Aquametrix 2200P Datasheet

Page 1
Aqua
Description
by
Water Analytics
Aqua
Metrix
Model 2200P pH analyzer/controller is a depend­able industrial grade instrument designed for con­trolling rising and falling pH in the most demanding environments.
Extremely easy to operate with a large bright LED display and intuitive user inter face, this micropro­cessor-based analyzer/controller is packaged in a NEMA 4X berglass reinforced polyester enclo­sure and used in a wide variety of industrial and municipal applications.
Frequently used functions are accessed through an intuitive step-through menu printed directly on the front panel for ease of navigation. Pressing the MENU key scrolls through the menu, illuminating a small LED beside the selected item and displaying the stored value. Arrow keys are used to change the value. Seldom used adjustments such as the temperature unit are set using DIP switches on the back of the swingout front panel.
INPUTS
The 2200P accepts two types of pH probes, the AquaMetrix P60 Series Differential (three elec­trode) pH probe and conventional combination pH probes.
CALIBRATION
Calibration is fast and easy. Press the MENU key until the CALIBR ATION LED illuminates. Place the probe in a buffer solution. Use the arrow keys to set the value. Press ENTER twice. Repeat the process for the second
bufer solution. The 2200P is now calibrated.
Aquametrix Model 2200P
pH Anayzer / Controller
OUTPUTS
Model 2200P provides three analog outputs, 4 -20 mA, 0 -5 VDC and 0-1 mA.
Each control relay has an independent cycle timer, with easy to adjust Cycle On and Cycle Off times. This feature enables a tighter control of batch pro­cesses by eliminating chemical overshoot.
Control Relay A has a manual override which places the relay in the AUTO, OFF, or ON mode. This feature enables the operator to override the controller from the front panel, turning the relay on and off. This is a useful feature when servicing auxiliar y devices.
An alarm relay has both high and low setpoints with a xed deadband. When any relay is ener­gized, a corresponding LED on the front panel is illuminated, clearly indicating alarm conditions and if auxilliary devices are energized.
DIAGNOSTICS
A self-diagnostics program clearly indicates any system errors or faults by turning the STATUS LED on the front panel from green to red. The operator then scrolls through the menu to STATUS to de­termine the fault code. The Manual indicates the source and likely remedy for the fault. When the status fault is corrected the STATUS LED will turn back to green.
ENCLOSURE
Model 2200D is packaged in a rugged NEMA 4X berglass reinforced polyester enclosure. It is ide­ally suited for heavy-duty outdoor applications such as municipal and industrial wastewater ap­plications.
Features
Simple step-through menu
design for quick & easy setup
Bidirectional control
through two control relays with independent setpoints for Relay On and Relay Off
Cycle-Time option mini-
mizes chemical overshoot
High/low alarm relay
ALARM LED on front panel
Accepts P60 Series differ-
ential pH probes with replace­able salt bridge or standard combination pH probes
Output signal is eld-scal-
able or maximum resolution
Automatic buffer recognition
STATUS errors illumi-
nate LED on front panel
Test feature allows simula-
tion of pH readings to test relay set-points and analog settings
Password protection and
watchdog timer provide security
Applications
Process Control
Industrial and Municipal Water
Treatment
Industrial and Municipal Waste
Treatment and Neutralization
Fume Scrubbers
Plating, Circuit Board Manu-
facturing, Food and Beverage, Chemical Processing, Pulp & Pa­per, Mining, Nuclear Energy and Pharmaceutical Industries
by
100 School Street
Andover, MA 01810
Toll free - 855-747-7623
www.Aquametrix.com
Metrix
Water Analytics
978-749-9949
Page 2
Model 2200P pH Analyzer / Controller
