ANKOM XT15 User Manual

Page 1
2052 O’Neil Rd, Macedon NY 14502
Telephone: (315) 986-8090 Fax: (315) 986-8091 www.ankom.com
Operator’s Manual
Rev F 12/08/14
Page 2
This page intentionally left blank
Page 3
Table of Contents
Introduction .................................................................................................................................................................. 5
Warranty ....................................................................................................................................................................... 5
Filter Bags ..................................................................................................................................................................... 5
Operating Environment ................................................................................................................................................ 5
Contact Information ..................................................................................................................................................... 6
Instrument Description ................................................................................................................................................. 7
Safety Precautions ........................................................................................................................................................ 8
Instrument Installation ................................................................................................................................................. 9
Fat Extraction Calculations ......................................................................................................................................... 10
Fat Extraction Support Items ...................................................................................................................................... 10
Sample Preparation .................................................................................................................................................... 11
Fat Extraction step-by-step procedure using the ANKOM
Periodic Maintenance ................................................................................................................................................. 20
Troubleshooting & Replacement Parts ....................................................................................................................... 22
Diagnostics .................................................................................................................................................................. 23
Appendix A – AOCS Official Fat Extraction Procedure ................................................................................................ 27
Appendix B – Electrical Diagram (pg. 1 of 3) ............................................................................................................... 29
XT15
.................................................................................... 16
Page 4
This page intentionally left blank
Page 5
Operator’s Manual
Please review the entire contents of this manual before you begin operating this product.
Introduction
ANKOM Technology designs, manufactures, and markets instruments and support products used by analytical laboratories around the world in the environmental, agricultural, biomass, and food industries. ANKOM Technology can provide you with products for determining or monitoring dietary fiber, detergent fiber, fat, digestibility, microbial fermentation (anaerobic or aerobic) and more.
Committed to Total Customer Satisfaction, ANKOM designs every product based on a thorough assessment of customer needs.
Congratulations on your purchase of the ANKOM effectively serve your needs.
The ANKOM technician variation, increasing precision. By carefully following the operating instructions in this manual you will understand the details of sample and filter bag handling as well as the instrument controls, helping you to achieve the best possible results.
XT15
Extraction System was designed to automate the process of fat extraction. This reduces
XT15
Extraction System. We are confident that this product will
Warranty
ANKOM Technology warrants the ANKOM or material for one year after the original date of purchase. This warranty does not include damage to the instrument resulting from neglect or misuse. During the warranty period, should any failure result from defects in workmanship or materials ANKOM Technology will, at its discretion, repair or replace the instrument free of charge.
Extended warranties are available upon request.
XT15
Extraction System against any defects due to faulty workmanship
Filter Bags
Use only ANKOM Technology filter bags (part # XT4) in your ANKOM filter mechanism will void the warranty. Filter bags can be purchased from ANKOM Technology or from your local authorized ANKOM distributor.
XT15
Extraction System. Use of any other
Operating Environment
Your ANKOM
Ambient Temperature Range: 15°−35°C Power: 100V120V ~ 50/60Hz 10A
XT15
Extraction System is designed to operate within the following environments:
220V240V ~ 50/60Hz 5A
Rev F 12/08/14 pg. 5
Page 6
Operator’s Manual
Contact Information
At ANKOM Technology we are committed to your total satisfaction and therefore always available to help you get the most from your ANKOM products. We are also very interested in any comments or suggestions you may have to help us improve.
For any questions or suggestions regarding your instrument, please contact us at:
Telephone: (315) 986-8090 Fax: (315) 986-8091 Email: service@ankom.com www.ankom.com
pg. 6 Rev F 12/08/14
Page 7
Operator’s Manual
Keypad
Display
Power Switch (not visible)
Safety Valve Handle
Pressure Gauge
Extraction Vessel
Solvent Fill Port
Main Handle
Solvent Sight Level
Reflux Water Flow Control
Instrument Description
General Description
The ANKOM petroleum ether. The compounds extracted are primarily triacylglycerols together with a small quantity of related lipids, traditionally termed “Crude Fat.” The analysis is achieved by measuring the loss of mass due to the extraction of fat/oil from the sample encapsulated in a filter bag. The quantitative isolation of the sample is accomplished by surrounding the sample in a sealed filter bag with a filtering capacity in the 2 to 3 micron range. The filter bag has sufficient porosity to permit rapid solvent passage and is composed of polymeric material that is resistant to the higher temperatures and solvents used in the instrument.
The quantitative isolation achieved by the filter bag permits the ANKOM process samples in batches of up to 15. With the solvent reservoir filled, simply place filter bags with sample in the Extraction Vessel, lock the vessel in place, select the extraction time and temperature, and press “Start.” The instrument will automatically fill the Extraction Vessel with solvent, extract the fat from the sample, and recycle the solvent.
Below is a detailed view of the ANKOM
XT15
Extraction System is designed to extract Crude (Free) Fat using conventional solvents, typically
XT15
Extraction System to automatically
XT15
Extraction System.
Rev F 12/08/14 pg. 7
Page 8
Operator’s Manual
Flammable Hazard: Static electricity is to be avoided as it is a spark hazard.
Hot Surfaces: Do not touch the extraction vessel during operation. The surface
can exceed 70°C (158°F). Failure to observe this caution may result in injury.
