1
Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
Safety Information
Understand and follow operating instructions carefully. Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.
WARNING
Identifies hazardous conditions and actions that could cause BODILY HARM or DEATH
CAUTION
Identifies conditions and actions that could DAMAGE the meter or equipment under test
WARNING
˙ When using test leads or probes, keep your fingers behind the finger guards. ˙ Remove test lead from Meter before opening the battery door or Meter case. ˙ Use the Meter only as specified in this manual or the protec­ tion by the Meter might be impaired. ˙Always use proper terminals, switch position, and range for measurements. ˙Verify the Meter’s operation by measuring a known voltage. If in doubt, have the Meter serviced. ˙ Do not apply more than the rated voltage, as marked on Meter, between terminals or between any terminal and
earth ground. ˙ Only replace the blown fuse with the proper rating as speci­ fied in this manual. ˙ Use caution with voltages above 30 Vac rms, 42 Vac peak , or 60 Vdc. These voltages pose a shock hazard. ˙ To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator =
appears. ˙ Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or
capacitance.
˙ Do not use Meter around explosive gas or vapor.
˙To reduce the risk of fire or electric shock do not expose this
product to rain or moisture.
Page 3
2
Risk of electric shock
See instruction manual
DC measurement
Equipment protected by double or reinforced insulation
=
Battery

Fuse
Earth
AC measurement
Conforms to EU directives
Do not discard this product or throw away.
CAUTION
˙ Disconnect the test leads from the test points before chang­ ing the position of the function rotary switch. ˙ Never connect a source of voltage with the function rotary switch in ,50V,100V,250V,500V,1000V position. ˙ Do not expose Meter to extremes in temperature or high humidity. ˙ Never set the meter in Ω,50V,100V,250V,500V,1000V function to measure the voltage of a power supply circuit
in equipment that could result in damage the meter and the equipment under test.
Symbols as marked on the Meter and Instruction manual
Unsafe Voltage
To alert you to the presence of a potentially hazardous voltage, when the Tester detects a voltage 2 V in resistance, 30 V in insulation test, or a voltage overload (OL), the ”“ symbol is
displayed and High voltage indicator is turned on.
Maintenance
Do not attempt to repair this Meter. It contains no userservice­able parts. Repair or servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.
Page 4
3
The Meter Description
Front Panel Illustration
1. LCD display 4000 counts .
2. Push-buttons for features.
3. Rotary switch to turn the Power On or Off and to select a function.
4. Input Terminal for function.
5. Input Terminal for V Insulation functions.
6. Common (Ground reference) Input Terminal for all functions.
Making Basic Measurements
Preparation and Caution Before Measurement
: Observe the rules of Warnings and Cautions
When connecting the test leads to the DUT (Device Under Test) connect the common (COM) test lead before
connecting the live lead ; when removing the test leads removing the test live lead before removing the common test lead.
The figures on the following pages show how to make basic measurements.
Page 5
4
Measuring ACV/DCVAuto sense function
Auto sense mode: The meter displays ACV or DCV whichever is higher (>0.3V). If the measured voltage is above 660Vac/dc, "> 660Vac/dc" will appear on the display.
Measuring ACV/DCV : Auto sense +LPF function
Page 6
5
Measuring Earth-Bond Resistance
1. Before starting the test : (a) The circuit under test must be completely de-energized.
Check the fuse before testing. Connecting to an energized circuit while the test is active will blow the fuse. See Testing the fuse later in this manual. If the detected voltage is above 2V, ">2V" will appear on the display.
In this condition, the test is inhibited. (b) Short the test leads before measurement, and press the blue button to compensate the probe resistance for <2Ω. If the probe resistance is stored, “->0<-” symbol will be displayed.
2. Lock mode : Press the Lock button to enter the Lock Mode. Then press the TEST button to start the test. The test volt -
age will continue to be applied until the TEST/LOCK button is pressed again.
3. The meter displays the "" symbol and the maximum resis­ tance for the range when measured resistance is higher than
the maximum display range.
Page 7
6
Measuring Insulation Resistance
1. Before starting the test :
The circuit under test must be completely de-energized. If the voltage detected is above 30V, ">30V" will appear on the display. In this condition, the test is inhibited.
2. Press the blue button to display insulation resistance or Leakage current during the test or when the test stops.
3. Lock mode: Press the Lock button to enter the Lock Mode.
Then press TEST button to start the test. The test voltage will continue to be applied until the TEST/LOCK button is pressed again.
