arkray AE-4022 User guide

Page 1
SEMI-AUTOMATED URINE ANALYZER
AE-4022
AUTION ELEVEN
Quick Reference Guide
Page 2
Table of Contents
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
AE-4022
System Overview ........................ 1
Sample Preparation .................... 2
Control Measurement ............... 6
Check Measurement ................. 7
Maintenance ......................... 8 -12
Cleaning the Feeder .............8-10
Cleaning the Waste Box ..........11
Replacing Thermal Paper ....... 12
Warnings .................................. 13
Errors .................................... 14-15
This Quick Reference Guide is a brief summary of the operation and routine maintenance of the AUTION ELEVEN Semi-Automated Urine Analyzer.
Read the Operating Manual before using the instrument.
Wear protective gloves to prevent
exposure to pathogenic microbes.
Page 3
System Overview
System Overview
Page 4
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Setup (test strip, measurement No., patient ID)
Measurement
with
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT Do NOT wet the black mark. After dipping the test strip in the sample, remove excess urine by letting the strip touch the edge of the collection cup.
Set the measurement number
Set the patient ID
(up to 13 digits)
,
(up to 4 digits)
AE-4022
System Overview
System Overview
1
Page 5
Sample Preparation
Sample Preparation
Page 6
SEMI-AUTOMATED URINE ANALYZER
Power switch
Turn off the instrument.
Open the maintenance cover.
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Test strip feed tray
Test strip tray
Detach the test strip tray.
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Test strip detection
window
Setup (test strip, measurement No., patient ID)
Measurement
with
buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT Do NOT wet the black mark. After dipping the test strip in the sample, remove excess urine by letting the strip touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test strip detection window, crossing the two white rails of the test strip tray. The carrying arm sends the test strip into the analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 . Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds for two seconds to signal the timing for dipping the next test strip. Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Waste box
Do NOT wet the black mark.
Test strip
Rails (white)
Test strip detection window
Detection area
(up to 13 digits)
,
(up to 4 digits)
2.3 Preparation for measurement
AUTION ELEVEN
AE-4022
1. Turn on power switch at back of instrument.
2. After a 20 second warm up the standby screen will appear.
3. Prepare suffi cient volume of sample in urine collection cup so entire test strip pad area can be dipped in a single movement.
4. Mix well.
IMPORTANT:
Do not centrifuge samples.
Centrifugal force may cause undesirable sedimentation of blood cells and/or other sample contents, thus causing several measurement items to have incorrect results.
Sample Preparation
Test strip
Test strip pad area
2
Page 7
Sample Measurement
Sample Measurement
Page 8
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
Select the type of test strip
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Set the measurement number
(up to 4 digits)
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Set the measurement number
Set the patient ID
(up to 13 digits)
,
(up to 4 digits)
2.3 Preparation for measurement
2.3 Preparation for measurement
AE-4022
1. Enter operator ID number (if function is ON).
Sample Measurement
• From the Standby Screen, press and hold
• Enter operator ID number using numeric and [-] keys or using barcode reader.
• Press
2. Confi rm display shows 10EA as test strip type.
3. Set measurement number (optional).
• If using 0001 for fi rst measurement number, go to step 4.
• Enter new number using keypad
• Press
to confi rm entry.
to confi rm entry.
for 3 seconds.
(up to 4 digits).
4. Set patient ID.
• Press
• Enter up to 13 digits using numeric and [-] keys or scan barcoded sample with barcode reader.
• Press
• The measurement number and patient ID number will display alternately on screen.
to confi rm entry.
Continued on next page…
3
Page 9
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Setup (test strip, measurement No., patient ID)
with
buttonsSelect
[ Select test strip type ] [ Menu screen ]
Select the type of test strip
Do NOT wet the black mark.
of the bottle, immediately close the cap or the test strips in the bottle will absorb moisture or dirt in the air, which may deteriorate reagents and make the test strips useless.
stage are required, depending on the setting of the “opera- tional mode when turning ON” parameter. (See “Operational mode” on page 2-3.)
standard setting is “Auto start” mode. To change the “operational mode when turning ON” parameter, see “3-7-8. No.005: Operational mode when turning ON” on page 3-18.
intervals to signal the dipping timing for the test strips.