Hazardous Voltages: Do not operate the instrument with the electrical compartment open. Hazardous voltages are present during operation. Failure to
observe this caution may result in electrical shock or electrocution.
Hazardous Materials: The ANKOM
XT15
Extraction System comes equipped with a charcoal filter which traps any residual vapor in the vent system. By properly maintaining the charcoal filter, volatiles will not escape the vent system. We recommend replacement of the filter every 3 months or after 100 extraction runs. Trace amounts of vapor may also be present when samples are removed. It is recommended that the ANKOM
XT15
Extraction System be placed in a well ventilated environment, away from any heat or ignition sources. Organic solvents may be flammable. Follow safe laboratory practices according to both local and federal regulations when installing this instrument and when handling any organic solvents.
WARNING: Attempts to override safety features or to use this instrument in a manner not specified by ANKOM Technology voids the warranty and may result in serious injury or even death.
This system is designed to meet and/or exceed the applicable standards of CE, CSA, NRTL and OSHA.
Please review the entire contents of this manual before you begin operating this instrument.
Safety Precautions
pg. 8 Rev F 12/08/14
Page 9
Operator’s Manual
1.
Remove the instrument from the shipping container and place in an area that is free of dust and excessive moisture, and is well ventilated (away from any heat or ignition sources).
Because the ANKOM
XT15
Extraction System is a closed system, there is very limited exposure to solvents. When used in conjunction with a charcoal filter on the vent line the instrument produces less than 1ml of volatiles. It is the user’s responsibility to determine whether venting of this instrument requires placement in a hood. Please follow both local and federal regulations when installing this instrument.
2.
Connect the tube from the top fitting to the water supply and the one from the bottom fitting to the drain.
Ensure that there are no kinks or bends in any of the tubing connected to the instrument.
3.
Connect the vent flask to the Vent Flask Connector on the instrument and run the tubing to the inline charcoal filter. Periodically check the vent line to verify that it is free from any blockage or kinks. Generally, there should not be any significant solvent in the vent flask. Any solvent visible in the flask may indicate a problem with the water supply to the instrument.
Change the Charcoal Filter every 3 months or after 100 extraction runs.
4.
Plug the power cord into a grounded receptacle.
Charcoal Filter
Water Supply
Drain
Vent Flask Connector
Instrument Installation
Site Requirements
To install and operate the ANKOM
Adjustable wrench Chilled or tap water supply not to exceed 23°C Adequate power (see “Operating Environment” section) Drain
Instrument Installation Procedure
To install the ANKOM
XT15
Extraction System, follow the procedure detailed below.
XT15
Extraction System you will need the following:
Rev F 12/08/14 pg. 9
Page 10
Operator’s Manual
Crude Fat contained within a food or feed sample can be calculated using the following formula:
% Crude Fat
=
100 x (W2 – W3)
W1
Where:
W1 = Original weight of sample
W2 = Weight of pre-extraction dried sample and filter bag
W3 = Weight of dried sample and filter bag after extraction
For Liquid Samples, W2 and W3 in the equation above will include the weight of Diatomaceous Earth (DE). For Meat Samples with > 15% fat and plant samples with > 20% fat:
% Crude Fat
=
100 x ((W2 – W4) – W3)
W1
Where:
W4 = Weight of the weigh tin
If hydrolysis is done prior to fat extraction, then Total Fat contained within a food or feed sample can be calculated using the following formula:
% Total Fat
=
100 x (W2 – (W3 + (C1 – C2)))
W1
Where:
W1 = Original weight of sample
W2 = Weight of dried sample, filter bag, and Diatomaceous Earth (DE) after hydrolysis
W3 = Weight of dry extracted sample, filter bag, and DE
C1 = Blank filter bag dry weight after hydrolysis
C2 = Blank filter bag weight after extraction
The Hydrolysis procedure can be found in the ANKOM HCl Operator's Manual.
Item
Recommended Product
Electronic Balance with four-place readout
ANKOM #TB Balance Hardware ANKOM #TBS Balance Software
Filter Bags
ANKOM #XT4
Bag Holder (used for adding sample to an empty filter bag)
ANKOM #101.2
Heat Sealer for sealing the filter bags
ANKOM #1915 (120V), #1920 (220V)
Solvent Resistant Marker
ANKOM #F08
Desiccant Pouch or Weigh Tin Desiccator for drying
ANKOM #X45, #X49
Dryer (capable of temperatures from 60-120°C) - optional
ANKOM #RD (120V), #RDI (220V)
Oven for drying (capable of maintaining 102°C±2°)
Ventilated oven trays to allow for sufficient oven air flow
Weigh Tins (used with samples of > 20% fat)
Sample
Spoon
Fat Extraction Calculations
Crude Fat Calculation
Total Fat Calculation
Fat Extraction Support Items
The following support items are needed to perform the fat extraction procedure:
pg. 10 Rev F 12/08/14
Page 11
Operator’s Manual
To prepare meat samples with ≤ 15% fat by weight or plant samples with ≤ 20% fat by weight for fat extraction, follow the procedure detailed below.
1.
Using a Solvent Resistant Marker, number all of the empty filter bags you will use during the fat extraction procedure. Although it is not necessary in the determination of fat, it is recommended that you weigh and record the weights of all the filter bags prior to filling them in order to calculate the moisture value for the samples.
