Arjo Freedom Back Quick Reference Guide - Site Preparations QUICK REFERENCE GUIDE

Page 1
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
FREEDOM BATH
SITE PREPARATIONS
02.AF.00/2GB May 2007
Page 2
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
© ARJO Hospital Equipment AB 2007. ARJO products are patented or patent pending. Patent information is
available by contacting ARJO Hospital Equipment AB. Our policy is one of continous development, and we therefore reserve
the right to make technical alterations without notice. The content of this publication may not be copied either whole or in part without the consent of ARJO Hospital Equipment AB.
Page 3
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
User Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measurements and Weights. . . . . . . . . . . . . . . . . . . . . 6
Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Site Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Available Kits (For special installations) . . . . . . . . . . 10
Water Supply (Internal Mixer) . . . . . . . . . . . . . . . . . . . 11
Water Supply (External Mixer) . . . . . . . . . . . . . . . . . . 15
Drain Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Check Points prior to the Installation. . . . . . . . . . . . . 22
3
Page 4
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Introduction
Freedom Bath
ARJO Freedom Bath is a sophisticated piece of medical equipment and has to be handled as such. The Freedom Bath does require installation, however we do not want the Freedom Bath tiled in as certain components need regular service and maintenance.
Site Preparations
This document has been developed to supply you with accurate and targeted information in order to make your site preparation for the ARJO Freedom Bath ideal. Combining good engineering practice with a well prepared installation site, will ensure a worry free installation the day your Freedom Bath is delivered. Please read through the complete manual prior to preparing your site as it will show different installation scenarios.
Please take extra care to identify your specific installation scenario. We suggest that you use the tic box in the lower right hand corner of each scenario for future reference as well as simple identification within different workgroups. (plumbing, electrical, drywall, tiling etc.)
For installation sites that require a seal between their tub and wall; we have an accessory in the form of a Tile-in kit comprising of a seal that can be placed between tub and wall, allowing for easy removal.
Approvals
UL classification
UL marking (US-models 120V-60Hz) Patient Bathing Units Classified in accordance
with the standards for Plastic Bathtub and Shower Units, ANSI Z124.1.2-2005; Whirlpool Bathtub Appliances, ANSI/ASME A112.19.7M-1995; Suction Fittings, ANSI/ASME A112.19.8-1987; Pipe Applied Vacuum Breakers, ANSI/ASSE 1001-1988; Hand- HeldShowers, ANSI/ASSE 1014-1989; Individual Thermostatic, Pressure Balancing and Combination Pressure Balancing and Thermostatic Control Valves for Individual Fixtures, ANSI/ASSE 1016-1996.
Medical Equipment with respect to electrical shock, fire and mechanical hazards only in accordance with Standards no. UL60601-1
CAN/CSA-C22.2No.601.1-M90 Patient bathing units Classified in accordance with
CAN/CSA-B45.0-2004 and B45.5-2004, “Plumbing Fixtures”, and CAN/CSA-B64.0-1994 and B64.1.1-1994, “Backflow Preventers and Vacuum Breakers.”
CE marking
CE-marking in accordance with the directive 93/42/EEC (Medical Devices Directive).
The equipment has been tested and fulfils all requirements with respect to electrical and mechanical hazards according to IEC 60 601-1 and for EMC according to IEC 60 601-1-2.
User Guidelines
In order to ensure that we accomplish your demands, we are grateful if this information is read. This will facilitate the installation procedure and give the correct and optimal working conditions of all involved persons.
WRAS (GB only)
WRAS approved Thermostatic Mixing Valves, according to National Health Service
Model Engineering Specification D 08- TMV3. Thermostatic Mixing Valves (Healthcare Premises). Designation Code: HP-S / HP-T44.
4
Page 5
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Safety Precautions
All electrical and plumbing work must be carried out by authorized personnel:
In accordance with local codes
In accordance with local regulations It is the responsibility of the customer to prepare
the site before the installation procedure starts.
To avoid personal damage due to heavy lifts, suitable transporting aids must be available when installing the product.
Attention:
Massachusetts customers With reference to Massachusetts state plumbing
regulations, customers shall install proper back­flow prevention devices in compliance with 248 CMR (Code of Massachusetts Regulations).
The drainage system shall also comply with 248 CMR.
