Aquametrix 2300 User Manual

Last revised: February 11, 2014
AquaMetrix 2300 Controller Installation & Operation Manual
V1.3
©Water Analytics 2014 All rights reserved.
by
Water Analytics
Aqua
Metrix
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 2
For Your Safety
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Statement of Limited Warranty
WATER ANALYTICS warrants that this product is free from defects in workmanship and material for a period of 24 months for electronics from the date of delivery from the factory or authorized distributor under normal use. WATER ANALYTICS’s liability under this warranty shall be limited to replacement or repair of any defective equipment or part which, having been returned to WATER ANALYTICS, transportation charges prepaid, has been inspected and determined by WATER ANALYTICS to be defective.
Copyright 2014 Water Analytics, Inc. All rights reserved.
Water Analytics, Inc 100 School St. Andover, MA 0810 Information: info@wateranalytics.net
Support issues can be emailed to support@wateranalytics.net.
Warning: The AC power input presents an electrical hazard.
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 3
2300 Multi-Parameter Controller
Table of Contents
AquaMetrix+2300+Controller+Installation+&+Operation+Manual+...............................+1!
1.! Overview+.............................................................................................................................+5!
"#"#! $%&'( ) * +& ,( %!##########################################################################################################################!-! "#.#! /01+,2,+3&,(%4!########################################################################################################################!-! "#5#! 6'(%&!73%18!9,4083:!###############################################################################################################!;! "#<#! 6'(%&!73%18!=1%*!#################################################################################################################!>! "#-#! ?&@1'%1&!A(%%1+&,(%!&(!3!%1&B('C!4B,&+@!('!'(*&1'!############################################################!D! "#;#! 7'(E1!/&3&*4!/+'11%!##############################################################################################################!D! "#>#! F@1!G1E!=1%*!###################################################################################################################!"H!
2.! Installation+and+Housekeeping+................................................................................+11!
.#"#! 7344B(')!A@3%I1!################################################################################################################!".! .#.#! F,J1!/1&*0!##########################################################################################################################!".! .#5#! ?K =3,8!/1&*0!########################################################################################################################!".! .#<#! A(%2,I*'1!93&3!L(II,%I!#######################################################################################################!".!
3.! Connecting+and+Configuring+Probes+.......................................................................+13!
5#"#! A(%2,I*',%I!M%38(I!/1%4('4!################################################################################################!"5! 5#.#! A(%%1+&,%I!7*841!N3&1!O*&0*&!/1%4('4!##############################################################################!"5! 5#5#! 7'(E1!A(%2,I*'3&,(%!/1&*0!#################################################################################################!"<! 5#<#! F@1!7'(E1!A(%2,I*'3&,(%!G,P3')!########################################################################################!"-! 5#-#! FB(K7(,%&!/+38,%I!('!A38,E'3&,(%!#########################################################################################!">! 5#;#! Q(%KL,%13'!/+38,%I!##############################################################################################################!"D! 5#>#! /+381!63+&('!#########################################################################################################################!.H! 5#D#! RK63+&('!/+38,%I!###################################################################################################################!.H!
The K-Factor option allows for a constant factor to be used to scale a frequency input into units of flow, ie volume per unit time. The K Factor is provided by the flow meter manufacturer in units of pulses per volume, where volume is typically gallons or liters. In order to display and record the proper units for the flow, the user must twice enter the volume units and the time units. The first time is on the same wizard page as the K Factor, where the choices are limited to gallons and liters for the volume, and seconds, minutes, hours, and days for the time. These entries are used by the 2300 to calculate the proper readings. The second time is on the next wizard page for the units to display. If the desired units are not on the pull down menu, choose Enter new units and type in what you want to see. Try to keep the total number of characters to less than 6.!##########################################################################################################!.H!
4.! Configuring+Virtual+Probes+........................................................................................+20!
<#"#! A(%2,I*',%I!3!F(&38,P1'!/1%4('!############################################################################################!.H! <#.#! A(%2,I*',%I!9,221'1%&,38!/1%4('4!#########################################################################################!."! <#5#! A(%2,I*',%I!3!68(B!/1%4('!2('!S3&+@!A(%&'(8!######################################################################!."!
5.! Configuring+Alarms,+Relays,+and+Notifications+....................................................+22!
-#"#! F:014!(2!=(%,&('4!###############################################################################################################!..!
by
Water Analytics
Aqua
Metrix
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 4
-#.#! A(%2,I*',%I!2('!?J3,8!Q(&,2,+3&,(%!######################################################################################!.;!
6.! Logs+and+Data+Logging+.................................................................................................+27!