% Moisture = 100 x ((Filter Bag Weight + Sample Weight) - Weight after drying)
Sample Weight
2.
Place an empty filter bag in the Bag Holder in an open position.
3.
Tare the weight of the empty filter bag and the holder together.
4.
Add 1.5 – 2g of sample to the filter bag. Keep all particles away from the sealing area of the filter bag.
5.
Record the weight of the sample (W1) and tare.
6.
Set the Heat Sealer dial to 6. (The setting may vary from sealer to sealer.)
7.
Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of the filter bag (as shown to the right). If the seal is not strong, re-seal the bag.
8.
Repeat steps 1 – 7 for all filter bags that will be used in the ANKOM
XT15
Extraction System. (Up to 15
bags can be processed during one procedure.)
9.
Dry all samples at 102°C±2° for 3 hours to remove moisture prior to the extraction.
10.
Remove the samples from the oven and place them in a Desiccant Pouch.
11.
Allow the samples to cool to room temperature. This should take about 10 – 15 minutes.
12.
Re-weigh each bag (W2) immediately after removing from the Desiccant Pouch.
The samples are now ready for the fat extraction procedure.
Seal
Sample Preparation
Sample Preparation Procedure for Crude Fat – ffoorr LLOOWW ffaatt ssaammpplleess
Rev F 12/08/14 pg. 11
Page 12
Operator’s Manual
To prepare meat samples with > 15% fat by weight or plant samples with > 20% fat by weight for fat extraction, follow the procedure detailed below.
1.
Using a Solvent Resistant Marker, number all of the weigh tins and empty filter bags you will use during the fat extraction procedure. Use one weigh tin per filter bag. Although it is not necessary in the determination of fat, it is recommended that you weigh and record the weights of all the filter bags prior to filling them in order to calculate the moisture value for the samples.
% Moisture = 100 x ((Filter Bag Weight + Sample Weight) - (Weight after drying - Tin Weight))
Sample Weight
2.
Dry all weigh tins in an oven at 102°C±2° for 15 minutes.
3.
Remove weigh tins from the oven and place them in a Weigh Tin Desiccator or in a Desiccant Pouch.
4.
Allow weigh tins to cool to room temperature.
5.
Weigh and record tin weights (W4).
6.
Place an empty filter bag in the Bag Holder in an open position.
7.
Tare the weight of the empty filter bag and the holder together.
8.
Add 1.5 – 2g of sample to the filter bag. Keep all particles away from the sealing area of the filter bag.
9.
Record the weight of the sample (W1) and tare.
10.
Set the Heat Sealer dial to 6. (The setting may vary from sealer to sealer.)
Sample Preparation Procedure for Crude Fat – ffoorr HHIIGGHH ffaatt ssaammpplleess
pg. 12 Rev F 12/08/14
Page 13
Operator’s Manual
11.
Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of the filter bag (as shown to the right). If the seal is not strong, re-seal the bag.
12.
Repeat steps 1 – 11 for all filter bags that will be used in the ANKOM
XT15
Extraction System. (Up to 15
bags can be processed during one procedure.)
13.
Place the sealed filter bags in their corresponding weigh tins for pre-drying.
14.
Dry all samples at 102°C±2° for 3 hours to remove moisture prior to the extraction.
15.
Remove the weigh tins and filter bags from the oven and place them together in a Weigh Tin Desiccator.
A Weigh Tin Desiccator is recommended because it will hold between 10 and 14 weigh tins with filter bags, depending on the size of the weigh tins. A Desiccant Pouch will hold about 4.
16.
Allow the samples to cool to room temperature. This should take about 10 – 15 minutes.
17.
Re-weigh and record the weights of each weigh tin (W4) and filter bag (W2) immediately after removing from the Weigh Tin Desiccator.
The samples are now ready for the fat extraction procedure.
Seal
Rev F 12/08/14 pg. 13
Page 14
Operator’s Manual
To prepare liquid samples for fat extraction, follow the procedure detailed below.
1.
Using a Solvent Resistant Marker, number all of the empty filter bags you will use during the fat extraction procedure. Although it is not necessary in the determination of fat, it is recommended that you weigh and record the weights of all the filter bags prior to filling them in order to calculate the moisture value for the samples.
% Moisture = 100 x ((Filter Bag Weight + Sample Weight) - Weight after drying)
Sample Weight
2.
Place an empty filter bag in the Bag Holder in an open position.
3.
Tare the weight of the empty filter bag and the holder together.
4.
Add Diatomaceous Earth (DE).
4.1 Add 1.0 – 1.25 g of DE to the filter bag.
4.2 Using a pencil eraser (or similar sized utensil) make an indentation or a
space in the DE into which the liquid sample can be added using a pipette.
4.3 Record the weight and tare.
5.
Pipette 1.5 – 2 g of liquid sample into the DE indentation in the filter bag. (Do NOT use more liquid sample than can be absorbed by the DE.)
6.
Record the weight of the sample (W1).
7.
Set the Heat Sealer dial to 6. (The setting may vary from sealer to sealer.)
8.
Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of the filter bag (as shown to the right). If the seal is not strong, re-seal the bag.
9.
Repeat steps 1 – 8 for all filter bags that will be used in the ANKOM
XT15
Extraction System. (Up to 15
bags can be processed during one procedure.)
