System Overview
Page 4
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Setup (test strip, measurement No., patient ID)
Measurement
with
[ Select test strip type ][ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANTDo NOT wet the black mark. After dipping the test strip in the sample, remove excess urine by letting the strip touch the edge of the collection cup.
Set the measurement number
Set the patient ID
(up to 13 digits)
,
(up to 4 digits)
AE-4022
™
System Overview
System Overview
1
Page 5
Sample Preparation
Sample Preparation
Page 6
SEMI-AUTOMATED URINE ANALYZER
Power switch
Turn off the instrument.
Open the maintenance cover.
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Test strip feed tray
Test strip tray
Detach the test strip tray.
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Test strip detection
window
Setup (test strip, measurement No., patient ID)
Measurement
with
buttonsSelect
[ Select test strip type ][ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANTDo NOT wet the black mark. After dipping the test strip in the sample, remove excess urine by letting the strip touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test strip detection window, crossing the two white rails of the test strip tray. The carrying arm sends the test strip into the analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 . Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds for two seconds to signal the timing for dipping the next test strip. Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Waste box
Do NOT wet the black mark.
Test strip
Rails (white)
Test strip detection window
Detection area
(up to 13 digits)
,
(up to 4 digits)
2.3 Preparation for measurement
AUTION ELEVEN
AE-4022
™
1. Turn on power switch at back of instrument.
2. After a 20 second warm up the standby screen will appear.
3. Prepare suffi cient volume of sample in urine collection cup so entire test strip pad
area can be dipped in a single movement.
4. Mix well.
IMPORTANT:
Do not centrifuge samples.
Centrifugal force may cause undesirable sedimentation of blood cells and/or other
sample contents, thus causing several measurement items to have incorrect results.
Sample Preparation
Test strip
Test strip
pad area
2
Page 7
Sample Measurement
Sample Measurement
Page 8
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
Select the type of test strip
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
[ Select test strip type ][ Menu screen ]
Select the type of test strip
Set the measurement number
(up to 4 digits)
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
[ Select test strip type ][ Menu screen ]
Select the type of test strip
Set the measurement number
Set the patient ID
(up to 13 digits)
,
(up to 4 digits)
2.3 Preparation for measurement
2.3 Preparation for measurement
AE-4022
1. Enter operator ID number (if function is ON).
™
Sample Measurement
• From the Standby Screen, press and hold
• Enter operator ID number using numeric and [-] keys or using barcode reader.
• Press
2. Confi rm display shows 10EA as test strip type.
3. Set measurement number (optional).
• If using 0001 for fi rst measurement number, go to step 4.
• Enter new number using keypad
• Press
to confi rm entry.
to confi rm entry.
for 3 seconds.
(up to 4 digits).
4. Set patient ID.
• Press
• Enter up to 13 digits using numeric and [-] keys or scan barcoded sample
with barcode reader.
• Press
• The measurement number and patient ID number will display alternately
on screen.
to confi rm entry.
Continued on next page…
3
Page 9
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
[ Select test strip type ][ Menu screen ]
Select the type of test strip
Do NOT wet the
black mark.
of the bottle, immediately closethe cap or the test strips in thebottle will absorb moisture or dirtin the air, which may deterioratereagents and make the teststrips useless.
stage are required, depending on the setting of the “opera-tional mode when turning ON” parameter. (See “Operational mode” on page 2-3.)
standard setting is “Auto start” mode. To change the “operational mode when turning ON” parameter, see “3-7-8. No.005: Operational mode when turning ON” on page 3-18.
intervals to signal the dipping timing for the test strips.
2.4 Measurement operation
AE-4022
™
Sample Measurement
5. Confi rm normal measurement icon
appears on standby screen.
6. Enter Clarity Value (If feature is ON).
• Press and hold
• Enter the clarity value of sample using numeric keys
• Enter a number between 1 and 5 using table below:
12345
ClearSlightly
• Press
to confi rm entry.
7. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
Slow beeping
Preparation for dipping:
3 seconds
for about 2 seconds.
CloudyTurbidOther
Cloudy
Fast beeping
Dipping period:
2 seconds
Test strip setting:
1.4 seconds
IMPORTANT:
• Do NOT wet the black
mark on test strip.
• After dipping test strip in
sample, gently remove
excess urine by drawing
the entire length of strip
against the rim of the
collection cup.
Continued on next page…
4
Page 10
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Do NOT wet the black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
AE-4022
8. Place test strip on test strip tray.
• Place strip in the detection area of the test strip detection window
with the black marker on right or left.