4. Release the TEST button before removing the test leads (to enable the tester to discharge capacitive circuits). If the screen displays volts, wait until it reaches zero.
5. The meter displays the "" symbol and the maximum
resistance for the range when measured resistance is
=
Page 8
7
Using the Compare function
1. Before starting the insulation test :
Press the COMP button to select the compare value among 100k,200k,500k,1M,2M,5M,10M,20M, 50M,100M,200M and 500M.
2. If the measured value is greater than the selected compare value, the green Pass indicator will be turned on.
Measuring PI/DAR
PI(Polarization Index)=R10-min/R1-min DAR(Dielectric Absorption Rations)=R1-min/R30-sec where
R10-min : the insulation resistance measured at the 10 minute after pressing the TEST button. R1-min : the insulation resistance measured at the 1 minute after pressing the TEST button. R30-sec : the insulation resistance measured at the 30 second after pressing the TEST button.
Page 9
8
The measured values after the PI test
PI=R10-min/R1-min
The measured values after the DAR test
DAR=R1-min/R30-sec
If the reading for DAR is bigger than 1.3 or PI is bigger than 2, it indicate a good insulation quality. When the measured resistance is higher than the maximum range, the screen will display "Err" symbol for PI/DAR value.
TEST button : Press once to start/interrupt the PI /DAR test.
Blue button : Press once during the PI/DAR test to display the
time left of the test.
Page 10
9
Using the Store function
Store/Recall button :
The screen will be flashing MEM symbol and the number of stored data twice when the Store/Recall button is pressed.
1. Press once to store the voltage value during voltage test
2. Press once to store the PI/ DAR, insulation or Earth-bond resistance value when the individual test is completed.
3. The memory is divided into five segmentseach segments has 100 maximum data. The method to store the data is first
in /first out when the memory is full.
Table: The stored values of the individual test
segments
Items
Voltage
Earth-Bond
resistance
Insulation resistance 50~1000V
DAR PI
1 Voltage Resistance Resistance
DAR
value
PI value
2
Leakage
current
R30-sec R1-min
3
Test voltage R1-min R10-min
Search the stored value under RECALL mode
NOTE :
Press the Store/Recall button 5 sec to clear the stored data. The screen will be flashing MEM symbol and “ cLr ” on
reading twice.
Using the Recall function
Press the Store/Recall button 2 sec to enter/exit the RECALL mode.
If the memory is empty, the meter will display the “nOnE ” symbol.
Page 11
10
Read the stored value of insulation test under RECALL mode
Read the stored value of DAR test under RECALL mode
Press the PI/DAR button 2 sec to choose the DAR
function, and Press the Store/Recall button 2 sec to enter the RECALL mode.
Page 12
11
Read the stored value of PI test under RECALL mode
Press the PI/DAR button 2 sec to choose the PI function,
and Press the Store/Recall button 2 sec to enter the RECALL mode.
Auto Power Off (Battery Saver)
Restore power by switching rotor or pressing any button.
Auto Backlight
The backlight is automatically turned on at dark environment.
BUZZER
The Meter beeps once for every valid key-press, and beeps twice for every invalid key-press.
Page 13
12
Testing the fuse
Battery and Fuse Replacement
Refer to the following figure to replace fuse and the batteries :
2 Caution
˙ Use only a fuse with the amperage, interrupt, volt­ age, and speed rating specified. ˙Fuse rating : Fast, 315mA, 1000V, Min Interrupt Rating 10000A
˙ Replace the batteries as soon as the low batteries indicator "=" appears, to avoid false reading.
˙ When the battery is too low for reliable operation , the meter displays “ bAtt”. The meter will not operate
at all until the battery is replaced.
˙1.5V x 4 alkaline batteries.
Power-up options :
Press button while turning meter on from OFF position.
Lock button : display of the software version. Store button : Reset all stored data
˙Remove the test leads before testing the fuse. ˙If the display reading is FUSE, the fuse is bad and should be
replaced.
Page 14
13
Specifications
General Specifications
Maximum voltage applied to any terminal : 600 V ac rms or dc. Display : 4000 counts. Polarity Indication :
Automatic, positive implied, negative indicated.