Seal
Sample Preparation Procedure for Crude Fat – ffoorr LLiiqquuiidd ssaammpplleess
pg. 14 Rev F 12/08/14
Page 15
Operator’s Manual
10.
Dry all samples at 102°C±2° for 3 hours to remove moisture prior to the extraction.
11.
Remove the samples from the oven and place them in a Desiccant Pouch.
12.
Allow the samples to cool to room temperature. This should take about 10 – 15 minutes.
13.
Re-weigh each bag (W2) immediately after removing from the Desiccant Pouch. The samples are now ready for the fat extraction procedure.
Rev F 12/08/14 pg. 15
Page 16
Operator’s Manual
1.
Turn the instrument power ON.
The Display will briefly show the software version number followed by the screen to the right.
The factory default solvent shown on the Display is "Pet Ether." However, the instrument computer will store in memory the last solvent selection made and use it as the default solvent shown on the Display screen every time the instrument is ready for a new run.
2.
Select the solvent you want to use.
2.1 Press the Down Arrow (v) key on the Keypad until the desired solvent is shown on the Display.
2.2 Fill the solvent reservoir as needed.
2.21 Unscrew the Fill Cap.
2.22 Pour solvent into the Fill Port up to the dark blue level mark
on the Solvent Sight Level. (About 500 ml of solvent is needed to reach the dark blue level mark.) Do not add solvent up to the top of the Solvent Sight Level tube or it will start flowing into the Vent Flask.
If using Diethyl Ether or a solvent mixture that includes Diethyl Ether (such as the Mojonnier Mixture), you will need the O-ring Kit for High Penetration Solvents (part # X95) to replace the standard O-rings on the top of the Extraction Vessel and the Valve Kit for High Penetration Solvents (part # X55).
Warning: Do NOT use Acetone in this instrument because it will cause damage
to the valves.
2.3 Press ENTER on the Keypad to complete the solvent selection.
The Display should now show the screen to the right.
Remove/Insert Samples <ENTER>
PET Ether v <ENTER>
Fat Extraction step-by-step procedure using the ANKOM
To perform fat extraction on prepared samples, follow the procedure detailed below.
XT15
pg. 16 Rev F 12/08/14
Page 17
Operator’s Manual
3.
Insert Filter Bags into the instrument.
3.1 Press down the Safety Valve Handle to release the Safety Valve
Pin and rotate the Main Handle upward until it stops.
3.2 Use the Extraction Vessel Handle to pull the Extraction Vessel
out.
3.3 Slide the Extraction Vessel toward you far enough to remove the
white Teflon Insert or to put the Bag Holder directly into the Teflon Insert while it remains in the Extraction Vessel.
It is not necessary to remove the white Teflon Insert from the Extraction Vessel in order to put the Bag Holder in place.
3.4 Put the Bag Holder (with up to 15 bags) into the Teflon Insert.
3.5 If you removed the Teflon Insert, put it back
into the Extraction Vessel, aligning the Teflon Insert slot with the Extraction Vessel pin.
The Teflon Insert O-ring and the Extraction Vessel O-ring must be in
place before running the instrument.
The Teflon Insert slot must be aligned with the Extraction Vessel pin
before running the instrument.
Teflon Insert O-ring
Extraction Vessel O-ring
Teflon Insert slot and Extraction Vessel pin
Teflon Insert
Bag Holder
Rev F 12/08/14 pg. 17
Page 18
Operator’s Manual
3.6 Slide the Extraction Vessel into the instrument ensuring the
Extraction Vessel O-ring stays in place.
3.7 Pull the Main Handle downward, closing and locking the Extraction
Vessel in place. You will hear a click noise when the Safety Valve Pin locks in place.
Before running the instrument, verify that the Extraction Vessel is locked in place by attempting to rotate the Main Handle upward. If it is not locked, repeat step #8 above.
3.8 Press ENTER on the Keypad.
The Display should now show the screen to the right. (The time shown will be the last time entered.)
4.
Set the Extraction Time.
4.1 Using the arrow keys on the Keypad, select the desired Extraction time. (The recommended time is
60 minutes.)
4.2 Press ENTER on the Keypad to set the Time.
The Display should now show the screen to the right. (The temperature shown will be the last temperature entered.)
5.
Set the Extraction Temperature.
5.1 Using the arrow keys on the Keypad, select the desired Extraction Temperature. (The
recommended maximum temperature is 90°C.)
5.2 Press ENTER on the Keypad to set the Temperature.
The Display should now show the screen to the right.
6.
Verify that the water supply is on.
7.
Run the instrument.
Press START on the Keypad. The instrument will run through its automatic operation, showing
its status on the Display. When the instrument has completed its automatic operation, the Display will show the screen to the right.
At this time, the only light shining on the Display should be the green power light.
PET Ether v <ENTER>
Close Vessel Water On <START>
Select temp 90 (^v)... <ENTER>
Select time 60 (^v)... <ENTER>
pg. 18 Rev F 12/08/14
Page 19
Operator’s Manual
8.
Determine the amount of fat that was extracted.
8.1 When the Pressure Gauge reads zero, remove the Extraction Vessel by pushing down on the Safety
Valve Handle to release the locking pin and then rotating the Main Handle to the open position.