• The carrying arm sends the test strip into the analyzer and
measurement starts automatically.
™
Sample Measurement
9. When buzzer setting is ON, the beeper sounds for two seconds to
When measuring continuously, repeat steps 2 to 4.
Set measurement numbers and patient IDs as necessary.
10. Read test results from printout.
11. Turn off instrument after completing all measurement procedures.
signal the timing for dipping the next test strip. Dip the test strip
while you hear the beep.
5
Page 11
Control Measurement
Control Measurement
Page 12
SEMI-AUTOMATED URINE ANALYZER
buttonsSelect
buttonsSelect
of the bottle, immediately closethe cap or the test strips in thebottle will absorb moisture or dirtin the air, which may deterioratereagents and make the teststrips useless.
stage are required, depending on the setting of the “opera-tional mode when turning ON” parameter. (See “Operational mode” on page 2-3.)
standard setting is “Auto start” mode. To change the “operational mode when turning ON” parameter, see “3-7-8. No.005: Operational mode when turning ON” on page 3-18.
intervals to signal the dipping timing for the test strips.
2.4 Measurement operation
Pad area
AUTION ELEVEN
AE-4022
1. Change measurement mode.
• From the [standby screen] press
2. Enter measurement number to identify control level.
3. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
Slow beeping
Preparation for dipping:
3 seconds
4. Place test strip on test strip tray.
™
Fast beeping
Dipping period:
2 seconds
or
Test strip setting:
1.4 seconds
until the
Control Measurement
icon is displayed.
• Place strip in the detection area of the test strip detection window with the black marker on right or left.
• The carrying arm sends the test strip into the analyzer and measurement starts automatically.
5. When buzzer setting is ON, the beeper sounds for two seconds to signal the timing for dipping the next test strip.
Dip the test strip while you hear the beep.
When measuring continuously, repeat steps 2 to 4.
Set measurement numbers as necessary.
6. Read test results from printout.
7. Turn off instrument after completing all measurement procedures.
6
Page 13
Check Measurement
Check Measurement
Page 14
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
buttonsSelect
buttonsSelect
Detection area
2.4 Measurement operation
AE-4022
™
1. Clean the feeder and waste box to prevent check strips from becoming soiled.
2. Change measurement mode.
Check Measurement
• From the [standby screen] press
3. Place white check strip on test strip tray and perform measurement.
4. Place gray check strip on test strip tray and perform measurement.
5. Remove check strips from waste box and return to check strip bottle.
6. Evaluate results
• Compare the refl ectance values on the test result printout to the ranges on the refl ectance label on the check strip bottle.
• If any of the measurement results are outside of the specfi ed ranges, repeat check measurement using the other set of
• If any of the measurement results are still outside the ranges, contact ARKRAY Technical Support at 1.855.646.3108.
check strips.
or
until the
icon is displayed.
7
Page 15
Maintenance
Maintenance
Page 16
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Turn off the instrument.
Open the maintenance cover.
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Test strip feed tray
Test strip tray
Detach the test strip tray.
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used test strips and protective gloves separately from
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat-ing Manual.
Clean the area circled in red below.
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Open the maintenance cover.
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Suction port
Test strip tray
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat-ing Manual.
Clean the area circled in red below.
Test strip tray
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used test strips and protective gloves separately from general waste according to local regulations on biohazardous waste.
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat-ing Manual.
Clean the area circled in red below.
Suction port
Test strip tray
Test strip tray
AE-4022
™
Maintenance – Daily
Clean the Feeder
1. Turn off instrument.2. Detach carrying arm.
4. Detach test strip tray.
• Slide the test strip tray to the
Power switch
front and lift up to detach.
Test strip tray
• Pull towards you and
lift up to remove.
5. Detach suction port.
• Pull suction port straight
up to detach.
Suction port
Test strip tray
3 Open maintenance cover.
• Press and hold the two open
buttons on either side of
instrument and lift up to open.
6. Detach test strip feed tray.
• Slide to front and lift up
to detach.
Continued on next page…
8
Page 17
SEMI-AUTOMATED URINE ANALYZER
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Suction port
Test strip feed tray
Test strip tray
Lever
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wear protective gloves to prevent exposure to pathogenic microbes.
Feed lever
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat-ing Manual.
AUTION ELEVEN
AE-4022
Clean the Feeder
7. Detach feed lever by sliding
the white lever in the center
towards you.