2.4 Measurement operation
AE-4022
Sample Measurement
5. Confi rm normal measurement icon
appears on standby screen.
6. Enter Clarity Value (If feature is ON).
• Press and hold
• Enter the clarity value of sample using numeric keys
• Enter a number between 1 and 5 using table below:
1 2 3 4 5
Clear Slightly
• Press
to confi rm entry.
7. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
Slow beeping
Preparation for dipping:
3 seconds
for about 2 seconds.
Cloudy Turbid Other
Cloudy
Fast beeping
Dipping period:
2 seconds
Test strip setting:
1.4 seconds
IMPORTANT:
• Do NOT wet the black mark on test strip.
• After dipping test strip in sample, gently remove excess urine by drawing the entire length of strip against the rim of the
collection cup.
Continued on next page…
4
Page 10
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Do NOT wet the black mark.
Test strip
Rails (white)
Test strip detection window
Detection area
AE-4022
8. Place test strip on test strip tray.
• Place strip in the detection area of the test strip detection window with the black marker on right or left.
• The carrying arm sends the test strip into the analyzer and measurement starts automatically.
Sample Measurement
9. When buzzer setting is ON, the beeper sounds for two seconds to
When measuring continuously, repeat steps 2 to 4. Set measurement numbers and patient IDs as necessary.
10. Read test results from printout.
11. Turn off instrument after completing all measurement procedures.
signal the timing for dipping the next test strip. Dip the test strip while you hear the beep.
5
Page 11
Control Measurement
Control Measurement
Page 12
SEMI-AUTOMATED URINE ANALYZER
buttonsSelect
buttonsSelect
of the bottle, immediately close the cap or the test strips in the bottle will absorb moisture or dirt in the air, which may deteriorate reagents and make the test strips useless.
stage are required, depending on the setting of the “opera- tional mode when turning ON” parameter. (See “Operational mode” on page 2-3.)
standard setting is “Auto start” mode. To change the “operational mode when turning ON” parameter, see “3-7-8. No.005: Operational mode when turning ON” on page 3-18.
intervals to signal the dipping timing for the test strips.
2.4 Measurement operation
Pad area
AUTION ELEVEN
AE-4022
1. Change measurement mode.
• From the [standby screen] press
2. Enter measurement number to identify control level.
3. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
Slow beeping
Preparation for dipping:
3 seconds
4. Place test strip on test strip tray.
Fast beeping
Dipping period:
2 seconds
or
Test strip setting:
1.4 seconds
until the
Control Measurement
icon is displayed.
• Place strip in the detection area of the test strip detection window with the black marker on right or left.
• The carrying arm sends the test strip into the analyzer and measurement starts automatically.
5. When buzzer setting is ON, the beeper sounds for two seconds to signal the timing for dipping the next test strip. Dip the test strip while you hear the beep.
When measuring continuously, repeat steps 2 to 4. Set measurement numbers as necessary.
6. Read test results from printout.
7. Turn off instrument after completing all measurement procedures.
6
Page 13
Check Measurement
Check Measurement
Page 14
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
buttonsSelect
buttonsSelect
Detection area
2.4 Measurement operation
AE-4022
1. Clean the feeder and waste box to prevent check strips from becoming soiled.
2. Change measurement mode.
Check Measurement
• From the [standby screen] press
3. Place white check strip on test strip tray and perform measurement.
4. Place gray check strip on test strip tray and perform measurement.
5. Remove check strips from waste box and return to check strip bottle.
6. Evaluate results
• Compare the refl ectance values on the test result printout to the ranges on the refl ectance label on the check strip bottle.
• If any of the measurement results are outside of the specfi ed ranges, repeat check measurement using the other set of
• If any of the measurement results are still outside the ranges, contact ARKRAY Technical Support at 1.855.646.3108.
check strips.
or
until the
icon is displayed.
7
Page 15
Maintenance
Maintenance
Page 16
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Turn off the instrument.
Open the maintenance cover.
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Test strip feed tray
Test strip tray
Detach the test strip tray.
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used test strips and protective gloves separately from
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat- ing Manual.