Technical Data
Display 4 x 7 segment 1/2” LED Display
Indicator Lights 4 LED indicators: Status, Alarm, Relay A,
Relay B
Measuring Range
pH 0.00 to 14.00 pH
Temperature 0 to 400C (32 to 1040F)
Power Requirements
Standard 98-132 VAC 50/60 Hz, (less than 10 VA)
Optional 196-26 4 VAC 50/60 Hz (less than 10 VA)
23-26 VDC (nominal current: 150 mA)
Ambient Conditions
Temperature -30 to 50°C (-22 to 122°F)
Humidity 0 to 90% RH non-condensing
Control Relays
Number Two Control Relays: Relay A, Relay B
Rating 5 A 115/230 VAC, 5A 30 VDC, SPDT
Bidirectional Relays factor y preset for bidirectional
control, can be individually eld set for ris­ing or falling process
On/ Off Setpoints Field selectable full scale
Fail-safe Normal or fail safe operation
Alarm Relay
Number One Alarm Relay
Rating 5A 115/230 VAC, 5A 30 VDC, SPDT
Alarm High Field selectable 0 -14 pH
Alarm Low Field selectable 0 -14 pH
Deadband Fixed at 2% of full scale
Fail Safe Normal or fail safe operation
Relay Indicators Three individual LEDs indicate status
of two control relays and alarm relay
Analog Outputs
Non-isolated 0-1 mA, 100 ohms maximum load
Non-Isolated 0-5 VDC, 1000 ohms minimum load
Isolated 4-20 mA, 800 ohms maximum load
Output is isolated from the input, the ground, line power and all other outputs.
Range Expand Analog outputs can be scaled to rep-
resent any segment of the instrument scale, minimum of 10% of full scale.
Output Hold Analog outputs are placed on hold
when the instrument is in menu mode.
Temperature The 0-1 mA and 0-5 VDC outputs can
be selected to track either the pH or the temperature of the process.
Sensor to Controller Distance
Differential P60 Series Differential pH Probe: maximum
Combination 500 Series: maximum distance 25 feet (7.62 m)
Calibration 2 point, manual or automatic buf fer recognition
Temperature Compensation
Automatic 0-100°C (32 to 212°F) if probe has temperature
Diagnostics
Invalid Entries Identied by ashing LEDs
System Error Two color STATUS LED indicates system sta-
Alarm Relay Alarm relay can be enabled to activate upon
Test Display value and analog outputs can be set to
Safety & Security
Operator Password protected (activated by DIP switch)
Memory Nonvolatile memory (EPROM)
Microprocessor Watchdog timer monitors microprocessor. In-
Sensitivity 0.1% of span
Stability 0.1% of span per 24 hours non-cumulative
Temperature Drift Zero - 0.01% of span per °C
Enclosure NEMA 4X molded berglass reinforced polyes-
Mounting
Standard Surface Mount
Optional Panel Mount kit P/N C35-68
Weight 3.5 lbs. (1.6 kg)
Related Products - Probes & Accessories
P60C-6 Industrial quality differential pH probe with replace-
P60C-8 Industrial quality differential pH probe with replace-
P575 Industrial quality combination style pH
P60-HTC CPVC Wet/Hot tap ball valve assembly
STC60 Tank mount hardware for pH/ORP differential probes
distance 3000 feet (914 m)
sensor.
tus. Green no fault, Red system fault. Scroll to Status in Menu to determine fault code. Refer to Manual for probable cause and remedy.
high and low process values and memory loss.
any value for testing and diagnostic purposes.
strument automatically returns to online opera­tion if left in menu mode for more than 10 min­utes and no key is pressed.
Span - 0.01% of span per °C
ter enclosure with four 1/2” conduit holes and mounting feet for sur face mount. A NEMA 4X plug is provided for one hole.
Pipe Mount kit P/N C35- 69
able salt bridge, temperature sensor and integral preamplier. Compression tting 1-1/4” NPT.
able salt bridge, temperature sensor and inte­gral preamplier. Fixed threads 1-1/2” NPT.
probe. Fixed threads 1” NPT.
Ordering Information
2200P pH Analyzer/Controller
1 120 VAC, 50/60 Hz Power
2 240 VAC, 50/60 Hz Power
A Sur face Mount
B Panel Mount
C Pipe Mount
2200P
www.Aquametrix.com 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)
Loading...