8.2 Use the Extraction Vessel Handle to remove the Extraction Vessel.
Warning: The Extraction Vessel interior and top will be HOT when the
extraction process is complete.
8.3 Remove the Teflon Insert from the Extraction Vessel.
8.4 Remove the Bag Holder from the Teflon Insert.
8.5 Remove the filter bags from the Bag Holder.
8.6 Dry the samples for 15 – 30 minutes in the oven at 102 ± 2°C or in the
ANKOMRD Dryer set at the factory setting of 110°C.
8.7 Remove the samples from the oven or dryer and place them in a Desiccant
Pouch.
8.8 Allow the samples to cool to room temperature. This should take about 10 –
15 minutes.
8.9 Re-weigh each filter bag (W3) immediately after removing from the Desiccant Pouch.
9.
Clean the fat from the bottom of the vessel using paper towel.
If too much fat accumulates in the bottom of the Extraction Vessel, an overtemp fault will occur.
Rev F 12/08/14 pg. 19
Page 20
Operator’s Manual
Do not use soap and/or water to clean the vessel. When cleaning the vessel, do not damage the level sensor tip. The tip may be cleaned with window cleaner or alcohol.
Do not use Acetone to clean the instrument.
1.
Remove the rear cabinet of the instrument.
Rear View of instrument with cabinet back removed
2.
Unscrew the compression fitting.
3.
Unscrew the filter body.
4.
Unscrew and clean the metal filter screen.
5.
Reassemble the filter and compression fitting. Make sure the valve body is tight and the Teflon washer has not come out of the filter body.
Compression Fitting
Filter body
Periodic Maintenance
Cleaning the Extraction Vessel
Some compounds extracted during the fat extraction procedure can create a black residue in the bottom of the Extraction Vessel. The vessel can most easily be cleaned by using a drill with a long shafted stiff wire brush (ANKOM part # X107). This treatment will remove a lot of the black coating that accumulates on the bottom of the vessel. When the vessel bottom is clean, put some solvent on a cloth or paper towel and wipe down the inside of the vessel in an effort to remove any oil residue and particles that have been loosened from the vessel bottom.
Cleaning the exterior of the instrument
Wipe the outside of the instrument with window cleaner or a damp sponge as necessary.
Cleaning the solvent filter
If filter bags are not sealed properly and sample is lost from the bag, it will be recovered in the solvent filter in the exhaust line.
To clean the solvent filter, follow the procedure detailed below.
pg. 20 Rev F 12/08/14
Page 21
Operator’s Manual
1.
Unscrew the filter from the instrument.
2.
Unscrew the filter clamp from the water supply line.
3.
Use compressed air to blow backwards through the filter (blow opposite the arrow on the filter). If the filter is clogged, replace it with ANKOM part # X97.
4.
Reconnect the filter and fittings.
5.
Turn on the water and check for leaks.
1.
Turn off the instrument power and unplug the Power Cord from the outlet.
2.
Remove the two screws on the Top Cover.
3.
Swing the Top Cover down exposing the electrical compartment.
4.
Locate the Fuse Holders on the terminal strip. There are two fuses within the instrument; one for each pole (100 – 120V 10A or 220 – 240V 5A).
5.
Press the tabs to lift the Fuse Holders up.
6.
Replace the fuses and close the Fuse Holders completely.
7.
Close the Top Cover and secure it with the two screws that you previously removed.
Fuse Holders
Tab to open Fuse Holder
Fuse
Filter
Cleaning the water filter
To clean the water filter (located on the water supply line), follow the procedure detailed below.
Replacing the Fuses
To replace the fuses in the ANKOM
XT15
Extraction System, follow the procedure detailed below.
Replacing the Charcoal Filter
The Charcoal Filter should be replaced every 3 months, or after 100 extraction operations using the instrument. To replace the filter, remove the old one from the vent line and install the new one.
Rev F 12/08/14 pg. 21
Page 22
Operator’s Manual
Troubleshooting & Replacement Parts
The ANKOM technology web site has the most current troubleshooting and replacement parts information. Therefore, if you have any questions about the operation of your ANKOM
XT15
Extraction System, or if you need
replacement parts, please visit our web site at www.ankom.com.
pg. 22 Rev F 12/08/14
Page 23
Operator’s Manual
When the instrument is powered up or when the instrument finishes a run, the screen to the right is displayed.
The factory default solvent shown on the Display is "Pet Ether." However, the instrument computer will store in memory the last solvent selection made and use it as the default solvent shown on the Display screen every time the instrument is ready for a new run.
To access the Diagnostics Screen, press the Down Arrow button (v) on the Keypad four times. The screen to the right will be displayed. From the Diagnostics Screen you can access all of the available diagnostics.
1. With the Diagnostics screen displayed, press the ENTER button on
the Keypad. The screen to the right will be displayed.
If you press the STOP button on the Keypad while this screen is displayed, you will go back to the Diagnostics screen. If you press the Down Arrow button (v) on the Keypad while this screen is displayed you will go to the next group of diagnostics (Analog).
2. With the Outputs screen displayed, press the ENTER button on the
Keypad. The screen to the right will be displayed.
3. Press the Down Arrow button (v) on the Keypad to select the
heater relay (Q0) or valve (Q1 - Q7) you want to toggle. With your selection made (in this example we selected the heater relay Q0), press the ENTER button on the Keypad. The screen to the right will be displayed.