;#"#! 93&3!L(II,%I!A(%2,I*'3&,(%!G,P3')!####################################################################################!.D! ;#.#! M83'J!L(I4!##########################################################################################################################!.D!
7.! User+Maintenance+and+Passwords+..........................................................................+29!
>#"#! G1E!7344B(')4!###################################################################################################################!.T! >#.#! 6'(%&!73%18!7344+()14!3%)!LA9!A(%2,I*'3&,(%!###################################################################!.T!
8.! Operating+the+2300+without+a+Network+................................................................+30!
D#"#! 6'(%&!73%18!A38,E'3&,(%!#######################################################################################################!5H! D#.#! 6'(%&!73%18!N183:!/1&!7(,%&4!###############################################################################################!5.! D#5#! 6'(%&!73%18!/3U1!3%)!N14&('1!A(%2,I*'3&,(%!######################################################################!55!
9.! Configuring+the+2300+for+a+Network+.......................................................................+33!
T#"#! 9:%3J,+!$7!/1&*0!################################################################################################################!55! T#.#! V,1B,%I!&@1!.5HH!(%!3!G,)1!M'13!Q1&B('C!########################################################################!55! T#5#! 91),+3&1)!B,'18144!+(%%1+&,(%!U,3!3!/J3'&!7@(%1!('!F3E81&!##############################################!5<! T#<#! /&3&,+!$7!M))'144!/1&*0!3%)!G,'1)!A(%%1+&,(%!##################################################################!5-!
10.! Network+Setup+.............................................................................................................+36!
"H#"#! ?&@1'%1&!/1&*0!##################################################################################################################!5;! "H#.#! G1E!/1'U1'!3%)!F6F7!/1'U1'!/1&*04!#################################################################################!5>! "H#5#! ?J3,8!/1&*0!#######################################################################################################################!5>! "H#<#! ?J3,8!R1:B(')!L,4&!############################################################################################################!5D! "H#-#! ?J3,8!F(!M))'14414!###########################################################################################################!5T! "H#;#! 9,3I%(4,%I!?J3,8!7'(E81J4!##############################################################################################!5T!
11.! Configuring+the+System+.............................................................................................+39!
""#"#! /:4&1J!Q3J1!3%)!F,J1!####################################################################################################!5T! ""#.#! S'(B41'!A(%2,I*'3&,(%!3%)!L(II,%I!O*&!##########################################################################!<H! ""#5#! LA9!A(%2,I*'3&,(%!#############################################################################################################!<H! ""#<#! ?''('!W3%)8,%I!###################################################################################################################!<H! ""#-#! X%,&!L,4&!#############################################################################################################################!<"!
12.! Advanced+Setup+...........................................................................................................+41!
13.! Available+I/O+Card+Types+.........................................................................................+42!
"5#"#! 5K+@3%%18!9,I,&38!$%0*&!Y!<K+@3%%18!9,I,&38!N183:!O*&0*&!+3')!#########################################!<.! "5#.#! ;K+@3%%18!9,I,&38!N183:!O*&0*&!+3')!#################################################################################!<5! "5#5#! >K+@3%%18!9,I,&38!$%0*&!+3')!#############################################################################################!<5! "5#<#! <K+@3%%18!M%38(I!O*&0*&!+3')!##########################################################################################!<5!
14.! Rebuilding+a+Corrupted+Internal+SD+Card+...........................................................+44!
15.! Panel+Mounting+Dimensions+...................................................................................+45!
!
!
! !
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 5
1. Overview
1.1. Introduction
The AquaMetrix 2300 Controller is a multi-input controller that is web-enabled such that it can be setup in minutes from any web browser. It contains a number of Set-up Wizards to aid the user in entering information, much of which is not reproduced in this manual. The Wizards and most other screens in the 2300 Controller are organized by a dialogue followed by a summary page with “Submit” and “Cancel” at the bottom. The user may familiarize himself/herself with the 2300 by going through most menu functions and just hitting “Cancel” and not “Submit”.
The standard configuration inputs up four 4-20mA analog sensors and three digital/frequency inputs for flow monitoring. Outputs include four relays and unlimited email/text notifications. The 2300 Controller provides remote viewing capability and configuring with password from any connected web browser. It also provides data logging with downloading or emailing of log files in CSV format. There is also one available slot for an optional card, which can expand digital inputs or relays outputs, or add analog outputs.
