AquaMetrix 2250 Specifications

Page 1
Aqua
Description
by
Water Analytics
Aqua
Metrix
The AM-2250 series is built on the legacy of the Shark and 2200 controllers. Like its predecessors it is designed to be the most flexible, easy to use, and easy to see multi­parameter controller on the market.
Aquametrix AM-2250 Controller
Zero Cards
The AM-2250 comes complete. There are no extra costs associated with buying boards for different sensors, or buying components to achieve NEMA 4X.
Features
• pH, ORP, conductivity & flow parameters available
• Highly visible large backlit LCD display
• Flexible and easy calibration, including
multi-point conductivity calibration for ac­ids and bases
• Two 4-20mA process output with range scaling and PID Control
• Universal mounting hardware for surface, panel and pipe mounting
• Compatible with AquaMetrix models 60-se­ries differential pH/ORP sensors, 500-series combination style pH/ORP sensors, AS/ AM-series conductivity sensors, and most pulsed flow sensors
• Temperature or flow totalizer output
• Three control/alarm relays with tempera-
ture or flow totalizer output option
Four Parameters
Select the parameter you wish to measure from the easy-to-use menu. Choose Conduc­tivity, pH, ORP or Flow. The user interface was designed under the principle that the user should not need to read the manual.
Three Relays
The AM-2250 provides control of external de­vices using its three independent control and alarm relays. Each relay has adjustable high, low, and in range set-points, cycle timer with adjustable on and off times. This feature en­ables tighter control of batch processes by eliminating chemical overshoot. Third relay can be activated by temperature or flow total­izer reading as well.
Analog Outputs
The AM-2250 provides two isolated, indepen­dent and scalable 4-20 mA outputs.
One 4-20 mA output can be configured for PID control. The second 4-20mA output can be set for process or temperature.
Enclosure
The AM-2250 is packaged in a rugged NEMA 4X polycarbonate enclosure making it ideally suited for indoor and outdoor heavy-duty ap­plications. A mounting kit is included for sur­face and panel mounting. The enclosure out­line makes panel-mount cutouts simple. Pipe mounting kits are available.
One Big Display
The AM-2250 features a backlit LCD display can be seen from a distance. The keypad allows easy entry of menu items and numeric values.
Calibration
No other controller offers the same combina­tion of flexibility and ease for calibration. The process value is visible during calibration so the user knows when it has settled down. Cali­bration of pH can be with 2 or 3 points. Cali­bration of conductivity can take as many as 16 points so acids and bases can be measured through their conductivity.
All Calibration data is stored.
Applications
• Industrial process control, e.g. plating, food and beverage, chemical processing, pulp & paper, mining, food and beverage
• Municipal water and wastewater treatment
• Industrial and municipal waste treatment
and Neutralization
• Fume scrubbers
• HVAC, cooling towers and boilers
by
100 School Street
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Metrix
Water Analytics
978-749-9949
Page 2
AM-2250 Controller for pH, ORP, Conductivity, Flow
Technical Data
Probe Parameters
pH ORP Conductivity Flow
Sensor
Temperature Elements
6-Wire Differential
or Combination
100, 1000 Ω RTD 300, 3000 Ω NTC
or none
6-Wire Differential
or Combination
100, 1000 Ω RTD 300, 3000 Ω NTC
or none
4-Wire contacting :
Any cell constant
between 0.01 and 100
100, 1000 Ω RTD 300, 3000 Ω NTC
Pulse output:
Paddle-wheel,
Magmeter
n/a
Sensor Input
Measurement Range
(Process)
Measurement Range (Temperature)
Temperature Compensation
Calibration modes
Analog
Relays
Relay Modes
Ingress Protection
Electrical
Max. Power Input
Temperature
Humidity
Mounting
Dimensions
Power
Weight
Protection
Panel Cut-out
-600 to +600 mV -999 to +999 mV 0 to 9999 Ω 0 to 2000 Hz
0 to 14 pH -1000 to +1000mV
-20 to 120 °C -20 to 120 °C -20 to 120 °C
Automatic
or none
pH: Automatic or Manual
2 or 3 points
Display temperature
ORP: Manual
1 point
0.055 to 500,000 µS/cm (Depending on the
cell constant)
Automatic or Manual
-20 to 120 °C
Up to 16 points K factor input
0 to 999 in any units
Flow Totalizer
0 to 999 in any units
Display Flow Totalizer
Outputs
Two 4-20 mA outputs
Scalable 4-20mA with PID (Process)
Scalable 4-20mA (Process or Temperature)
Optically isolated. Max Load - 800 Ω
3 Dry contact relay with snubber circuit
10A @ 120/240 VAC or 8A @ 30 VDC (Resistive Load)
5A @ 120/240 VAC or 4A @30 VDC (Inductive load)
Rising/Falling/In Range.
Options: Relay Delay, Cycle, Overfeed Timer, Override
One relay can be triggered by
temperature or flow totalizer.
Ratings
NEMA 4X
ETL (US and Canada) and CE pending
0.2 A @ 115 VAC or 15 W
-20 to 70 °C
0 to 90% Relative Humidity, non-condensing
Physical
Wall mount, panel mount with kit provided. pipe mount with optional kit
Front cover: 5.5”x5.5” (14 cm x 14 cm). Depth: 5” (13 cm)
120/240 VAC, 50 or 60 Hz
2 lbs
NEMA 4X
5.4” x 5.4” (138 x 138 mm) full DIN
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)