Clean the area circled in red below.
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Open the maintenance cover.
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Suction port
Test strip tray
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat- ing Manual.
Clean the area circled in red below.
Test strip tray
Wear protective gloves to prevent exposure to pathogenic microbes.
Discard used test strips and protective gloves separately from general waste according to local regulations on biohazardous waste.
Feed lever
Lever
Detach the feed lever by sliding the white lever in the center towards you.
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat- ing Manual.
Clean the area circled in red below.
Suction port
Test strip tray
Test strip tray
AE-4022
Maintenance – Daily
Clean the Feeder
1. Turn off instrument. 2. Detach carrying arm.
4. Detach test strip tray.
• Slide the test strip tray to the
Power switch
front and lift up to detach.
Test strip tray
• Pull towards you and lift up to remove.
5. Detach suction port.
• Pull suction port straight
up to detach.
Suction port
Test strip tray
3 Open maintenance cover.
• Press and hold the two open buttons on either side of instrument and lift up to open.
6. Detach test strip feed tray.
• Slide to front and lift up to detach.
Continued on next page…
8
Page 17
SEMI-AUTOMATED URINE ANALYZER
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Suction port
Test strip feed tray
Test strip tray
Lever
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wear protective gloves to prevent exposure to pathogenic microbes.
Feed lever
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before attaching them back on.
This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance. Before using the instrument, make sure to read the Operat- ing Manual.
AUTION ELEVEN
AE-4022
Clean the Feeder
7. Detach feed lever by sliding the white lever in the center towards you.
Lever
Feed lever
8. Clean the table.
• Clean the areas circled in red below with water.
Maintenance – Daily
Continued on next page…
9
Page 18
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Wash the detached carrying arm, waste box, test strip tray, suction port, test strip feed tray and feed lever with water to remove any dirt.
Carrying arm
Suction port
Test strip feed tray
Test strip tray
Suction port
Suction port
Test strip feed tray
Suction port
Test strip feed tray
Test strip tray
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
AE-4022
Clean the Feeder
9. Wash detached test strip tray, suction port, test strip feed tray and feed lever with water.
Maintenance – Daily
10. Sterilize the carrying arm using alcohol. Then wash with water.
11. Wipe the parts with a cloth and dry completely before reattaching to instrument.
10
Page 19
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Waste box
Waste box
4.1 Daily maintenance
AE-4022
Clean the Waste Box
Maintenance – Daily
1. Remove waste box and discard used test strips.
2. Sterilize and clean waste box.
• Sterilize with alcohol then
wash with water and wipe dry.
3. Attach waste box.
11
Page 20
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Printer cover
Take out.
Cut using scissors.
4.2 Replacing the thermal recording paper
4.2 Replacing the thermal recording paper
Printer cover
Take out.
Cut using scissors.
Printer cover
Take out.
Cut using scissors.
4.2 Replacing the thermal recording paper
AE-4022
Replacing thermal recording paper
Maintenance – As Needed
1. Check that the [Standby screen] is displayed,
2. Press
3. Cut the fi rst (outermost) layer of a new paper roll and trim
4. Place new roll of thermal recording paper in paper holder so
5. Close printer cover.
NOTE:
then open the printer cover.
to remove any remaining paper in
When paper is removed from printer, error message E004 appears and a beep sounds for 1 minute. The alarm can be stopped by pressing
the printer.
the end to make it straight. A skewed or odd-shaped end may cause the paper to jam.
that paper feeds from the bottom. Guide the trimmed end of the thermal recording paper into and through the slot that receives the recording paper. The thermal recording paper is automatically wound forward and fed into the unit.
NOTE:
Always press the paper. If you do not fully feed the thermal recording paper through the unit by pressing the measurement results.
button after replacing the thermal recording
button, the printer will fail to print the
.
Correct Incorrect Incorrect
Paper Holder
Set the new roll of thermal recording paper in the paper holder in the proper orientation so that the end feeds from the bottom.
12
Page 21
Warnings
Warnings
Page 22
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
Warning number
AE-4022
Warnings
A warning message appears when a normal measurement result is not obtained.
When a problem occurs during measurement, the system continues measurement and prints warning messages with the measurement results.
