Ando AQ6317 Data Sheet

Page 1
Optical Spectrum Analyzer
AQ6317
High-resolution optical spectrum analyzer for evaluating D-WDM systems and components.
Page 2
High-resolution optical spectrum ana D-WDM systems and components.
General
The AQ6317 is an advanced optical spectrum analyzer for a wide range of applications, including light source evaluation, measurement of loss wavelength character­istics in optical devices, and waveform analysis of WDM (Wavelength Division Multiplexing) systems.
Especially at 1550 nm band, the unit achieves high wavelength accuracy and wavelength linearity, and can evaluate optical devices for WDM. Analysis functions make operation and expandability simple.
The AQ6317 contains the latest of Ando’s technology for optical spectrum analyzers. Areference equipment for the next generation.
Features
50 GHz WDM-Signals
High-resolution measurement in a wide 50 GHz spec­trum dynamic range.
Versatile analysis functions
Analysis functions for WDM and other optical devices (LD, LED, FBG, etc.)
High wavelength accuracy
Provides ±0.05 nm wavelength accuracy in the 1550 nm band, with ±0.01 nm wavelength linearity, making it especially useful for high-precision loss wavelength characteristic and other evaluation of WDM devices. The wavelength scale indicates both in air and in vacu­um.
Synchronous sweep
In conjunction with a AQ4320 Tunable Laser Source, much higher wavelength resolution/wide dynamic range can be achieved by high-speed synchronous sweep.
High wavelength resolution
Achieves wavelength resolution of 0.015 nm
High sensitivity over a wide band
Covers from 600 nm to 1750 nm and high sensitivity allow measurement of light at down to -90 dBm.
Low polarization dependancy
Measurements such as gain of optical amplifier can be proceeded accurately because polarization dependen­cy is suppressed as low as ±0.05 dB.
High-level accuracy
Accurate within a ±0.3 dB.
High power measurement: Max. +20 dBm (100 mW)
Even high-power output from an optical amplifier can be measured directly without an optical attenuator.
9.4-inch color LCD
Pulsed light can be measured
Three individual trace memories
Page 3
Applications
Measures transmission characteristics of 16-channel AWG and displays results on AQ6317s screen.
Measurement example of wavelength dependency of gain/NF of EDFA (Measurement result on AQ8423Bs screen)
Measurement example using EDFAanalysis function (Measurement result on AQ6317s screen)
AQ2141/AQ4224(155)
Optical coupler
EDFA
Example of measurement with optical spectrum analyzer
Example of measurement in combination with optical amplifier analyzer
AQ6317
AQ6317
AQ8423A/8423B
AQ2141/AQ4224(155)
EDFA
Optical coupler
Optical fiber amplifier (EDFA) evaluation
The ASE interpolation method is used to measure gain and NF, key parameters for optical fiber evaluation. In conjunction with the AQ8423A/8423B optical amplifi­er analyzer, the system can accurately measure gain and NF with the pulse method, which is optimum for evaluation of WDM optical fiber amplifiers.
Optical Multiplexer evaluation
In conjunction with the AQ4320 Tunable Laser Source, the AQ6317 can achieve high wavelength resolution/wide dynamic range with high-speed syn­chronous sweep function, and result insertion loss, passed central wavelength and linearity as evaluation parameter of optical MUX/DEMUX.
AQ4320 Tunable Laser Source
AQ6317 Optical Spectrum Analyzer
Optical Multiplexer
(OMUX)
AQ3540
Optical Channel Selector
GP-IB
Configuration example for measurement with synchronous sweep function
Page 4
19-7, Kamata 4-chome, Ota-ku, Tokyo, 144-0052 Japan Phone: +81(0)3 3733 1151 Fax: +81(0)3 3739 7310
HEADQUARTERS: 2021 N. Capitol Avenue, San Jose, CA 95132, U.S.A. Phone: +1 408 941 0100 Fax: +1 408 941 0103 EAST OFFICE: 7617 Standish Place, Rockville, MD 20855, U.S.A. Phone: +1 301 294 3365 Fax: +1 301 294 3359
HEADQUARTERS: “Vijverdam”, Dalsteindreef 57, 1112XC Diemen, The Netherlands Phone: +31(0)20 698 1441 Fax: +31(0)20 699 8938 NIEDERLASSUNG DEUTSCHLAND: Nymphenburger Straße 119 B, D-80636 München, Germany Phone: +49(0)89 143 8150 Fax: +49(0)89 143 81555
19 Kim Keat Road #05-03, Jumbo Industrial Building, Singapore 328804 Phone: +65 251 1391 Fax: +65 251 1987
7F-1, No. 346 Pei-Ta Road, Hsin Chu, Taiwan Phone: +886 35 28 4168 Fax: +886 35 28 4110
Please visit our website for more information:
www.ando.com
Specifications are subject to change without notice.
