Aquametrix 2200C Datasheet

Aquametrix Model 2200C
Aqua
Aqua
by
Description
Model 2200C conductivity analyzer/controller is a dependable industrial grade instrument de­signed for controlling rising and falling conduc­tivity in the most demanding environments.
Extremely easy to operate with a large bright LED display and intuitive user interface, this microprocessor-based analyzer/controller is packaged in a NEMA 4X berglass reinforced polyester enclosure and used in a wide variety of industrial and municipal applications.
Frequently used functions are accessed through an intuitive step-through menu printed directly on the front panel for ease of navigation. Press­ing the MENU key scrolls through the menu, il­luminating a small LED beside the selected item and displaying the stored value. Arrow keys are used to change the value. Seldom used adjust­ments such as password activation are set us­ing DIP switches on the back of the swing-out front panel.
INPUTS
The 2200C accepts AquaMetrix MS (epoxy body) or AM (stainless steel body) conductivity sensors. These have graphite or stainless steel electrodes and can be ordered for use in stan­dard or high temperature/pressure applications.
To optimize resolution the 2200C can be or­dered pre-congured for any of seventeen standard conductivity ranges, plus two special ranges.
CALIBRATION
tion. The 2200C is now calibrated.
Metrix
Water Analytics
Conductivity Analyzer / Controller
OUTPUTS
Model 2200C provides three analog outputs, 4-20 mA, 0-5 Vdc and 0-1 mA.
The instrument also provides control of an ex­ternal device using the control relay. The control relay can be selected to activate in response to either rising or falling conductivity. Setpoints for Relay On and Relay Off are independently set.
The control relay has a builtin cycle timer, with easy-to-adjust Relay On and Relay Off cycle times. This feature enables tighter control of batch processes.
The Control Relay has a manual override, which overrides the controller, placing the relay in AUTO, OFF, or ON mode. This is a useful fea­ture when servicing auxiliary devices connected to the relay.
An alarm relay has both high and low setpoints with a xed deadband. When this relay is ener­gized, the Alarm LED on the front panel also il­luminates to clearly indicate an alarm condition.
DIAGNOSTICS
A self-diagnostics program clearly indicates any system errors or faults. When the unit is online, the RUN LED on the front panel will illuminate. In the event of a system error the FAIL LED will illuminate. When the FAIL LED is illuminated, the Status menu item will display a fault code. The Manual will indicate the source and likely remedy for the fault. When the fault is corrected the FAIL LED will no longer be illuminated.
ENCLOSURE
Model 2200C is packaged in a rugged NEMA 4X berglass reinforced polyester enclosure. It is ideally suited for heavy-duty outdoor applica­tions such as municipal and industrial wastewa­ter applications.
Features
Simple step-through menu
design for quick and easy setup
On/Off cycle timer feature
High/low alarm relay
ALARM LED on front panel
Output signal eld-scalable
for maximum resolution
Status errors illuminate
FAIL LED on front panel
Test feature allows sweep-
through range to test set­points and alarms or to simulate value for setup
Auto tempera-
ture compensation
Password protection
High temperature/pres-
sure and low temperature/ pressure cells available
Applications
• Process Control and Cooling Tower Control
• Water Quality Monitoring
• Industrial and Municipal Water Treatment, including Reverse Os­mosis, De-ionization, Desalina­tion, and High Purity Water Pro­cessing
• Plating, Computer Chip Fabri­cation, Circuit Board Manufactur­ing, Food and Beverage, Chemi­cal Processing, Mining, Nuclear Energy and Pharmaceutical In­dustries