Lever
Feed lever
™
8. Clean the table.
• Clean the areas circled in red below with water.
Maintenance – Daily
Continued on next page…
9
Page 18
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Test strip feed tray
Test strip tray
Suction port
Suction port
Test strip feed tray
Suction port
Test strip feed tray
Test strip tray
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
AE-4022
Clean the Feeder
9. Wash detached test strip tray, suction port, test strip feed tray and feed lever with water.
™
Maintenance – Daily
10. Sterilize the carrying arm using alcohol. Then wash with water.
11. Wipe the parts with a cloth and dry completely before reattaching to instrument.
10
Page 19
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Waste box
Waste box
4.1 Daily maintenance
AE-4022
Clean the Waste Box
™
Maintenance – Daily
1. Remove waste box and
discard used test strips.
2. Sterilize and clean waste box.
• Sterilize with alcohol then
wash with water and wipe dry.
3. Attach waste box.
11
Page 20
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Printer cover
Take out.
Cut using scissors.
4.2 Replacing the thermal recording paper
4.2 Replacing the thermal recording paper
Printer cover
Take out.
Cut using scissors.
Printer cover
Take out.
Cut using scissors.
4.2 Replacing the thermal recording paper
AE-4022
™
Replacing thermal recording paper
Maintenance – As Needed
1. Check that the [Standby screen] is displayed,
2. Press
3. Cut the fi rst (outermost) layer of a new paper roll and trim
4. Place new roll of thermal recording paper in paper holder so
5. Close printer cover.
NOTE:
then open the printer cover.
to remove any remaining paper in
When paper is removed from printer,
error message E004 appears and a
beep sounds for 1 minute. The alarm
can be stopped by pressing
the printer.
the end to make it straight. A skewed or odd-shaped end
may cause the paper to jam.
that paper feeds from the bottom. Guide the trimmed end of the
thermal recording paper into and through the slot that receives
the recording paper. The thermal recording paper is automatically
wound forward and fed into the unit.
NOTE:
Always press the
paper. If you do not fully feed the thermal recording paper through
the unit by pressing the
measurement results.
button after replacing the thermal recording
button, the printer will fail to print the
.
Correct Incorrect Incorrect
Paper Holder
Set the new roll of thermal recording
paper in the paper holder in the
proper orientation so that the end
feeds from the bottom.
12
Page 21
Warnings
Warnings
Page 22
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Warning number
AE-4022
™
Warnings
A warning message appears when a normal measurement result is not obtained.
When a problem occurs during measurement, the system continues
measurement and prints warning messages with the measurement results.
The table below explains the meaning of and measures (corrective action) to be
taken for each warning message.
W004No dippingThe entire pad section of the test strip was not dipped.
W005Wrong test stripA test strip other than that specifi ed was used.
W006Transport errorThe test strip may not have been transported properly.
W007
W008Elapse of QC deadlineThe QC deadline has elapsed. Perform control measurements.
Abnormally high
refl ectance
Incorrect test strip
position
Number of registered
operator ID numbers
The ambient light level around the instrument changed
rapidly, due to a photofl ash or other reasons.
An abnormal sample (e.g. a sample containing
medicine) was measured.
• The test strip was displaced from its correct position
due to vibration.
• The test strips were incorrectly placed.
150 operator ID numbers have been registered.
Maintain a constant level of illumination
around the instrument during measurement.
Check if the tested sample is abnormal or
not. If the sample is normal, test it again.
• Do not expose the instrument to any
vibration during measurement.
• Place the test strips correctly on the test
strip tray.
Dip a new test strip in the sample correctly
and test again.
Only use ARKRAY’s test strips designed for
the AUTION ELEVEN.
• Clean the test strip tray and suction ports.
• Attach the suction ports properly.
• Clean the incoming strip sensor window.
Delete unnecessary operator ID numbers
before registering new ones.
13
Page 23
Errors
Errors
Page 24
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
4.2 Replacing the thermal recording paper
4.2 Replacing the thermal recording paper
5.2 Error messages
AE-4022
™
Errors may occur when you operate the instrument incorrectly or make a mistake
during normal operation. When an error occurs, an error message appears on the
screen and an alarm sounds for approximately 1 minute.
Errors
Press
to cancel the error notice. The display returns to the [Standby screen].
To prevent the error from recurring, take appropriate measures (corrective action),
referring to the table below. If unable to clear the error, note error code and call
ARKRAY Technical Support at 1.855.646.3108.