In the Commonwealth of 248 Massachusetts, all water temperatures shall comply with 248 CMR.
5
Page 6
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Measurements and Weights
Box
W
H
L
Box 1
The Freedom Bath is delivered in box 1 Box 2 contains installation kit
Freedom Bath
27 1/4"
760 mm
e
690 mm
29 3/4"
b
Box 2
a
59 1/2"
1510 mm
Measurements:
L Length W Width H Height
64 3/4” (1645 mm) 36” (915 mm) 54” (1370 mm)
W
Shipping weights:
Model Weight with
Still bath 335 lb (152 kg) 313 lb (142 kg)
Air spa 357 lb (162 kg) 335 lb (152 kg)
built-in water control
Measurements:
Length
a
Width
b
Height
c
Height, spout
d
Recommended depth of alcove
e
Distance (c/c) drain-floor
f
Weight without
built-in water control
41 1/8"
1045 mm
c
Freedom Bath is available in right-hand version only!
10 1/2"
270 mm
44 5/8"
1134 mm
d
Drain connection
flexible drain hose included
Weights:
Model Weight with
Still bath 287 lb (130 kg) 264 lb (120 kg)
Air spa 309 lb (140 kg) 287 lb (130 kg)
built-in water control
6 2/8"
160 mm
f
Weight without
built-in water control
6
Page 7
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Space Requirements
Space requirements around the bath
Freedom Bath must always be installed according to recommended space requirements, this to pro­vide optimal working conditions for personnel and resident.
Note that space for furniture and other personal items are not included in these measurements.
tub 59 1/2" (1510 mm)
59 1/2" (1510 mm)
tub
30" (760 mm)
59 " (1500 mm)
Available area for lift integration
Freedom Bath in lift accessible version (LA) allows following area for lift integration, limited by the drain connections.
tub
59 1/2" (1510 mm)
32" (800 mm)
59 " (1500 mm)
16" (410 mm)
tub
30" (760 mm)
91 1/2" (2310 mm)
31 1/2" (800 mm)
42 1/8" (1070 mm)
6" (155 mm)
7
Page 8
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Site Preparations
Floor/Walls
Prior to installation of Freedom Bath the floor and walls within the required area must be finished.
Materials
The floor should be made of a slip-resistant material, free of bumps.
Levels
The doorway must not have a treshold or if necessary it should be of the depressible rubber type, this to facilitate the use of a mobile aid.
Floor Slope
When using a floor drain the slope towards the drain should not exceed 1:50.
Drain
Floor attachments
The sliding tracks provided in the installation kit have to be used when installing Freedom Bath. The floor construction must be suitable for anchor­ing the bolts.
For exact positioning of the two sliding tracks a template is provided. (See figures 1 and 2.)
Loads
Max. floor load (filled with water + resident): 384 psi (265 N/cm
2
)
Max. floor loading (filled with water + resident): 292 lbs pounds force (1300 N)/foot pad
Weights
Total weight of tub filled with water and typical 176 pound resident (80 kg).
Water capacity Tub weight Max. load = resident weight + water weight + tub
weight.
Caulking between tub and wall
Not recommended as it will make the service access difficult. NOTE ! The tub must never be built into the finished wall. In order to seal the space between tub and wall we strongly recom­mend use of the tile-in kit.
Tile-in kit
When a water protection is wanted between tub rim and wall the optional tile-in kit should be used. (See figure 3.) The tile-in kit consists of a flexible seal that can be placed between tub and wall. The tile-in kit can both be applied horizontally as well as vertically.
Slope max. 1:50
8
Page 9
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
3/8" hexagonal screw
Sliding track
Optional spacer
Concrete anchor
Te mplate
Figure 1. Floor attachment kit
Drilled holes (x4) for internally 3/8" threaded flush mount concrete anchors. Follow instructions from anchor manufacturer, for example www.hilti.com
Suitable anchor, example: HILTI HDI 3/8 SS 303
5 6/8"
146 mm
16 5/8"
423 mm
6 2/8"
158 mm
47"
1194 mm
Drilled hole in flooring
6 2/8"
158 mm
Figure 2. Sliding tracks mounted in alcove layout
Figure 3. Tile-in kit applied between Freedom Bath and wall in alcove layout
Tile-in kit
Tile-in kit
Tub
Wall
9
Page 10
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Available Kits (For special installations)
For the in this manual described alternatives for water and drainage installations, the following kits are available:
AFA0090
1 Drainage pump with accessories
AFA0025
A2 water hoses 94 4/8” (2400mm) B2 holders C4 ties
Kit AFA0090
Kit AFA0025
A
B
C
AFA0045
A1 coupling 90° 1 1/2” B1 drain pipe C2 holders D4 ties
Kit AFA0045
D
A
C
B
10
Page 11
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Water Supply (Internal Mixer)
Time necessary to fill tub will be dependent on actual delivery rates.