The table below explains the meaning of and measures (corrective action) to be taken for each warning message.
Message Error Possible cause Measures (Corrective Action)
W001 Drift
W002
W003
W004 No dipping The entire pad section of the test strip was not dipped.
W005 Wrong test strip A test strip other than that specifi ed was used.
W006 Transport error The test strip may not have been transported properly.
W007
W008 Elapse of QC deadline The QC deadline has elapsed. Perform control measurements.
Abnormally high
refl ectance
Incorrect test strip position
Number of registered operator ID numbers
The ambient light level around the instrument changed
rapidly, due to a photofl ash or other reasons.
An abnormal sample (e.g. a sample containing medicine) was measured.
• The test strip was displaced from its correct position due to vibration.
• The test strips were incorrectly placed.
150 operator ID numbers have been registered.
Maintain a constant level of illumination around the instrument during measurement.
Check if the tested sample is abnormal or not. If the sample is normal, test it again.
• Do not expose the instrument to any vibration during measurement.
• Place the test strips correctly on the test strip tray.
Dip a new test strip in the sample correctly and test again.
Only use ARKRAY’s test strips designed for the AUTION ELEVEN.
• Clean the test strip tray and suction ports.
• Attach the suction ports properly.
• Clean the incoming strip sensor window.
Delete unnecessary operator ID numbers before registering new ones.
13
Page 23
Errors
Errors
Page 24
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
4.2 Replacing the thermal recording paper
4.2 Replacing the thermal recording paper
5.2 Error messages
AE-4022
Errors may occur when you operate the instrument incorrectly or make a mistake during normal operation. When an error occurs, an error message appears on the screen and an alarm sounds for approximately 1 minute.
Errors
Press
to cancel the error notice. The display returns to the [Standby screen].
To prevent the error from recurring, take appropriate measures (corrective action), referring to the table below. If unable to clear the error, note error code and call ARKRAY Technical Support at 1.855.646.3108.
Message Error Possible cause Measures (Corrective Action)
E001 Power down
E002
E004
E005
Backup memory error
No paper in the printer
Waste box is full
The instrument was suddenly turned OFF during measurement.
As the instrument was unused for a long period, the memory backup battery was exhausted and the measurement result stored in the memory were lost.
• The thermal recording paper has run out.
• The thermal recording paper roll was not
At the end of the measurement, the test strip waste counter exceeded 90.
properly installed.
Retest the sample(s) currently being measured.
Keep the instrument powered for at least 11 hours to charge the backup battery. After charging the instrument, set the date and time.
• Install a new roll of thermal recording paper.
• Make sure the thermal recording paper is installed correctly.
Cancel the error notice by pressing used test strips from the waste box, and press ➀.
. Empty the
E006
Surplus urine is full
At the end of the measurement, the total number of measurements since turning ON the instrument exceeded 190.
Turn OFF the instrument and discard any surplus urine from the suction port.
14
Page 25
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
AE-4022
Message Error Possible cause Measures (Corrective Action)
Errors
E007 Data not
found
E008 Auto start
sensor error
E009 No test strip
on the feeder
E011 Elapse of QC
deadline
No data was found in the specifi ed range. Verify that the specifi ed range for re-printing,
re-sending, or list printing is appropriate.
• A test strip was placed while the carrying arm was moving to the suction ports.
• The detection section of the auto start sensor was soiled with urine.
• The auto start sensor is faulty.
• The feed lever is not attached.
• The incoming strip sensor is faulty.
• The sensor failed to detect a test strip.
The QC deadline elapsed while the instrument was OFF or on standby.
• Remove the test strip.
• Clean the detection section.
• If you fi nd no abnormality and the error recurs,
contact your local distributor.
• Turn OFF the instrument and open the maintenance cover.
• Mount the feed lever.
• If you fi nd any obstacles or scattered test strips
inside, remove them.
• If you fi nd no abnormality and the error recurs,
contact your local distributor.
Perform control measurements.
Situation Error Possible cause Measures
Buttons will not respond and/or testing does not start when strip is placed.
Elapse of operator ID effective time.
The operator ID effective time elapsed.
Enter operator ID.
15