1.2. Specifications
Table 1 - Specifications
Inputs
Analog
Four 4-20 mA powered or unpowered current loops
Pule Rate
Two optically isolated inputs suitable for flow sensors, up to 24 VDC or 120 VAC
Counter
One accumulator for tracking equipment on time
Contacts
Four contact sensing inputs for alarm acknowledgement or remote resets.
Outputs
Relays
Four 120V/240 V @ 10A/5A expandable to eight*
Alerts
Emails and/or texts for alarms, relay actions, errors, or reminders
Web
Remote viewing with a browser on a computer, tablet, or smart phone. Remote programming from any of these.
Data logging
Logging of CSV files directly to micro SD card
Digital Outputs
Modbus TCP over Ethernet connection
Analog Outputs
Optional: Four non-isolated 4-20 mA outputs proportional to any reading.*
Control
Web or LAN
Viewing with a browser on a computer, tablet, or smart phone. Setup and control from any of these. Multi-user password protected.
Front Panel
Status, Probe enable/disable/calibrate, System Configuration
Ratings
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 6
Current
0.3 A
Temperature
0 to 70
0C
Protection
NEMA 4X, 1/4 DIN panel mount or wall mount with optional hardware
Physical
Mounting
1/4 DIN panel mount or wall mount with optional hardware
Dimensions
Front cover: 6.5” wide x 5.5” tall; rear enclosure: 3.5” wide x 3.5” tall x 4.5” deep
Power
120 or 220 VAC, 60 or 50 Hz, auto-recognized. (Cable not included)
Weight
2 lbs.
*One expansion card slot is available
1.3. Front Panel Display
The 2300 front panel display shows a series of RUN screens and MENU screens. It is meant to be the secondary user interface, and has only select capabilities such as viewing status, calibrating probes, setting relay set that have already been defined by the primary, network interface.
There are eight navigation buttons below the screen, including four arrows. The functions of the others are:
RUN brings up the quad screen from anywhere
MENU brings up the front panel menu
BACK returns the user to the previous screen, and
ENTER selects the highlighted choice.
The Run screens include a summary screen showing the status and values of 4 probes at one time. If more than 4 probes are enabled in the controller, the Down and Up arrows may be used to see more than the first 4 probes. Detailed status screens are shown using the Right arrow button, which brings up the first probe status with monitor and relay settings. The Down and Up arrows cycle through screens of individual probes. Pressing the Left arrow button brings back the summary Run screen. When a probe activates an alarm the front panel display automatically switches to a flashing detailed status screen. To restore the screen to the full display simply click on (or, for the real controller, press) the Run button.
Each quadrant in the Run screen shows a probe label, current value with units, trend arrows for the reading, and a status light for the controller outputs. The status light is green for normal operation, yellow for an activated relay, and red for an activated alarm.
The Right Arrow brings up a detailed screen, shown on the right side of Figure 1. This screen lists the monitor outputs for this probe, including the monitor description and set point value, along with the Last Calibration date. The monitors are color coded, with green for normal operation, red for alarm condition, and yellow for an activated relay.
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 7
Pressing the Down arrow will cycle through all of the individual probe screens. Pressing the Enter button will bring up a Meter Action screen, allowing the user to acknowledge an alarm, if active, or setting relay trip points.
Run button will bring up the Run screen once again.
Figure 1 - The Run display on the left reports 4 process values at a time. If there are more than 4 pressing the down button reveals these. The identical image appears on the web browser. The Detail display on the right shows the detailed status of the Tank pH condition.
1.4. Front Panel Menu
The front panel menu is limited to a select group of functions. The outline of this menu structure is below, with descriptions.
Main Menu
View System Info
Unit name, IP address, software version, and system status information
Disable/Calibrate Probe
Manually disable, enable, or calibrate defined probes
Backup/Restore
Save or restore the unit configuration and setup to a removable SD card
Reboot System
Powers down and restarts controller
Prepare for Power Down
After a confirmation dialog is confirmed, all probes are disabled, all logs are closed, and the system is put in a state where it is safe to remove power from the controller.
Advanced Menu
View Raw Analog
Not used
View Raw Precision
Not used
Network Setup
Change the network between Dynamic (DHCP) and Static IP
Network Type
DHCP or Static IP; allows manual entry of static IP data
Web Server Port
Allows entry of port number
Commit and Restart
Confirm or cancel choices
Multi-Parameter Controller
Menu
Back
Enter
Run
AquaMetrix 2300
Tank pH
7.35
Tank ORP
386
Tank D.O.
3.92
Tank Level
88
pH
mV
ppm
cm
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 8
These functions are geared to ensuring that the unit may be connected to a computer or the web to allow setup and configuring. When not connected to a computer or web, the unit is still operating and all non-web-based features are still fully functional.