Actual delivery rates will be dependent on temper­ature of hot and cold water in facility water supply net and related supply pressure.
Plumbing
This manual contains sets of values for tempera­ture and flow; the min./max. values are there to give you the extremes and are showing the toler­ances of the Freedom Bath, we will however also give you a recommended temperature span and your target should be to end up within this span to achieve an optimal function of your product.
Please note that in some countries/states/cities; local variations of recommended values might dif­fer from our recommended values, in this case the local code/rules apply. If the local recommenda­tions fall outside the min./max. values, other solu­tions will have to be looked into and we ask you to contact your local ARJO representative.
Temperatures
The Freedom Bath contains (optional) a thermo­static mixer that responds very well to differences in the supply line, both temperature and pressure differences.
This mixer will safely ensure that the resident never is exposed to temperatures exceeding 108˚ F (42˚ C) even if the maximum specified hot water temperature is supplied. Please note that a few degrees are always "lost" when using a ther­mostatic mixer and it is therefore recommended to have a higher supply temperature than the wanted bath water temperature.
An important reason for keeping hot supply water really hot is that high temperatures prevents the rapid growth of Legionella in hot water systems. Of the same reason it is important that the cold water temperature is really cold. Of course other meth­ods of reducing the risk of Legionella than regulat­ing temperature can be used.
Pressure and Flow
When designing a bath system with an integrated mixer it is essential to plan for a low filling time and a high temperature accuracy. In the Freedom Bath we did this by installing a combined 3x1/2" pres­sure and temperature compensating mixer. This mixer allows us to fill the Freedom Bath at a high rate and at the same time ensuring a very stable low flow situation when taking a shower. In order to achieve this we require you follow the recommended values/rates on the supply side. Please see to that the Freedom Bath is fed accord­ing to the recommended specifications.
Water connections
G 3/4", Stub-ups must be well sealed not protruding more than 4" above the finished floor. (See figure 1 on next page.) Stub-ups with shut-off valves are recommended.
Flexible water hoses are included in delivery.
Restrictions
Pressure:
Max static: 87 psi (600kPa, 6 bar) Min. dynamic: 14.5 psi (100kPa, 1 bar)
Flow/supply pipe:
Min.: 6.6 gallons/min (25 l/min) CW+HW: 13 gallons/min (50 l/min)
Temperature:
Cold water: 36-77˚F (2-25˚C) Hot water: 100-176˚F (38-80˚C)
Recommendations
Pressure:
Dynamic: 29-43.5 psi (200-300kPa, 2-3 bar)
Temperature:
Hot water temperature 114-140˚F (45-60˚C) Cold water temperature 41-59˚F (5-15˚C)
11
Page 12
3/4" BSP MALE
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
max.4"(100 mm)
Figure 1
20 1/2"
520 mm
18 3/4"
480 mm
CW 3/4"
HW 3/4"
15 6/8"
400 mm
26 3/8"
670 mm
Figure 2
12
Page 13
ARJO recommended water supply - floor
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
2 3/8"
60 mm
2 3/8"
60 mm
2 3/8"
60 mm
Water supply - wall, left side of tub
TUB
90° Elbow needed (supplied by others)
17 3/4"-25 5/8"
450-650 mm
Maximum 2 3/8" (60 mmm)
CW HW
9 7/8"-15 3/4" (250-400 mm)
2" (50 mm)
SITE SPECIFIC
CHOICE
Note! Tub can only be put in place from this direction!
SITE SPECIFIC
CHOICE
13
Page 14
Water supply - wall, right side of tub. Use kit AFA0025
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
2 3/8"
60 mm
2 3/8"
60 mm
2 3/8"
60 mm
Holders mounted under rear side of each sliding track.
