Applied Biosystems GeneAmp 9700 User manual

GeneAmp® PCR System 9700
0.5mL Sample Block Module
User’s Manual
© Copyright 2006, 2010 Applied Biosystems. All rights reserved.
Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that may appear in this document.
NOTICE TO PURCHASER:LIMITED LICENSE
The combination of the GeneAmp® PCR System 9700 Sample Block Module with a GeneAmp® PCR System 9700 Base Unit is licensed under patent
claims for thermal cycler apparatus and counterpart patent claims in other countries. No other rights are conveyed expressly, by implication, or by
estoppel. Further information on purchasing licenses may be obtained by contacting the Director of Licensing, Aplied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.
TRADEMARKS:
AB (Design), Applied Biosystems, and MicroAmp are registered trademarks, Applera and Celera Genomics are trademarks of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries.
GeneAmp is a registered trademark of Roche Molecular Systems, Inc.
All other trademarks are the sole property of their respective owners.
Part Number 4307808 Rev. D 06/2010
Contents
1 0.5-mL Sample Block Module
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
General Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Electrical Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Instrument Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Operating Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Grounding and Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Voltage Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Input/Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Pollution Category. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Instrument Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Communautes Europeennes (CE) Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
FCC Compliance (U.S.). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Routine Maintenance for Safe Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Technical Support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
To Reach Us On the Web. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Hours for Telephone Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
To Reach Us by Telephone or Fax in North America . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Documents-on-Demand. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
To Reach Us by E-Mail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Regional Offices Sales and Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
About the 0.5-mL Sample Block Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Sample Block. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Heated Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Peltier Heating/ Cooling Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
Modules, Accessories, and Disposables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Ordering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Installing the 0.5-mL Sample Block Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Installing the Sample Block Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Loading Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Loading Samples: Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Reaction Tube/Frame Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-17
i
Loading Samples into the Reaction Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Placing the Reaction Tubes into the Sample Block . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
Cleaning the Sample Block Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Cleaning: Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Cleaning Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Cleaning the Heated Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Cleaning the Sample Block Wells. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
Running the Calibration Verification Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Setting Up the 0.5-mL Probe Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
Configuring the System 9700 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
Running the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Evaluating the Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
Running the Temperature Non-Uniformity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Setting Up the 0.5-mL Probe Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-28
Configuring the System 9700 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-29
Running the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Evaluating the Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-33
Running System Performance Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-34
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-34
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-34
Running the Rate Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-34
Running the Cycle Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-36
Data Sheet: Calibration Verification Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-38
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-38
Data Sheet: Temperature Non-Uniformity Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-39
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-39
Index
ii
0.5-mL Sample Block Module 1
Safety and Regulatory Information
General Symbols
IMPORTANT Read this section before you install the GeneAmp
the 0.5-mL Sample Block Module.
The following symbols are used on Applied Biosystems instruments. Whenever such symbols appear on instruments, please observe appropriate safety procedures.
Electrical Symbols
This symbol indicates the on position of the main power switch.
This symbol indicates the off position of the main power switch.
This symbol indicates the on/off position of a push-push main power switch.
This symbol indicates that a terminal may be connected to another instrument’s signal ground reference.
This is not a protected ground terminal. This symbol indicates that this is a protected ground terminal that must be
connected to earth ground before any other electrical connections are made to the instrument.
®
PCR System 9700 and
~
~
This symbol indicates that this terminal either receives or delivers alternating current or voltage.
This symbol indicates that this terminal either receives or delivers alternating and direct current or voltage.
0.5-mL Sample Block Module 1-1
Non-electrical Symbols
This symbol indicates the presence of high voltage and warns the user to proceed with caution.
This symbol alerts you to consult the manual for further information and to proceed with caution.
This symbol illustrates a heater hazard. Proceed with caution when working around these areas to avoid being burned by hot components.
Electrical Safety
Testing
Instrument Labels
Routine safety testing of analytical instruments ( e.g. , high potential voltage testing) may be required by various safety agencies.
Testing should only be carried out by qualified personnel after seeking advice from the Applied Biosystems Service Department.
The following instrument labels are on the system 9700 with the 0.5-mL Sample Block Module installed.
! WARNING !
circuit continuity is vital for safe operation of equipment. Ne ver operate equipment with grounding conductor disconnected.
AVERTISSEMENT: Debrancher le cordon d’alimentation avant d’ouvrir la continuite des masses est essentielle.
Pour un fonctionnement sans danger. Ne jamais utiliser l’equipment si le fil de terre n’est pas raccorde.
\
! WARNING !
area to avoid being burned by hot components.
Attention. Surface chaude.