by
100 School Street
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Metrix
Water Analytics
978-749-9949
Model 2200C Conductivity Analyzer / Controller
Technical Data
Display 4 x 7 segment 1/2” LED Display
Indicator Lights 4 LED indicators: Status, Alarm, Relay
Measuring Range
Range The range is factory set but may be eld
Temperature 0 to 100°C (32 to 212°F)
Power Requirements
Standard 98-132 VAC 50/60 Hz, (less than 10 VA)
Optional 196-26 4 VAC 50/60 Hz (less than 10 VA)
Ambient Conditions
Temperature -30 to 50°C (-22 to 122°F) Humidity 0 to 90% RH non-condensing
Control Relays
Number Two Control Relays: Relay A, Relay B Rating 5 A 115/230 VAC, 5A 30 VDC, SPDT On/ Off Setpoints Field selectable full scale Cycle Timer Field selectable 0 - 600 seconds Manual Override AUTO/OFF/ON Fail-safe Normal or fail safe operation
Alarm Relay
Number One High and Low Alarm Relay Rating 5 A, 115/230 VAC, 5A 30 VDC, SPDT
Alarm High Field selectable full scale Alarm Low Field selectable full scale
Deadband Fixed at 2% of full scale Fail Safe Normal or fail safe operation
Relay Indicators Two individual LEDs indicate status
Analog Outputs
Non-isolated 0-1 mA, 100 ohms maximum load
Non-Isolated 0-5 VDC, 1000 ohms minimum load
Isolated 4-20 mA, 800 ohms maximum load
Range Expand Analog outputs can be scaled to rep-
Output Hold Analog outputs are placed on hold
Temperature The 0-1 mA and 0-5 VDC outputs
Automatic 0-100°C (32 to 212°F)
Calibration 1 point, manual Temperature Compensation
Automatic
A, Relay B
changed
23-26 VDC (nominal current: 150 mA)
of two control relays and alarm relay
Output is isolated from the input, the ground, line power and all other outputs.
resent any segment of the instrument scale, minimum of 10% of full scale.
when the instrument is in menu mode.
can be selected to track either the pH or the temperature of the process.
0-100°C (32 to 212°F) if probe has tem­perature sensor.
Diagnostics
Invalid Entries Identied by ashing LEDs
Fail LED Illuminates red to indicate Status Fault. Scroll
Auto LED Illuminates red when Alarm Relay is energized,
Test Display value and analog outputs can be set to
Safety & Security
Operator Password protected (activated by DIP switch)
Memory Nonvolatile memory (EPROM)
Microprocessor Watchdog timer monitors microprocessor. In-
Sensitivity 0.1% of span
Stability 0.1% of span per 24 hours non-cumulative
Temperature Drif t Zero - 0.01% of span per °C
Enclosure NEMA 4X molded berglass reinforced polyes-
Mounting
Standard Surface Mount
Optional Panel Mount kit P/N C35- 68
Weight 3.5 lbs. (1.6 kg)
to Status in Menu to determine fault code. Re­fer to Manual for probable cause and remedy
indicating high/low process value or memory loss.
any value for testing and diagnostic purposes
strument automatically returns to online opera­tion if left in menu mode for more than 10 min­utes and no key is pressed.
Span - 0.01% of span per °C
ter enclosure with four 1/2” conduit holes and mounting feet for surface mount. A NEMA 4X plug is provided for one hole.
Pipe Mount kit P/N C35-69
Related Products - Probes & Accessories
MS Cells Epoxy, low/high pressure, compression tting 3/4” NPT.
MHMSC Tank Mount Hardware for MS Cells.
MS-HTC CPVC Wet/Hot Tap Mount for MS Cells.
Conductivity Range
(µSiemens/cm)
0-2 0.01 0-10 10
0-5 0.01 0-20 10
0-10 0.05 0-50 10
0-20 0.05 0-100 20
0-50 0.05 0-500 50
0-100 0.05 0-1000 50
0-200 0.5
0-500 0.5 Special Ranges
0-1000 0.5 0-15% H2SO
0-2000 1 0-15% NAOH 50 tt
0-5000 1
Cell
Constant
Conductivity Range
(mSiemens/cm)
4
Cell
Constant
50
Ordering Information
2200C Conductivity Analyzer/Controller
2200C
1 120 Vac, 50/60 Hz Power
2 240 Vac, 50/60 Hz Power
A Sur face Mount
B Panel Mount
C Pipe Mount
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)