Rear side of tub
Water hoses fastened with two ties at each holder.
SITE SPECIFIC
CHOICE
14
Page 15
Water Supply (External Mixer)
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Unassistad bathing - recommended locations of externally mounted water control
1
W
S
F
2
W
S
C
F
3
Corner Back wall Recessed
Figure 1
Assisted bathing - recommended locations of externally mounted water control
W
S
= Shower
S
= Water mixing valve
W
= Fill spout
F
F
F
1
Figure 2
2
S
W
C
S
F
W
F
3
Corner Back wall Recessed
4
S
F
W
Back wall
S
W
F
15
F
F
= Shower
= Water mixing valve
= Fill spout
S
W
Page 16
Drain Installation
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Connections must be done in accordance with local regulations.
Restrictions
Min.drainage capacity: 24.6 US gallons/min (100 l/min).
Recommendations
8" floor sink with trap, indirect connection with 3" drain pipe is preferred. A floor sink facilitates maintenance of the tub and eliminates the risk of water damage in case of emergency draining of the tub.
Recommended location of floor sink drain accord­ing to measurements in figure on next page.
If alternative area for floor sink is chosen be sure the floor sink drain will not interfere with water supply area.
16
Page 17
ARJO recommended floor sink drain
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
3 15/16" 100 mm
300 mm
11 13/16"
SITE SPECIFIC
CHOICE
17
Page 18
Alternative stub-up connection
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
DRAIN STUB-UP
1 1/2"
(DN 40)
2 3/8"
60 mm
max. 4" (100 mm)
8 1/4"
210 mm
SITE SPECIFIC
CHOICE
18
Page 19
Drain connection with stub-up on right hand side. Use kit AFA0045
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
8 1/4"
210 mm
2 3/8"
60 mm
Holders mounted under rear side of each sliding track.
Rear side of tub
Rear side of tub
Drain pipe fastened with two ties at each holder.
SITE SPECIFIC
CHOICE
19
Page 20
Power
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
The unit must be permanently connected to the mains.
When used in USA and Canada: All electrical installations must comply with the US National Electrical Code and the Canadian Electrical Code respectively.
Requirements
The unit shall be permanently connected using the following equipment (not included in delivery):
Separate external fuse.
•Ground fault circuit interrupter (10mA).
Safety switch, according to regulations and local codes.
Mains cable (3x16AWG / 3x1.5 mm2).
Market Voltage Frequency Power consumption
Europe N.America
230/240V single phase AC
120V single phase AC
Potential equalization.
Freedom Bath is equipped with a potential equalization connection. Connect to the potential equalization connection according to local/national requirements. (See figure 2.)
50Hz 60Hz
max. 800 VA max. 800 VA
20
Page 21
Figure 1
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
49 3/4"
1265 mm
Warning! - moving parts in this area
Place no cables or connections under the tub
22"
560 mm
7 7/8"
200 mm
Field wiring box (US)
Figure 2
Potential equalization
21
Page 22
Check Points prior to the Installation
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
Always make sure that the following check points are fulfilled and completed prior to the installation. Reference to values and other information in this manual.
Completed
Water cold/hot pressure, temperature, location, local/national requirements
Drain location, capacity, local/national requirements
Power fusing, capacity, local/national requirements
Potential equalization local/national requirements
Attachments floor/wall fittings, local/national requirements
Floor loads construction, materials, local/national requirements
Materials slip/moisture protection, local/national requirements
Floor slope, evenness, thresholds, local/national requirements
Installation accessibility doors, measurements
Space requirements for personal and patient
Service accessibility for maintenance staff and annual service
Make sure to include these check points in your internal installation activities.
22
Page 23
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
ARJO Hospital Equipment AB, PO Box 61, Verkstadsvägen 5, SE-241 21 Eslöv, Sweden
Page 24
AUSTRALIA
Published
Check occasionally
for new editions.