Overrange Indication : OL Batteries Life :
Resistance Measurements : Tester can perform at least 2600 earth-bond resistance measurements with new alkaline batter­ies at room temperature. These are standard tests of 1Ω with a duty cycle of 5 seconds on and 25 seconds off. Insulation test: Tester can perform at least 1100 insulation tests with new alkaline batteries at room temperature. These are standard tests of 1 MΩ at 1000 V with a duty cycle of 5 seconds on and 25 seconds off. Low Batteries Indication : "=" is displayed when the batter- ies voltage drops below operating voltage.
Low battery voltage : Approx. 4.5V Auto Power Off : Approx 20 minutes.
Operating Ambient : Non-condensing 10°C,
11°C ~ 30°C (80% RH),
30°C ~ 40°C (75% RH),
40°C ~ 50°C (45%RH)
Storage Temperature :
-20°C to 60°C , 0 to 80% R.H. (batteries not fitted)
Temperature Coefficient :
0.15 x (Spec.Accy)/°C, < 18°C or > 28°C .
Measure : Samples 2 times per second nominal. Altitude : 6561.7 ft (2000m) Safety : Complies with EN61010-1, UL61010-1,IEC 61010-1, V/ : CAT.IV. 600V. CAT Application field
Compliance to EN 61557 : IEC61557-1, IEC61557-2,
IEC61557-4, IEC61557-10
Weight : (630g) including battery.
The circuits not connected to mains.
The circuits directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
Page 15
14
Electrical Specifications
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C < 80%RH.
Voltage Measurement
Function Range Accuracy
DCV 600.0V
±(1%+5dgt)
ACV 600.0V
±(1.5%+5dgt)(50~60Hz)
±(2%+5dgt)(61~500Hz)
LPF ACV 600.0V
±(1.5%+5dgt)(50~60Hz)
±(5%+5dgt)(61~400Hz)
Start measuring voltage : AC 0.6V.
Over voltage protection : 600V rms or dc. The cut-off frequency of the low pass filter : 1 kHz. Input Impedance : 3M// less than 100pF. CMRR / NMRR : (Common Mode Rejection Ratio) (Normal Mode Rejection Ratio)
V
AC
: CMRR > 60dB at DC, 50Hz / 60Hz
V
DC
: CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type :
AC conversions are ac-coupled, true rms responding, calibrated to the sine wave input. For non-sine wave add the following Crest Factor corrections: For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy. For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy. For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy. CF 3 @ 330V 2 @ 500V
Dimensions (W x H x D) :
95mm x207mm x 52mm with holster.
Accessories : Battery (installed), Test leads and user manual. Power Requirements : 1.5V x 4 IEC LR6 or AA size. Pollution degree : 2 EMC : EN 61326-1 Shock vibration : Sinusoidal vibration per MIL-T- 28800E
(5 ~ 55 Hz, 3g maximum).
Drop Protection :
4 feet drop to hardwood on concrete floor.
Indoor Use.
Page 16
15
Insulation resistance Measurement
Function Range Accuracy *
4.000M
40.00M
±(1.5%+5dgt)
400.0M 4000M
±(3%+5dgt)
4.1G~20.0G
±(10%+3dgt)
Insulation
Resistance
* Above specifications only apply when high quality silicone leads with test clips are being used with no hands touch.
Test Voltage vs. Maximum resistance range :
50V/50.0M,100V/100.0M,250V/250.0M,500V/500M and 1000V/20.0G.
Test Voltage vs. Minimum resistance (with test current=1mA) :
50V/50k,100V/100k,250V/250k,500V/500k and 1000V/1M. Test Voltage Accuracy : -0%,+20% Short Circuit Test Current:1mA(nominal)
Auto discharge function : discharge time<1 sec for C≦1
µF
Maximum Capacitive load : Operable with up to 1
µF load
Live Circuit Detection : if30V ac/dc at inputs, test inhibited
Function Range Accuracy
40.00
±(1.5%+5dgt)*
400.0 4000
40.00k
±(1.5%+3dgt)
Earth-bond Resistance
*<1.00 add 3dgt Open Circuit Test Voltage : >4.0V,<8V Short Circuit Current : >200.0mA
Live Circuit Detection: if2V ac/dc at inputs, test inhibited.
Earth-bond resistance Measurement
Page 17
16
Limited Warranty
This meter is warranted to the original purchaser against defects in material and workmanship for 1 year from the date of purchase. During this warranty period, Manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling. Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.