1.5. Ethernet Connection to a network switch or router
Connecting the 2300 to an existing network with the Ethernet cable allows the 2300 to be automatically assigned an IP address. You only need to set the Ethernet connection on your computer or smart device to configure via DHCP. The IP address is shown on the front panel during power up or can be seen via the Main Menu under View System Info (hit Menu then Enter from the front panel). The IP address may look like 192.168.1.7: save this address to access the 2300 from the network. Refer to Section 9.4 to use a static IP, used for direct connection to a computer.
Connect a CAT-5 cable into the Ethernet connector, shown in Figure 2, and connect the other end to a network switch or router. Type the address into the address window of your browser, hit enter, and a login window should appear. The default username is “admin” and the default password for admin is “aquametrix”. Note that there are two other levels of password protection, that don’t access everything that the admin level does. Users may be added via Setup/User Maintenance to uniquely identify login identities that access the controller. See Section 7.
1.6. Probe Status Screen
When first logged into the 2300 Controller from the web, the Probe Status screen appears showing all of the probe readings. An example of this screen showing four probes is shown in Figure 2. The order of probes shown is user determined via the Setup menu. Each has a label, a reading, and a graph of recent values, Last Calibration date, and a graphic of the alarm and relay states for each probe. The red (Alarm triggered) and green (not triggered) bars to the right of each graph depict the alarm (A) state and relay (R) state with set point and reset values. If using the optional analog output card, this value will also be shown under an “O” label to the right of the trend chart. A yellow bar indicates that an action is taken but non­alarm, such as a relay activated or analog output generated.
Clicking on a Config button brings up a new window that displays the probe description and allows the user to acknowledge an alarm, or to set high and low axis values. An example is shown is Figure 3. Alternatively, clicking on the graph brings up a new window with just the graph. The Previous and Next buttons allow the user to view data from the entire day’s log in 15-minute increments. The line represents the actual reading, while the light shading around the line indicates the high and low range of the data for that interval. The bar at the bottom is a time scale, and the line in it can be dragged to the time period desired. The left and right arrows on this bar are used to move the time frame by 45 minutes per click.
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 9
Figure 2 - Probe status screen. The screen lists all enabled sensors.
Figure 3 - Config screen for setting up axes of graph shown in the Probe Status screen.
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 10
1.7. The Web Menu
Clicking on one of the five top buttons brings up new screens with submenus. The overall menu structure and description for ADMIN access via the network is shown below in Table 2.
Table 2 - Menu structure of the web interface
Probe Status
Displays the current values of the sensors along with recent history and monitor status.
Config Button
Allows reformatting of the history, alarm acknowledgement.
Trend Chart
Allows viewing of longer histories.
Remote Display
Displays a simulated image of the current front panel of the 2300 with active buttons. Clicking on one of the 8 navigation buttons changes the screens and moves the cursor just as if these buttons were being pushed on the actual unit. These actions update the unit accordingly, which in turn updates the simulated display. Try to remember to leave the Remote Display in RUN mode!
Logs
Data Logs
Brings up the list of daily data logs in memory, allowing data to be emailed, downloaded, or deleted. Also does setting of the period for logging.
Alarm Logs
Brings up the list of alarm logs in memory, can be saved or deleted
System Logs
Brings up the list of system logs in memory, including web server errors, javascript errors, and system access. These can be viewed, saved, or deleted.
Set-up
Probe Configuration
Set up sensors, alarms, relays, and notifications. See Sections 4 and 5 for details.
User Maintenance
Set up users and passwords.
Data Logging Config
Set up data logging parameters
Network Setup
Set up web server, TFTP server, and emails
System Config
Miscellaneous configuration options
System Documentation
System doc file
Advanced
Miscellaneous functions
Logout
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 11
2. Installation and Housekeeping
Mount the 2300 Controller through the front panel, with the gasket on the outside of the mounting panel. Hook the sliding mounting blocks on each side of the back of the 2300 enclosure and tighten the setscrews against the mounting panel with a torque of 4 in-lbs as shown in Figure 4. Connect an AC power cord to the power connector labeled in Figure 5.
Figure 4 - Mounting hardware attached to panel with 4 in-lb of torque.
Figure 5 - 2300 back panel. All connections are made via connectors on the back.
Ethernet Connection
Fault LED
System Operating LED
Power On LED
Power 120-220 VAC
Pulse Rate Input
Analog Probe
inputs
Contact Inputs
Relay Outputs
Optional Card Slot
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 12
Apply power on to the 2300 Controller. The front panel will first display a splash screen with the AquaMetrix logo. If the 2300 Controller is operating properly, the LED's on the back panel indicate proper operation. The Power ON should be lit, the Fault should also be lit, and the System Operating should be blinking. The green LED on the Ethernet connector indicates a valid network link. The yellow LED will blink when there is network activity.