1:11 pm, Jun 05, 2007
ARJO Hospital Equipment Pty Ltd 154 Lytton Road BULIMBA Brisbane QLD 4171 Australien Tel: 07-3395 6311 Fax: 07-3395 6712
AUSTRIA
ARJO-Sic GmbH Föhrenweg 5 A-6065 THAUR Tel: 05223-49 33 50 Fax: 05223-49 33 50 - 75
BELGIUM
ARJO Hospital Equipment nv-sa Ternesselei 248 B-2160 WOMMELGEM Tel: 03 353 91 00 Fax: 03 353 91 01 E-mail: info@arjo.be
CANADA
ARJO Canada Inc. 1575 South Gateway Road Unit "C" MISSISSAUGA, ON, L4W 5JI Tel: 1-800-665-4831 Fax: 1-800-309-1116 E-mail: info@arjo.ca Web page: www.arjo.com.
CZECH REPUBLIC
ARJO Hospital Equipment s.r.o. Hlinky 118 CZ- 603 00 BRNO Tel.: 549 254 252 Fax: 541 213 550
DENMARK
ARJO Scandinavia Postboks 51 DK-4632 BJAEVERSKOV Tel: 45 93 27 37 Fax: 45 93 27 39
FAR EAST
ARJO Far East Limited
Unit 3A, 4/F., block B Hoi Luen Industrial Centre
55 Hoi Yuen Road, Kwun Tong, Kowloon HONG KONG Tel: 2508 9553 Fax: 2508 1416
FINLAND
OY Vestek AB Martinkuja 4 FI-02270 ESPOO Tel: 9 8870120 Fax: 9 88701291
FRANCE
ARJO Equipements Hospitaliers S.A. 45, Avenue de l’Europe Eurocit BP 133 F-59436 RONCQ CEDEX Tel: 03 20 28 13 13 Fax: 03 20 28 13 14 E-mail: info@arjo.fr
GERMANY
ARJO GmbH Christof - Ruthof - Weg 6 D-55252 MAINZ-KASTEL Tel: 06134-186-0 Fax: 06134 186 160 E-mail: info@arjo.de
GREECE
C. Psimitis Co Ltd Dimitriou Andr. 59 GR-16121 KAISARIANI ATTIKIS Tel: 21 0724 36 68 Fax: 21 0721 55 53
ITALY
ARJO Italia S.p.A. Via Tor Vergata 432 I-00133 ROMA Tel: 06-87426211 Fax: 06-87426222 E-mail: promo@arjo.it
THE NETHERLANDS
ARJO Nederland BV De Blomboogerd 8 4003 BX TIEL Postbus 6116 NL-4000 HC TIEL Tel: 0344-64 08 00 Fax: 0344-64 08 85 E-mail: info@arjo.nl
NORWAY
ARJO Scandinavia Ryenstubben 2 N-0679 OSLO Tel: 98 28 11 70 Fax: 22 57 06 52
POLAND
ARJO Poland Sp.z o.o. Ul. Lirowa 27 PL-02-387 WARSZAWA Tel: 22 882 06 26 Fax: 22 882 06 29
PORTUGAL ARJO International AB
Rua das Valas 82 PT 4510-154 JOVIM GONDOMAR Tel: 22 450 32 48 Fax: 22 450 03 27
SPAIN
ARJO Spain, S.A C/ San Rafael, No. 6 E-28108 ALCOBENDAS Tel: 902 300 126 Fax: 902 300 191 E-mail: arjospain@arjospain.es
SWEDEN
ARJO Scandinavia AB Verkstadsvägen 5 Box 61 SE-241 21 ESLÖV Tel: 0413-645 00 Fax: 0413-645 83 E-mail: kundservice@arjo.se
SWITZERLAND
ARJO-Sic AG Florenzstr. 1d Postfach CH-4023 BASEL Tel: 061-337 97 77 Fax: 061-311 97 42
UNITED KINGDOM
ARJO Med AB Limited St Catherine Street GLOUCESTER GL1 2SL Tel: 08702 430 430 Fax: 01452-525 207
USA
ARJO Inc. 50 North Gary Avenue ROSELLE, IL 60172 Tel: 1-800-323-1245 Fax: 1-888-594-2756
www.arjo.com
...with people in mind.
info@arjo.com
R
D
E
E
G
R
I
S
E
T
FM 24321 ISO 9001 MD 87841 ISO 13485
If your country is not listed here, please contact your local distributor or: ARJO INTERNATIONAL AB, Box 61, S-241 21 ESLÖV, SWEDEN Tel: +46 413 645 00; Fax: +46 413 645 63; www.arjo.com
7000202
MEMBER OF THE GETINGE GROUP
Loading...