Heater (Q0) OFF ENTER to toggle
Heater (Q0) v <ENTER>
Outputs <ENTER> Exit <STOP> v
Diagnostics ^ <ENTER>
PET Ether v <ENTER>
Diagnostics
To help solve problems that may occur while running the ANKOM Diagnostics are available and accessible through the Display and Keypad.
1. Toggle Outputs (Q0 Heater Relay and Q1 - Q7 valves)
2. Check Temperature and Level Sensors
3. Check the state of the ETS (Emergency Temperature Shut-off switch) relay
XT15
Extractor, the following computer controlled
Access Diagnostics Screen
Toggle Outputs
To toggle outputs, follow the procedure below.
Rev F 12/08/14 pg. 23
Page 24
Operator’s Manual
4. Press the ENTER button on the Keypad to turn ON the heater relay
(Q0). The screen to the right will be displayed.
5. Press the ENTER button on the Keypad to turn OFF the heater
relay (Q0). The screen to the right will be displayed.
Warning: If the heater relay (Q0) is left ON, the Extraction Vessel interior will
become very HOT and within five minutes the instrument ETS switch will trip resulting in the Fault screen shown here.
From this screen press ENTER to get back to the main Diagnostics screen.
6. Press the STOP button on the Keypad to go back to the main
Outputs screen (shown to the right).
7. Press the STOP button on the Keypad to go back to the main
Diagnostics screen (shown to the right).
8. Press the Up Arrow button (^) on the Keypad to exit Diagnostics.
Press the Up Arrow button (^) on the Keypad up to three more times to return to your solvent screen (example shown to the right).
When you exit Diagnostics, heater relay Q0 turns OFF and valves Q1 - Q7 return to the closed position.
PET Ether v <ENTER>
Diagnostics ^ <ENTER>
Outputs <ENTER> Exit <STOP> v
Diagnostics ETS Fault <ENTER>
Heater (Q0) OFF ENTER to toggle
Heater (Q0) ON ENTER to toggle
pg. 24 Rev F 12/08/14
Page 25
Operator’s Manual
1. Access the Diagnostics screen (see the "Access Diagnostics
Screen" section for details).
2. With the Diagnostics screen displayed, press the ENTER button on
the Keypad. The screen to the right will be displayed.
3. Press the Down Arrow button (v) on the Keypad. The screen to
the right will be displayed.
4. Press the ENTER button on the Keypad. The screen to the right
will be displayed showing the current Temperature and Level Sensor readings.
When the instrument has not been run for more than two hours, the temperature should read the ambient temperature of the room. If the temperature reading does not show ambient temperature when the instrument is cool, contact ANKOM Technology.
When the Level Sensor in the instrument Vessel is not covered by liquid, the Level Sensor should read
10.00v. If the reading is less than 10.00v and the Level Sensor is not covered by liquid, clean the Level Sensor Tip inside the Vessel with a cotton swab and alcohol. If the Level Sensor reading is between
5.00 - 10.00v after cleaning the tip, the instrument will run, but you should consider replacing the tip. Contact ANKOM Technology for more information.
When the instrument Vessel is filled with liquid that covers the Level Sensor, the Level Sensor should read less than 3.00v. If the Level Sensor reading is more than 3.00v when the sensor is covered by liquid, contact ANKOM Technology.
If you fill the Vessel with water during Diagnostics, wipe the Vessel dry before running the instrument.
5. Press the STOP button on the Keypad to go back to the Analog
screen (shown to the right).
6. Press the STOP button on the Keypad to go back to the main
Diagnostics screen (shown to the right).
Diagnostics ^ <ENTER>
Analog <ENTER> Exit <STOP> ^ v
Temp Lvl Sense 20°C 10.00v
Analog <ENTER> Exit <STOP> ^ v
Outputs <ENTER> Exit <STOP> v
Diagnostics ^ <ENTER>
Check Temperature and Level Sensors
To check the Temperature and Level Sensors, follow the procedure below.
Rev F 12/08/14 pg. 25
Page 26
Operator’s Manual
7. Press the Up Arrow button (^) on the Keypad to exit Diagnostics.
Press the Up Arrow button (^) on the Keypad up to three more times to return to your solvent screen (example shown to the right).
1. Access the Diagnostics screen (see the "Access Diagnostics
Screen" section for details).
2. With the Diagnostics screen displayed, press the ENTER button on
the Keypad. The screen to the right will be displayed.
3. Press the Down Arrow button (v) on the Keypad. The screen to the
right will be displayed.
4. Press the Down Arrow button (v) on the Keypad. The screen to
the right will be displayed.
5. Press the ENTER button on the Keypad. The screen to the right
will be displayed showing the state of the ETS relay. If the ETS relay is "OFF", contact ANKOM Technology.
6. Press the STOP button on the Keypad to go back to the Inputs
screen (shown to the right).
7. Press the STOP button on the Keypad to go back to the main
Diagnostics screen (shown to the right).
8. Press the Up Arrow button (^) on the Keypad to exit Diagnostics.
Press the Up Arrow button (^) on the Keypad up to three more times to return to your solvent screen (example shown to the right).
