Apera Intruments PC60 User Manual

PC60 Premium Multi-Parameter Tester
(pH/EC/TDS/Salinity/Temp.)
Instruction Manual
APERA INSTRUMENTS, LLC
V6.2
!
2!
Thank!you!for!purchasing!Apera!Instruments!PC60!Premium!Multi-Parameter!Tester.!Please! carefully!read!this!instruction!manual!before!using!the!product!in!order!to!have!an!accurate!and! reliable!test!result,!and!avoid!unnecessary!damages!to!the!meter!or!probe.! ! For!video!tutorials,!please!go!to!support.aperainst.com!
Contents!
1.!Battery!Installation!.................................................................................................................................!3!
2.!Keypad!Functions!....................................................................................................................................!3!
3.!Complete!Kit!...........................................................................................................................................!4!
4.!Preparation!Before!Use!..........................................................................................................................!4!
5.!pH!Calibration!.........................................................................................................................................!5!
6.!pH!Measurement!....................................................................................................................................!6!
7.!Conductivity!Calibration!.........................................................................................................................!7!
8.!Conductivity!Measurement!....................................................................................................................!7!
9.!Parameter!Setting!...................................................................................................................................!9!
10.!Technical!Specifications!......................................................................................................................!10!
11.!Icons!and!Functions!............................................................................................................................!11!
12.!Probe!Replacement!............................................................................................................................!11!
13.!Warranty!.............................................................................................................................................!12!
!
3!
1. Battery Installation
Please install batteries according to the following steps. *Please note direction of batteries: All
POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the
tester and potential hazards)
2. Keypad Functions
Short press------ < 2 seconds ,Long press------- > 2 seconds
1.Short press to turn on the tester and long press to turn off the tester.
2.When turned off, long press to enter parameter setting.
3. In measurement mode, short press to turn on backlight.
1.In measurement mode, short press to switch parameter pHCONDTDSSAL
2.In mode setting, short press to change parameter (Unidirectional)
1. Long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
3.When measured value is locked, short press to
unlock;
Pull!the!battery!cap!up!
Slide!the!battery!cap!along!to!the!direction!of!arrow! Open!the!battery!cap! Insert!the!batteries!(ALL$POSITIVE$SIDES$FACING$UP)!(see!graph)! Close!the!battery!cap! Slide!and!lock!the!battery!cap!along!to!the!direction!of!arrow! Fit!the!testers!cap!while!making!sure!to!push!all!the!way!down.!The!!testers!
waterproof!design!may!be!compromised!if!the!cap!is!not!fitted!correctly.!
Battery!
Buttons
Probe!
Probe
Probe!Cap
LCD!
pH!Sensor
BPB!Sensor
!
+
+ + +
- - -
-
!
4!
3. Complete Kit
4. Preparation Before Use
If it is first-time use or the tester hasn’t been used for a long time, pour some 3M KCL solution in
the probe cap (about 1/5 of the probe cap) and soak the probe for 15-30 minutes.
When not in use, we recommend store the pH probe in the storage 3M KCL solution in the probe
cap to keep the sensor’s accuracy. But even if stored dry, it won’t do any permanent damage to
the sensor. It will only temporarily cause the probe to lose its sensitivity, which can always be
restored by soaking in the storage solution.
The!storage!solution!is!3M!KCL!(potassium!chloride).!One!bottle!of!10mL! storage! solution!com es! with! the!
tester!kit.!If!the!soaking!solution!was!contaminated,!please!replace!with!new!ones!timely.!
*!DO!NOT!use!any!other!brand’s!storage!solutions!because!different!chemicals!may!be!used!and!
potential!permanent!damage!could!be!caused!to!the!meter.!
Diagram!-!2!