Disconnect supply cord before opening. Grounding
Hot Surface. Use care when working around this
1-2 0.5-mL Sample Block Module
Operating Precautions
Precautions
The following precautions should be taken whenever you operate the system 9700 with the 0.5-mL Sample Block Module installed.
Note This instrument is able to withstand transient overvoltage according to Installation
Category II as defined in IEC 1010-1.
General Use
\
CAUTION The instrument should be used according to the
instructions provided in this manual. If used otherwise, the protection provided by this instrument may be impaired.
Environment, Humidity, and Temperature
CAUTION This instrument is designed for indoor use.
CAUTION Do not operate in a Cold Room or a refrigerated area.
The system 9700 will operate safely when the ambient temperature is 5–40 °C (41–104 °F). The system 9700 will meet performance specifications when the ambient temperature is 15–30 °C and the ambient relative humidity is 20–80%. These specifications have been calculated for altitudes between 0 and 2,000 meters.
CAUTION This instrument is not designed for operation in an
explosive environment. Do not place the instrument close to potentially explosive materials or objects.
CAUTION This instrument is not designed for operation with the
heated cover retracted when running at 4 °C. If the heated cover is retracted and the instrument runs at 4 °C, water condensation may be excessive in the sample block area.
0.5-mL Sample Block Module 1-3
Sample Block Module
! WARNING !
heated cover or touch the sample block module when the word Hot displays on the screen. This indicates a sample block temperature above 50 °C.
The hot areas for the 0.5-mL Sample Block Module are shown below.
\
CAUTION To protect your samples and to guarantee the best
temperature uniformity, keep the heated cover closed at all times, except when loading or unloading samples.
To protect yourself against burns, do not open the
Grounding and
Electrical Safety
Voltage Quality
Input/Output
Connections
Physical
Specifications
The system 9700 must be grounded for protection against electrical shock.
! WARNING !
this does not provide positive ground protection.
Refer to the GeneAmp
®
PCR System 9700 Base Module User’s Manual
Do not use an adapter to a two-terminal outlet since
(P/N 0993-6247) for more information on electrical safety.
Line voltage must be within
±
10% of the nominal value. High or low voltages may hav e adverse effects on the electronic components of the instrument. In areas where the supplied power is subject to fluctuations exceeding these limits , a pow er line regulator may be required.
Refer to the GeneAmp
®
PCR System 9700 Base Module User’s Manual
(P/N 0993-6247) for information on the input and output connections of this instrument.
Refer to the GeneAmp
®
PCR System 9700 Base Module User’s Manual
(P/N 0993-6247) for information on the physical specifications of this instrument.
1-4 0.5-mL Sample Block Module
Pollution Category
Instrument Storage
Communautes
Europeennes (CE)
Compliance
FCC Compliance
(U.S.)
Routine
Maintenance for
Safe Operation
The system 9700 will operate safely in environments that contain nonconductive foreign matter up to Pollution Degree 2 in IEC 1010-1.
The system 9700 must be stored at a temperature between -20 °C and 60 °C (-4 °F and 140 °F) at altitudes ranging from 0–12,000 meters above sea level.
IMPORTANT The system 9700 is guaranteed to meet performance specifications only
when the ambient temperature is 15–30 °C and the ambient relative humidity is 20–80% (between altitudes of 0 and 2,000 meters).
\
All instruments shipped to the European Union (EU; formerly known as the European Community) have the CE label on the bac k of the instrument. This label signifies that these instruments comply with the Electromagnetic Compatibility and Low Voltage Directives.
This product is classified as a digital device used exclusively as industrial, commercial, or medical test equipment. It is exempt from the technical standards specified in Part 15 of the FCC Rules and Regulations, based on Section 15.103(c).
Before using any cleaning or decontamination method, e xcept those recommended in the manual, the user should check with Applied Biosystems to ensure that the proposed method will not damage the equipment.
Maintain your instrument in good working order. In the event that the instrument has been subjected to adverse environmental conditions (such as fire, flood, earthquake, etc.), a service inspection of the instrument should be made to ensure safe operation.
0.5-mL Sample Block Module 1-5
Technical Support
To Reach Us on the
Web
Hours for Telephone
Technical Support
To Reach Us by
Telephone or Fax in
North America
Applied Biosystems web site address is: http://www.appliedbiosystems.com/techsupport
We strongly encourage you to visit our web site for answers to frequently asked questions, and to learn more about our products. You can also order technical documents and/or an index of available documents and have them faxed or e-mailed to you through our site (see the “Documents on Demand” section below).
In the United States and Canada, technical support is available at the following times.