PET Ether v <ENTER>
Diagnostics ^ <ENTER>
Inputs <ENTER> Exit <STOP> ^
ETS relay - ON
Inputs <ENTER> Exit <STOP> ^
Analog <ENTER> Exit <STOP> ^ v
Outputs <ENTER> Exit <STOP> v
Diagnostics ^ <ENTER>
PET Ether v <ENTER>
Check the state of the ETS (Emergency Temperature Shut-off switch) relay
To check the state of the ETS relay, follow the procedure below.
pg. 26 Rev F 12/08/14
Page 27
Operator’s Manual
ANKOM Technology Method
AOCS Official Procedure Am 5-04
Rapid Determination of Oil/Fat Utilizing High Temperature Solvent Extraction
Definition
This method determines crude fat by extracting with petroleum ether. The compounds extracted are predominantly triacylglycerides. Small amounts of other lipids along with minor compounds having some solubility in petroleum ether are also extracted.
Scope
This method is applicable to solid processed foods with 0-100% fat content.
Apparatus
1. Analytical Balancecapable of weighing 0.1 mg.
2. Ovencapable of maintaining a temperature of 102 ± 2°C.
3. Extraction instrumentcapable of extracting at 90 ± 2°C.
(XT10, XT15, XT20, ANKOM Technology).
4. Filter Bagsconstructed from chemically inert and heat
resistant filter media, capable of being heat sealed closed and able to retain 1 micron particles while permitting solution penetration (XT4, ANKOM Technology).
5. Heat sealersufficient for sealing the filter bags closed to
ensure complete closure (1915, ANKOM Technology).
6. Desiccant Pouchcollapsible sealable pouch with desiccant
inside that enables the removal of air from around the filter bags (MoistureStop weigh pouch, ANKOM Technology).
7. Marking penSolvent and acid resistant (F08, ANKOM
Technology).
8. Dryer (optional)forced air dryer capable of maintaining a
temperature between 60-120°C (ANKOMRD Dryer, ANKOM Technology).
Reagents
1. Petroleum Ether (B.P. 35-65º C). Depending on the method
used, other solvents such as hexanes and ethyl ether may be used. CAUTION: Most fat solvents are extremely flammable. Avoid static electricity and follow safe laboratory practices according to your local regulations when handling solvents.
Precision
Results from the collaborative study (see Table 1) indicate the precision (Sr, RSDr, r) that the analyst should use as a benchmark for evaluation replication in the same laboratory.
Procedure (see the Fat Extraction Process section of the Operator’s
Manual for more detail)
1. Place a labeled filter bag on the balance and zero the balance.
2. Weigh 1.5 – 2g of sample into the bag and record the weight
(W1).
3. Heat seal the filter bag closed within 4mm of the top to
encapsulate the sample.
4. It is good practice to include a blank bag periodically,
especially when changing solvents, in order to measure any effect on the filter bag.
5. If a meat sample has an expected fat value of > 15% or a plant
sample has an expected fat value of > 20%, then place the filter bag with sample in a tared weigh tin (W4) during the oven drying procedure in order to capture fat that may migrate out of the bag during heating.
6. Place samples in the drying oven at 102 ± 2°C for 3 hrs (see
Numbered Note 1).
7. Cool the dried samples in the Desiccant Pouch, weigh and
record the weight (W2).
8. Place filter bags with sample into the bag holder and place in
the extractor.
9. Follow the extraction instrument instructions.
10. When the extraction process is complete dry the samples for
15 – 30 minutes in the oven at 102 ± 2°C or in the ANKOMRD Dryer set at the factory setting of 110°C. (See Numbered Note
2.)
11. Cool in the Desiccant Pouch and weigh (W3).
Calculations
Meat samples ≤ 15% fat and plant samples ≤ 20% fat
Meat samples > 15% fat and plant samples > 20% fat
% Crude Fat
=
100 x (W2 – W3)
% Crude Fat
=
100 x ((W2 – W4) - W3)
W1
W1
Where:
W1 = Original weight of sample
Where:
W4
=
Weight of the weigh tin
W2 = Weight of pre-extracted dried sample with filter
bag
W3 = Weight of dry sample with filter bag after
extraction
NOTE: For Liquid Samples, W2 and W3 in the above equations will include the weight of Diatomaceous Earth (DE).
Appendix A – AOCS Official Fat Extraction Procedure
Rev F 12/08/14 pg. 27
Page 28
Operator’s Manual
Numbered Notes
1. Pre-extraction drying time of the samples in the filter bags can be greatly reduced by using a microwave oven. However,
microwave ovens vary in energy output and samples vary in water and fat composition. You will need to experiment with your microwave oven to determine the proper conditions for use. Contact ANKOM Technology for further information.
2. Some samples with large amounts of carbohydrates (starches and cellulose) may bind small quantities of solvent and require
additional post drying. To remove bound solvent, dry at 100°C for 1 – 3 hrs or at 125°C for 30 minutes.
