PART # PHOENIX GENOME TYPE I WATER SYSTEM
PHX-AD-S-1 Benchtop or wall mount base system, 120V/60Hz
PHX-AD-R-1 Remote Dispense Configuration base system, 120V/60Hz
PHX-AD-S-2 Benchtop or wall mount base system, 220V/50Hz
PHX-AD-R-2 Remote Dispense Configuration base system, 220V/50Hz
PART # ACCESSORIES
PHA-FM Flow Totalizer Meter
PHA-DF Direct Feed Port
PHA-DG Recirculating Dispensing Gun
PHA-WB Wall Mount Bracket
PART # CARTRIDGES & FILTERS
PX-13-5001 Dual Conditioning Cartridge (RO Feed)
PX-13-5002 Dual Conditioning Cartridge (SDI / TAP Feed)
PX-11-5103 Polishing Ultrapure Cartridge
PX-11-5104 Polishing Ultrapure Low Organics Cartridge
PF-00-6402 0.2 Micron Capsule Filter - 1/4” MNPT x 1/4” hose barb
PF-00-7101 0.2 Micron Endotoxin Removal Capsule Filter - 1/4” MNPT x 1/4” hose barb
PHOENIX ACADEMY
Type I, Ultra-High Purity Water for General Lab Applications
PART # OTHER CONSUMABLES
HPA-017 Sanitization Kit (includes cartridges, test strips, and gloves)
NOTE: Conditioning cartridge, polishing cartridge and final filters are sold separately.
SYSTEM SPECIFICATIONS
Dimensions (H x W x D)
Height: 21.6 in. (54.8cm)
Weight: 14.7 in. (37.0cm)
Depth: 14.4 in. (36.5cm)
Weight
Dry 42 lbs. (22.7 kg)
Operating 47 lbs. (25.0 kg)
Inlet Connection
O.D. Tubing 3/8” (9.53 mm)
Power Requirements
120 VAC/ 60 Hz @ 1.0 amp
Pressure
20 PSIG min. — 90 PSIG max.
Max Temperature 100°F (30°C)
Flow Rate (Typical)
2.5 LPM (0.67 GPM)
<2.0 LPM with Endotoxin filter installed
Ultra-High Purity Lab Water for use with:
General Chemistry
Academia
Buffer Solutions
CALL 856.626.1550
OR EMAIL info@ariesfilterworks.com
Aries FilterWorks is a division of ResinTech, Inc., a global leader ion exchange technology.
All media, cartridges and systems are lot-numbered and manufactured in state-of-the-art facilities the U.S.A under ISO standards.
THE PHOENIX academy system
SYSTEM HIGHLIGHTS
Designed for general laboratory
applications, the Phoenix Academy
delivers 2.5 LPM of Type I, 18.2 Megohm
water quality on demand. The water is
purified using a multi-stage purification
process. In the two-step deionization stage,
source water passes through high-purity
ion exchange resins to remove dissolved
minerals. A subsequent stage uses a 0.2
micron filter to remove particulates and
bacteria before dispensing. The result is
water that meets universally recognized
Type I specifications.
The Phoenix Academy’s intuitive,
“touch-screen” display monitors
system health and the remaining life of your
consumables. Intermittent recirculation
ensures that the water inside the unit is
continuously moving when not in use,
mitigating any chance of bacterial growth.
Convenient, programmable presets allow
you to dispense frequently used volumes
with speed and precision consistently.
Applications include general water
chemistry, reagent preparation and buffers.
CONDITIONING CARTRIDGE
POLISHING CARTRIDGE
0.2 MICRON
Performance Specications
Resistivity
Bacteria
Particulates
TOC
Performance based upon RO / DI or Distillation Feed
(< 20 uS/cm, < 50 ppb of TOC, 0.2 micron ltration)
Includes Endotoxin Removal Capsule Filter
18.2 MΩ-cm
< 1 cfu/ml
< 0.2 µm ltration
< 15 ppb
18.2
100%
75%
50%
THE BEST VALUE IN REPLACEMENT CONSUMABLES
5
4
INTUITIVE TOUCH SCREEN
CONTROLS
PROGRAMMABLE
VOLUME DISPENSE
REAL-TIME QUALITY
MONITORING & ALERTS
REMOTE DISPENSE OPTION
18.2
100%
75%
50%
QUICK CHANGE
CARTRIDGE DESIGN
LARGE CAPACITY
CARTIRDGES FILLED WITH
PREMIUM ResinTech® MEDIA
SYSTEM FLOW
Direction of Flow
Solenoid Valve
Pump
Check Valve
Intermediate Resistivity Sensor
Final Resistivity Sensor
Pressure Regulating Valve
Final Filter
Optional Flow Sensor
Inlet
Conditioning
Cartridge
3
2
1
Filtration
or
Endotoxin Filter
(Optional)
Polishing
Cartridge