KT-1806-2 9910 KL230
Printed in Japan
Specifications
Applicable fibers SM, GI (50/125 µm) Measurement wavelength range1)600 to 1750 nm Wavelength accuracy
1, 3)
±0.05 nm (1510 to 1580 nm, after calibration with built-in reference light source) ±0.5 nm (600 to 1750 nm)
Wavelength linearity
1, 3)
±0.01 nm (1510 to 1580 nm, resolution setting: 0.01 nm)
Wavelength repeatability
1, 3)
±0.005 nm (1 min.)
Wavelength resolution
1, 3)
Max. resolution: 0.015 nm or better (1510 to 1580 nm, resolution setting: 0.01 nm) Resolution setting: 0.01, 0.02, 0.05, 0.1, 0.2,
0.5, 1.0, 2.0 nm
Resolution accuracy
1, 3)
±5 %: (1300 to 1650 nm, resolution: 0.05 nm or more, resolution correction: ON)
Measurement level range
2, 3)
-90 to +20 dBm (1200 to 1650 nm, sensitivity: HIGH3)
-80 to +20 dBm (600 to 1200 nm, sensitivity: HIGH3)
-60 to +20 dBm (600 to 1750 nm, sensitivity: HIGH3)
Level accuracy
2, 3)
±0.3 dB (1310/1550 nm, input: -30 dBm, sensitivity: HIGH 1-3)
Level linearity
2, 3)
±0.05 dB (Input: +10 to -50 dBm, sensitivity: HIGH 1-3)
Level flatness
2, 3)
±0.1 dB (1510 to 1580 nm)
Polarization dependency
2, 3)
±0.05 dB (1550 nm), ±0.05 dB typ.(1310 nm)
Dynamic range
3)
60 dB (1523 nm, peak: ±0.2 nm, resolution: 0.01 nm) 70 dB (1523 nm, peak: ±0.4 nm, resolution: 0.01 nm)
Sweep time Approx. 500msec. (Span: 100 nm or less,
sensitivity: NORMAL HOLD, ave.: 1, 501 samples, resolution correction: OFF)
Function Automatic Program function (20 programs, 200 steps),
measurement Long-term measurement function Setting of Span setting: 0 to 1200 nm measuring Measuring sensitivity setting: NORMAL conditions HOLD/AUTO, HIGH 1/2/3
Number of averaging setting: 1 to 1000 times Sample number setting: 11 to 20001, AUTO Automatic setting function of measuring conditions Sweep-between-marker function 0nm sweep function Pulse light measurement function Air/vacuum wavelength measurement function TLS synchronized measurement function
Trace display Level scale setting: 0.1 to 10 dB/div, linear
Simultaneous display of 3 independent traces Max./Min. hold display Roll averaging display Calculation-between-traces display Normalized display
Notes
1) Horizontal scale: wavelength display mode
2) Vertical scale: absolute power display mode, resolution: 0.05 nm or more, resolution correction: OFF
3) At 23±3 °C, with 10/125 µm single mode fiber, after 2 hours of warm-up.
Function Trace display Curve-fit display
3D display Split display Power density display, % display, dB/km display Frequency display of horizontal axis scale
Data analysis WDM waveform analysis (Wavelength/
Level/SNR list display) Optical fiber amplifiers analysis (GAIN/NF), PMD analysis DFB-LD analysis, FP-LD analysis, LED analysis, SMSR analysis Peak search, bottom search, spectral width search, notch width search Delta marker (max. 100), line marker (analysis range specification) Graph display of long-term measurement results
Others Self-wavelength calibration function (using
built-in reference light source) Wavelength/Level compensation function, label function, help function
Memory Built-in FDD 3.5-inch 2HD
File format Trace file, program file, measuring
conditions file Text file (trace, analysis data, etc.) Graphics file (BMP, TIFF)
Internal memory
32 traces, 20 programs Data output Printer Built-in high speed printer Interface
Remote control G
P-IB (2 ports)
TLS control interfaces (TTL)
Others Sweep trigger input (TTL)
Sample enable input (TTL)
Sample trigger input (TTL)
Analog output (0 to 5V)
Video output (VGA) Display 9.4-inch color LCD
(Resolution: 640 x 480 dots) Optical connector FC (Standard) Power requirements AC 100 to 120/200 to 240 V, 50/60 Hz,
approx. 200 VA Environmental conditions Operating temperature: 5 to +40 °C
Storage temperature: -10 to +50 °C
Humidity: 80 %RH or less (No condensation) Dimensions and mass Approx. 425(W) x 222(H) x 450(D) mm
Approx. 30 kg
Loading...