Before configuring the probes, you have some housekeeping tasks to do which, although optional, do allow you to use the 2300’s full capabilities like emailing.
2.1. Password Change
Yes, you get requests for password changes all the time, but if you want security and don’t want all the other purchasers of 2300’s to know your username and password, now’s the time to change them. Select Setup, User Maintenance, and then add a user. Make sure that you click on Admin for the user level, and write down the username and password. You can also add multiple users at the Admin, Editor, or User levels in order to track via the logs which users have done what to the 2300. Logout, then log back in under your new admin name, and delete the default name “admin”.
2.2. Time Setup
The time of day sets automatically if the 2300 is connected to the web via an NTP time server. To manually set the time, select Setup, System Configuration > Name > Time Configuration. Time zone, date, and time are set here. This page also allows you to customize the name of the 2300, identifying the unit by location or function or zodiac sign.
2.3. E-Mail Setup
The 2300 is preconfigured with an SMTP server entry suitable for sending emails and alerts along a network. The email sender is aquametrix2300@juno.com. If you want to set up sending from your own email account, talk to your IT Department or follow the directions below. Note that the 2300 internal web server does not support SSL security protocol, so you may have to establish a third party email provider such as juno.com to generate the emails.
To change the email server, select Setup > Network Setup > Email Setup > SMTP Connection Setup. The two required configuration values are the SMTP server address and the SMTP port. The SMTP server address can be entered as a DNS name (e.g. SMTP.mycompany.com) or an IP address (192.168.2.200). The SMTP port is a number that is assigned by your IT department and is normally 25. You should get both of these values from your IT department. The Global From Address will be used in all email messages as the sending email address (e.g. myemail@mycompany.com). You also need to enter email addresses, optionally enter form emails to use later, and send a test email.
2.4. Configure Data Logging
If you will be using the data logging feature of the 2300, check that the default settings for the data file are what you want. Default values are:
1. Data log files will be kept for 15 days
2. Data will be logged every 5 seconds
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 13
3. File delivery is by manual download from browser, and
4. Data is stored for configured probes only.
To change these settings, go to Setup > Data Logging Configuration, and follow the Wizard.
3. Connecting and Configuring Probes
3.1. Configuring Analog Sensors
In its standard configuration, the 2300 has four 4-20 mA (analog) inputs. The rear-side block connectors are shown in Figure 5 above. Wiring of a sensor depends on whether it is loop powered or externally powered; variations are shown in Figure 6.
1. If the probe is to be loop-powered, then its black (negative) wire connects to the
chosen AN-IN-x (where x is 1, 2,3, or 4) terminal and the red (positive) wire connects to one of the 24V terminal.
2. If the probe is externally powered such as from a transmitter, then its red (positive)
wire connects to the externally to the chosen AN-IN-x terminal and the black (negative) wire connects to the ground (GND) terminal.
3. If the probe has three wires, it needs an external power supply: refer to the probe
directions for hookup.
Figure 6 - Possible analog probe wiring diagrams. There are other possible three-wire probe setups, so please refer to your specific probe’s instructions for proper hook up.
3.2. Connecting Pulse Rate Output Sensors
The 2300 has inputs for 3 pulse rate sensors. These are typically flow sensors that output a sinusoidal or square wave, e.g. paddlewheel and magnetic flow meters. The contacts D1 and D2 are high precision optically isolated frequency counters, while D3 is not isolated. D3 uses a one-second count of the pulse rates to determine frequency.
N116-133 AquaMetrix 2300 Installation and Operation Manual Page 14
Wiring of a pulse rate sensor is shown below in Figure 7. The three wires from a typical unpowered probe are +, -, and signal.
Figure 7 - Configuring a flow sensor (paddle wheel or magmeter) to one of the pulse input connections.
3.3. Probe Configuration Setup
The Probe Configuration menu screen is shown in Figure 8. This screen appears in the web menu at Setup/Probe Configuration. This screen is used to enable or disable individual probes, to set up the order that the probes are displayed on both the front panel and web, to delete old probes, and to add or substitute new probes to the controller. Eight slots are listed, which is the order that the 2300’s software keeps track of the probes. Each slot may be configured as analog, digital, or virtual probes. Slot # refers to the position on the front panel and web Probe Status displays.
Loading...
+ 31 hidden pages