Product Hours
Chemiluminescence 9:00 a.m. to 5:00 p.m. Eastern Time LC/MS 9:00 a.m. to 5:00 p.m. Pacific Time All Other Products 5:30 a.m. to 5:00 p.m. Pacific Time
See the “Regional Offices Sales and Service” section below for how to contact local service representatives outside of the United States and Canada.
Call Technical Support at 1-800-831-6844, and select the appropriate option (below) for support on the product of your choice at any time during the call. (To open a service call for other support needs, or in case of an emergency, press 1 after dialing 1-800-831-6844.)
For Support On This Product Dial 1-800-831-6844, and...
®
RISM
ABI P
ABI P
3700 DNA Analyzer
®
RISM
3100 Genetic Analyzer
Press FAX 8 650-638-5891
Press FAX 26 650-638-5891
BioInformatics (includes BioLIMS BioMerge
DNA Synthesis
Fluorescent DNA Sequencing
Fluorescent Fragment Analysis (includes GeneScan
Integrated Thermal Cyclers
1-6 0.5-mL Sample Block Module
, and SQL GT
®
applications)
applications)
,
Press FAX 25 505-982-7690
Press FAX 21 650-638-5981
Press FAX 22 650-638-5891
Press FAX 23 650-638-5891
Press FAX 24 650-638-5891
For Support On This Product Dial 1-800-831-6844, and...
PCR and Sequence Detection
Press FAX 5, or call
1-800-762-4001, and press 1 for PCR, or 2 for Sequence Detection
240-453-4613
FMAT
Peptide and Organic Synthesis
Protein Sequencing
Chemiluminescence
LC/MS
Telephone FAX
1-800-899-5858, and press 1, then press 6
Press FAX 31 650-638-5981
Press FAX 32 650-638-5981
Telephone FAX
1-800-542-2369 (U.S. only), or
1-781-271-0045 (Tropix)
Telephone FAX
1-800-952-4716 650-638-6223
508-383-7855
781-275-8581 (Tropix)
9:00 a.m. to 5:00 p.m. ET
9:00 a.m. to 5:00 p.m. PT
Documents on
Demand
Free 24-hour access to Applied Biosystems technical documents, including MSDSs , is available by fax or e-mail.
You can access Documents on Demand through the internet or by telephone:
If you want to order... Then...
through the internet
by phone from the United States or Canada
Use http://www.appliedbiosystems.com/techsupport You can search for documents to order using keywords. Up to five documents can be faxed or e-mailed to you by title.
a. Call 1-800-487-6809 from a touch-tone phone. Have your fax
number ready.
b. Press 1 to order an index of available documents and have it faxed
to you. Each document in the index has an ID number. (Use this as
your order number in step “d” below.) c. Call 1-800-487-6809 from a touch-tone phone a second time. d. Press 2 to order up to five documents and have them faxed to you.
0.5-mL Sample Block Module 1-7
If you want to order... Then...
by phone from outside the United States and Canada
a. Dial your international access code, then 1-858-712-0317 from a
touch-tone phone.
Have your complete fax number and country code ready (011 precedes the country code).
b. Press 1 to order an index of available documents and have it faxed
to you. Each document in the index has an ID number. (Use this as
your order number in step “d” below.) c. Call 1-858-712-0317 from a touch-tone phone a second time. d. Press 2 to order up to five documents and have them faxed to you.
To Reach Us by
E-Mail
Regional Offices
Sales and Service
Contact technical support by e-mail for help in the following product areas.
For this product area Use this e-mail address
Chemiluminescence info@appliedbiosystems.com Genetic Analysis galab@appliedbiosystems.com LC/MS apisupport@sciex.com PCR and Sequence Detection pcrlab@appliedbiosystems.com Protein Sequencing, Peptide and DNA
Synthesis
corelab@appliedbiosystems.com
If you are outside the United States and Canada, you should contact your local Applied Biosystems service representative.