Sample ID
oat meal
brownie mix
soybean A
canola
soybean meal
corn A
poultry starter
cattle feed
pig starter
alfalfa
cat food
cookies
bkft cereal
tortilla chips
Number of laboratories 12 12 11 9 12 12 11 10 11 11 12 11 12 12 Number of replicates 24 24 22 18 24 24 22 20 22 22 24 22 24 24
Filter Bag Average, % Crude Fat 5.8 8.7 20.9 39.0 1.6 3.3 3.3 3.2 5.6 2.4 6.3 22.7 2.3 19.9
Certified Labs Average
a
5.7 8.7 21.1 39.7 1.6 3.6 3.5 3.0 5.5 2.2 6.2 23.1 2.3 20.0
AOCS Am 2-93 Avg. for Oilseeds 21.9 44.8 3.6
S
r
0.36 0.20 0.35 0.23 0.14 0.31 0.24 0.18 0.20 0.39 0.27 0.20 0.26 0.39
S
R
0.54 0.31 0.63 0.68 0.27 0.42 0.42 0.20 0.28 0.50 0.30 0.20 0.36 0.48
RSDr, %
6.2 2.3 1.7 0.6 8.5 9.5 7.3 5.6 3.6 16.1 4.2 0.9 11.4 2.0
RSDR, %
9.4 3.5 3.0 1.7 16.3 12.7 12.6 6.1 5.0 20.7 4.7 0.9 15.7 2.4
r 0.99 0.56 0.98 0.65 0.39 0.88 0.68 0.51 0.56 1.08 0.75 0.56 0.72 1.09 R 1.52 0.86 1.76 1.90 0.75 1.18 1.16 0.55 0.78 1.39 0.83 0.56 1.00 1.35
HORRAT 3.05 1.21 1.19 0.76 4.38 3.87 3.77 1.83 1.61 5.92 1.56 0.35 4.45 0.95
Sample ID
dog food
crackers
turkey
ham
beef ground
chicken breast
soybean B
safflower
potato chips
hot dog
sausage
corn B
cheese curls
corn silage
Number of laboratories 12 10 12 9 11 11 11 9 11 12 11 12 12 12 Number of replicates 24 20 24 18 22 22 22 18 22 24 22 24 24 24
Filter Bag Average, % Crude Fat 6.8 23.8 3.2 11.6 23.8 2.8 19.4 22.5 32.0 39.5 25.7 3.4 30.6 2.3
Certified Labs Average
a
6.9 24.0 3.2 11.3 23.5 2.7 19.7 23.2 32.0 39.0 25.0 3.7 30.5 2.3
AOCS Am 2-93 Avg. for Oilseeds 20.8 24.7
S
r
0.35 0.23 0.21 0.30 0.24 0.33 0.38 0.53 0.48 0.35 0.34 0.39 0.48 0.23
S
R
0.35 0.23 0.34 0.30 0.36 0.33 0.62 0.83 0.52 0.59 0.51 0.41 0.69 0.51
RSDr, %
5.23 0.96 6.57 2.59 1.01 11.89 1.97 2.36 1.49 0.89 1.33 11.48 1.59 9.87
RSDR, %
5.23 0.96 10.84 2.59 1.49 11.89 3.19 3.69 1.61 1.49 1.98 11.93 2.27 22.45
r 0.99 0.64 0.58 0.84 0.67 0.94 1.07 1.49 1.34 0.98 0.96 1.10 1.36 0.63 R 0.99 0.64 0.96 0.84 0.99 0.94 1.73 2.33 1.45 1.65 1.43 1.14 1.94 1.44
HORRAT 1.74 0.39 3.22 0.94 0.6 3.47 1.25 1.48 0.68 0.65 0.81 3.59 0.95 6.36 aAOCS Official Methods Ba 3-38, AOAC 920.39 or equivalent
Table 1. A summary of the statistical analysis of the international collaborative study of the Filter Bag Technique for the proposed Official AOCS Method, “Rapid Determination of Oil/Fat Utilizing High Temperature Solvent Extraction” is presented in this table. Included in the summary is a comparison of the Filter Bag Technique with the results of analysis by
AOCS certified laboratories using official AOCS or AOAC methods
pg. 28 Rev F 12/08/14
Page 29
Operator’s Manual
Appendix B – Electrical Diagram (pg. 1 of 3)
Rev F 12/08/14 pg. 29
Page 30
Operator’s Manual
Appendix B – Electrical Diagram (pg. 2 of 3)
pg. 30 Rev F 12/08/14
Page 31
Operator’s Manual
Appendix B – Electrical Diagram (pg. 3 of 3)
Rev F 12/08/14 pg. 31
Page 32
Automation saves time and money!
ANKOM Technology is an international company with products that include…
TDF Dietary Fiber Analyzer
Automates AOAC 991.43/AACC 32.07.01 IDF/SDF and TDF values
Faster, Technician-free FilteringComputer controlled operationReduced per assay costs
A2000 Fiber Analyzer
Crude Fiber (AOCS Ba 6a-05), ADF, NDF Automatically adds solutions and rinses
Batch process - up to 24 samples at one time
XT15 Fat Extractor
Official Method AOCS Am 5-04 Fully automatic Solvent recovery at 97% or greater
Batch process - up to 15 samples at one time
RF Gas Production System
High sensitivity pressure measurement Applications include: Ruminant Nutrition, Human Digestion, Yeast Activity,
Beer/Wine Fermentation, Biomass-to-Energy analysis (e.g., Ethanol), Biodegradability, Soil respiration, BOD, etc.
Wireless Computer control and data storage
Chemicals
A wide variety of chemicals used for many different lab operations
Pre-mixed solutions available
Please visit our web site at www.ankom.com for more information.
2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 986-8090 Fax: (315) 986-8091 www.ankom.com
Loading...