The Americas
United States Applied Biosystems 850 Lincoln Centre Drive Foster City, California 94404
Tel: (650) 570-6667
(800) 345-5224
Fax: (650) 572-2743
Europe
Austria (Wien) Tel: 43 (0)1 867 35 75 0 Fax: 43 (0)1 867 35 75 11
Belgium Tel: 32 (0)2 712 5555
Fax: 32 (0)2 712 5516 Czech Republic and Slovakia (Praha)
Tel: 420 2 61 222 164 Fax: 420 2 61 222 168
Denmark (Naerum) Tel: 45 45 58 60 00
Fax: 45 45 58 60 01
Latin America (Del.A. Obregon, Mexico) Tel: (305) 670-4350
Fax: (305) 670-4349
Hungary (Budapest) Tel: 36 (0)1 270 8398
Fax: 36 (0)1 270 8288 Italy (Milano)
Tel: 39 (0)39 83891 Fax: 39 (0)39 838 9492
The Netherlands (Nieuwerkerk a/d IJssel) Tel: 31 (0)180 331400
Fax: 31 (0)180 331409 Norway (Oslo)
Tel: 47 23 12 06 05 Fax: 47 23 12 05 75
1-8 0.5-mL Sample Block Module
Europe
Finland (Espoo) Tel: 358 (0)9 251 24 250
Fax: 358 (0)9 251 24 243
France (Paris) Tel: 33 (0)1 69 59 85 85
Fax: 33 (0)1 69 59 85 00 Germany (Weiterstadt)
Tel: 49 (0) 6150 101 0 Fax: 49 (0) 6150 101 101
Spain (Tres Cantos) Tel: 34 (0)91 806 1210
Fax: 34 (0)91 806 1206 Sweden (Stockholm)
Tel: 46 (0)8 619 4400 Fax: 46 (0)8 619 4401
Switzerland (Rotkreuz) Tel: 41 (0)41 799 7777
Fax: 41 (0)41 790 0676 Middle Eastern Countries and North Africa
(Monza, Italia) Tel: 39 (0)39 8389 481
Fax: 39 (0)39 8389 493 All Other Countries Not Listed
(Warrington, UK) Tel: 44 (0)1925 282481
Fax: 44 (0)1925 282509
Poland, Lithuania, Latvia, and Estonia (Warszawa)
Tel: 48 (22) 866 40 10 Fax: 48 (22) 866 40 20
Portugal (Lisboa) Tel: 351 (0)22 605 33 14
Fax: 351 (0)22 605 33 15 Russia (Moskva)
Tel: 7 095 935 8888 Fax: 7 095 564 8787
South Africa (Johannesburg) Tel: 27 11 478 0411
Fax: 27 11 478 0349 United Kingdom (Warrington, Cheshire)
Tel: 44 (0)1925 825650 Fax: 44 (0)1925 282502
South East Europe (Zagreb, Croatia) Tel: 385 1 34 91 927
Fax: 385 1 34 91 840 Africa (English Speaking) and West Asia
(Fairlands, South Africa) Tel: 27 11 478 0411
Fax: 27 11 478 0349
Japan
Japan (Hatchobori, Chuo-Ku, Tokyo) Tel: 81 3 5566 6100 Fax: 81 3 5566 6501
Eastern Asia, China, Oceania
Australia (Scoresby, Victoria) Tel: 61 3 9730 8600
Fax: 61 3 9730 8799 China (Beijing)
T el: 86 10 6238 1156 Fax: 86 10 6238 1162
Malaysia (Petaling Jaya) Tel: 60 3 758 8268
Fax: 60 3 754 9043 Singapore
Tel: 65 896 2168 Fax: 65 896 2147
0.5-mL Sample Block Module 1-9
Eastern Asia, China, Oceania
Hong Kong Tel: 852 2756 6928
Fax: 852 2756 6968 Korea (Seoul)
Tel: 82 2 593 6470/6471 Fax: 82 2 593 6472
Taiwan (Taipei Hsien) Tel: 886 2 2698 3505
Fax: 886 2 2698 3405 Thailand (Bangkok)
Tel: 66 2 719 6405 Fax: 66 2 319 9788
1-10 0.5-mL Sample Block Module
About the 0.5-mL Sample Block Module
Overview
Diagram
The GeneAmp
®
PCR System 9700 amplifies nucleic acids using the polymerase chain reaction (PCR) process. The 0.5-mL Sample Block Module attaches to the top of the GeneAmp® PCR System 9700 Base Module and allows you to perf orm up to 60 reactions.
You can remove the 0.5-mL Sample Block Module and replace it with another sample block module as needed. Interchangeability lets you change sample well formats as well as throughput capacity.
IMPORTANT Interchanging sample block modules may require a firmware change. If the sample block module is placed on a base module running an earlier software version, the instrument will not operate correctly. To upgrade the base module software, use the 3.5” floppy disk that is shipped with the 0.5-mL Sample Block Module. For instructions on upgrading the software, see the GeneAmp or visit our web site (www.appliedbiosystems.com/techsupport).
0.5-mL Sample Block Module
System 9700 Base Module
®
PCR System 9700 Base Module User’s Manual (P/N 0993-6247)
Heated cover
Sample block
0.5-mL Sample Block Module 1-11